OLYMPUS IGA REAGENT

K073489 · Olympus America, Inc. · CFN · Feb 11, 2008 · Immunology

Device Facts

Record IDK073489
Device NameOLYMPUS IGA REAGENT
ApplicantOlympus America, Inc.
Product CodeCFN · Immunology
Decision DateFeb 11, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 866.5510
Device ClassClass 2

Indications for Use

System reagent for the quantitative determination of IgA immunoglobulins in human serum and plasma on OLYMPUS analyzers. For in vitro diagnostic use.

Device Story

The Olympus IgA reagent (OSR6X171) is an in vitro diagnostic reagent used on Olympus AU400/400e, 600/640/640e, and 2700/5400 analyzers. The device utilizes an immunoturbidimetric principle: human IgA in patient serum or plasma samples reacts with goat anti-human IgA antibodies to form insoluble aggregates. These immune complexes scatter light; the analyzer measures the reduction in transmitted light intensity (increase in absorbance) to quantify IgA concentration. The system is operated by laboratory technicians in a clinical setting. Results are provided to clinicians to assist in diagnosing abnormal protein metabolism and immune system deficiencies. The device offers improved prozone capacity and expanded specimen types (serum, EDTA plasma, Li-Heparinized plasma) compared to the predicate.

Clinical Evidence

Bench testing only. Precision studies (n=80) across multiple platforms showed total CVs ranging from 1.50% to 4.01%. Linearity demonstrated from 10–700 mg/dL. Method comparison (n=111-115) against predicate showed high correlation (r=0.998-0.999). Matrix comparison (n=45) confirmed equivalence between serum and Li-heparin/EDTA plasma (r=0.999-1.000). Interference testing for bilirubin, lipids, hemoglobin, and RF showed acceptable performance.

Technological Characteristics

Immunoturbidimetric assay; liquid reagents (Tris buffer, PEG 6000, goat anti-IgA antiserum). Analyzers: Olympus AU series. Calibration: Olympus Serum Protein Multicalibrator (traceable to CRM 470). Measuring range: 10-700 mg/dL. On-board stability: 90 days. Sterilization: Not applicable (reagent).

Indications for Use

Indicated for the quantitative determination of IgA immunoglobulins in human serum and plasma to aid in the diagnosis of abnormal protein metabolism and the body’s lack of ability to resist infectious agents. For prescription use only.

Regulatory Classification

Identification

An immunoglobulins A, G, M, D, and E immunological test system is a device that consists of the reagents used to measure by immunochemical techniques the immunoglobulins A, G, M, D, an E (serum antibodies) in serum. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infectious agents.

Predicate Devices

Submission Summary (Full Text)

{0} 1 # 510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION DECISION SUMMARY A. 510(k) Number: k073489 B. Purpose for Submission: Addition of plasma (Li heparin/EDTA) matrix claim to the predicate device C. Measurand: Immunoglobulin A (IgA) D. Type of Test: Quantitative immunoturbidimetric assay E. Applicant: Olympus America, Inc. F. Proprietary and Established Names: Olympus IgA reagent (OSR6X171) G. Regulatory Information: | Product Code | Classification | Regulation Section | Panel | | --- | --- | --- | --- | | CFN Method, Nephelometric, Immunoglobulins (G, A, M) | Class II | 21 CFR 866.5510 Immunoglobulins A, G, M, D, E Immunological Test System | Immunology (IM82) | H. Intended Use: 1. Intended use(s): System reagent for the quantitative determination of IgA immunoglobulins in human serum and plasma on OLYMPUS analyzers 2. Indication(s) for use: The spectrum of abnormalities in serum immunoglobulin concentration is broad. Abnormal concentrations range from a virtual absence of one or more of the three major classes of immunoglobulins (IgA, IgG and IgM) to polyclonal increases in one or more immunoglobulins. Measurement of these immunoglobulins aids in the diagnosis of abnormal protein metabolism and the body’s lack of ability to resist infectious agents. 3. Special conditions for use statement(s): For prescription use only 4. Special instrument requirements: OLYMPUS analyzers: AU400/400ᵉ, 600/640/640ᵉ and 2700/5400 I. Device Description: The device consists of two reagents: R1 buffer (Tris buffer pH 7.2, polyethylene glycol 6000) and R2 (goat anti-IgA antiserum). The reagents contain sodium azide as preservative. J. Substantial Equivalence Information: 1. Predicate device name(s): Olympus IgA reagent (OSR6X44) {1} 2. Predicate 510(k) number(s): k951055 3. Comparison with predicate: | Similarities | | | | --- | --- | --- | | Item | Device | Predicate | | | Olympus IgA reagent (OSR6X171) | Olympus IgA reagent (OSR6X44) | | Intended Use | System reagent for the quantitative determination of IgA immunoglobulins in human serum and plasma on Olympus analyzers | Same but in serum only | | Indications for Use | Aid in the diagnosis of abnormal protein metabolism and the body's lack of ability to resist infection | Same | | Test principle | Immunoturbidimetric | Same | | Antibody | Goat anti-IgA | Same | | Reagent form and storage | Liquid, on-board storage | Same | | On-board stability | 90 days | Same | | Calibrator | Olympus Serum Protein Multicalibrator | Same | | Calibrator traceability | International Reference Preparation CRM 470 | Same | | Expected values | 66-433 mg/dL | Same | | Differences | | | | --- | --- | --- | | Item | Device | Predicate | | | Olympus IgA reagent (OSR6X171) | Olympus IgA reagent (OSR6X44) | | Matrix | Serum, plasma (Li heparin or EDTA) | Serum only | | Calibration frequency | 90 days | 7 days | # K. Standard/Guidance Document Referenced (if applicable): EN14971 (2000) ISO Medical Devices – Application of Risk Management to Medical Devices; EP7-A2 (2005) CLSI Interference Testing in Clinical Chemistry; EP5-A2 (2004) CLSI Evaluation of Precision Performance of Clinical Chemistry Devices; EP9-A2 (2002) CLSI Method Comparison and Bias Estimation Using Patient Samples; CEN 13640 (2002) Stability Testing of In Vitro Diagnostic Reagents; C28-A2 (2000) CLSI How to Define and Determine Reference Intervals in the Clinical Laboratory; EP6-A (2003) CLSI Evaluation of the Linearity of Quantitative Measurement Procedures: A Statistical Approach; FDA: Draft Guidance document for 510(k) Submission of Immunoglobulins A, G, M, D and E Immunoglobulin Test System In Vitro Devices # L. Test Principle: {2} When a sample is mixed with R1 buffer and R2 antiserum solution, human IgA reacts specifically with anti-human IgA antibodies to yield insoluble aggregates. Immune complexes formed in solution scatter light in proportion to their size, shape and concentration. The Olympus analyzer measures the decrease in intensity of light transmitted (increase in absorbance) through particles suspended in solution as a result of complexes formed during the antigen-antibody reaction. ## M. Performance Characteristics (if/when applicable): ### 1. Analytical performance: #### a. Precision/Reproducibility: Assays of three human serum pools (low, medium and high) sera were performed in duplicate 2 runs per day for 20 days (n=80) on the AU400/400ᵉ, AU600/640/640ᵉ, and AU2700/5400. The acceptance criteria for within-run and total precision were <4.2% and <10% respectively. The within-run precision covering the platforms ranged from 1.05–2.18% and the total precision ranged from 1.50–4.01%. AU400/400ᵉ | N=80 | Within-run | | Total | | | --- | --- | --- | --- | --- | | Mean (mg/dL) | SD | %CV | SD | %CV | | 102 | 1 | 1.20 | 2 | 2.43 | | 244 | 4 | 1.48 | 6 | 2.52 | | 504 | 11 | 2.18 | 15 | 2.95 | AU600/640/640ᵉ | N=80 | Within-run | | Total | | | --- | --- | --- | --- | --- | | Mean (mg/dL) | SD | %CV | SD | %CV | | 102 | 1 | 1.41 | 3 | 3.39 | | 240 | 4 | 1.52 | 9 | 3.85 | | 479 | 10 | 2.18 | 19 | 4.01 | AU2700/5400 | N=80 | Within-run | | Total | | | --- | --- | --- | --- | --- | | Mean (mg/dL) | SD | %CV | SD | %CV | | 102 | 1 | 1.05 | 2 | 1.50 | | 237 | 4 | 1.56 | 5 | 1.91 | | 486 | 10 | 1.20 | 9 | 1.83 | ### Auto dilution: To validate the accuracy and precision of automated sample dilution protocol, three auto-dilution samples were diluted manually and run on the instrument. The same samples were diluted automatically by the AU640. Accuracy (%difference) and precision (%CV) were determined. Accuracy 1:5 | Level | Automatic dilution (mg/dl) | Manual dilution (mg/dl) | % Difference | | --- | --- | --- | --- | | 1 | 996 | 948 | -5.1 | | 2 | 817 | 805 | -1.5 | | 3 | 522 | 512 | -2.0 | {3} Accuracy 1:10 | Level | Automatic dilution (mg/dl) | Manual dilution (mg/dl) | % Difference | | --- | --- | --- | --- | | 1 | 970 | 961 | -0.9 | | 2 | 767 | 780 | 1.7 | | 3 | 577 | 584 | 1.2 | Precision (within run) 1:5 | Level | Mean (mg/dL) | SD (mg/dL) | CV (%) | Essential Specification | | | --- | --- | --- | --- | --- | --- | | 1 | 93 | 1 | 1.38 | ≤4.2% CV | Pass | | 2 | 125 | 1 | 0.91 | | | | 3 | 157 | 2 | 1.42 | | | Precision (within run) 1:10 | Level | Mean (mg/dL) | SD (mg/dL) | CV (%) | Essential Specification | | | --- | --- | --- | --- | --- | --- | | 1 | 42 | 1 | 1.24 | ≤4.2% CV | Pass | | 2 | 58 | 1 | 1.12 | | | | 3 | 75 | 1 | 1.29 | | | # b. Linearity/assay reportable range: The measuring range for the assay is $10 - 700\mathrm{mg / dL}$ . The procedure used to demonstrate linearity was based on CLSI EP6-A. A series of at least ten analyte concentrations, covering the linear dynamic range were prepared by dilution of a high pool sample. Each dilution was assayed in quadruplicate and the mean analytical results were plotted versus the relative analyte concentrations (\% dilution). Studies were performed on the AU400, AU640 and AU2700 analyzers. The acceptance criteria for deviation from the regression line for the $10 - 40\mathrm{mg / dL}$ and $40 - 700\mathrm{mg / dL}$ ranges were $\pm 4\mathrm{mg / dL}$ and $\pm 10\%$ respectively. The studies showed the assay was linear from $10 - 700\mathrm{mg / dL}$ . There was no high dose hook effect up to $10,000\mathrm{mg / dL}$ . # c. Traceability, Stability, Expected values (controls, calibrators, or methods): The calibrator is traceable to the International Reference Preparation CRM470 (US designation RPPHS lot 91/0619). # Calibration frequency: The calibration stability period change from 7 days to 90 days was validated during the on-board reagent stability studies. Linearity and in-house control recovery were checked after 90 days of reagent storage. The linearity is displayed at day 90 and the $\%$ drift from Day 0 from control recovery (with calibration at day 0) was calculated. Results were acceptable. Reagent on-board stability was demonstrated according to internal procedures where the linearity displayed at day 90 and the $\%$ drift from Day 0 from control recovery were calculated. A change of $\leq 10\%$ was demonstrated over the 90 days. # d. Detection limit: The Limit of Quantitation (LoQ) for the new assay was determined by testing {4} 3 patient pools, 40 fold at an analyte concentration below the lower end of the measuring range on the AU400, AU640 and AU2700. The analyte level at which a CV of less than 20% was determined to be <10 mg/dL. This was determined using a method based on the CLSI protocol EP17-A. | | Mean Concentration (mg/dL) | SD | CV (%) | | --- | --- | --- | --- | | AU400 | 6.3 | 1.1 | 18.0 | | AU640 | 3.2 | 0.21 | 6.4 | | AU2700 | 2.9 | 0.26 | 8.8 | The Limit of Detection (LoD) or the concentration of analyte which is significantly different from zero was determined by testing an analyte free sample twenty-fold on the AU400, AU640 and AU2700. LoD was calculated as the absolute mean + 3SD. The lowest detectable level was determined to be ≤1 mg/dL. | | Mean Concentration (mg/dL) | SD | LoD (mg/dL) | | --- | --- | --- | --- | | AU400 | 0.0 | 0.0 | 0.00 | | AU640 | -0.02 | 0.162 | 0.51 | | AU2700 | -0.02 | 0.107 | 0.34 | e. Analytical specificity: The impact of bilirubin, lipids and hemoglobin were assessed in accordance with CLSI EP7-A2. The RF interference studies were carried out following in-house procedures. | Substance | Levels up to | % Interference | | | | --- | --- | --- | --- | --- | | | | AU400/400^{e} | AU600/640/640^{e} | AU2700/5400 | | Bilirubin | 40 mg/dL | ≤ 2% | ≤ 3% | ≤ 3% | | Lipids | 1000 mg/dL | ≤ 10% | ≤ 6% | ≤ 4% | | Hemoglobin | 500 mg/dL | ≤ 1% | ≤ 5% | ≤ 4% | | RF | 600 IU/mL | ≤ 8% | ≤ 8% | ≤ 4% | f. Assay cut-off: See reference range 2. Comparison studies: a. Method comparison with predicate device: | Y method (new) | AU2700 | AU2700/5400 | AU2700/5400 | | --- | --- | --- | --- | | X method (predicate) | AU2700 | AU400 | AU640/640^{e} | | Slope | 0.923 | 1.029 | 0.948 | | Intercept | 15.1 | -2.8 | 9.1 | | Correlation coefficient (r) | 0.999 | 0.998 | 0.999 | | Number of samples | 111 | 115 | 115 | | Range (mg/dL) Y method | 38-672 | 21-672 | 21-672 | | Range (mg/dL) X method | 32-684 | 23-659 | 21-697 | b. Matrix comparison: Studies were performed based on CLSI EP9-A2. {5} | Y method | Li-heparin plasma | EDTA plasma | | --- | --- | --- | | X method | Serum | Serum | | Slope | 0.944 | 0.942 | | Intercept | +3.984 | +2.227 | | Correlation coefficient | 1.000 | 0.999 | | Number of samples | 45 | 45 | | Patient mean value – serum mg/dL | 245.14 | 245.14 | | Patient mean value – plasma mg/dL | 235.50 | 233.05 | | Reference range – serum mg/dL | 42.88 – 614.44 | 42.88 – 614.44 | | Reference range - plasma mg/dL | 40.61 – 581.67 | 40.54 – 598.41 | 3. Clinical studies: a. Clinical Sensitivity: Not determined b. Clinical specificity: Not determined c. Other clinical supportive data (when a. and b. are not applicable): Not applicable 4. Clinical cut-off: Not applicable 5. Expected values/Reference range: Expected values may vary with age, sex, diet and geographical location. The reference range of 66-433 mg/dL established for the predicate device was reverified according to CLSI C28-A2 on the Olympus AU400, 600 and 5400. N. Proposed Labeling: The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10. O. Conclusion: The submitted information in this premarket notification is complete and supports a substantial equivalence decision.
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...