The BPCard is a Non-Invasive Blood Pressure Monitor that measures blood pressure (systolic and diastolic) using the auscultatory method in both male and female, adult subjects. The BPCard also measures heart rate. Patient Population: Male/Female, Adults . Environment of Use: Hospital, Clinic and Home Use . Prescription Device by a Physician .
Device Story
BPCard is a non-invasive blood pressure monitor using auscultatory method; measures systolic/diastolic pressure and heart rate. System comprises PCMCIA card, microphone assembly, blood pressure cuff, and software. Used in hospital, clinic, or home settings by physician-prescribed patients. Device integrates with QRS Office Medic software on Windows PCs. Input consists of acoustic signals from cuff; processed via microcontroller firmware (v1.05) and software algorithms including automatic microphone gain control, dynamic pulse detection, and pulse proximity evaluation. Output displayed on PC screen; provides clinical data for provider decision-making. Benefits include automated, accurate blood pressure monitoring and digital record-keeping. Hardware updates include bayonet-style air connector, Lee deflation valve, and improved EMC protection.
Clinical Evidence
Bench testing only. Compliance verified against IEC/EN 60601-1 (safety), IEC/EN/UL 60601-1-2 (EMC), ISO 10993-1 (biocompatibility), and ANSI/AAMI SP10:2002 (sphygmomanometer performance). No clinical trial data presented.
Indicated for non-invasive measurement of systolic/diastolic blood pressure and heart rate in adult male/female patients. Prescription use only. Contraindications: None stated.
Regulatory Classification
Identification
A noninvasive blood pressure measurement system is a device that provides a signal from which systolic, diastolic, mean, or any combination of the three pressures can be derived through the use of tranducers placed on the surface of the body.
{1}------------------------------------------------
# 510(k) Summary of Safety and Effectiveness
Date:12/14/07
K073630
#### Submitter:
QRS Diagnostic, LLC Street Address: 14755 27th Ave N. City: Plymouth State: MN Zip Code: 55447 Telephone: 763-559-8492 Facsimile: 763-559-2961
JAN 1 1 2008
Contact: Mary Kay Jensen Phone: 763-559-8492 Ext 958 Facsimile: 763-559-2961 e-mail: mkjensen@grsdiagnostic.com
Device Name: Trade Name: BPCard Common Name: Blood Pressure Monitor Classification Name: System, Measurement, Blood-Pressure, Non-invasive Classification: Class II Panel Code: DXN Regulation Number: 21CFR. 870.1130 Noninvasive blood pressure measurement system.
### Identification of Legally Marketed (Unmodified) Device (Predicate Device):
| Name of<br>Predicate | Manufacturer | Use | 510K) | Date<br>Cleared |
|----------------------|---------------------------|---------------------------|---------|-----------------|
| BPCard™ | QRS<br>Diagnostic,<br>LLC | Blood Pressure<br>Monitor | K031964 | 9/5/03 |
### Device Description:
### Indications for Use
Blood Pressure Monitor that measures blood pressure (systolic and diastolic) using the ausculatory method. The BPCard also measures heart rate.
- . Patient Population: Male/Female Adults
- . Environment of Use: Hospital, Clinic and Home Use
- . Prescription Device by a Physician
{2}------------------------------------------------
### Technological Comparison to (Unmodified) Predicate Device:
The following summary table of comparisons compares the Modified BPCard Device to the Previously Cleared BPCard (Predicate) Device.
| # | Area | Modified Device | Previously Cleared Device | Same | Different |
|-----------------------------------|-------------------------------------------------------|-------------------------------------------------------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------|------|-----------|
| Indications for Use | | | | | |
| 1 | Patient Population | Male/Female<br>Adult | Male/Female<br>Adult | X | |
| 2 | Environment of Use | Hospital, Clinic, Home Use | Hospital, Clinic, Home Use | X | |
| Fundamental Scientific Technology | | | | | |
| 3 | Measurements taken by Device | • Systolic Pressure<br>• Diastolic Pressure<br>• Pulse Rate | • Systolic Pressure<br>• Diastolic Pressure<br>• Pulse Rate | X | |
| 4 | Measurement Method | Auscultatory | Auscultatory | X | |
| 5 | Device Components | • BPCard (PCMCIA)<br>• Microphone Assembly<br>• Blood Pressure Cuff<br>• Software | • BPCard (PCMCIA)<br>• Microphone Assembly<br>• Blood Pressure Cuff<br>• Software | X | |
| Contraindications | | | | | |
| 6 | Contraindications | No contraindications | Contraindicated in patients who are not ambulatory, have heart failure, arrhythmia or cardiac valve abnormality. | | X |
| Sterility/Expiration Dating | | | | | |
| 7 | Sterile/Non Sterile | Non-Sterile/No Expiration Date | Non-Sterile/No Expiration Date | X | |
| Energy Type | | | | | |
| 8 | AC/DC | Both | Both | X | |
| Environmental Specifications | | | | | |
| 9 | Temperature/<br>Humidity/EMC<br>Requirements | Same | Same | X | |
| Performance Standards | | | | | |
| 10 | AAMI SP10 | Meets AAMI | Meets AAMI | | X |
| # | Area | Modified Device | Previously Cleared<br>Device | Same | Different |
| | | SP10:2002 | SP10:1992 | | |
| | Device Hardware | | | | |
| 11 | Dimensional and<br>Physical<br>Specifications | Same | Same | X | |
| 12 | PCB Board | Same | Same | X | |
| 13 | Air Coupling | Bayonet Style<br>Connector | Luer Slip | | X |
| 14 | Deflation Valve | Lee Valve | Pneutronic Valve | | X |
| 15 | EMC Protection | Added measures for<br>EMC protection to<br>meet IEC/EN<br>60601-1-2:2001 as a<br>result of other<br>changes | | | X |
| 16 | Microphone Gain<br>(Sensitivity) | X40 | X10 | | X |
| 17 | Microcontroller<br>Firmware | V 1.05 | V 0.03 | | X |
| | Materials/Components | | | | |
| 18 | Blood Pressure<br>Cuff | Trimline Kuff Link.<br>Pre-amendment<br>device. | Pymah Corporation<br>K884421 | | X |
| 19 | Microphone<br>Material | Medical Grade ABS<br>Plastic | ABS Plastic | | X |
| 20 | Microphone Cable<br>Assembly | Dual Lumen Tubing<br>enclosing<br>microphone cable | No tubing enclosing<br>microphone cable | | X |
| | Algorithms | | | | |
| 21 | Automatic<br>Microphone Gain<br>Control | Available | Not available | | X |
| 22 | Level Based Pulse<br>Detection | Dynamic | Static | | X |
| 23 | Pulse Proximity<br>Evaluation | Available | Not available | | X |
| 24 | Pulse Rate<br>Calculation | Calculated from<br>median pulse<br>intervals. | Calculated from<br>minimum pulse<br>intervals. | | X |
| | Software | | | | |
| 25 | Print Test Feature | Available | Not Available | | X |
| 26 | Save and Review<br>Tests Feature | Available | Not Available | | X |
| 27 | Accept/Reject or | Available | Not Available | | X |
| # | Area | Modified Device | Previously Cleared<br>Device | Same | Different |
| | | | Device | | |
| | Edit Test Feature | | | | |
| 28 | Operating System | Windows | Windows | X | |
| 29 | Software Platform | Integrated with QRS<br>Office Medic | Stand alone software | | X |
| 30 | Deflation Rate | 2.5 mmHg and 5.0<br>mmHg | 2.5 mmHg | | X |
| 31 | Deflation Start | Starts when pressure<br>and pulse detection<br>criteria met | Starts when start<br>button pushed | | X |
| 32 | Maximum Pressure<br>Exceeded | Test will stop and<br>cuff will deflate | Software provides<br>warning | | X |
| 33 | Zero Function and<br>Open and Close<br>Valve function | Performed<br>Automatically at<br>start and end of test | User selects Zeroing<br>Button and End Test<br>Button | | X |
| 34 | Calibration | User can check<br>pressure calibration | User can check<br>pressure calibration | X | |
| 35 | Pulse Detection<br>Beep and Display | System does not<br>provide audible or<br>display indication<br>when pulse is<br>detected | System provides<br>audible indication<br>when pulse is<br>detected and displays<br>number of pulses | | X |
| 36 | Raw Data<br>Processing | Raw Data file can<br>only be reprocessed<br>with QRS<br>development<br>software. | Raw Data File can be<br>reprocessed | X | |
| 37 | Pressure Display | Pressure displayed<br>after test has begun | Pressure<br>continuously<br>displayed | | X |
| 38 | Start of Test | Test begins after<br>auto-zeroing<br>completed and user<br>begins manual<br>inflation | Test begins after user<br>completed manual<br>inflation | | X |
| 39 | Battery Test | OS/Computer<br>Battery warning | BPCard software<br>provides low battery<br>warning | | X |
{3}------------------------------------------------
{4}------------------------------------------------
{5}------------------------------------------------
### Summary of Performance Testing:
The modified BPCard has been tested or found otherwise to comply with the following standards:
- IEC/EN 60601-1, Medical electrical equipment Part 1: General requirements . for basic safety and essential performance
- IEC/EN/UL 60601–1-2, Medical Electrical Equipment Part 1-2: General . requirements for safety - collateral standard: Electromagnetic compatibility -Requirements and Test
- ISO 10993-1, Biological Evaluation Of Medical Devices Part 1: Evaluation . And Testing
- . ANSI/AAMI SP10:2002, Manual, Electronic or Automated Sphygmomanometers
- EN1060-1 "Non-Invasive Sphygmomanometers General Requirements" .
- EN1060-3 "Non-Invasive Sphygmomanometers Supplementary . Requirements"
### Conclusions:
The results of the tests discussed above, indicate that the modified QRS Diagnostic BPCard is as safe, as effective, and performs as well as or better than the non modified device.
{6}------------------------------------------------
Public Health Service
Image /page/5/Picture/2 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and is smaller than the caduceus symbol.
JAN 1 1 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
QRS Diagnostic, LLC c/o Ms. Mary Kay Jensen QA/RA Manager 14755 27th Ave. N. Plymouth, MN 55447
Re: K073630
BPCard, Model Z-7000-0700 Regulation Number: 21 CFR 870.1130 Regulation Name: Noninvasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Code: DXN Dated: December 14, 2007 Received: December 26, 2007
Dear Ms. Jensen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{7}------------------------------------------------
Page 2 -- Ms. Mary Kay Jensen
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements nft he Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/odrh/industry/support/index.html.
Sincerely yours.
B. Bemmer for
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): To be determined
K073630
Device Name: BPCard
Indications for Use:
The BPCard is a Non-Invasive Blood Pressure Monitor that measures blood pressure (systolic and diastolic) using the auscultatory method in both male and female, adult subjects. The BPCard also measures heart rate.
- Patient Population: Male/Female, Adults .
- Environment of Use: Hospital, Clinic and Home Use .
- Prescription Device by a Physician .
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
## Concurrence of CDRH, Office of Device Evaluation (ODE)
Bhumima
Page 1 of 1
(Division Sign-Off)
Division of Cardiovascular Devices
510(k) Number K073630
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.