ELECSYS PROBNP II CALCHECK

K080147 · Roche Diagnostics Corp. · JJX · Feb 27, 2008 · Clinical Chemistry

Device Facts

Record IDK080147
Device NameELECSYS PROBNP II CALCHECK
ApplicantRoche Diagnostics Corp.
Product CodeJJX · Clinical Chemistry
Decision DateFeb 27, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1660
Device ClassClass 1

Intended Use

For use in the verification of the calibration established by the Elecsys proBNP II reagent on the Elecsys and cobas e immunoassay analyzers. The electrochemiluminescence immunoassay “ECLIA” is intended for use on the Elecsys and cobas e immunoassay analyzers.

Device Story

Elecsys proBNP II CalCheck is a lyophilized quality control material; consists of NT-proBNP (1-76) amide human serum and potassium phosphate buffered matrix. Used in clinical laboratory settings to verify calibration of Elecsys proBNP II reagent on Elecsys and cobas e immunoassay analyzers. Product requires reconstitution with distilled or deionized water before use. Healthcare professionals use the output to confirm the accuracy of the immunoassay analyzer's calibration, ensuring reliable patient test results for proBNP levels.

Clinical Evidence

No clinical data. Bench testing only. Stability testing protocols and acceptance criteria were reviewed and found acceptable. Values are assigned via multiple analyses on multiple analyzers calibrated against reference standards.

Technological Characteristics

Lyophilized product; NT-proBNP (1-76) amide in human serum and potassium phosphate buffered matrix. Three concentration levels. Storage 2-8°C. Reconstituted stability: 4 hours at 25°C. Analyte concentration levels achieved via spiking during manufacture.

Indications for Use

Indicated for use as a quality control material to verify calibration of the Elecsys proBNP II reagent on Elecsys and cobas e immunoassay analyzers. Prescription use only.

Regulatory Classification

Identification

A quality control material (assayed and unassayed) for clinical chemistry is a device intended for medical purposes for use in a test system to estimate test precision and to detect systematic analytical deviations that may arise from reagent or analytical instrument variation. A quality control material (assayed and unassayed) may be used for proficiency testing in interlaboratory surveys. This generic type of device includes controls (assayed and unassayed) for blood gases, electrolytes, enzymes, multianalytes (all kinds), single (specified) analytes, or urinalysis controls.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K080147 ## 510(k) Summary FEB 27 2009 - | Introduction | According to the requirements of 21 CFR 807.92, the following information provides sufficient detail to understand the basis for a determination of substantial equivalence. | |----------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Submitter name, address, contact | Roche Diagnostics 9115 Hague Road Indianapolis, IN 46250 317-521-2458 | | | Contact Person: Kay A. Taylor | | | Date Prepared: January 18, 2008 | | Device Name | Proprietary name: Elecsys proBNP II CalCheck | | | Common name: proBNP II CalCheck | | | Classification name: Single (specified) analyte controls (assayed and unassayed) | | Predicate device | The Elecsys proBNP II CalCheck is substantially equivalent to other products in commercial distribution intended for similar use. We claim equivalency to the currently marketed Elecsys ProBNP CalCheck (K020883). | | Device Description | The Elecsys proBNP II CalCheck is a lyophilized product consisting of NT-proBNP (1-76) amide human serum and potassium phosphate buffered matrix. During manufacture, the analytes are spiked into the matrix at the desired concentration levels. | | Intended use | For use in the verification of the calibration established by the Elecsys proBNP II reagent on the Elecsys and cobas e immunoassay analyzers. | | | Continued on next page | . {1}------------------------------------------------ ## 510(k) Summary, Continued The table below compares Elecsys proBNP II CalCheck with the predicate Comparison Table device, Elecsys ProBNP CalCheck (K020883). . . . . . ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ | Characteristic | Elecsys proBNP CalCheck<br>(Predicate) | Elecsys proBNP II CalCheck | |----------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------| | Intended Use | For use in the verification of the<br>calibration established by the<br>Elecsys proBNP reagent on the<br>indicated Elecsys and cobas e<br>immunoassay analyzers. | For use in the verification of the<br>calibration established by the Elecsys<br>proBNP II reagent on the Elecsys and<br>cobas e immunoassay analyzers. | | Levels | Three | same | | Format | Lyophilized | same | | Handling | Reconstitute with exactly 1.0 mL<br>distilled or deionized water and allow<br>standing closed for 15 minutes, then<br>mixing gently. | Reconstitute with exactly 1.0 mL<br>distilled or deionized water and allow<br>standing closed for 15 minutes, then<br>mixing gently. | | Stability | Unopened:<br>· Store at 2-8℃ until expiration date<br>Reconstituted:<br>● 20-25℃:4 hrs | same | | Matrix | Level 1: NT-proBNP free human<br>serum matrix<br>Levels 2/3: synthetic NT-proBNP in<br>human serum / buffer matrix | same | The Elecsys proBNP II CalCheck was evaluated for value assignment and Performance Characteristics stability. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wing. The eagle is facing right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 2098 Gaither Road Rockville MD 20850 FEB 2 7 2008 Roche Diagnostics c/o Ms. Kay Taylor, (ASCP) Regulatory Affairs Principal 9115 Hague Road P.O. Box 50416 Indianapolis, IN 46250-0416 k080147 Re: > Trade Name: Elecsys proBNP II CalCheck Regulation Number: 21 CFR 862.1660 Regulation Name: Quality control material (assayed and unassayed) Regulatory Class: Class I reserved Product Code: JJX Dated: January 18, 2008 Received: January 22, 2008 Dear Ms. Taylor: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21. Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {3}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97), You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Jean M. Cooper, M.S., D.V.M. Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for Use - Elecsys proBNP II CalCheck 510(k) Number (if known): 2080147 Device Name: Elecsys proBNP II CalCheck Indication For Use: For use in the verification of the calibration established by the Elecsys proBNP II reagent on the Elecsys and cobas e immunoassay analyzers. The electrochemiluminescence immunoassay "ECLIA" is intended for use on Elecsys and cobas e immunoassay analyzers. Prescription Use XXX (21 CFR Part 801 Subpart D) And/Or Over the Counter Use _________________________________________________________________________________________________________________________________________________________ (21 CFR Part 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE; CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of In Vitro Diagnostic Device Evaluation and Safety (OIVD) Signature Divisign Sign-Off Office of In Vitro Diagnostic Device Evaluation and Safety 510(k) K080147
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...