VIRTUAL PORTS ENDOCLEAR SYSTEM

K080051 · Virtual Ports, Ltd. · GCJ · Mar 21, 2008 · Gastroenterology, Urology

Device Facts

Record IDK080051
Device NameVIRTUAL PORTS ENDOCLEAR SYSTEM
ApplicantVirtual Ports, Ltd.
Product CodeGCJ · Gastroenterology, Urology
Decision DateMar 21, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2

Indications for Use

EndoClear™ Laparoscopes Accessory is intended to be used by qualified physicians to provide endoscope lens cleaning for uninterrupted visualization of internal structures in a wide variety of diagnostic and therapeutic laparoscopic procedures.

Device Story

EndoClear™ system is a sterile, single-patient-use accessory for laparoscopes. It consists of an internally anchored, hands-free lens cleaner and an introducer. The lens cleaner is inserted into the abdominal cavity via a cannula using the introducer and anchored to the internal abdominal wall. It remains in position throughout the surgery, allowing the surgeon to clean the laparoscope lens of blood, fat, fog, and secretions without removing the scope from the cavity. The introducer is also used to remove the cleaner at the end of the procedure. Used in clinical settings by surgeons to maintain uninterrupted visualization, potentially reducing procedure time and improving surgical efficiency.

Clinical Evidence

Bench testing and an animal study were conducted. Bench tests confirmed the device performs as intended. The animal study evaluated safety and effectiveness as an internal lens cleaner, demonstrating that the system functions as designed with no safety or effectiveness concerns identified.

Technological Characteristics

Sterile, single-patient-use system. Materials are biocompatible per ISO 10993-1. Mechanical device consisting of an internally anchored lens cleaner and an introducer cannula. No electronic, software, or energy-based components.

Indications for Use

Indicated for use by qualified physicians for endoscope lens cleaning during diagnostic and therapeutic laparoscopic procedures to maintain visualization.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ p. 1 of 3 Image /page/0/Picture/1 description: The image shows the logo for "Virtual Ports Laparoscopy Systems". The logo features a stylized checkmark-like symbol on the left, followed by the words "Virtual Ports" in a bold, sans-serif font. Below "Virtual Ports", the words "Laparoscopy Systems" are written in a smaller, less bold font. Virtual Ports Ltd. # 510(K) SUMMARY # 510(k) Number K Q8 005 | MAR 2 1 2008 | Applicant's Name: | Virtual Ports<br>Teradion Industrial Park<br>6 Kahol Street<br>D.N Misgav 20179<br>ISRAEL<br>Tel: (972)4-999-0189<br>Fax: (972)4-999-1901 | |---------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Yoram Levy, Qsite<br>31 Haavoda Street<br>Binyamina, Israel 30500<br>Tel (972)4-638-8837; Fax (972)4-638-0510<br>Yoram@qsitemed.com | | Trade Name: | EndoClear™ system | | Common name: | Accessory to a Laparoscope | | Classification: | Name: Endoscope and accessories<br>Product Code: GCJ<br>Regulation No: 876.1500<br>Class: II<br>Classification Panel: General and Plastic Surgery | | Predicate Devices: | Substantial equivalence to the following predicate device is claimed: | | | 1. Laparoscope and Monopolar laparoscopic instruments; Instrumed<br>International, Inc. K040855.<br>2. g-Lix™ Tissue Grasper; USGI Medical K061268 | | Device Description: | The Virtual Ports EndoClear™ system is a sterile, single patient use<br>system consisting of: EndoClear™ Lens Cleaner and the EndoClear™<br>Introducer. The EndoClear Lens Cleaner is an internally anchored<br>hands-free, laparoscope lens cleaning device which is attached to the<br>internal abdominal cavity wall and remains in position until | Virtual Ports Ltd. | EndoClear™ 510k Notification {1}------------------------------------------------ Ko8oosi f 2 of 3 Image /page/1/Picture/2 description: The image shows the logo for Virtual Ports Laparoscopy Systems. The logo features a stylized figure on the left, followed by the words "Virtual Ports" in a bold, sans-serif font. Below "Virtual Ports" are the words "Laparoscopy Systems" in a smaller font. # Virtual Ports Ltd. completion of the surgery, enabling the surgeon to effectively clean the lens of blood, fat, fog, and secretions without removing it from the cavity. The Virtual Ports EndoClear™ Lens Cleaner is introduced via a cannula using the EndoClear™ Introducer, which also removes the EndoClear™ Lens Cleaner at the end of the surgical procedure. #### Intended Use Statement: EndoClear™ Laparoscopes Accessory is intended to be used by qualified physicians to provide endoscope lens cleaning for uninterrupted visualization of internal structures in a wide variety of diagnostic and therapeutic laparoscopic procedures. ### Performance Validation: Performance Testing - bench tests Series of bench tests were performed to ensure that the device performs as intended. All testing results demonstrated satisfactory performance. #### Performance Testing - animal An animal study was performed in order to evaluate the safety and effectiveness of using the Virtual Ports EndoClear™ System as an internal endoscope lens cleaner. This study demonstrated that the EndoClear™ system performs as intended and that no safety and effectiveness questions were raised. #### Tests conclusion: Both bench tests and the animal study were shown that the device performs safely and efficiently in accordance with its intended use. #### Materials: {2}------------------------------------------------ jc040051 . 30 r. Image /page/2/Picture/2 description: The image shows a logo with the words "Virtual Ports" in a bold, sans-serif font. Below "Virtual Ports" is the text "Laparoscopy Systems" in a smaller font. To the left of the text is a stylized graphic that resembles a check mark or a stylized letter V. There is a small dot above the graphic. # Virtual Ports Ltd. Materials of the EndoClear ™ device that are in contact with the human body are biocompatible in accordance with ISO 10993-1. ## Substantial Equivalence: Preclinical and bench performance data was supplied to demonstrate that the EndoClear ™ meets its labeled performance claims, and to demonstrate substantial equivalence to the predicate devices. {3}------------------------------------------------ Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle with three stripes extending from its back, enclosed within a circle. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged around the circumference of the circle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## MAR 2 1 2008 Virtual Ports % Osite Yoram Levy General Manager 31 Haavoda Street Binyamina, Israel Re: K080051 Trade/Device Name: EndoClear™ Regulation Number: 21 CFR 876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Code: GCJ Dated: March 5, 2008 Received: March 10, 2008 Dear Yoram Levy: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set {4}------------------------------------------------ Page 2 - Yoram Levy forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark N Millman Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ 1080051 p. lot l Image /page/5/Picture/2 description: The image shows the logo for "Virtual Ports Laparoscopy Systems". The logo features a stylized "V" shape on the left, with a dot above the center of the "V". To the right of the "V" is the text "Virtual Ports" in a bold font. Below "Virtual Ports" is the text "Laparoscopy Systems" in a smaller font. # Virtual Ports Ltd. ### INDICATIONS FOR USE STATEMENT 510(k) Number (if known): Device Name: EndoClear™ System Indications for Use: EndoClear™ Laparoscopes Accessory is intended to be used by qualified physicians to provide endoscope lens cleaning for uninterrupted visualization of internal structures in a wide variety of diagnostic and therapeutic laparoscopic procedures. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) ### (PLEASE DO NOT WRITE BELOW THIS LINE -CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) (Division Sign-off) Division of Division of General, Restorative and Neurological Devices 510(k) Number Mark A. Melhem Division Sig - Off Division of General. Restorative, and Neurological Devices *510(k) Number K080051 1-2 Virtual Ports Ltd. | EndoClear™ 510k Notification
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...