INTOUCH REMOTE PRESENCE ROBOTIC SYSTEM, MODEL RP-7
Applicant
Intouch Technologies
Product Code
DRG · Cardiovascular
Decision Date
Apr 11, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 870.2910
Device Class
Class 2
Indications for Use
The Remote Presence Robotic System is a communications tool that provides a means of transmitting, receiving, and storing real-time audio and video, and patient data. The Remote Presence Robotic also can be used in conjunction with 510(k)-cleared devices that transmit patient biometric data including vital signs information. The Remote Presence Robotic System transmits and receives information over high speed connection between patients and health professional, and is intended to be used in a hospital or clinic environment. Clinical judgment and experience are required to review and interpret the information transmitted.
Device Story
Mobile robotic telecommunications platform (RP-7) enabling real-time videoconferencing; consists of ControlStation (PC) and mobile robot. Inputs: real-time audio/video feeds, biometric data from integrated/external 510(k)-cleared devices (e.g., electronic stethoscope, vital signs monitors). Operation: physician at remote ControlStation uses joystick to navigate robot within healthcare facility; two-way audio/video allows interaction with patients and on-site staff. Output: real-time audiovisual link and patient data display. Benefit: enables remote physician presence, examination, and consultation. Context: hospital/clinic environment; operated by healthcare professionals. Redundant safety measures manage robotic mobility risks.
Clinical Evidence
Bench testing only. Validation testing performed to demonstrate the effectiveness of the videoconferencing link and the redundant safety measures associated with robotic mobility. Independent testing performed to verify the communication channel for the integrated electronic stethoscope.
Technological Characteristics
Mobile robotic platform; 802.11 wireless and broadband internet connectivity; integrated cameras, displays, microphones, and speakers. Includes optional integrated electronic stethoscope. System architecture facilitates data transfer from external 510(k)-cleared devices. No control or manipulation of external devices occurs through the system.
Indications for Use
Indicated for use in hospital or clinic environments to facilitate real-time audio/video communication and transmission of patient biometric data between patients and healthcare professionals. Requires clinical judgment for interpretation of transmitted information.
Regulatory Classification
Identification
A radiofrequency physiological signal transmitter and receiver is a device used to condition a physiological signal so that it can be transmitted via radiofrequency from one location to another, e.g., a central monitoring station. The received signal is reconditioned by the device into its original format so that it can be displayed.
Predicate Devices
BL Healthcare Remote Care Management System (K051470)
CareStation 126S Videophone Motion Media Technology, Inc. (K031863)
Submission Summary (Full Text)
{0}------------------------------------------------
073710
P1/2
510(k) Submission
Remote Presence Robotic System
APR 1 1 2008
## ATTACHMENT 2 510(k) SUMMARY
## InTouch Health® 510(k) Owner: 90 Castilian Drive, Suite 200 Goleta, CA 93117
| Contact: | Bill Stout |
|----------|--------------------------|
| | Phone: 805-862-8686 x112 |
| | Fax: 805-562-8663 |
Date Summary 12/28/07 Prepared:
Device:
Remote Presence Robotic System
Physiological signal transmitter and receiver (21 C.F.R. § 870.2910, Product Code DRG); Class II electronic stethoscope (21 C.F.R. § 870.1875. Product Code DQD)
Classification:
Trade Name:
Class II
Predicate Devices:
BL Healthcare Remote Care Management System BL Healthcare, Inc. K051470
Common/Classification Name:
Telecare D.R. HealthCare Vision, Inc. K033133
CareStation 126S Videophone Motion Media Technology, Inc. K031863
Device Description: The Remote Presence Robotic System is a mobile, robotic telecommunications platform that enables real-time videoconferencing. The Remote Presence Robotic System consists of a ControlStation (a desktop or laptop computer) and the RP-7 mobile robot, which is most often linked via broadband Internet and an 802.11 wireless network. The Robot and ControlStation are each equipped with cameras, displays, microphones and speakers, allowing two-way audio-video communication. Additionally, the ControlStation is equipped with a joystick that the operator uses to control the movement of the Robot in the remote location. As a result, the user can move through a remote facility and engage in conversations as if physically present while seated at a ControlStation in a remote facility with broadband Internet access.
{1}------------------------------------------------
As an option, the Remote Presence Robotic System includes an integrated electronic stethoscope. The stethoscope is a FDA-cleared device which is used for the same purpose for which it received 510(k) clearance, i.e., to transmit auscultation sounds from a patient at one location to a clinic at a different location. The System provides a channel for the digital data that the stethoscope transmits.
The Remote Presence Robotic System is a communications tool that provides a Intended Use: means of transmitting, receiving, and storing real-time audio and video, and patient data. The Remote Presence Robotic also can be used in conjunction with 510(k)-cleared devices that transmit patient biometric data including vital signs information. The Remote Presence Robotic System transmits and receives information over high speed connection between patients and health professional, and is intended to be used in a hospital or clinic environment. Clinical judgment and experience are required to review and interpret the information transmitted.
Like the predicate devices, the Remote Presence Robotic System provides a Technological real-time link between the patient and the healthcare professional. This link Characteristics: occurs over broadband Internet and includes real-time audio and video to facilitate communication between the patient, patient-side healthcare professionals, and remote healthcare professionals. Also like the predicate devices, the Remote Presence Robotic System transfers data from 510(k)cleared devices between the patient and the healthcare professional. Like the predicate devices, the devices are not controlled or manipulated through the Remote Presence Robotic System, and consequently, no additional risk is presented.
> Unlike the predicate devices, the Remote Presence Robotic System provides mobility to the physician by allowing the physician to control the movement of the mobile robotic platform within the healthcare facility. The real-time audio video link allows the driver to see and hear effectively within the remote environment.
Redundant safeguards are designed into the Remote Presence Robotic System Performance to address risks associated with the mobility of the robotic platform. The Data: effectiveness of these measures, and the videoconferencing link were demonstrated by the validation testing performed on the system. The communication channel used by the electronic stethoscope was also proven effective by independent tests.
The performance data discussed in this 510(k) application demonstrate that the Conclusions: Remote Presence Robotic System is as safe and effective, and performs as well as or better than the predicate devices.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle with its wings spread, facing to the right. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
APR 1 1 2008
InTouch Technologies Inc. c/o Mr. Bill Stout Director of Operations 90 Castilian Drive, Suite 200 Goleta, CA 93117
Re: K073710
Trade Name: Remote Presence Robotic System, Model RP-7 Regulation Number: 21 CFR 870.2910 Regulation Name: Transmitters And Receivers, Physiological Signal, Radiofrequency Regulatory Class: Class II (two) Product Codes: DRG Date: February 28, 2008 Received: February 29, 2008
Dear Mr. Stout:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
Page 2 -- Mr. Bill Stout
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807): labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Blymmmmmow for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
510(k) Submission Remote Presence Robotic System
## ATTACHMENT 1 Indications for Use
510(k) Number Not assigned. (if known):
Device Name:
Remote Presence Robotic System
Indications for Use: The Remote Presence Robotic System is a communications tool that provides a means of transmitting, receiving, and storing real-time audio and video, and patient data. The Remote Presence Robotic also can be used in conjunction with 510(k)-cleared devices that transmit patient biometric data including vital signs information. The Remote Presence Robotic System transmits and receives information over high speed connection between patients and health professional, and is intended to be used in a hospital or clinic environment. Clinical judgment and experience are required to review and interpret the information transmitted.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
PLEASE DO NOT WRITE BELOW THIS LINE -- CONTINUE ON ANOTHER PAGE IF NEEDED
Concurrence of CDRH, Office of Device Evaluation (ODE)
B.Bimmerman
(Division Sign-O
Division of Cardiovascular Devices
510(k) Number K0233710
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.