K073614 · Foosin Medical Supplies, Inc. · NEW · Apr 30, 2008 · General, Plastic Surgery
Device Facts
Record ID
K073614
Device Name
WG-SURGICAL SUTURES WITH NEEDLE
Applicant
Foosin Medical Supplies, Inc.
Product Code
NEW · General, Plastic Surgery
Decision Date
Apr 30, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4840
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
Absorbable polyglycolic acid suture with needle is indicated for use in general soft tissue approximation and/or ligation, but not for use in ophthalmic, cardiovascular and neurological tissues. Absorbable Polydioxanone Suture with Needle is indicated for use in all types of soft tissue approximation, including use in cardiovascular tissue where growth is expected to occur, PDO suture is not indicated in adult cardiovascular tissue, microsurgery, ophthalmic and neural tissue.
Device Story
WG-Surgical Sutures with Needle consist of absorbable surgical sutures (polyglycolic acid or polydioxanone) attached to stainless steel needles. Used by surgeons for soft tissue approximation and ligation. Devices are EO sterilized and pyrogen-free. Performance verified via USP 30 physical testing (diameter, attachment, tensile strength) and ISO 10993 biocompatibility studies. In vivo testing confirms tensile and mass loss rates.
Clinical Evidence
No clinical data. Evidence consists of bench testing including USP 30 physical testing (suture diameter, attachment, tensile strength) and in vivo strength testing for tensile and mass loss rates. Biocompatibility testing performed per ISO 10993.
Technological Characteristics
Materials: Polyglycolic acid or polydioxanone suture with stainless steel needle. Sterilization: Ethylene Oxide (EO). Standards: USP 30 (diameter, attachment, tensile strength), ISO 10993 (biocompatibility).
Indications for Use
Indicated for soft tissue approximation and/or ligation. Polyglycolic acid sutures: contraindicated for ophthalmic, cardiovascular, and neurological tissues. Polydioxanone sutures: indicated for soft tissue approximation, including pediatric cardiovascular tissue; contraindicated for adult cardiovascular tissue, microsurgery, ophthalmic, and neural tissue.
Regulatory Classification
Identification
An absorbable polydioxanone surgical suture is an absorbable, flexible, sterile, monofilament thread prepared from polyester polymer poly (p-dioxanone) and is intended for use in soft tissue approximation, including pediatric cardiovascular tissue where growth is expected to occur, and ophthalmic surgery. It may be coated or uncoated, undyed or dyed, and with or without a standard needle attached.
Special Controls
*Classification.* Class II (special controls). The special control for the device is FDA's “Class II Special Controls Guidance Document: Surgical Sutures; Guidance for Industry and FDA.” See § 878.1(e) for the availability of this guidance document.
{1}------------------------------------------------
K073614 page 1/3
Premarket Notification 510(k) Submission -510(k) Summary Report No .: A20071012
### APR 3 0 2008
# 510(k) Summary
As Required by 21 CFR 807.92
The Assigned 510(k) Number is: K073614
Submitter Information:
Manufacturer Name: Foosin Medical Supplies Inc., Ltd. No.312, Shichang Road Weihai, Shandong, China, 264209
Contact Person of the Submission: Ms. Diana Hong Mr. Eric Chen Shanghai Mid-Link Business Consulting Co., Ltd Suite 8D, Zhongxin Zhongshan Mansion, No.19, Lane 999, Zhong Shan Nan Er Road Shanghai, China 20020 Phone: +86-21-64264467 x 152 Fax: +86-21-64264468 x 809 Email: Diana.hong(@mid-link.net eric.chen(a)mid-link.net
#### Applicant Device Information
1. Absorbable Polyglycolic Acid Suture with Needle
#### 1.1 Applicant Device Information
Trade/Proprietary Name: WG-Surgical Sutures with Needle Common Name: Absorbable Polyglycolic Acid Suture with Needle Classification Name: Suture, Absorbable, Synthetic, Polyglycolic Acid Device Class: II Product Code: GAM Regulation Number: 878.4493 Review Panel: General & Plastic Surgery
Shanghai Mid-Link Business Consulting Co., Ltd
{2}------------------------------------------------
K073614 p
Premarket Notification 510(k) Submission -510(k) Summary Report No .: A20071012
#### Intended Use:
Absorbable polyglycolic acid suture with needle is indicated for use in general soft tissue approximation and/or ligation, but not for use in ophthalmic, cardiovascular and neurological tissues.
#### 1.2Predicate Device
K-number: K063536
#### 1.3Device Description
The applicant devices of absorbable polyglycolic acid suture with needle consist of a polyglycolic acid surgical suture made of polyglycolic acid and a stainless steel needle. It is EO sterilized, and prygon-free.
#### 1.4 Performance Data:
Physical testing was performed on the subject device to USP 30, including <861> Suture Diameter, <871> Suture Attachment, <881> Tensile Strength. Biocompatibility studies were performed in accordance with ISO10993. In vivo strength test was also conducted on the subject device to demonstrate rates of tensile and mass loss.
#### 1.5 Substantially Equivalence Determination
The applicant device of Absorbable Polyglycolic Acid Suture with Needle is substantially equivalent to the predicate device.
#### 2. Absorbable Polyglycolic Acid Suture with Needle
#### 2.1Applicant Device Information
Trade/Proprietary Name: WG-Surgical Sutures with Needle Common Name: Absorbable Polydioxanone Suture with Needle Classification Name: Suture, Absorbable, Synthetic, Polydioxanone Device Class: II Product Code: NEW Regulation Number: 878.4840 Review Panel: General & Plastic Surgery Intended Use:
Absorbable Polydioxanone Suture with Needle is indicated for use in all types of soft tissue approximation, including use in cardiovascular tissue where growth is expected to occur, PDO suture is not indicated in adult cardiovascular tissue, microsurgery, ophthalmic and neural tissue.
#### 2.2Predicate Device
K-number: K053380
{3}------------------------------------------------
K073614
#### 2.3Device Description
The applicant devices of absorbable polydioxanone suture with needle consist of a polydioxanone surgical suture made of polyester and a stainless steel needle. It is unbraided. It is EO sterilized, and prygon-free.
#### 2.4 Performance Data:
Physical testing was performed on the subject device to USP 30, including <861> Suture Diameter, <871> Suture Attachment, <881> Tensile Strength. Biocompatibility studies were performed in accordance with ISO10993. In vivo strength test was also conducted on the subject device to demonstrate rates of tensile and mass loss.
#### 2.5 Substantially Equivalence Determination
The applicant device of Absorbable Polydioxanone Suture with Needle is substantially equivalent to the predicate device.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle or bird-like symbol with three curved lines representing its wings or body. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol.
#### Public Health Service
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## APR 3 0 2008
Foosin Medical Supplies, Inc. % Shanghai Midlink Business Consulting Co., Ltd. Ms. Diana Hong Suite 8D, Zhongxin Zhongshan Mansion No. 19, Lane 999, Zhong Shan No. 2 Road(S) Shanghai, China 200030
Re: K073614
Trade/Device Name: WG-Surgical Sutures with Needle Regulation Number: 21 CFR 878.4840 Regulation Name: Absorbable polydioxanone surgical suture Regulatory Class: II Product Code: NEW, GAM Dated: April 4, 2008 Received: April 4, 2008
Dear Ms. Hong:
We have reviewed vour Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{5}------------------------------------------------
Page 2 - Ms. Diana Hong
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket survéillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark M. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number: K073614
WG-Surgical Sutures with Needle Device Name:
Indications for Use:
Absorbable polyglycolic acid suture with needle is indicated for use in general soft tissue approximation and/or ligation, but not for use in ophthalmic, cardiovascular and neurological tissues.
Absorbable Polydioxanone Suture with Needle is indicated for use in all types of soft tissue approximation, including use in cardiovascular tissue where growth is expected to occur, PDO suture is not indicated in adult cardiovascular tissue, microsurgery, ophthalmic and neural tissue.
Prescription Use Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nil Rr Ggl for AKm
(Division Sign-Off)
Page 1 of 1
Division of General, Restorative, and Neurological Devices
510(k) Number K073614
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.