The Ambulo 2400 Ambulatory Blood Pressure Monitor is designed to measure systolic and diastolic blood pressure and pulse rate of adults who are eighteen (18) years and older, using the oscillometric method on a cuffed arm.
Device Story
Ambulo 2400 is a compact, battery-powered, non-invasive ambulatory blood pressure monitoring system. Device uses an air micropump, release valves, and a standard arm cuff to measure systolic/diastolic blood pressure and pulse rate via oscillometric method. Intended for 24-hour monitoring to capture fluctuations during daily activities and sleep. Operated by patients in home/clinic settings; physician configures device and interprets downloaded data. Device features an LCD for status/results and a single button for manual measurement initiation or pause. Data is downloaded to a computer for clinical analysis to support diagnosis and treatment effectiveness evaluation. Benefits include improved diagnostic accuracy compared to single-point measurements by accounting for daily BP variability.
Clinical Evidence
Bench testing only. Performance validated against AAMI/ANSI SP10:2002 standards for automated sphygmomanometers. Additional testing included electrical safety (EN/IEC 60601-1), electromagnetic compatibility (EN/IEC 60601-1-2), programmable electrical medical systems (EN/IEC 60601-1-4), usability (EN/IEC 60601-1-6), and automated NIBP safety (EN/IEC 60601-2-30). Environmental and mechanical robustness verified via MIL-STD 810E (temperature, humidity, altitude, shock, vibration), IEC 68-2-29 (mechanical bump), and ISTA Series 2 AB (packaging vibration).
Technological Characteristics
Battery-powered electromechanical device; microcontroller-based; air micropump and release valves; oscillometric sensing principle; standard cuff interface. Dimensions: compact/lightweight. Connectivity: USB cable for computer interface. Display: LCD. Standards: EN/IEC 60601-1, 60601-1-2, 60601-1-4, 60601-1-6, 60601-2-30, AAMI/ANSI SP10:2002, MIL-STD 810E, RoHS compliant.
Indications for Use
Indicated for adults 18+ years for measurement of systolic/diastolic blood pressure and pulse rate via oscillometric method on a cuffed arm.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
A&D Medical model TM-2430 Ambulatory Blood Pressure Monitoring system (K992808)
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
APR 2 3
### Submitter Information:
Tiba Medical, Inc. 2701 NW Vaughn St., Suite 470 Portland, Oregon 97210 USA
Phone: 503-222-1500 FAX: 800-985-8422 www.tibamedical.com
#### USA Contact:
Merat Bagha, President
E-Mail: merat@tibamedical.com Phone: 503-222-1500 FAX: 800-985-8422 www.tibamedical.com
### Device Name:
Trade Name: Ambulo 2400 Ambulatory Blood Pressure Monitoring System
Common Name: Ambulatory Blood Pressure Monitoring System Classification Name: System, Measurement, Blood-Pressure, Non-Invasive
Classification Product code: DXN
#### Predicate Device(s):
The Ambulo 2400 is substantially equivalent to the A&D Medical model TM-2430 Ambulatory Blood Pressure Monitoring system cleared for marker under K992808.
### Device Description:
The Ambulo 2400 is a compact, lightweight, non-invasive ambulatory blood monitoring system. The Ambulo 2400 can measure systolic and diastolic blood pressure, mean arterial pressure and pulse rate over a 24 hour period. It is easily configured and individually-fitted for each patient by a physician or health care professional. Measurements are automatically captured, without medical supervision, by the Ambulo 2400 and later downloaded to a computer for analysis and interpretation by a physician.
Since blood pressure naturally fluctuates throughout a full day according to various factors, such as sleep patterns, medication, diet, exercise and stress, a single measurement is not sufficient to make a sound diagnosis. Analyzing a person's blood pressure over an extended period of time can improve the diagnosis and ultimately the treatment. The Ambulo 2400 can collect blood pressure data under a variety of daily activities over a 24 hour period and then downloaded to a computer for analysis and treatment. The Ambulo 2400 may also be use on an ongoing basis to evaluate the effectiveness of a prescribed treatment.
Page 1 of 3
{1}------------------------------------------------
### Intended Use:
The Ambulo 2400 Ambulatory Blood Pressure Monitor is designed to measure systolic and diastolic blood pressure and pulse rate of adults who are eighteen (18) years and older, using the oscillometric method on a cuffed arm.
# Technology Characteristics:
The Ambulo 2400 Ambulatory Blood Pressure Monitor is a batterypowered electromechanical device that utilizes a microcontroller along with air micropump and release valves to pump up and measure and store blood pressure and pulse rate via a standard cuff and using an oscillometric algorithm. The device is compact and lightweight, intended for everyday use by subjects (typically over a 24-hour period) to determine fluctuations in blood pressure during day and night. The device interfaces to a computer for configuration and download of measurements. The device has an LCD for displaying time of day, current status, measurement results, and/or any error conditions. The device also has a single button that allows the subject to initiate or stop a measurement, and to toggle the device between automatic measurement or PAUSE modes. The device is packaged with accessories including multiple cuffs, a carrying pouch, shoulder strap, application software, and a USB communication cable.
| Standard | Title and Comments |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| EN/IEC60601-1 | EN/IEC 60601 Medical Electrical Equipment - Part 1 General<br>Requirements for Safety, 1988: Amendment 1, 1991, Amendment<br>2, 1995 Note: EN/IEC60601-1, UL2601-1, and CSA, No. 22.2<br>#601.1 M90 cover the same electrical safety requirements |
| EN/IEC60601-1-2:2001 | EN/IEC 60601-1-2:2001 Medical Electrical Equipment - Collateral<br>Standard; General Requirements for Safety: Electromagnetic<br>Compatibility - Requirements and Test. |
| EN/IEC60601-1-4:1996 | EN/IEC 60601-1-4:1996 Medical Electrical Equipment, - Collateral<br>Standard - General Requirements for programmable electrical |
| EN/IEC 60601-1-6:2004 | EN/IEC 60601-1-6:2004 Medical Electrical - Collateral standard;<br>Usability |
| EN/IEC 60601-2-30:2000 | EN/IEC 60601-2-30:2000 Medical Electrical Equipment - Part 2-<br>30: Particular requirements for the safety including essential<br>performance of automated cycling NIBP monitoring equipment |
### Summary of Performance Testing:
{2}------------------------------------------------
ﮐﯿﺴ
| Standard | Title and Comments |
|----------------------|--------------------------------------------------------------------------------------|
| AAMI/ANSI SP10;2002 | AAMI/ANSI SP10;2002 manual, electronic or automated<br>sphygmomanometers |
| EU 2002/95/EC | RoHS-compliance (Pb-free components and solder; mercury-free<br>measurement device). |
| MIL - STD 810E, 1989 | Temperature, Humidity, Altitude, Shock and Vibration |
| IEC 68-2-29 | Operating Mechanical Bump Test |
| EN IEC 60601-1:2000 | Operating Shock |
| ISTA Series 2 AB | Packaging Vibration |
# Conclusion:
The Ambulo 2400 Ambulatory Blood Pressure Monitoring System is as safe and effective as the predicate device when used according to the instructions in the User's Guide supplied with the device.
Page 3 of 3
. .
{3}------------------------------------------------
Public Health Service
Image /page/3/Picture/2 description: The image shows the seal for the Department of Health & Human Services - USA. The seal is circular with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized caduceus, a symbol often associated with medicine and healthcare.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 2 3 2008
Tiba Medical, Inc. c/o Mr. Merat Bagha President 2701 NW Vaughn St., Suite 470 Portland, OR 97201
Re: K080274
Trade/Device Name: Ambulo™ 2400 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Monitoring System Regulatory Class: Class II (two) Product Code: DXN Dated: January 30, 2008 Received: February 1, 2008
Dear Mr. Bagha:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 – Mr. Merat Bagha
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Survoillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
B. Zimmerman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
### INDICATIONS FOR USE
510(k) Number: __ K080274
### Device Name:
Ambulo 2400 Ambulatory Blood Pressure Monitoring System
#### Indication for use:
The Ambulo 2400 Ambulatory Blood Pressure Monitor is designed to measure systolic and diastolic blood pressure and pulse rate of adults who are eighteen (18) years and older, using the oscillometric method on a cuffed arm.
Prescription Use X (Per 21 CFR 801.109)
OR Over-The-Counter_
(Please do not write below this line - continue on another page if needed)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B. Bimmolmo
Vision Sian-O Division of Cardiovascular Devices, 510/ki Numb
Page 1 of 1
Contains Proprietary Information
Page 35 of 648
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.