K073275 · Nuvasive, Inc. · KWQ · Jan 29, 2008 · Orthopedic
Device Facts
Record ID
K073275
Device Name
NUVASIVE HELIX MINI ACP SYSTEM
Applicant
Nuvasive, Inc.
Product Code
KWQ · Orthopedic
Decision Date
Jan 29, 2008
Decision
SESE
Submission Type
Special
Regulation
21 CFR 888.3060
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The NuVasive HELIX Mini Anterior Cervical Plating System is intended for anterior screw fixation of the cervical spine. These implants have been designed to provide stabilization as an adjunct to cervical fusion. Indications for the use of this implant system include degenerative disc disease defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma, spinal stenosis, deformity, tumor, pseudarthrosis or failed previous fusion. WARNING: The NuVasive HELIX Mini Anterior Cervical Plating System is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.
Device Story
NuVasive HELIX Mini ACP System; cervical plate and screw system for anterior spinal stabilization. Used as adjunct to cervical fusion. Surgeon-operated in clinical/surgical settings. Provides mechanical fixation to stabilize cervical spine segments. Benefits patient by facilitating fusion in cases of degenerative disease, trauma, or deformity. Mechanical device; no software or electronic components.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Spinal intervertebral body fixation orthosis consisting of metallic plates and screws. Mechanical fixation device. No software, electronics, or connectivity.
Indications for Use
Indicated for anterior cervical spine stabilization as an adjunct to fusion in patients with degenerative disc disease (discogenic neck pain confirmed by history/radiography), spondylolisthesis, trauma, spinal stenosis, deformity, tumor, pseudarthrosis, or failed previous fusion. Contraindicated for posterior element (pedicle) fixation.
Regulatory Classification
Identification
A spinal intervertebral body fixation orthosis is a device intended to be implanted made of titanium. It consists of various vertebral plates that are punched into each of a series of vertebral bodies. An eye-type screw is inserted in a hole in the center of each of the plates. A braided cable is threaded through each eye-type screw. The cable is tightened with a tension device and it is fastened or crimped at each eye-type screw. The device is used to apply force to a series of vertebrae to correct “sway back,” scoliosis (lateral curvature of the spine), or other conditions.
Predicate Devices
HELIX ACT System
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/0 description: The image shows the logo for Nuvasive, a medical device company. The logo features a stylized "N" with a curved line above it, followed by the word "NUVASIVE" in bold, sans-serif font. Below the company name is the tagline "Creative Spine Technology" in a smaller, lighter font.
# VII. 510(k) Summary
In accordance with Title 21 of the Code of Federal Regulations, Part 807, and in particular §807.92, the following summary of information is provided:
### Submitted by: A.
Laetitia Cousin Director of Regulatory and Clinical Affairs, and Quality Assurance NuVasive, Incorporated 4545 Towne Centre Court San Diego, California 92121 Telephone: (858) 909-1868 Fax: (858) 909-2068
JAN 29 2002
### Device Name B.
| Trade or Proprietary Name: | NuVasive HELIX Mini ACP System |
|----------------------------|----------------------------------------------|
| Common or Usual Name: | Cervical Plate and Screw System |
| Classification Name: | Spinal Intervertebral Body Fixation Orthosis |
| Device Class: | Class II |
| Classification: | §888.3060 |
| Product Code: | KWQ |
### C. Predicate Devices
The subject HELIX Mini ACP System is substantially equivalent to the HELIX ACT System currently distributed commercially in the U.S. by NuVasive.
## D. Device Description
The NuVasive HELIX Mini ACP System consists of a variety of plates and screws designed to provide stabilization as an adjunct to cervical fusion.
#### E. . Intended Use
The NuVasive HELIX Mini ACP System is intended for anterior screw fixation of the cervical spine. These implants have been designed to provide stabilization as an adjunct to cervical fusion. Indications for the use of this implant system include degenerative disc disease defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma, spinal stenosis, deformity, tumor, pseudarthrosis or failed previous fusion.
WARNING: The NuVasive HELIX Mini ACP System is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for "NUVASIVE Creative Spine Technology". The logo features the company name in bold, sans-serif font, with the tagline "Creative Spine Technology" in a smaller, lighter font underneath. To the left of the company name is a curved line that appears to be a stylized representation of a spine.
# F. Comparison to Predicate Devices
The subject device has indications for use identical to those of its predicate, and employs the same principles of operation.
## G. Summary of Non-Clinical Tests
Mechanical testing was presented.
# H. Summary of Clinical Tests
(Not Applicable).
{2}------------------------------------------------
DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/2/Picture/12 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the circumference. Inside the circle is a stylized emblem featuring three abstract, bird-like shapes stacked vertically, creating a sense of upward movement.
Public Health Service
JAN 2 9 2008
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NuVasive, Incorporated % Ms. Laetitia Cousin Director of Regulatory and Clinical Affairs and Ouality Assurance 4545 Towne Centre Court San Diego, CA 92121
Re: K073275
Trade/Device Name: Helix Mini Anterior Cervical Plate System Regulation Number: 21 CFR 888.3060 Regulation Name: Spinal intervertebral body fixation orthosis Regulatory Class: Class II Product Code: KWQ Dated: January 7, 2007 Received: January 8, 2007
Dear Ms. Cousin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
# Page 2 - Ms. Laetitia Cousin
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark H. Millman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
K073275
Page 1 of 1
# Indications for Use
510(k) Number (if known): ____________________________________________________________________________________________________________________________________________________
Device Name: HELIX Mini Anterior Cervical Plating System
Indications For Use:
The NuVasive HELIX Mini Anterior Cervical Plating System is intended for anterior screw fixation of the cervical spine. These implants have been designed to provide stabilization as an adjunct to cervical fusion. Indications for the use of this implant system include degenerative disc disease defined as neck pain of discogenic origin with the degeneration of the disc confirmed by history and radiographic studies, spondylolisthesis, trauma, spinal stenosis, deformity, tumor, pseudarthrosis or failed previous fusion.
WARNING: The NuVasive HELIX Mini Anterior Cervical Plating System is not intended for screw attachment or fixation to the posterior elements (pedicles) of the cervical, thoracic or lumbar spine.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 807 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Barbara Budup for you
(Division Sign-Off)
Page 1 of
(Division Sign-Off)
Page 1 of 1
**Division of General, Restorative, and Neurological Devices**
**510(k) Number** KD2325
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.