K080004 · Sicel Technologies, Inc. · NZT · Jan 23, 2008 · Radiology
Device Facts
Record ID
K080004
Device Name
DOSE VERIFICATION SYSTEM PATIENT DOSIMETRY SYSTEM
Applicant
Sicel Technologies, Inc.
Product Code
NZT · Radiology
Decision Date
Jan 23, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 892.5050
Device Class
Class 2
Indications for Use
The DVS (Dose Verification System) is intended for use in radiation therapy to verify treatment planning and radiation dose to tissue and organs in or near the irradiated areas of a patient.
Device Story
DVS is an in vivo dosimetry system for radiotherapy. Components: implantable MOSFET dosimeter, insertion tool, reader system (wand/base station), and data system (software/database). Dosimeter implanted percutaneously; uses MOSFET sensing to measure radiation dose. Reader wand powers dosimeter via electromagnetic energy; dosimeter transmits threshold voltage readings to reader. Dosimeter is radioopaque, visible on CT scans for point-of-interest dose determination. Clinicians use system pre- and post-therapy fractions to calculate daily/cumulative fractional doses. Data system stores patient therapy info, calibration files, and dose history. Server-based software allows multi-user access and report generation. Benefits: provides real-time verification of prescribed dose to tumor/normal tissue, aiding treatment planning accuracy.
Clinical Evidence
No clinical data. Substantial equivalence established via bench testing, risk analysis, and verification/validation of performance specifications to ensure modified components perform as well as the predicate.
Technological Characteristics
Implantable MOSFET dosimeter; biocompatible glass capsule; potting epoxy; electromagnetic energy powering; radioopaque for CT visualization. System includes reader wand, base station, and server-based software. Sterilization method unchanged from predicate.
Indications for Use
Indicated for breast and prostate cancer patients undergoing photon beam radiation therapy to measure in vivo radiation dose at tumor periphery, tumor bed, and surrounding tissues for treatment planning validation.
Regulatory Classification
Identification
A medical charged-particle radiation therapy system is a device that produces by acceleration high energy charged particles (e.g., electrons and protons) intended for use in radiation therapy. This generic type of device may include signal analysis and display equipment, patient and equipment supports, treatment planning computer programs, component parts, and accessories.
Predicate Devices
Sicel Technologies, Inc. DVS, Dose Verification System (K052118)
Sicel Technologies, Inc. DVS, Dose Verification System (K061051)
Sicel Technologies, Inc. DVS, Dose Verification System (K071399)
Submission Summary (Full Text)
{1}------------------------------------------------
510(k) Summary
K080004
pg 1 of 2
Predicate Device(s):
Sicel Technologies, Inc. K052118. K061051. K071399 DVS, Dose Verification System
JAN 2 3 2008
Device Description:
The DVS, Dose Verification System consists of four sub-systems: the DVS Implantable Dosimeter for measuring radiation dose in vivo, the DVS Insertion Tool for implanting the dosimeter during percutaneous procedures, the DVS Reader System (Wand and Base Station) for powering the dosimeter and providing a user interface when taking dose measurements, and the DVS Data System (Plan and Review Software and Dosimetery Database) for storing and reporting patient data and for storing dosimeter information. The dosimeters use a MOSFET, Metal Oxide Semiconductor Field Effect Transistor, as a sensing mechanism. The dosimeter is factory calibrated and powered by the Reader Wand utilizing electromagnetic energy. The dosimeter contains a transmitter, to transmit threshold voltage readings to the reader. It is radioopaque and thus registers on computed tomography scans as a point of interest whereby a point dose may be determined. Patients are implanted prior to radiotherapy. Information on the patient's therapy, dose planning, point dose at the dosimeter, dosimeter serial number and calibration files are entered into the Plan and Review software and stored in the Dosimetry Database. At each therapy fraction the dosimeter is read pre- and post-therapy using the Reader Wand and Base Station. This translates into a daily fractional dose. The patient's daily and cumulative dose may be reviewed via the Plan and Review software. Because the Plan and Review software and Dosimetry Database are designed to be stored on a server, multiple users may be logged into the system at any one time. Reports on the patient's daily and cumulative dose history may be printed using the Plan and Review software.
Indication for Use:
The DVS (Dose Verification System) is intended for use in radiation therapy to verify treatment planning and radiation dose to tissue and organs in or near the irradiated areas of a patient.
The DVS system is specifically indicated for breast and prostate cancer to measure photon beam therapy and as an adjunct to treatment planning to permit measurement of the in vivo radiation dose received at the tumor periphery, tumor bed and/or surrounding normal tissues for validation of the prescribed dose.
{2}------------------------------------------------
K080004
Pg. 2 of 2
Comparison with Predicate Device:
The intended use of this SICEL DVS is identical to the predicate device, the DVS Dose Verification System in K052118, K061051, and K071399. The indications for use of the SICEL DVS also are the same.
The technological features of the SICEL DVS Dosimeter are the same as the predicate including the use of MOSFET technology, the calibration method, dose range, and energy sources measured. The dosimeter performance specifications are the same and have not changed. The principle of operation is the same as the predicate. The tissue contacting materials are the same and the sterilization method is the same. The insertion tool design, materials, and sterilization are the same. The Reader design and performance specifications are the same and have not changed. The Data System is the same and has not changed.
The primary difference between the predicate device and the modified device is the change to the dosimeter's potting epoxy used to secure and position the electronics within the biocompatible glass capsule. Other changes include minor dimensional changes, minor changes to the electronics components, and minor changes to the Reader software and accompanying Operator's Manual. Details of the substantial equivalence comparison are provided.
Furthermore, verification and validation testing based on the risk analysis, provided information sufficient to determine that the modifications did not have an effect on safety or efficacy and demonstrated that the device met pre-determined acceptance criteria based on performance specifications. The testing demonstrated that the modified device is substantially equivalent to the predicate device and performs as well as the predicate device. The verification and validation results are provided within the 510(k).
Thus, the DVS, Dose Verification System is substantially equivalent to the DVS, Dose Verification System (K052118, K061051, K071399).
{3}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
JAN 2 3 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Tammy B. Carrea Vice President, Regulatory Affairs Sicel Technologies, Inc. 3800 Gateway Centre Boulevard, Suite 308 MORRISVILLE NC 27560
Re: K080004
Trade/Device Name: Dose Verification System Patient Dosimetry System Regulation Number: 21 CFR 892.5050 Regulation Name: Medical charged-particle radiation therapy system Regulatory Class: II Product Code: NZT Dated: December 31, 2007 Received: January 2, 2008
Dear Ms. Carrea:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1796, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0126 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Nancy Brogdon
Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
510(k) Number (if known):
Device Name: DVS, Dose Verification System
Indications for Use:
Intended Use
The DVS (Dose Verification System) is intended for use in radiation therapy to verify treatment planning and radiation dose to tissue and organs in or near the irradiated areas of a patient.
Indications for Use
The DVS system is specifically indicated for breast and prostate cancer to measure photon beam therapy and as an adjunct to treatment planning to permit measurement of the in vivo radiation dose received at the tumor periphery, tumor bed and/or surrounding normal tissues for validation of the prescribed dose.
## (PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED)
| Concurrence of CDRH, Office of Device Evaluation (ODE) | |
|------------------------------------------------------------------|---------------------------------------------------------------|
| Prescription Use | <div style="display:inline-block; vertical-align:top;"></div> |
| (Per 21 C.F.R. 801.109) | OR |
| | Over-The-Counter Use ______ |
| | (Optional Format 1-2-96) |
| (Division Sign-Off) | |
| Division of Reproductive, Abdominal, and<br>Radiological Devices | |
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.