NEOCOIL 3.0T 32-CHANNEL TORSO ARRAY COIL FOR GE

K073236 · Neocoil · MOS · Nov 30, 2007 · Radiology

Device Facts

Record IDK073236
Device NameNEOCOIL 3.0T 32-CHANNEL TORSO ARRAY COIL FOR GE
ApplicantNeocoil
Product CodeMOS · Radiology
Decision DateNov 30, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1000
Device ClassClass 2
Attributes3rd-Party Reviewed

Intended Use

The NeoCoil 3.0T 32-Channel Torso Array Coil is a receive only phased array RF coil used to produce diagnostic images of the chest, abdomen, and pelvic area in Magnetic Resonance Imaging systems that can be interpreted by a trained physician. The NeoCoil 3.0T 32-Channel Torso Array Coil is designed for use with the 3.0T HD Series MR System manufactured by General Electric Healthcare (GEHC).

Device Story

Multi-element phased array receive-only RF coil; captures MR signals from chest, abdomen, and pelvic regions. Comprises two antenna arrays: posterior (17 elements, 16 channels, rigid housing) and anterior (17 elements, 16 channels, flexible housing). Wrapped around patient torso using hook and loop straps. Used in clinical MRI settings; operated by trained radiology staff. Output consists of raw RF signal data processed by GE 3.0T HD Series MR System to produce diagnostic images. Physician interprets images to aid clinical decision-making. Benefits include high-channel count imaging for improved diagnostic clarity in torso applications.

Clinical Evidence

Bench testing only. No clinical data provided. Testing demonstrated that the device design met predetermined acceptance criteria.

Technological Characteristics

Receive-only phased array RF coil. 34 total antenna elements (17 posterior, 17 anterior). 32-channel output. Rigid posterior housing; flexible anterior housing. Hook and loop strap fixation. Compatible with 3.0T GE HD series MR systems. Non-invasive, passive sensing device.

Indications for Use

Indicated for 2D and 3D magnetic resonance imaging and parallel body imaging of the abdomen, pelvis, and chest regions on 3.0T GE HD series Magnetic Resonance scanners.

Regulatory Classification

Identification

A magnetic resonance diagnostic device is intended for general diagnostic use to present images which reflect the spatial distribution and/or magnetic resonance spectra which reflect frequency and distribution of nuclei exhibiting nuclear magnetic resonance. Other physical parameters derived from the images and/or spectra may also be produced. The device includes hydrogen-1 (proton) imaging, sodium-23 imaging, hydrogen-1 spectroscopy, phosphorus-31 spectroscopy, and chemical shift imaging (preserving simultaneous frequency and spatial information).

Special Controls

*Classification.* Class II (special controls). A magnetic resonance imaging disposable kit intended for use with a magnetic resonance diagnostic device only is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 892.9.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K073236 # 510(k) Summary | Proprietary Name: | NeoCoil 3.0T 32-Channel Torso Array Coil | |-----------------------|------------------------------------------------------------------------------------------------------------------| | Common Device Name: | Magnetic Resonance Specialty Coil | | Regulation Number: | 892.1000 | | Review Panel: | Radiology Device Panel | | CDRH Product Code: | MOS | | Device Class: | II | | Reason for 510(k): | New device | | Applicant: | NeoCoil<br>N27 W23910A Paul Rd<br>Pewaukee, WI 53072 | | Contact | Brian Brown<br>Executive Director<br>262-347-1250 x 12 (office)<br>261-347-1251 (fax)<br>brian.brown@neocoil.com | | Preparation date: | 04/02/2007 | | Est. Registration No: | | NOV 3 0 2007 | Intended Use: | The NeoCoil 3.0T 32-Channel Torso Array Coil is a receive only phased array RF coil used to produce diagnostic images of the chest, abdomen, and pelvic area in Magnetic Resonance Imaging systems that can be interpreted by a trained physician. The NeoCoil 3.0T 32-Channel Torso Array Coil is designed for use with the 3.0T HD Series MR System manufactured by General Electric Healthcare (GEHC). | |--------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | Indications for Use | The NeoCoil 3.0T 32-Channel Torso Array Coil is intended to be used in the abdomen, pelvis, and chest regions for 2D and 3D magnetic resonance imaging and parallel body imaging for use on 3.0T GE HD series Magnetic Resonance scanners. | | Performance Standards: | No applicable performance standards have been issued under Section 514 of the Food, Drug and Cosmetic Act. | | Device Description: | The NeoCoil 3.0T 32-Channel Torso Array Coil is a multi-element phased array receive only coil used for obtaining diagnostic images of the chest, abdomen, and pelvic area in Magnetic Resonance Imaging Systems. The coil consists of 2 antenna arrays; a posterior and an anterior. The posterior array contains 17 elements to achieve 16 channels and is enclosed in a rigid housing. The anterior array also contains 17 elements to achieve the later 16 channels but is enclosed in a flexible housing. The coil is wrapped around the patient's torso with hook and loop straps. The NeoCoil 3.0T 32-Channel Torso Array Coil is for use with the GE 3.0T HD Series MR System (K052293) | | Predicate Devices: | GE 3.0T Torso Phased Array Coil (K030495) GE 1.5T 8 Channel Torso Coil (K031209) | | Comparison to Predicate: | The NeoCoil 3.0T 32-Channel Torso Array Coil is most similar to the GE 1.5T 8 Channel Torso Coil (K031209) with the main differences being the number of elements and the field strength. The NeoCoil Torso Coil uses 34 | {1}------------------------------------------------ elements compared to 12. It also operates at 3.0T like the GE 3.0T Torso Phased Array Coil (K030495) instead of 1.5T. Testing was performed to demonstrate to that the design of the NeoCoil 3.0T Summary of Studies: 32-Channel Torso Array Coil met predetermined acceptance criteria. It is the opinion of NeoCoil that the NeoCoil 3.0T 32-Channel Torso Array Conclusion: Coil is substantially equivalent to the predicate devices. Use of the NeoCoil 3.0T 32-Channel Torso Array Coil does not result in any new potential hazards and does not alter the safety of the MRI scanner. {2}------------------------------------------------ ### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized eagle with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular pattern around the eagle. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 ## NOV 3 0 2007 NeoCoil % Mr. Daniel W. Lehtonen Responsible Third Party Official Intertek Testing Services 2307 E. Aurora Rd., Unit B7 TWINSBURG OH 44087 Re: K073236 Trade/Device Name: NeoCoil 3.0T 32-Channel Torso Array Coil Regulation Number: 21 CFR 892.1000 Regulation Name: Magnetic resonance diagnostic device Regulatory Class: II Product Code: MOS Dated: November 15, 2007 Received: November 16, 2007 Dear Mr. Lehtonen We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device ean be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ### Indications for Use K073236 510(k) Number (if known): __ Device Name: NeoCoil 3.0T 32-Channel Torso Array Coil Indications For Use: The NeoCoil 3.0T 32-Channel Torso Array Coil is intended to be used in the abdomen, pelvis, and chest regions for 2D and 3D magnetic resonance imaging and parallel body imaging for use on 3.0T GE HD series Magnetic Resonance scanners. Prescription Use × (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Page 1 of 1 lomikhing (Division Sign Off) (Division Sign-Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number Page 24 of 112
Innolitics
510(k) Summary
Decision Summary
Classification Order
Enter a record ID and click Load to view the document.
100%