OMNIGUIDE STERILE WAVEGUIDE ADAPTER SYSTEM
K073409 · Omniguide, Inc. · GEX · Dec 14, 2007 · General, Plastic Surgery
Device Facts
| Record ID | K073409 |
| Device Name | OMNIGUIDE STERILE WAVEGUIDE ADAPTER SYSTEM |
| Applicant | Omniguide, Inc. |
| Product Code | GEX · General, Plastic Surgery |
| Decision Date | Dec 14, 2007 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4810 |
| Device Class | Class 2 |
| Attributes | Therapeutic, 3rd-Party Reviewed |
Intended Use
The OmniGuide Sterile WaveGuide Adapter System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy. The indications for use for which the delivery system is used for are dependent upon the cleared indications for use of the laser system and those laser system accessories to which it is attached.
Device Story
Adapter system connecting hospital CO2 laser to OmniGuide WaveGuide fibers; enables sterile surgical use via filtered, autoclavable components. Device acts as conduit for laser energy; performance parameters (power output, beam quality) match predicate. Used by surgeons in OR/clinical settings for soft tissue procedures. Benefits include sterile delivery of laser energy for minimally invasive and open surgeries.
Clinical Evidence
No clinical data. Bench testing only; evaluated laser power output and beam quality per FDA guidance for medical lasers.
Technological Characteristics
Adapter system for CO2 laser to WaveGuide fibers. Includes output tube, gas hose, and filter element. Components are autoclavable or gamma-sterilized. Mechanical interface for laser energy delivery.
Indications for Use
Indicated for incision, excision, ablation, vaporization, and coagulation of soft tissues (including intra-oral) across multiple surgical specialties (general, plastic, oral/maxillofacial, dental, derm, gyn, ENT, GI, neuro, urology, pulmonary). Applicable for open and endoscopic procedures.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
- OmniGuide Beam Delivery System (K062423)
Related Devices
- K062423 — OMNIGUIDE BEAMPATH LASER BEAM DELIVERY SYSTEM · Omniguide, Inc. · Sep 21, 2006
- K070157 — OMNIGUIDE BEAM PATH CO2 MARK III WAVEGUIDE FIBER · Omniguide, Inc. · Feb 1, 2007
- K992472 — SUREGUIDE CO2 LASER BEAM DELIVERY SYSTEM · Clinicon Corp. · Sep 8, 1999
- K130164 — FIBERLASE ENDURE CO2 FIBER · Lumenis, Ltd. · Mar 24, 2013
- K050541 — OMNIGUIDE BEAMPATH CO2 MARK I LASER BEAM DELIVERY · Omniguide Communications, Inc. · May 9, 2005
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the word "OmniGuide" in a sans-serif font. The word is partially obscured by two concentric circles. The circles are thick and black, and the word is centered within them.
:
DEC 1 4 2007
# K073409
# 510(k) Summary
| Submitter: | OmniGuide, Inc.<br>One Kendall Square, Building 100 3rd Floor<br>Cambridge, MA 02139 | |
|-------------------------------|-----------------------------------------------------------------------------------------|--|
| Contact Person:<br>Telephone: | Douglas W. Woodruff<br>617-551-8404 | |
| Fax: | 617-551-8445 | |
| Proprietary Name: | OmniGuide Sterile WaveGuide Adapter System | |
| Common Name: | CO2 Laser Powered Surgical Instrument | |
| Classification: | 878.4810 | |
| Product Code: | GEX | |
| Classification Name: | Laser surgical instrument for use in general and plastic<br>surgery and in dermatology. | |
### Substantial Equivalence Claimed To:
K062423, OmniGuide Beam Delivery System
{1}------------------------------------------------
#### Description:
#### Intended Use:
The OmniGuide Sterile WaveGuide Adapter System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, gastroscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy.
The indications for use for which the delivery system is used for are dependent upon the cleared indications for use of the laser system and those laser system accessories to which it is attached.
#### Summary of Technological Characteristics:
The OmniGuide Sterile WaveGuide Adapter System connects a hospitals CO2 laser to the OmniGuide WaveGuide Fibers. The additional filter and autoclave sterilization of the adapter and hose allows for use in sterile surgical procedures.
The output tube and gas hose will be sold non-sterile and the hospital will autoclave sterilize. The filter element may be provided as a gamma sterilized device (high Pressure option) or as a non sterile device for autoclave sterilization. The low pressure filter may be autoclave sterilized or provided sterile at the end users request. See section S for sterilization validation activites.
#### Performance Data:
#### Non-clinical Performance Data:
The OmniGuide Sterile WaveGuide Adapter System performance characteristics have been evaluated through testing and analysis of laser power output and beam quality. This type of testing complies with the respective section of the FDA Guidance on the Content and Organization of a Premarket Notification for a Medical Laser (1995) and is similar to the predicate device tests. The performance of the OmniGuide Sterile WaveGuide Adapter System and the related parameters of the predicate device is comparable.
#### Clinical Performance Data:
Formal clinical trials were not deemed necessary as the device is using the same technology and intended use as the predicate device.
{2}------------------------------------------------
#### Conclusions Drawn from Tests and Analysis:
The intended use and major performance parameters (energy transmission levels and beam quality) of the OmniGuide Sterile WaveGuide Adapter System are similar or equivalent to the characteristics of above mentioned legally marketed devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a serpent entwined around it. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus symbol.
Public Health Service
DEC 1 4 2007
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OmniGuide, Inc. % Regulatory Technology Services, LLC Mr. Mark Job 1394 25th Street, Northwest Buffalo, Minnesota 55313
Re: K073409
Trade/Device Name: OmniGuide Sterile WaveGuide Adapter System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: December 3, 2007 Received: December 4, 2007
Dear Mr. Job:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Mark Job
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitts your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliary, at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events, (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistances of at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark N. Mulhearn
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## Indications for Use
510(k) Number (if known): Not yet assigned
Device Name: OmniGuide Sterile WaveGuide Adapter System
Indications For Use:
The OmniGuide Sterile WaveGuide Adapter System is indicated for the incision, excision, ablation, vaporization and coagulation of body soft tissues including intra-oral tissues. It is indicated in the medical specialties of general and plastic surgery, oral/maxillofacial surgery, dentistry, dermatology, gynecology, otorhinolaryngology, gastroenterology, neurosurgery, urology, and pulmonology, and can be used in open surgical procedures as well as endoscopic minimally invasive procedures in conjunction with rigid or flexible endoscopes, such as in laryngoscopy, colonoscopy, laparoscopy, thoracoscopy, hysteroscopy and bronchoscopy.
The indications for use for which the delivery system is used for are dependent upon the cleared indications for use of the laser system and those laser system accessories to which it is attached.
| Prescription Use (Part 21 CFR 801 Subpart D) | X |
|----------------------------------------------|---------------------------------------------|
| AND/OR | Over-The-Counter Use (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
(Division Sign-Off)
Division of General, Restorative, and Neurological Devices
| Page | 1 of 1 |
|---------------|---------|
| 510(k) Number | 1073469 |
Page 15