The CMS network transfers information between HYPERVISOR VI Central Monitoring System and other networked devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and/or WLAN connections. CMS can be used for remote monitor management, storing, printing, reviewing or processing of information from networked devices, and it is operated by medical personnel in hospitals or medical institutions. Telemetry Monitoring System is a sub-system of CMS, intended to obtain ECG and SpO2 physiological information from adult and pediatric patients, and send it to CMS via WMTS frequency within a defined coverage area.
Device Story
The HYPERVISOR VI Central Monitoring System (CMS) is a PC-based medical information system that aggregates, stores, prints, and reviews physiological data from networked medical devices. The system includes a Telemetry Monitoring System (TMS) sub-system that collects ECG and SpO2 signals via sensors, transmitting them wirelessly over the WMTS band to receivers, which then forward the data to the CMS. The system is operated by medical personnel in hospitals or medical institutions to facilitate remote monitor management and centralized clinical oversight. Key features include ViewOtherBed functionality and real-time printing. By centralizing patient data, the device assists clinicians in monitoring patient status, potentially improving clinical decision-making and patient care efficiency.
Clinical Evidence
Bench testing only. Laboratory testing was conducted to validate and verify that the CMS and TMS meet all design specifications and comply with applicable standards. No clinical data was provided.
Technological Characteristics
PC-based platform; network connectivity via hardwired cables and WLAN; wireless transmission via WMTS band; physiological sensors for ECG and SpO2; software-based data processing, storage, and display.
Indications for Use
Indicated for adult and pediatric patients requiring monitoring of ECG, heart rate, arrhythmia, ST segment, SpO2, and pulse rate in hospital or medical institution settings.
Regulatory Classification
Identification
A cardiac monitor (including cardiotachometer and rate alarm) is a device used to measure the heart rate from an analog signal produced by an electrocardiograph, vectorcardiograph, or blood pressure monitor. This device may sound an alarm when the heart rate falls outside preset upper and lower limits.
Spacelabs Medical Ultraview Digital Telemetry System (K983996)
Submission Summary (Full Text)
{0}------------------------------------------------
APR - 3 2008
# 510(K) SUMMARY
This summary of 510(k) safety and effectiveness information is submitted in accordance with the requirements of SMDA 1990 and 21 CFR § 807.92.
08019-The assigned 510(k) number is:
#### Submitter:
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China Tel: +86 755 2658 2888 Fax: +86 755 2658 2680
- Contact Person:
- Li Dongling
Shenzhen Mindray Bio-medical Electronics Co., LTD Mindray Building, Keji 12th Road South, Hi-tech Industrial Park, Nanshan, Shenzhen, 518057, P. R. China
Date Prepared: .
January 4, 2008
### Name of the Device:
- Trade/Proprietary Name: HYPERVISOR VI Central Monitoring System . (including TMS)
- Common Name: Central Monitoring System (CMS) ●
- Classification: ●
MSX, 21 CFR Part 870.2300 MHX, 21 CFR Part 870.1025
DQA, 21 CFR Part 870.2700
DPZ, 21 CFR Part 870.2710
DRG, 21 CFR Part 870.2910
System, Network and Communication, Physiological Monitors Monitor, Physiological, Patient (With Arrhythmia Detection or Alarms) Oximeter Oximeter, Ear Transmitters and Receivers, Physiological Signal, Radiofrequency
ge 1 OF 3
{1}------------------------------------------------
#### Legally Marketed Predicate Device:
HYPERVISOR VI Central Monitoring System, Shenzhen Mindray Bio-Medical K062194 Electronics Co., LTD.
Spacelabs Medical Ultraview Digital Telemetry System, Spacelabs Medical, Inc. K983996
#### Description:
Central Monitoring System network is a kind of medical information system, which consists of different networked devices (which have separate 510(k) clearance). Hypervisor VI Central Monitoring System (CMS) is the primary maintainer of communication between other networked devices. It can store, print, review or process information from networked devices. It can also realize remote monitor management function to free doctors from clinical monitoring work and conduct centralized monitoring management.
In this submission, CMS adds the function of supporting communication with Telemetry Monitoring System. Besides this, CMS is improved to include some new functions, such as, add the function of ViewOtherBed and add the function of real-time print.
By using radiofrequency signal, TMS is intended to monitor Electrocardiogram (ECG), Heart Rate (HR), Arrhythmia Detection, ST Segment Analysis, Saturation of Pulse Oxygen (SpO2) and Pulse Rate (PR) for adult and pediatric patients. Physical signals are collected by sensors and wirelessly transmitted by transmitters in WMTS band, receivers get the signals and forward to CMS for processing, displaying, storing, printing, etc. It can be used within a defined coverage area in hospitals or medical institutions.
CMS is based on the PC platform, and the PC can be configured by users according to the requirements on the operator's manual.
#### Statement of Intended Use:
The CMS network transfers information between HYPERVISOR VI Central Monitoring System and other networked devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and/or WLAN connections. CMS can be used for remote monitor management, storing, printing, reviewing or processing of information from networked devices, and it is operated by medical personnel in hospitals or medical institutions.
Telemetry Monitoring System is a sub-system of CMS, intended to obtain ECG and SpO2 physiological information from adult and pediatric patients, and send it to CMS via WMTS frequency within a defined coverage area.
#### Comparison of Technological Characteristics:
The CMS is substantially equivalent to the previously cleared CMS (K062194) in technological characteristics. Both of them apply the same design principles, and the
Page 2 of 3
{2}------------------------------------------------
modifications are concentrated on software consummating and network testing. The TMS is substantially equivalent to Ultraview Digital Telemetry System (K983996). Both of them monitor similar parameters of patients, and apply similar design principles.
## Testing:
Laboratory testing has been conducted to validate and verify that the CMS (including TMS) meet all design specifications. Results of these tests demonstrate compliance to the requirements of all applied standards.
: 上
### Conclusion:
The Central Monitoring System is substantially equivalent to the predicate devices.
Page 3 of 3
{3}------------------------------------------------
Image /page/3/Picture/12 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized caduceus, a symbol of medicine and health. The caduceus has three snakes winding around a staff.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR - 3 2008
Shezhen Mindray Bio-Medical Electronics Co., Ltd c/o Ms. Susan D. Goldstein-Falk MDI Consultants, Inc. 55 Northern Blvd, Suite 200 Great Neck, NY 11021
Re: K080192
HYPERVISOR VI Central Monitoring System (including Telemetry Monitoring System, Model TMS-6016) Regulation Number: 21 CFR 870.2300 Regulation Name: Cardiac Monitor (including Cardiotachometer and Rate Alarm) Regulatory Class: Class II (two) Product Code: MSX Dated: January 4, 2008 Received: January 25, 2008
Dear Ms. Goldstein-Falk:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Page 2 - Ms. Susan D. Goldstein-Falk
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.
Sincerely yours,
B.J. Zimmerman for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
KO80x92
Device Name: HYPERVISOR VI Central Monitoring System (including TMS)
#### Indications for Use:
The CMS network transfers information between HYPERVISOR VI Central Monitoring System and other networked devices. It also allows information transfer between several CMS. Network connections consist of hardwired network cables and/or WLAN connections. CMS can be used for remote monitor management, storing, printing, reviewing or processing of information from networked devices, and it is operated by medical personnel in hospitals or medical institutions.
Telemetry Monitoring System is a sub-system of CMS, intended to obtain ECG and SpO2 physiological information from adult and pediatric patients, and send it to CMS via WMTS frequency within a defined coverage area.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
B. Vettumtor
Page
stan Sign-Of scular Devices
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.