SWIFT LT

K073638 · Resmed, Ltd. · BZD · Mar 20, 2008 · Anesthesiology

Device Facts

Record IDK073638
Device NameSWIFT LT
ApplicantResmed, Ltd.
Product CodeBZD · Anesthesiology
Decision DateMar 20, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.5905
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Swift LT channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system. The Swift LT is: to be used by adult patients (> 66 lb/30 kg) for whom positive airway pressure has been prescribed. intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment.

Device Story

Swift LT is a vented nasal mask interface for positive airway pressure therapy (CPAP/bilevel). Device channels airflow from external pressure source to patient nose via silicone interface; held in place by adjustable fabric/nylon headgear. Incorporates vent holes to flush CO2. Connects to standard air delivery hoses via ISO 5356-1:2004 conical connectors. Used in home or hospital settings by patients; prescribed by clinicians. Output is delivery of pressurized air to patient airway; facilitates therapy compliance and CO2 clearance. Design differences from predicates involve component geometry and interface configuration.

Clinical Evidence

Bench testing only. No clinical data provided. Safety and efficacy supported by established performance of vented nasal masks with CPAP/bilevel therapy and comparative bench testing against predicate devices regarding pressure-flow, impedance, and CO2 performance.

Technological Characteristics

Materials: molded plastic, silicone, fabric/nylon headgear (ISO 10993-1 compliant). Sensing/Actuation: passive airflow delivery via vent holes for CO2 flushing. Connectivity: standard conical connectors (ISO 5356-1:2004). Form factor: nasal mask with adjustable headgear. Sterilization: supplied nonsterile. No software or electronic components.

Indications for Use

Indicated for adult patients (> 66 lb/30 kg) prescribed positive airway pressure therapy via CPAP or bilevel systems. No specific contraindications listed.

Regulatory Classification

Identification

A noncontinuous ventilator (intermittent positive pressure breathing-IPPB) is a device intended to deliver intermittently an aerosol to a patient's lungs or to assist a patient's breathing.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 073638 ## MAR 2 0 2008 # RESMED . Swift LT Traditional 510k | 510(k) SUMMARY<br>[As required by 21 CFR 807.92(c)] | | | |-----------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----| | Date Prepared | December 18th, 2007 | | | Official Contact | Mr. David Thomson,<br>Senior Regulatory Affairs Manager | | | Device Trade Name | Swift™ LT | | | Device Common Name/<br>Classification Name | Vented Nasal Mask;<br>Accessory to Noncontinuous Ventilator (IPPB) | | | Classification | 21 CFR 868.5905, 73 BZD (Class II) | | | Predicate Devices | Mirage Swift (K042403)<br>Ultra Mirage II Mask (K050359) | | | Description | The Swift LT provides an interface such that airflow from a<br>positive pressure source is directed to the patient's nose. The<br>mask is held in place with adjustable headgear that straps the<br>mask to the face. | | | | Swift LT is safe when used under the conditions and purposes<br>intended as indicated in the labeling provided with the product. | | | | Swift LT is a prescription device supplied nonsterile. | | | Intended Use | The Swift LT channels airflow noninvasively to a patient from a<br>positive airway pressure device such as a continuous positive<br>airway pressure (CPAP) or bilevel system.<br>The Swift LT is:<br>to be used by adult patients (> 66 lb/30 kg) for whom positive airway pressure has been prescribed. intended for single-patient re-use in the home environment and multipatient re-use in the hospital/institutional environment. | | | Technological<br>Characteristics<br>comparison | Comparison with predicate Mirage Swift<br>The new device and the predicate mask, provide a seal via silicone interface. Both masks are offered in various sizes to ensure adequate fit over the extended patient population. | | | | Both the masks incorporate vent holes to provide continuous air leak to flush out and minimize the amount of CO2 rebreathed by the patient. The design of the mask components is such that the incorporation of these vent-holes does not interfere with the intended performance of the masks. | | | | Both the masks connect to a conventional air delivery hose between the mask and the positive airway-pressure source via standard conical connectors (ref: ISO 5356-1:2004) | | | | Both the masks are constructed using molded plastic and silicone components and fabric / nylon headgear. All the | | | December 18, 2007 | | 17 | | | components of both the masks are fabricated using materials<br>deemed safe. (ref: ISO 10993-1). | | | | Both the new device and the predicate are designed to operate<br>on the same Mirage or Swift ResMed flow generator settings.<br>The pressure-flow characteristics and flow impedance of both<br>the new device and the predicate device are substantially<br>equivalent. | | | | Both the new device and the predicate device can be reused in<br>the home and hospital / institution environment. | | | | The main differences between Swift LT and Mirage Swift is in<br>the number of components, their design/geometry and how<br>individual components interface with each other. Both the<br>masks are designed and constructed under ResMed's 21 CFR<br>Part 820 compliant Quality Management System. | | | Clinical Data | Use of vented nasal masks with CPAP or Bilevel therapy is<br>proven technology and is well accepted by the medical<br>community. Bench testing is sufficient to demonstrate safety<br>and efficacy of the Swift LT, as was the case with the predicate<br>device. | | | Performance Data | Comparison with predicate Ultra Mirage II<br>The CO2 performance of the new device and the predicate<br>device are substantially equivalent. | | | Substantial Equivalence<br>Conclusion | Swift LT is substantially equivalent to the predicate devices:<br>- it has the same though reworded intended use;<br>- it has similar technological characteristics to both<br>predicates:<br>- it does not raise new questions of safety and<br>effectiveness;<br>- it is at least as safe and effective as the predicate<br>devices Mirage Swift and Ultra Mirage II | | {1}------------------------------------------------ ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------ : {2}------------------------------------------------ Image /page/2/Picture/2 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States of America. The seal features a stylized eagle with its wings spread, symbolizing protection and service. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. ### MAR 2 @ 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ResMed Limited C/O Mr. David D' Cruz Vice President, Clinical & Regulatory Affairs ResMed Corporation 14040 Danielson Street Poway, California 92064-6857 Re: K073638 Trade/Device Name: Swift™ LT Regulation Number: 21 CFR 868.5905 Regulation Name: Noncontinuous Ventilator (IPPB) Regulatory Class: II Product Code: BZD Dated: March 6, 2008 Received: March 10, 2008 Dear Mr. D' Cruz: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Mr. D' Cruz Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Syutte Y. Michie Dats Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### RESMED Swift LT Traditional 510k #### Indication for Use 510(k) Number (if known): Swift™ LT Device Name: Indication for Use The Swift LT channels airflow noninvasively to a patient from a positive airway pressure device such as a continuous positive airway pressure (CPAP) or bilevel system. The Swift LT is: - to be used by adult patients (> 66 lb/30 kg) for whom positive airway pressure has been prescribed. . - intended for single-patient re-use in the home environment and multipatient re-use in the . hospital/institutional environment. Min Thul (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 07363 510(k) Number: J Prescription Use X AND/OR Over-The-Counter Use_ (Part 21 CFR 801 Subpart D) (Part 21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH; Office of Device Evaluation (ODE) Page 1 of__1_ December 18, 2007 15
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...