WRIST BLOOD PRESSURE MONITOR, MODEL WS-1100/WS-1100PV
Applicant
Nihon Seimitsu Sokki Co., Ltd.
Product Code
DXN · Cardiovascular
Decision Date
May 9, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.1130
Device Class
Class 2
Indications for Use
The WS-1100/WS-1100PV system is intended for noninvasive measurement of systolic and diastolic blood pressures and determination of pulse rate in adults in a homecare environment. The device features include the display of irregular pulse rhythm detection, the classification display of measured blood pressure values against the guideline by World Health Organization or equivalent guidelines, the personal setting for individual blood pressure target values, the two memory banks to save the measurement results with date and time of measurement and the transferring the saved results to personal computers.
Device Story
Automatic wrist sphygmomanometer for home use; measures systolic/diastolic blood pressure and pulse rate via oscillometric method. Device consists of main unit and preformed cuff (4.9-8.5 inch circumference). User wraps cuff around wrist; device inflates/deflates automatically. Microprocessor-controlled pressure-frequency converter senses arterial oscillations; converts amplitude to mmHg. Features include: irregular pulse rhythm detection (compares pulse intervals to mean; flashes if >25% difference); WHO-based blood pressure classification; user-defined target value alerts; two memory banks (60 readings each) with date/time stamping; PC connectivity via USB for data review. Operated by patient; provides digital LCD output. Benefits include convenient self-monitoring of cardiovascular metrics and tracking of trends over time.
Clinical Evidence
Bench testing performed in accordance with IEC60601 and ANSI/AAMI SP-10 standards. Clinical performance evaluated to confirm accuracy across expanded wrist circumference range (up to 21.5 cm). No specific clinical trial results (e.g., sensitivity/specificity) provided in summary; substantial equivalence supported by adherence to consensus standards and comparison to predicate performance.
Technological Characteristics
Oscillometric blood pressure measurement; microprocessor-controlled. Materials: ABS (housing), Nylon (cuff), PMMA (display panel). Power: 2x AAA alkaline batteries. Connectivity: USB for PC data transfer. Standards: IEC60601, ANSI/AAMI SP-10. Form factor: Wrist-mounted unit with integrated cuff.
Indications for Use
Indicated for noninvasive measurement of systolic/diastolic blood pressure and pulse rate in adults in homecare settings.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
Predicate Devices
Nissei Model WS-500 Digital Wrist Blood Pressure Monitor (K003444)
Digital Blood Pressure Monitor, Model DS-1901 (K071384)
Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U (K061403)
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the text "K080177" on the top line and "Pi/4" on the second line. The text is written in a handwritten style. The numbers are clearly visible, and the letters are also easily readable.
### 5. 510(K) Summary
This document was prepared in accordance with 21 CFR 807.92.
MAY - 9 2008
# Section (a)
| (1) | Name of the submitter: | Nihon Seimitsu Sokki Co., Ltd. |
|-----|------------------------------------|-------------------------------------------------------|
| | Address of the submitter: | 2508-13 Nakago, Shibukawa, Gunmma 377-0293, Japan |
| | Telephone number of the submitter: | 81-279-20-2311 |
| | Contact person: | Mitsuo Kanai |
| | Date of documentation: | January 22, 2008 |
| (2) | Trade name of the device: | Wrist Blood Pressure Monitor, Model WS-1100/WS-1100PV |
| | Common name: | Blood pressure monitor |
| | Classification: | Class II, 74 DXN, 21 CFR 870.1130 - Cardiovascular |
| | | devices panel |
| (3) | The predicate devices: | Nissei Model WS-500 Digital Wrist Blood Pressure |
| | | Monitor, K003444 |
| | | Digital Blood Pressure Monitor, Model DS-1901, |
| | | K071384 |
| | | Microlife Wrist Watch Blood Pressure Monitor, |
### Description of the device: (4)
Wrist Blood Pressure Monitor, Model WS-1100/WS-1100PV is an automatic sphygmomanometer to be used in a homecare environment. Blood pressure, systolic and diastolic, and pulse rate are measured in the wrist artery using the oscillometric method. WS-1100WS-1100PV is a single-mounted device of the main unit and the cuff unit. ABS and PMMA are used for the outer housing of the main unit and the Lenticular image is printed as the positioning guide to remind the user to take correct posture for blood pressure measurement. The preformed cuff unit, which is applicable to wrist circumference approximately between 4.9 and 8.5 inches (between 125 and 215 mm), includes the inflatable bladder and the nylon shell. The device consists of the microprocessor, the
Model BP3AX1-4U, K061403
{2}------------------------------------------------
pressure-frequency converter, the operation keys, the pump, the electromagnetic deflation control valve (ECV) and the display. The device is powered by two AAA alkaline batteries. The cuff, wrapped over the user's wrist, is inflated and deflated during the course of a measurement. Circuits within the cuff sense the oscillations in pressure against the cuff produced by the expansion and contraction of the arteries in the wrist in response to each heart beat. The amplitude of each pressure waves is measured, converted to millimeters of mercury, and displayed on the LCD as a digital value.
Measured blood pressure values are classified against the guideline by World Health Organization or equivalent guidelines and the classified levels are displayed in the reading display. The device compares the measured blood pressure value to the individual target value set by the user and flashes the value when it exceeds the target value.
The device also compares the longest and the shortest time intervals of the detected pulse waves to the mean time interval and flashes heart mark to indicate the detection of irregular pulse rhythm when the difference of the time intervals is over 25 %.
The user of WS-1100/WS-1100PV can choose to save the measurement results in either one of two memory banks, each of which has the capacity for sixty readings, or not to save the results at the end of a measurement. The results are saved along with the date and time of the measurement and the user can review the saved readings. The device also has feature to display the saved readings taken in the morning and the evening separately. The user can also review the saved readings on a personal computer by downloading saved readings using the designated USB cable included in the product package. The application software is downloaded over the internet.
### Intended use of the device: (5)
The WS-1100/WS-1100PV system is intended for noninvasive measurement of systolic and diastolic blood pressures and determination of pulse rate in adults in a homecare environment.
The device features include the display of irregular pulse rhythm detection, the classification display of measured blood pressure values against the guideline by World Health Organization or equivalent guidelines, the personal setting for individual blood pressure target values, the two memory banks to save the measurement results with date and time of measurement and the transferring the saved results to personal computers.
{3}------------------------------------------------
The subject device is basically a modification of the Nissei Model WS-500 Digital Wrist Blood Pressure Monitor, K003444, to include the product features of Digital Blood Pressure Monitor, Model DS-1901, K071384 and Microlife Wrist Watch Blood Pressure Monitor, Model BP3AX1-4U, K061403. Although the specified age of intended patient population is not exactly the same due to the amendment of SP-10, one of the consensus standard for the non-invasive blood pressure monitors, the intended use of the predicate device, Nissei Model WS-500, to measure blood pressure and pulse rate in adults, is the principal intended diagnostic use of the device shared by the other two predicate devices, Model DS-1901 and Model BP3AX1-4U. These two devices, however, have extra indications for use, which are more related to product features than the scientific technology. The intended use of the subject device also includes these extra indications and further indications related to product features in the principle intended use. Therefore the difference of the intended use is not critical to the intended diagnostic use and does not affect the safety and effectiveness of the device.
#### Technological characteristics of the subject device and the predicate device: (6)
The predicate device which the subject device is to be compared to in respect of the substantial equivalency of technological characteristics is Nissei Model WS-500 Digital Wrist Blood Pressure Monitor.
The subject device and Nissei Model WS-500 are automatic blood pressure monitor with the main unit and the inflatable cuff unit singly-mounted and powered with two AAA batteries, operated by the user with push button controls. With both devices, blood pressure is taken at the wrist, using the oscillometric method.
The subject device and the predicate mainly uses the same materials for the outer housing of the main unit and the surface of the cuff unit, where become in contact with the users. ABS is used for the upper and bottom cases of the main unit and Nylon is for the surface of the cuff unit. The only difference is the material used for the display panels. While the display panel of the predicate device is made of polycarbonate (PC), that of the subject device is made of polymethyl methacrylate (PMMA). PMMA is new with the subject device when compared to the predicate device. PMMA, however, is used with one of the other predicate device, Model DS-1901, which is also our product, as well the material is used in as wide range of fields and applications as the PC, the material used with the predicated device.
{4}------------------------------------------------
The subject device and the predicate device share the same principal design construction and the same energy source and mainly use the same materials. Although the materials used for the display panels raise the difference between two devices, it neither raises new questions regarding the safety and the effectiveness of the subject device nor affects the substantial equivalency of the device.
## Section (b)
The predicate devices were tested to evaluate their safety and effectiveness, including electrical safety, electromagnetic compatibility, environmental safety and clinical performance, in accordance with IEC60601 and ANSI/AAMI SP-10 standards. With the clinical studies performed with the subject device, however, there is a difference in the test protocol when it is compared to that of the predicate device, Nissei Model WS-500. The protocol for the predicate device states to exclude patients with wrist circumference greater than 20.5 cm while the exclusion criteria for wrist circumference is changed to 21.5 cm (8.5 inches) with the subject device. This is to expand the applicable wrist circumference range of the subject device.
The subject device was also tested and fulfilled the requirements from these standards. Although there is a difference in test protocol as described above, it is concluded that the subject device is substantially equivalent to the predicate devices.
{5}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
UAY - 9 2008
Nihon Seimitsu Sokki Co., Ltd. c/o Mr. Koji Kobo c/o Cosmos Corporation, Tokyo Office 3F, 2-17-6 Akebono- Cho Tachikawa- Shi TOKYO 190-0012 JAPAN
Re: K080177
Wrist Blood Pressure Monitor, Model WS-1100, WS-1100PV Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive blood pressure measurement system Regulatory Class: Class II (two) Product Code: DXN Dated: April 24, 2008 Received: April 28, 2008
Dear Mr. Kobo:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Page 2 - Mr. Koji Kobo
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050, This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807,97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Bram/D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and · Radiological Health
Enclosure
{7}------------------------------------------------
510(K) Number: K080177
Device Name: Wrist Blood Pressure Monitor, Model WS-1100/WS-1100PV/
# Indications for Use:
WS-1100WS-1100PV system is intended for noninvasive measurement of systolic and diastolic blood pressure and determination of pulse rate in adults in a homecare environment.
The device features include the display of irregular pulse rhythm detection, the classification display of measured blood pressure values against the guideline by World Health Organization or equivalent guidelines, the personal setting for individual blood pressure target values, the two memory banks to save the measurement results with date and time of measurement and the transferring the saved results to personal computers.
Prescription Use______________________________________________________________________________________________________________________________________________________________ AND / OR Over-The Counter Use ________________________________________________________________________________________________________________________________________________ × (Per 21 CFR 801.109 Subpart D) (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
| (Division Sign-Off) | |
|------------------------------------|---------|
| Division of Cardiovascular Devices | |
| 510(k) Number | 1080177 |
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.