K073704 · J. Morita USA, Inc. · LLZ · Feb 21, 2008 · Radiology
Device Facts
Record ID
K073704
Device Name
I-DIXEL
Applicant
J. Morita USA, Inc.
Product Code
LLZ · Radiology
Decision Date
Feb 21, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.2050
Device Class
Class 2
Attributes
Software as a Medical Device
Indications for Use
i-Dixel is a software for management of images obtained from x-ray imaging equipment and other imaging modalities. i-Dixel is used for medical and dental image examination and diagnosis. The software is to be operated and used by doctor, dentists and other legally qualified professionals.
Device Story
i-Dixel is a medical/dental diagnostic application for processing, archiving, and managing patient images and records; operates in network environments. Inputs include 2D/3D X-ray images from J. Morita modalities, TWAIN-compliant devices, and video camera imports. System features include database management, image viewing, printing, and import/export functions. Operated by doctors, dentists, and qualified professionals in medical/dental settings. Includes password-protected operator authentication to prevent data loss or operational errors. Output consists of displayed or printed diagnostic images used by clinicians to support diagnostic decision-making.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Software-based PACS/image viewer. Operates on networked systems. Supports TWAIN-compliant modalities and J. Morita proprietary imaging hardware. Includes database management, password-based user authentication, and image import/export capabilities. No specific hardware materials or sterilization required as it is a software-only device.
Indications for Use
Indicated for medical and dental professionals to manage, examine, and diagnose images obtained from X-ray and other imaging modalities.
Regulatory Classification
Identification
A medical image management and processing system is a device that provides one or more capabilities relating to the review and digital processing of medical images for the purposes of interpretation by a trained practitioner of disease detection, diagnosis, or patient management. The software components may provide advanced or complex image processing functions for image manipulation, enhancement, or quantification that are intended for use in the interpretation and analysis of medical images. Advanced image manipulation functions may include image segmentation, multimodality image registration, or 3D visualization. Complex quantitative functions may include semi-automated measurements or time-series measurements.
Special Controls
*Classification.* Class II (special controls; voluntary standards—Digital Imaging and Communications in Medicine (DICOM) Std., Joint Photographic Experts Group (JPEG) Std., Society of Motion Picture and Television Engineers (SMPTE) Test Pattern).
{1}------------------------------------------------
# 073704
### 510(k) SUMMARY J. Morita USA Inc.'s i-Dixel
11:0 Z 1 2
# 1. Submitter Name and Address with Phone/Fax
Registration No. 2081055 Registration No. 3002807636 Initial Distributor: Manufacturer: J. Morita USA, Inc. J. MORITA MFG. CORP. 9 Mason 680 Higashihama Minami-cho Irvine, CA 92618 Fushimi-ku, Kyoto USA Japan 612-8533 Telephone: 949-581-9600 +81-75-611-2141 Facsimile: 949-581-9688 +81-75-605-2354
#### 2. Contact Person
Keith A. Barritt Fish & Richardson P.C. 1425 K Street, N.W. Suite 1100 Washington, DC 20005 Phone: (202) 783-5070 Facsimile: (202) 783-2331
#### 3. Date Summary Prepared
November 12, 2007
#### 4. Device Name
Trade or Proprietary Name:
i-Dixel
Common Name:
Product Code:
Image viewer software
Classification Name:
Picture archiving and communications system (21CFR892.2050) LLZ
#### 5. Substantial Equivalency
Substantial equivalency is claimed against the following devices:
{2}------------------------------------------------
i-Dixel is practically the same as 3DX Integrated Information System, which is incorporated in our legally marketed device, 3D Accu-I-tomo X YZ Slice View Tomograph (K030450). i-Dixel is also presumably the same as N-Liten which is incorporated to ProMax 3D (Planmeca Ov, Finland).
#### 6. Description of the Device
i-Dixel is a medical and dental diagnostic application for processing, archiving, and managing diagnostic images from multiple imaging modalities and other patient information. i-Dixel processes acquired 2D/3D images for diagnostic use. Any patient record, name, ID, diagnosis, or image can be retrieved by simply clicking the mouse on the desired item in the easy-to-manage filing system made up of patient folders and their images.
i-Dixel is used in the medical environment and has the following functions as the entire system.
- 1) An image is acquired from X-ray equipment and saved in the database.
- 2) It saves data in the patient database besides the image.
- 3) Information in the image etc. saved in the patient data base can be seen,
- 4) It prints the image with applying information.
- 5) It makes maintenance of the patient database.
- 6) The image is imported and exported by various forms.
- 7) It is operated under the network environment.
- 8) It has the function of the input authentication by the password with operator ID to prevent patient's fault and data being lost by the operational mistake.
- 9) It is operated with 3D imaging modalities manufactured by J. Morita Mfg. Co.
- 10) It is operated with 2D imaging modalities or other panoramic/cephalometric x-ray device manufactured by J. Morita.
- 11) It can be connected with modalities which comply with TWAIN specifications.
- 12) It can incorporate images of device which is compatible to drivers including DixelD or 3DXD by J. Morita Mfg. Co.
- 13) It can import data from video camera.
#### 7. Intended Use
i-Dixel can be used as viewer or database of medical imaging to support the diagnostic by medical imaging.
#### 8. Safety and Effectiveness of the Device
i-Dixel is slightly revised version of 3DX Integrated Information System which is incorporated to FDA- cleared device (See below) in J. MORITA MFG. CORP.
{3}------------------------------------------------
# Information of the device which 3DX Integrated Information System is incorporated:
ਾ
| No. | Device | 510(k) number | Functions of the device | NOTE |
|-----|-----------------|---------------|-------------------------------------|------|
| 1 | 3D accu-i- tomo | K030450 | Dental x-ray computed<br>tomography | 3D |
The i-Dixel is a slightly modified model from 3DX Integrated Information System but all changes are minor and don't pose any additional safety concerns. Safety charges are min has not changed between these two models and we judge these two software ane substantially equal.
Consequently, it is self-explanatorily evident that this i-Dixel is substantially equivalent to 3DX Integrated Information System which is incorporated to 3D Accu-I-tomo XYZ Slice View Tomograph (K030450) in the sense of similar characteristics.
3/3
{4}------------------------------------------------
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular arrangement of text that reads "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA". Inside the circle is a stylized symbol that resembles an eagle or bird with three curved lines representing its wings or feathers. The symbol is black and the text is also in black.
FEB 21 2008 Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850
J. Morita USA, Inc. % Mr. Keith A. Barritt Attorney Fish & Richardson P.C. 1425 K Street, N.W., Suite 1100 WASHINGTON DC 20005
Re: K073704
Trade/Device Name: i-Dixel Image Viewer Software Regulation Number: 21 CFR 892.2050 Regulation Name: Picture archiving and communications system Regulatory Class: II Product Code: LLZ Dated: December 28, 2007 Received: December 31, 2007
Dear Mr. Barritt:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Image /page/3/Picture/10 description: The image is a circular logo commemorating the centennial of the FDA from 1906-2006. The letters FDA are in bold, stacked on top of the word "Centennial". The logo is surrounded by a circular border with text, but the text is not legible.
Protecting and Promoting Public Health.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html.
Sincerely yours,
Nancy C. Brogdon
Nancy C. Brogdon . Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
K673704
510(K) Number: anknown Device Name: i-Dixel Indications for Use:
i-Dixel is a software for management of images obtained from x-ray imaging equipment and other imaging modalities. i-Dixel is used for medical and dental image examination and diagnosis.
The software is to be operated and used by doctor, dentists and other legally qualified professionals.
Prescription Use AND/OR (Part21CFR801 Subpart D)
Over-The-Counter Use (Part21CFR807 Subpart C).
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED Concurrence of CDRH, Office of Device Evaluation(ODB)
(Division Sign-Off).
Division of Anesthesiology, General Hospital, Infection Control, Dental Devices
510(K) Number: L078704
Prescription Use (Part21CFR801.109) Over-The-Counter Use (Optional Format 1-2-96)
(Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.