K073238 · Siemens AG Medical Solutions · JAK · Jan 3, 2008 · Radiology
Device Facts
Record ID
K073238
Device Name
SYNGO VOLUME PERFUSION-CT NEURO
Applicant
Siemens AG Medical Solutions
Product Code
JAK · Radiology
Decision Date
Jan 3, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1750
Device Class
Class 2
Attributes
Software as a Medical Device, 3rd-Party Reviewed
Indications for Use
The Siemens syngo® Volume Perfusion-CT Neuro software package has been designed to evaluate areas of brain perfusion. The software can calculate cerebral blood flow (CBF), cerebral blood volume (CBV), local bolus timing (i.e. time to start, time to peak), mean transit time (MTT), and vascular permeability (blood brain barrier disturbances) from sets of images or volumes reconstructed from continuously acquired CT data after the injection of contrast media. The package also allows the calculation of mirrored regions or volumes of interest and the visual inspection of time attenuation curves. One clinical application is to visualize the apparent blood perfusion in brain tissue affected by acute stroke. Areas of decreased perfusion, as is observed in acute cerebral infarcts, appear as areas of changed signal intensity (lower for CBF and CBV, higher for time to peak and MTT). Syngo® Volume Perfusion-CT Neuro supports the physician in identifying areas of decreased perfusion which indicate the occurrence of acute stroke during the first 6 hours after onset of symptoms. A second application is the visualization of the permeability. It is used for the modeling of extra-vascular leakage of blood into the interstitial space. This additional capability can display blood brain barrier disturbances and thus may improve the differential diagnosis of brain tumors and be helpful in therapy monitoring.
Device Story
Post-processing software for Intel-based PCs; processes dynamic CT image data acquired from SOMATOM CT scanners following contrast bolus injection. Calculates cerebral blood flow (CBF), cerebral blood volume (CBV), mean transit time (MTT), bolus timing, and vascular permeability. Used by physicians in clinical settings to visualize perfusion deficits in acute stroke and blood-brain barrier disturbances in tumors. Output includes quantitative maps and time-attenuation curves; assists physicians in identifying ischemic tissue and monitoring therapy. Benefits include improved diagnostic assessment of stroke and tumor pathology.
Clinical Evidence
No clinical data provided. Safety and effectiveness supported by hazard analysis, software verification, and validation testing.
Technological Characteristics
Post-processing software package; runs on Intel-based PC platform; compatible with SOMATOM CT scanners. Performs mathematical modeling of contrast kinetics (CBF, CBV, MTT, permeability).
Indications for Use
Indicated for use by physicians to evaluate brain perfusion in patients, including those with suspected acute stroke (within 6 hours of symptom onset) or brain tumors, using dynamic CT data acquired after contrast media injection.
Regulatory Classification
Identification
A computed tomography x-ray system is a diagnostic x-ray system intended to produce cross-sectional images of the body by computer reconstruction of x-ray transmission data from the same axial plane taken at different angles. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
{1}------------------------------------------------
# 510(k) - Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR §807.92.
#### 1. GENERAL INFORMATION
#### 1. Device Name and Classification
syngo® Volume Perfusion-CT Neuro Product Name: Classification Name: Accessory to Computed Tomography System Classification Panel: Radiology CFR Section: 21 CFR §892.1750 Device Class: Class II 90 JAK Product Code:
## 2. lmporter/Distributor Establishment: Registration Number: 2240869
Siemens Medical Solutions, Inc. 51 Valley Stream Pkwy Malvern, PA 19355
### 3. Manufacturing Facility:
Siemens AG Medical Solutions Henkestrasse 127
D-91052 Erlangen, Germany
### Contact Person: 4.
Mr. Ralf Hofmann Regulatory Affairs Specialist Siemensstr.1; D-91301 Forchheim Phone: +49 9191 18-8170 Fax: +49 9191 18-9782
### Date of Preparation of Summary: Oct 12th 2007 5.
510(k)
2073238
Section 8
pg. 1 of 5
JAN - 3 2006
is being subr
{2}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the word "SIEMENS" in a distressed, textured font. The letters are bold and appear to have a rough, almost eroded surface, giving the word a vintage or weathered look. The text is horizontally aligned and fills most of the frame, making it the central focus of the image. The overall effect is a strong, industrial aesthetic.
Image /page/1/Picture/1 description: This image contains handwritten text and numbers. The text includes "K073238" which is likely a reference or identification code. Below that, "PG. 2 of 3" indicates a page number within a document, specifically page 2 of a 3-page document. Finally, "510(k)" is present, which is a reference to a section of the Food, Drug, and Cosmetic Act in the United States.
# ll. SAFETY AND EFFECTIVENESS INFORMATION SUPPORTING THE SUBSTANTIAL EQUIVALENCE DETERMINATION
#### General Safety and Effectiveness Concerns: 6.
The device labeling contains instructions for use and any necessary cautions and warnings, to provide for safe and effective use of the device.
Risk management is ensured via a hazard analysis, which is used to identify potential hazards. These potential hazards are controlled via software development, verification and validation testing. To minimize electrical, and radiation hazards, Siemens adheres to recognized and established industry practice and standards.
#### 7. Substantial Equivalence:
The syngo® Volume Perfusion-CT Neuro software package that is addressed in this premarket notification, is substantially equivalent to the following commercially available software package
| Manufacturer | Product | 510(k) | Clearance date |
|---------------------|---------------------|---------|----------------|
| 1. Siemens | syngo Perfusion-CT | K033832 | 12/23/2003 |
| 2. Siemens | syngo Perf. CT Body | K050867 | 04/14/2005 |
| 3. General Electric | CT Perfusion 4 | K052839 | 03/10/2006 |
#### 8. Device Description and Intended Use:
syngo® Volume Perfusion-CT Neuro is a post-processing software package, which runs on an Intel-based PC platform designed to post-process images acquired with SOMATOM CT scanners, which meet certain minimal requirements (i.e. Siemens Definition, Sensation 64, ... }. It is a package containing evaluation software that supports the evaluation of Dynamic CT data gathered after the injection of a compact bolus of contrast media, where the contrast media acts as a pure intravascular tracer.
The Siemens syngo® Volume Perfusion-CT Neuro software package has been designed to evaluate areas of brain perfusion. The software can calculate cerebral blood flow (CBF), cerebral blood volume (CBV), local bolus timing (i.e. time to start, time to peak), mean transit time (MTT), and vascular permeability (blood brain barrier disturbances) from sets of images or volumes reconstructed from continuously acquired CT data after the injection of contrast media. The package also allows the calculation of mirrored regions or volumes of interest and the visual inspection of time attenuation curves.
{3}------------------------------------------------
Image /page/2/Picture/0 description: The image shows the word "SIEMENS" in a bold, sans-serif font. The letters are textured with a grainy, speckled pattern, giving them a distressed or aged appearance. The word is presented in all capital letters and is horizontally aligned.
Image /page/2/Picture/1 description: The image shows a handwritten document with the text "K073238" at the top. Below a horizontal line, the text "Pg. 3 of 3" is written in a cursive style. The text "510(k)" is at the bottom of the image.
One clinical application is to visualize the apparent blood perfusion in brain tissue affected by acute stroke. Areas of decreased perfusion, as is observed in acute cerebral infarcts, appear as areas of changed signal intensity (lower for CBF and CBV, higher for time to peak and MTT).
Syngo® Volume Perfusion-CT Neuro supports the physician in identifying areas of decreased perfusion which indicate the occurrence of acute stroke during the first 6 hours after onset of symptoms.
A second application is the visualization of the permeability. It is used for the modeling of extra-vascular leakage of blood into the interstitial space. This additional capability can display blood brain barrier disturbances and thus may improve the differential diagnosis of brain tumors and be helpful in therapy monitoring.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo is circular and contains the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the circle is an abstract symbol that resembles three human profiles facing to the right, with flowing lines beneath them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN - 3 2008
Siemens AG Medical Solutions % Mr. Olaf Teichert Responsible Third Party Official TÜV SÜD America 1775 Old Hwy 8 NW. Ste 104 NEW BRIGHTON MN 55112-1891
Re: K073238
Trade/Device Name: syngo® Volume Perfusion-CT Neuro Regulation Number: 21 CFR 892.1750 Regulation Name: Computed tomography x-ray system Regulatory Class: II Product Code: JAK Dated: December 10, 2007 Received: December 14, 2007
Dear Mr. Teichert:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1796, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy Brogdon
Nancv C. Brogdon Director. Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
{6}------------------------------------------------
Indication for use
510(k) Number (if known):
K073238
Device Name:
syngo® Volume Perfusion-CT Neuro
The Siemens syngo® Volume Perfusion-CT Neuro software package has been designed to evaluate areas of brain perfusion. The software can calculate cerebral blood flow (CBF), cerebral blood volume (CBV), local bolus timing (i.e. time to start, time to peak), mean transit time (MTT), and vascular permeability (blood brain barrier disturbances) from sets of images or volumes reconstructed from continuously acquired CT data after the injection of contrast media. The package also allows the calculation of mirrored regions or volumes of interest and the visual inspection of time attenuation curves.
One clinical application is to visualize the apparent blood perfusion in brain tissue affected by acute stroke. Areas of decreased perfusion, as is observed in acute cerebral infarcts. appear as areas of changed signal intensity (lower for CBF and CBV, higher for time to peak and MTT).
Syngo® Volume Perfusion-CT Neuro supports the physician in identifying areas of decreased perfusion which indicate the occurrence of acute stroke during the first 6 hours after onset of symptoms.
A second application is the visualization of the permeability. It is used for the modeling of extra-vascular leakage of blood into the interstitial space. This additional capability can display blood brain barrier disturbances and thus may improve the differential diagnosis of brain tumors and be helpful in therapy monitoring.
Prescription Use × Over-The-Counter Use AND/OR (Part 21 CFR 801 Subpart D) (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Nancy Brogdon
Division Sign-Off
Division of Reproductive, Abdominal, and
Radiological Devices
510(k) Number K073238
Section 2
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.