BLUE LATEX EXAM GLOVES NON STERILE CHLORINATED POWDER FREE
K073299 · Betel Care Sdn. Bhd. · LYY · Mar 5, 2008 · General Hospital
Device Facts
Record ID
K073299
Device Name
BLUE LATEX EXAM GLOVES NON STERILE CHLORINATED POWDER FREE
Applicant
Betel Care Sdn. Bhd.
Product Code
LYY · General Hospital
Decision Date
Mar 5, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 880.6250
Device Class
Class 1
Indications for Use
A medical gloves is worn on the hand of healthcare and or similar personnel to prevent contamination between healthcare personnel and the patient. A patient examination gloves is a disposable device intended for medical purposes worm on the examiner, s hand to prevent contamination between patient and examiner. This gloves is intended for medical and dental use.
Device Story
Disposable, non-sterile, blue, powder-free natural rubber latex examination gloves. Worn by healthcare and dental personnel to provide a protective barrier against cross-contamination between examiner and patient. Manufactured via standard latex dipping processes. Performance verified against ASTM standards for water leak, residue, tensile strength, elongation, and extractable protein. Used in clinical and dental settings. Provides physical barrier protection; does not claim hypoallergenic properties. Contains natural rubber latex, posing potential allergic reaction risk to sensitive individuals.
Clinical Evidence
Bench testing only. Device performance verified against ASTM standards: ASTM D5151-06 (water leak), ASTM D6124-06 (residue), ASTM D3578-05 (tensile strength/elongation), and ASTM D5712-05e1 (extractable protein). Biocompatibility testing included primary skin irritation and dermal sensitization studies, both of which passed.
Indicated for use by healthcare and dental personnel as a disposable barrier on the examiner's hand to prevent cross-contamination between the examiner and the patient. Contains natural rubber latex; not for use by latex-sensitized individuals.
Regulatory Classification
Identification
A non-powdered patient examination glove is a disposable device intended for medical purposes that is worn on the examiner's hand or finger to prevent contamination between patient and examiner. A non-powdered patient examination glove does not incorporate powder for purposes other than manufacturing. The final finished glove includes only residual powder from manufacturing.
Submission Summary (Full Text)
{1}------------------------------------------------
Image /page/0/Picture/0 description: The image shows a logo with the letters "RELE" arranged around a stylized image of three plants. Below the letters and plants, there is a rectangular box with the words "natural rubber product" written inside. The logo has a vintage or retro appearance due to its black and white color scheme and simple design. The text is small and somewhat difficult to read, but the overall design is clear and recognizable.
lcare Son
Image /page/0/Picture/2 description: The image shows two logos. The logo on the left is a circular logo with the letters "BSI" at the top and the word "REGISTERED" around the bottom. The logo on the right is a square logo with a crown at the top and the words "UKAS QUALITY MANAGEMENT" in the middle. The number "003" is at the bottom of the logo.
Lot 3342, Jalan Keretapi Lama, 7½ Mile, Off Jalan Kapar, 42200 Kapar, Selangor Darul Ehsan, West Malaysia.
Tel: (603) 3290 3301, (603) 3290 3302 Fax: (603) 3290 3303
510K Summary for BETELCARE SDN BHD-Latex Exam gloves Blue PowleteriFreet@pd.jaring.my
- 1. Submission Applicant:
K073299
Name: Betelcare Sdn Bhd Address : Lot 3342,Jalan Keretapi Lama 7.5 Miles Off Jalan Kapar,Kapar Selangor Darul Ehsan West Malaysia Telephone No: 603 3290 3301 Fax No: 603 3290 3303 Email address : kbsiah@betel.com.my Contact Person: K.B .Siah
Registration Form 2891 number: 8040910 Device Listing Number : A 653345 Activity : Manufacturer
Applicant 510k number: K893166,C047750
- 2. Device Particulars: Device Name : Powder Free Latex Exam gloves Blue Common Name : Chlorinated Blue Powder Free Latex Exam Gloves Classification: Patient Examination Gloves
- 3. Device Classification Device Class: Class II Product code: LYY
- 4. Device Classification Device Class: Class II Product code: LYY
- 5. Summary of Intended Use A medical gloves is worn on the hand of healthcare and or similar personnel to prevent contamination between healthcare personnel and the patient.
MAR - 5 2008
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows a logo with the letters "REEL" arranged vertically on the left and right sides of a semi-circular arch. Inside the arch, there are three stylized plants with leaves. Below the arch, there is a rectangular box with the words "natural rubber product" written in a smaller font. The logo appears to be for a company that produces natural rubber products.
are Sdn
Image /page/1/Picture/2 description: The image contains two logos. The logo on the left is the BSI Registered logo, which features a heart shape inside a circle with the letters BSI at the top. The logo on the right is the UKAS Quality Management logo, which features a crown above a checkmark with the words UKAS Quality Management below it. The number 003 is printed below the UKAS logo.
Lot 3342, Jalan Keretapi Lama, 71/2 Mile, Off Jalan Kapar, 42200 Kapar, Selangor Darul Ehsar, West Malaysia.
Tel: (603) 3290 3301. (603) 3290 3302 Fax: (603) 3290 3303 Email: betellapd jaring my www.betel.com.my
- 6. Comparison Characteristic:
Characteristic Reference Doc Betel Device performance Meets or Exceeds Water Leak ASTM D 5151-06 ASTM D6124-06 Meets or Exceeds Residue Powder Unaged ASTM D 3578-05 Meets or Exceeds Tensile :- Aged Meets or Exceeds ASTM D 3578 -05 Unaged ASTM D 3578-05 Meets or Exceeds Elongation@break ASTM D 3578-05 Meets or Exceeds Unaqed Meets or Exceeds Aged ASTM D 3578 -05
Extractable Protein Test ASTM D 5712-05e1 Meets or Excceds
The above device does NOT include "hypoallergenic"labelling in packing.
- 7. Labeling and Attributes
Safe use of this gloves by or on latex sensitized individuals has not been established.This product contains natural rubber latex which may cause allergic reactions in some individual.
- 8. Assesment of Non clinical Performance Data The device meets or exceeds the ASTM Std or Equivalent Standards and FDA pin hole requirements
- 9. Assesment of Biocompatibility
Primary Skin Irritant Test: Pass Dermal Sensitization Study: Pass
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image is a logo with the letters "R", "E", "L", and "E" arranged around a central design. The central design features three plant-like figures. Below the letters is a rectangular box containing the words "natural rubber product" in a stylized font. The logo appears to be for a company or organization related to natural rubber products.
care Sdn
Image /page/2/Picture/2 description: The image contains two logos. The logo on the left is a circular seal with the word "REGISTERED" around the bottom half of the circle and "QSI" at the top. Inside the circle is a heart shape pointing downwards. The logo on the right is a square with a crown on top of a check mark. Below the check mark is the text "UKAS QUALITY MANAGEMENT" and the number "003".
Lot 3342, Jalan Keretapi Lama, 71/2 Mile, Off Jalan Kapar, 42200 Kapar, Selangor Darul Ehsan, West Malaysia.
Tel: (603) 3290 3301, (603) 3290 3302 Fax: (603) 3290 3303 Email betellapd jaring my www.betel.com.my
- 10. Conclusion of non clinical and biocompability performance The device has been carefully compared to legally marketed devices in the 510K. The data summaries indicates that the proposed device meets or exceeds all acceptable requirements for Powder Free Latex Exam gloves blue non sterile
I,Keng Beng Siah,the Chief Executive Officer of BETELCARE SDN BHD,certify That to the best of my knowledge and belief and based on the data and Information submitted to me in the course of my responsibilities as CEO of Betelcare Sdn Bhd and in reliance thereupon the data and information Submitted in the pre market notification are truthful and accurate no facts Materials to a review of the substantial equivalence of the device have been Knowingly omitted from this submission.
Image /page/2/Picture/8 description: The image shows a signature. The signature is illegible and appears to be a series of overlapping lines and curves. The word "Signature" is written above the signature.
KENG BENG SIAH Name
22nd Feb 2008 Date
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is a circular seal with the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract image of an eagle.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. Keng Beng Siah Director Betelcare Sdn Bhd Lot 3342 Jalan Keretapi Lama 7 1/2 Mile, Off Jalan Kapar, 42200 Kapar Selangor Darul Ehsan, WEST MALAYSIA
Re: K073299
Trade/Device Name: Betel Latex Exam Gloves, Blue, Powder Free, Non Sterile Regulation Number: 21 CFR 880.6250 Regulation Name: Patient Examination Glove Regulatory Class: I Product Code: LYY Dated: February 22, 2008 Received: February 27, 2008
MAR - 5 2008
Dear Mr. Siah:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
## Page 2 - Mr. Siah
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Snyte Y. M. Michie MD.
Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
## Indications for Use
510(k) Number (if known): 510K number is K073299
Betel Hatex Exam Gloves,Blue ,Powder Free,Non sterile Device Name: Indications For Use:
> A patient examination gloves is a disposable device intended for medical purposes worm on the examiner, s hand to prevent contamination between patient and examiner.
This gloves is intended for medical and dental use.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Stile H. Murphy ho
Division Sign Off
(Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices
Page 1 of
510(k) Number: K073999
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.