← Product Code [LLZ](/productcode/LLZ) · K080013

# REGISTRATION AND FUSION (K080013)

_Agfa Healthcare Corporation · LLZ · Jan 17, 2008 · Radiology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K080013

## Device Facts

- **Applicant:** Agfa Healthcare Corporation
- **Product Code:** [LLZ](/productcode/LLZ.md)
- **Decision Date:** Jan 17, 2008
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 892.2050
- **Device Class:** Class 2
- **Review Panel:** Radiology
- **Attributes:** Software as a Medical Device, 3rd-Party Reviewed

## Indications for Use

Registration and Fusion (Basic and Extended) are intended for the simultaneous visualization of multiple medical images of the same patient. The application can assist the user in visually matching and comparing anatomical studies taken at different times or acquired via different imaging modalities (Basic: CT-CT, CT-MR, MR-MR, Extended: CT-PET, MR-PET) as well as assist in making measurements. It is primarily used in the fields of diagnostic radiology, neurology and oncology.

## Device Story

Registration and Fusion is a PACS plug-in software for image analysis. It inputs DICOM 3.0 compliant 3-D medical data sets from PET, MR, and CT imaging. The device performs rigid registration, computing an explicit mathematical transformation to align images geometrically. It produces fused outputs via semitransparent overlays or side-by-side displays. Used in radiology, oncology, and neurology departments by trained professionals (radiologists, referring physicians, radiation therapists, physicists, radiographers). The software runs on Agfa Impax PACS workstations. Clinicians use the output to navigate to identical anatomical locations across different studies, facilitating comparison of lesions and treatment follow-up. It also supports SUV and image intensity quantification.

## Clinical Evidence

Bench testing only. The device was tested for compatibility with Agfa Impax PACS systems. No clinical data provided.

## Technological Characteristics

PACS plug-in software; runs on Windows-based workstations (Impax 5.3/6.3). Processes DICOM 3.0 compliant 3-D image data (CT, MR, PET). Employs rigid registration algorithms and provides SUV/image intensity quantification. Connectivity via PACS network.

## Regulatory Identification

A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.

## Special Controls

*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).

## Predicate Devices

- Mirage 5.5 ([K043441](/device/K043441.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K080013
pg. 1 of 3

# 510(k) Summary

# Agfa Registration and Fusion

Common/Classification Name: Picture Archiving and Communications System (PACS), 21 CFR 892.2050 Proprietary Name: Registration and Fusion

Agfa HealthCare Corporation 10 South Academy Street Greenville, SC 29602-9048

Contact: Tom Holbrook, Prepared: November 9, 2007 Telephone: (519) 746-6210 ext. 3297 Facsimile: (519) 746-3745

JAN 17 2008

### A. LEGALLY MARKETED PREDICATE DEVICES

This is a 510(k) for Agfa's Registration and Fusion software. The predicate device, Mirage 5.5 (K043441), manufactured by Segami Corporation.

The new device is similar in indications and intended use as the predicate device.

#### DEVICE DESCRIPTION B.

Medical Data Registration and Fusion R 1.0 is a PACS plug-in (accessory). It is an image analysis software package that establishes the geometrical relationship between different DICOM 3.0 compliant 3-D medical data sets from PET, MR and CT imaging (registration").

The matched images are displayed (fusion') by either the use of semitransparent overlays or by displaying them side-by-side.

Registration and fusion facilitates the comparison of PET/CT, PET/MRI, CT/CT, MRI/MRI, CT/MRI image data sets for use in:

- The general radiology department for various lesions. ●
- The oncology department for various cancerous lesions. .

Definitions

Image Registration: The alignment of one or more [medical] images to a reference image in order to facilitate geometric comparison. This is a numerical operation that results in the computation of an explicit mathematical transformation between every point in the registered image sets.

mage Fusion: Registration forms the basis of image fusion in the sense that the geometrical alignment of images is a prerequisite. The notion of "fusion" takes this a step further by considering how to visualize the ttent of different images representing the same object [organ, anatomical region, etc.]. Such techniques include the use of overlays, semi-transparent renderings, etc.

{1}------------------------------------------------

K080013
pg-2 of 3

- The neurology department for various lesions. .
The rigid registration used in the application, aims to help the clinician navigate to the same anatomical location in both image sets.

The Medical Data Registration and Fusion software runs on Agfa's PACS workstations (Impax 5.3 and 6.3 workstations).

Typical users of this system are trained professionals, including but not limited to radiologists, referring physicians, radiation therapists, and physicists. Registration may also be used by radiographers for use of the images by radiologists. Registration and fusion has application in general radiology, oncology and neurology.

#### C. INTENDED USE

Registration and Fusion (Basic and Extended) are intended for the simultaneous visualization of multiple medical images of the same patient. The application can assist the user in visually matching and comparing anatomical studies taken at different times or acquired via different imaging modalities (Basic: CT-CT, CT-MR, MR-MR, Extended: CT-PET, MR-PET) as well as assist in making measurements. It is primarily used in the fields of diagnostic radiology, neurology and oncology.

#### SUBSTANTIAL EQUIVALENCE SUMMARY D.

Agfa's Registration and Fusion Software has similar indications and intended use as the legally marketed predicate devices.

The differences do not modify the intended diagnostic or therapeutic effect. It is intended for registering and fusing volumetric medical image data sets for the purpose of diagnosis and treatment follow-up.

Registration and Fusion is used primarily in the fields of general radiology department for various lesions, oncology for various cancerous lesions and neurology for various lesions.

Descriptive characteristics and data provided in this submission are sufficiently precise to assure substantial equivalence.

#### ய் TECHNOLOGICAL CHARACTERISTICS

The technological characteristics are the same in the proposed and predicate device. All devices use similar, commercially available, computers with Windows® operating systems. All operate on DICOM images. All provide rigid registration of CT, MR and PET image data sets

{2}------------------------------------------------

K080013
PG. 3 of 3

and SUV / Image intensity quantification.

## F. TESTING

Registration and Fusion has been tested for compatibility with Agfa's Impax® PACS Systems.

## G. CONCLUSIONS

This 510(k) has demonstrated Substantial Equivalence as defined and understood in the Federal Food Drug and Cosmetic Act and various guidance documents issued by the Center for Devices and Radiological Health.

{3}------------------------------------------------

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS). The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized representation of three wavy lines, which are meant to symbolize the department's mission of promoting health and well-being. The seal is simple and recognizable, and it is used on official documents and materials produced by the HHS.

### Public Health Service

JAN 17 2008

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Agfa HealthCare Corporation % Mr. Mark Job Responsible Third Party Official Regulatory Technology Services LLC 1394 25th Street NW BUFFALO MN 55313

Re: K080013

Trade/Device Name: Registration and Fusion Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: January 2, 2007 Received: January 3, 2007

### Dear Mr. Job:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely vours.

Nancy C. Brogdon

Nancy C. B rogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

{5}------------------------------------------------

## Indications for Use

510(k) Number (if known): Koloo13

Device Name: Registration and Fusion Indications for Use:

Registration and Fusion (Basic and Extended) is indicated for the simultaneous visualization of multiple medical images of the same patient. The application can assist the user in visually matching and comparing anatomical studies taken at different times or acquired via different imaging modalities (Basic: CT-CT, CT-MR, MR-MR, Extended: CT-PET, MR-PET) as well as assist in making measurements. It is primarily used in the fields of diagnostic radiology, neurology and oncology.

Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Arrih Whang
(Division Sign-Off)

Division of Reproductive, Abdominal, and Radiological Devices 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K080013](https://fda.innolitics.com/device/K080013)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
