The Kimberly-Clark* MIC* G Introducer Kits are intended to facilitate primary placement of Kimberly-Clark* and Kimberly-Clark* MIC* brand of Gastrostomy Feeding Tubes. The Kimberly-Clark* MIC-KEY* G Introducer Kits are intended to facilitate primary placement of Kimberly-Clark* and Kimberly-Clark* MIC-KEY* brand of Gastrostomy Feeding Tubes. The Kimberly-Clark* MIC* Jejunal and Kimberly-Clark* MIC* Transgastric-Jejunal Introducer Kit is intended to facilitate primary placement of Kimberly-Clark* MIC* Jejunal Feeding Tube and the Kimberly-Clark* MIC* Transgastric-Jejunal Feeding Tube. The Kimberly-Clark* MIC-KEY* Jejunal and Kimberly-Clark* MIC-KEY* Transgastric-Jejunal Introducer Kit is intended to facilitate primary placement of Kimberly-Clark* MIC-KEY* Low-Profile Jejunal Feeding Tube and the Kimberly-Clark* MIC-KEY* Low-Profile Transgastric-Jejunal Feeding Tube.
Device Story
Introducer kits facilitate primary placement of balloon-retained enteral feeding tubes. Used in interventional radiology or endoscopy suites by clinicians. Kit components include gastropexy device, dilator with peel-away sheath, syringe, hemostat, introducer needle, seeking catheter, scalpel, guidewire, and stoma measuring device. Device functions by dilating the tract and attaching the gastric wall to the anterior abdominal wall to allow tube insertion. Output is the successful placement of the feeding tube, enabling enteral nutrition delivery. Benefits include standardized, essential components for clinical procedure efficiency.
Clinical Evidence
Bench testing only. Biocompatibility, sterilization, and functional testing demonstrated safety and effectiveness. No clinical data presented.
Technological Characteristics
Kit containing mechanical components for percutaneous tube placement: gastropexy, dilator, sheath, syringe, hemostat, needle, catheter, scalpel, guidewire, and measuring device. Materials have established history in medical devices. Non-powered, manual surgical instrumentation.
Indications for Use
Indicated for facilitating primary placement of various Kimberly-Clark brand gastrostomy, jejunal, and transgastric-jejunal feeding tubes in patients requiring enteral nutrition.
Regulatory Classification
Identification
A gastrointestinal tube and accessories is a device that consists of flexible or semi-rigid tubing used for instilling fluids into, withdrawing fluids from, splinting, or suppressing bleeding of the alimentary tract. This device may incorporate an integral inflatable balloon for retention or hemostasis. This generic type of device includes the hemostatic bag, irrigation and aspiration catheter (gastric, colonic, etc.), rectal catheter, sterile infant gavage set, gastrointestinal string and tubes to locate internal bleeding, double lumen tube for intestinal decompression or intubation, feeding tube, gastroenterostomy tube, Levine tube, nasogastric tube, single lumen tube with mercury weight balloon for intestinal intubation or decompression, and gastro-urological irrigation tray (for gastrological use).
Special Controls
*Classification.* (1) Class II (special controls). The barium enema retention catheter and tip with or without a bag that is a gastrointestinal tube and accessory or a gastronomy tube holder accessory is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I (general controls) for the dissolvable nasogastric feed tube guide for the nasogastric tube. The class I device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to § 876.9.
{0}------------------------------------------------
This 510(k) summary of safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
| APPLICANT | Kimberly-Clark*<br>1400 Holcomb Bridge Road<br>Roswell, GA 30076 | FEB 26 2008 |
|-------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------|-------------|
| OFFICIAL<br>CORRESPONDENT | Sherry Saurini<br>Associate Director, Regulatory Affairs<br>Tel: 770.587.8502<br>Fax: 920.969.3455<br>email: sherry.saurini@kcc.com | |
| TRADE NAME: | KIMBERLY-CLARK* Introducer Kits | |
| CLASSIFICATION<br>NAME: | Gastrointestinal tube and accessories | |
| DEVICE<br>CLASSIFICATION<br>AND PRODUCT<br>CODE | Class II per 21 CFR §876.5980<br>Product Code - KNT | |
### SUBSTANTIAL EQUIVALENCE:
The new Kimberly-Clark* Introducer Kits are substantially equivalent to the MIC* Gastrostomy Tube Percutaneous Insertion Kit cleared under K852363, the MIC* Gastro-Enterostomy Tube Modification cleared under K921370 and the MIC* Transgastric Jejunal Tube Kit cleared under K926581. Both the Kimberly-Clark* Introducer Kits and the predicate kit have the same intended use and basic scientific technology.
Both kits have the same intended use, and utilize the same scientific principles, dilation and attachment of the gastric wall to the anterior abdominal wall. Bench testing has demonstrated that the new Kimberly-Clark* Introducer Kits performs the same function as the predicate kits, and that any minor differences between the modified device and the predicate device do not affect safety or efficacy.
8.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the text "K080253" in a bold, sans-serif font. The characters are large and take up most of the frame. The text appears to be a code or identifier, with a mix of letters and numbers.
SUMMARY OF SAFETY & EFFECTIVE
#### DESCRIPTION OF THE DEVICE:
The Introducer Kits consists of a group of essential components required to facilitate placement of a balloon-retained enteral feeding tube. The kit contains these essential components in an easy to use tray and is configured for use in an interventional radiology and /or endoscopy suite.
This device consists of the following components and accessories:
- 1. Gastropexy
- 2. Dilator with Peel-away sheath
- 3. Syringe
- 4. Hemostat
- 5. Introducer Needle
- 6. Seeking Catheter
- 7. Scalpel
- 8. Guidewire
- 9. Stoma Measuring Device
The device consists of the following main components and accessories: list components and accessories.
{2}------------------------------------------------
Image /page/2/Picture/0 description: The image shows a sequence of alphanumeric characters, seemingly handwritten. The sequence starts with the letter 'K', followed by the number '0', then '80253'. The characters are bold and dark, contrasting with the background.
SUMMARY OF SAFETY & EFFECTIVENESS
# INDICATIONS FOR USE:
| Kit Configuration | Indications for Use Statement |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| For the Kimberly-<br>Clark* MIC*G<br>Introducer Kit | The Kimberly-Clark* MIC* G Introducer Kits are intended to<br>facilitate primary placement of Kimberly-Clark* and Kimberly-<br>Clark* MIC* brand of Gastrostomy Feeding Tubes. |
| For the Kimberly-<br>Clark* MIC-KEY*G<br>Introducer Kit | The Kimberly-Clark* MIC-KEY* G Introducer Kits are intended<br>to facilitate primary placement of Kimberly-Clark* and<br>Kimberly-Clark* MIC-KEY* brand of Gastrostomy Feeding<br>Tubes. |
| For the Kimberly-<br>Clark* MIC*J-TJ<br>Introducer Kit | The Kimberly-Clark* MIC* Jejunal and Kimberly-Clark* MIC*<br>Transgastric-Jejunal Introducer Kit is intended to facilitate<br>primary placement of Kimberly-Clark* MIC* Jejunal Feeding<br>Tube and the Kimberly-Clark* MIC* Transgastric-Jejunal<br>Feeding Tube. |
| For the Kimberly-<br>Clark* MIC-KEY*J-<br>TJ Introducer Kit | The Kimberly-Clark* MIC-KEY* Jejunal and Kimberly-Clark*<br>MIC-KEY* Transgastric-Jejunal Introducer Kit is intended to<br>facilitate primary placement of Kimberly-Clark* MIC-KEY*<br>Low-Profile Jejunal Feeding Tube and the Kimberly-Clark*<br>MIC-KEY* Low-Profile Transgastric-Jejunal Feeding Tube. |
## PERFORMANCE DATA:
All components that come in direct contact with the patient have a history of use in medical devices and are biocompatibility, sterilization and functional test results demonstrate that the device is safe and effective for use in humans. Biocompatibility summaries can be found in Attachment D.
{3}------------------------------------------------
Image /page/3/Picture/0 description: The image contains the number 8 followed by a period. The number is written in a bold, sans-serif font. The number is black and is set against a white background. The period is the same color and font as the number.
Summary of Safety & E
# CONCLUSION:
Based on the performance testing, it can be concluded that the new Kimberly-Clark* Introducer Kits are equivalent to the predicates MIC* Gastrostomy Tube Percutaneous Insertion Kit (K852363), MIC* Enterostomy Tube Modification (K921370) and MIC* Transgastric Jejunal Tube Kit (K926581) with respect to intended use and technological characteristics.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular, with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. In the center of the seal is a stylized image of an eagle, with its wings spread and its head turned to the left. The eagle is a symbol of the United States, and its presence on the seal indicates the department's role in protecting the health and well-being of Americans.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB 2 6 2008
Ms. Sherry Saurini Associate Director, Regulatory Affairs Kimberly-Clark* Corporation 1400 Holcomb Bridge Road ROSWELL GA 30076
Re: K080253
Trade/Device Name: Kimberly-Clark* MIC* G, MIC-KEY* G, MIC* Jejunal, MIC* Transgastric-Jejunal, MIC-KEY* Jejunal and MIC-KEY* Transgastric-Jejunal Introducer Kits Regulation Number: 21 CFR §876.5980 Regulation Name: Gastrointestinal tube and accessories Regulatory Class: II Product Code: KGC Dated: January 30, 2008 Received: January 31, 2008
Dear Ms. Saurini:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act). You may, therefore, market the device, subject to the general controls provisions of the Act. However, you are responsible to determine that the medical devices you use as components in the kit have either been determined as substantially equivalent under the premarket notification process (Section 510(k) of the act), or were legally on the market prior to May 28, 1976, the enactment date of the Medical Device Amendments. Please note: If you purchase your device components in bulk (i.e., unfinished) and further process (e.g., sterilize) you must submit a new 510(k) before including these components in your kit. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, and labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
#### Page 2 -- Ms. Sherry Saurini
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on the labeling regulation, please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification''(21 CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll free number (800) 638-2041 or (240) 276-3150, or at its Internet address http://www.fda.gov/cdrh.dsmamain.html.
Sincerely yours,
Nancy C Brogdon
Nancy C. B. rogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
K080253
1 of 1
# Indications for Use
510(k) Number (if known):_____________________________________________________________________________________________________________________________________________________ K080253
Device Name: _________________________________________________________________________________________________________________________________________________________________
Indications For Use:
| Kit Configuration | Indications for Use Statement |
|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| For the Kimberly-<br>Clark* MIC*G<br>Introducer Kit | The Kimberly-Clark* MIC* G Introducer Kits are intended to<br>facilitate primary placement of Kimberly-Clark* and Kimberly-<br>Clark* MIC* brand of Gastrostomy Feeding Tubes. |
| For the Kimberly-<br>Clark* MIC-KEY*G<br>Introducer Kit | The Kimberly-Clark* MIC-KEY* G Introducer Kits are intended<br>to facilitate primary placement of Kimberly-Clark* and<br>Kimberly-Clark* MIC-KEY* brand of Gastrostomy Feeding<br>Tubes. |
| For the Kimberly-<br>Clark* MIC*J-TJ<br>Introducer Kit | The Kimberly-Clark* MIC* Jejunal and Kimberly-Clark* MIC*<br>Transgastric-Jejunal Introducer Kit is intended to facilitate<br>primary placement of Kimberly-Clark* MIC* Jejunal Feeding<br>Tube and the Kimberly-Clark* MIC* Transgastric-Jejunal<br>Feeding Tube. |
| For the Kimberly-<br>Clark* MIC-KEY*J-<br>TJ Introducer Kit | The Kimberly-Clark* MIC-KEY* Jejunal and Kimberly-Clark*<br>MIC-KEY* Transgastric-Jejunal Introducer Kit is intended to<br>facilitate primary placement of Kimberly-Clark* MIC-KEY*<br>Low-Profile Jejunal Feeding Tube and the Kimberly-Clark*<br>MIC-KEY* Low-Profile Transgastric-Jejunal Feeding Tube. |
Prescription Use × (Part 21 CFR 801 Subpart D) ANDYOR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy C. Brogdon
(Division Sign-t Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number,
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.