ATLANTIS 018 PERIPHERAL IMAGING CATHETER

K073623 · Boston Scientific Corp · ITX · Mar 13, 2008 · Radiology

Device Facts

Record IDK073623
Device NameATLANTIS 018 PERIPHERAL IMAGING CATHETER
ApplicantBoston Scientific Corp
Product CodeITX · Radiology
Decision DateMar 13, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1570
Device ClassClass 2

Indications for Use

The Atlantis™ 018 Peripheral Imaging Catheter is intended for intravascular ultrasound examination of peripheral vessels only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures.

Device Story

Sterile, single-use 40MHz ultrasonic imaging catheter; designed for intravascular ultrasound (IVUS) examination of peripheral vessels. Device comprises imaging core with rotating drive cable and distal transducer, plus catheter body with telescoping section for 15 cm linear movement. Connects to Boston Scientific IVUS System Motor Drive Unit (MDU) via electro-mechanical interface. Operated by clinicians during interventional procedures to visualize vessel morphology. Output displayed on IVUS system monitor; assists clinicians in assessing vessel status and guiding interventional decisions. Benefits include real-time intraluminal imaging to support procedural success.

Clinical Evidence

No human clinical trials. Evidence includes bench testing (dimensional/functional verification), acoustic output testing (within FDA Track 1 limits), biocompatibility testing (ISO 10993), and animal testing (in-vivo deliverability and imaging characteristics). Results demonstrate device meets performance specifications and is clinically acceptable for intended use.

Technological Characteristics

40MHz radial ultrasonic transducer; catheter body with distal, proximal, and telescoping sections; hi-torque flexible rotating drive cable. Connects to external MDU. Biocompatibility per ISO 10993. Acoustic output compliant with FDA 1997 guidance. Single-patient use.

Indications for Use

Indicated for patients who are candidates for transluminal interventional procedures requiring intravascular ultrasound examination of peripheral vessels.

Regulatory Classification

Identification

A diagnostic ultrasonic transducer is a device made of a piezoelectric material that converts electrical signals into acoustic signals and acoustic signals into electrical signals and intended for use in diagnostic ultrasonic medical devices. Accessories of this generic type of device may include transmission media for acoustically coupling the transducer to the body surface, such as acoustic gel, paste, or a flexible fluid container.

Predicate Devices

Reference Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # 510(k) Summary of Safety and Effectiveness ## Boston Scientific Corporation # Atlantis™ 018 Peripheral Imaging Catheter | Submitted By | Boston Scientific Corporation | |--------------|-------------------------------| | | 47900 Bayside Parkway | | | Fremont, CA 94538 | | Contact Person | Emilly Tojima<br>Regulatory Affairs Specialist II<br>Tel: (510) 624-2546<br>Fax: (510) 624-2569<br>emilly.tojima@bsci.com | |----------------|---------------------------------------------------------------------------------------------------------------------------| |----------------|---------------------------------------------------------------------------------------------------------------------------| | Date Prepared | December 21, 2007 | |------------------------|------------------------------------------------------------------------------------------------------------------------------| | Proprietary Name | Atlantis™ 018 Peripheral Imaging Catheter | | Common Name(s) | Ultrasonic Diagnostic Imaging Catheter<br>Ultrasonic Diagnostic Transducers (ITX)<br>Intravascular Ultrasound Catheter (OBJ) | | Classification Name(s) | 21 CFR Part 892. 1570 (ITX, OBJ)<br>Diagnostic Ultrasonic Transducer | {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows a fraction with "52736.23" as the numerator and "13,265" as the denominator. The numerator is written above a horizontal line, and the denominator is written below the line. The numbers appear to be handwritten. The image is in black and white. ### Predicate Devices The Atlantis™ 018 Peripheral Imaging Catheter is substantially equivalent to the following devices: | Product | 510(k) | Clearance Date | |-----------------------------------------------|---------|-------------------| | Atlantis™ SR Pro Coronary Imaging<br>Catheter | K063312 | November 30, 2006 | | Sonicath Ultra™ Imaging Catheter<br>(3.2F) | K060947 | April 19, 2006 | #### Description of the Device The Atlantis™ 018 Peripheral Imaging Catheter is a sterile, single patient use 40MHz transducer imaging catheter, intended for intravascular ultrasound examination of peripheral vessels. This imaging catheter is designed to be operated with Boston Scientific Corporation (BSC) Intravascular Ultrasound (IVUS) Systems: iLab™ (K072517), Galaxy2/Galaxy (K980851), and Clearview™M Ultra (K921750). The Atlantis™ 018 Peripheral Imaging Catheter consists of two main assemblies: - . Imaging core - . Catheter body The catheter body comprises three sections: - . Distal lumen - . Proximal single lumen - . Telescoping section The distal lumen and proximal single lumen sections compose the "working length" of the catheter, and the telescoping section remains outside of the guiding catheter and/or introducer sheath. The telescoping shaft section allows the imaging core to be advanced and retracted for 15 cm of linear movement. The corresponding movement {2}------------------------------------------------ of the transducer occurs from the proximal end of the wire exit port to the proximal end of the distal lumen. The imaging core is composed of a hi-torque, flexible, rotating drive cable with a radial looking 40 MHz ultrasonic transducer at the distal tip. An electro-mechanical connector interface at the proximal end makes the connection to the BSC IVUS System Motor Drive Unit (MDU). ### Intended Use / Indications for Use The Atlantis™ 018 Peripheral Imaging Catheter is intended for intravascular ultrasound examination of peripheral vessels only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. ### Device Technology Characteristics and Comparison to Predicate Device The Atlantis™ 018 Peripheral Imaging Catheter and the predicate devices are comprised of a catheter body and an imaging core. The Atlantis™ 018 Peripheral Imaging Catheter is equivalent in terms of design, operational characteristics, and material composition to the Atlantis™ SR Pro Coronary Imaging Catheter. In terms of intended use and operational characteristics, the Atlantis™ 018 Peripheral Imaging Catheter is equivalent to the Sonicath Ultra™ 3.2F Peripheral Imaging Catheter. The use of the device in intravascular ultrasound imaging is the same among the 2 predicate devices and Atlantis™ 018; however both Atlantis™ 018 and Sonicath Ultra™ 3.2F are indicated for peripheral use, while Atlantis™ SR Pro is indicated for coronary use. {3}------------------------------------------------ #### Non-Clinical Test Results The performance test results of the Atlantis™ 018 Peripheral Imaging Catheter demonstrate the device meets or exceeds the performance requirements for the specified intended use. Bench, acoustic output, animal, and biocompatibility tests were conducted on the device. #### Bench Testing Bench testing was conducted to evaluate the physical integrity and functionality of the catheter sheath and imaging core. The testing consisted of dimensional and functional verification. The Atlantis™ 018 Peripheral Imaging Catheter met or exceeded all required specifications. #### Acoustic Output Acoustic output testing was conducted as specified in the FDA guidance document Information for Manufacturers Seeking Market Clearance of Diagnostic Ultrasound Systems and Transducers, issued in 1997. The Atlantis™ 018 Peripheral Imaging Catheter test results are below the FDA Track 1 specified limits. ### Animal Testing The Atlantis™ 018 Peripheral Imaging Catheter was evaluated through animal testing to assess the in-vivo deliverability and imaging characteristics of the catheter. The device performance demonstrated clinical acceptability for the intended use. ### Biocompatibility Testing Biocompatibility tests were selected in accordance with Blue Book Memo, G95-1, Use of International Standard ISO-10993, and Biological Evaluation of Medical Devices Part 1: Evaluation and Testing, dated May 1, 1995. The biocompatibility test results demonstrate the device is acceptable for its intended use. {4}------------------------------------------------ #### Conclusion The Atlantis™ 018 Peripheral Imaging Catheter is substantially equivalent to the Atlantis™ SR Pro Coronary Imaging Catheter and the Sonicath Ultra™ 3.2F Peripheral Imaging Catheter. The Atlantis™ 018 Peripheral Imaging Catheter test results support the determination of substantial equivalence to predicate devices, and confirm the device is safe and effective for its intended use. {5}------------------------------------------------ Public Health Service Image /page/5/Picture/11 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 3 2008 Boston Scientific Corporation c/o Ms. Emilly Tojima Regulatory Affairs Specialist II 47900 Bayside Parkway Fremont, CA 94538 Re: K073623 Trade Name: Atlantis 018 Peripheral Imaging Catheter Regulation Number: 21 CFR 892.1570 Regulation Name: Non-invasive Blood Pressure Measurement System Regulatory Class: Class II (two) Product Codes: ITX, OBJ Date: December 21, 2007 Received: December 26, 2007 Dear Ms. Tojima: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Trug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {6}------------------------------------------------ Page 2 -- Ms. Emilly Tojima Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance. please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Bfommemasfor Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {7}------------------------------------------------ 510(k) Number: K073623 Device Name: Atlantis™ 018 Peripheral Imaging Catheter Indications for Use: The Atlantis™ 018 Peripheral Imaging Catheter is intended for intravascular ultrasound examination of peripheral vessels only. Intravascular ultrasound imaging is indicated in patients who are candidates for transluminal interventional procedures. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Blummer n Sian-Off Vices 510
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...