MICROSTAAR MSI INJECTOR FOAM TIP SYSTEM

K073591 · Staar Surgical Co. · MSS · Mar 13, 2008 · Ophthalmic

Device Facts

Record IDK073591
Device NameMICROSTAAR MSI INJECTOR FOAM TIP SYSTEM
ApplicantStaar Surgical Co.
Product CodeMSS · Ophthalmic
Decision DateMar 13, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.4300
Device ClassClass 1
AttributesTherapeutic

Indications for Use

The MicroSTAAR ™ MSI Injector Foam Tip System is a device intended to fold and insert STAAR Surgical one and three piece Collamer® lenses for surgical placement in the human eye.

Device Story

MicroSTAAR MSI Injector Foam Tip System; manual surgical instrument; used for folding and inserting STAAR Surgical one-piece and three-piece Collamer lenses into human eye. Device facilitates lens delivery during ophthalmic surgery. Operated by ophthalmologists in clinical/OR settings. Provides mechanical assistance for lens implantation; reduces risk of lens damage during insertion. No electronic or software components.

Clinical Evidence

Bench testing only; no clinical data provided.

Technological Characteristics

Manual surgical instrument; foam tip design for lens handling; intended for use with STAAR Surgical Collamer lenses. Class I (reserved) device; product code MSS. No energy source, software, or electronic components.

Indications for Use

Indicated for folding and inserting STAAR Surgical one and three piece Collamer® lenses into the human eye during surgical procedures.

Regulatory Classification

Identification

An intraocular lens guide is a device intended to be inserted into the eye during surgery to direct the insertion of an intraocular lens and be removed after insertion is completed.

Submission Summary (Full Text)

{1}------------------------------------------------ Image /page/0/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 MAR 1 3 2008 STAAR Surgical Co. c/o Ms. Diana Smith Director of Regulatory Affairs 1911 Walker Avenue Monrovia, CA 91016 Re: K073591 Trade Name: MicroSTAARTM MSI Injector Foam Tip System Regulation Number: 21 CFR 886.4300 Regulation Name: Intraocular lens guide Regulatory Class: Class I (reserved) Product Code: MSS Dated: February 12, 2008 Received: February 19, 2008 Dear Ms. Smith: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {2}------------------------------------------------ This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Maling B. Egleston, und Malvina B. Eydelman, M.I Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {3}------------------------------------------------ Page 3 – Ms. Smith ## Indications for Use 510(k) Number (if known): K0703591 Device Name: MicroSTAAR™ MSI Injector Foam Tip System Indications For Use: The MicroSTAAR ™ MSI Injector Foam Tip System is a device intended to fold and insert STAAR Surgical one and three piece Collamer® lenses for surgical placement in the human eye. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Ay (Division'Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devises Page 1 of 1 **510(k) Number** Ko 70359)
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...