← Product Code [HGM](/productcode/HGM) · K073221

# CADENCE II (K073221)

_Edan Instruments, Inc. · HGM · Dec 28, 2007 · Obstetrics/Gynecology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K073221

## Device Facts

- **Applicant:** Edan Instruments, Inc.
- **Product Code:** [HGM](/productcode/HGM.md)
- **Decision Date:** Dec 28, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 884.2740
- **Device Class:** Class 2
- **Review Panel:** Obstetrics/Gynecology

## Indications for Use

The CADENCE II fetal monitor is used to monitor fetal well being during the antepartum period what is commonly called the non stress test. It is to be used by trained medical personnel in hospitals, clinics, physicians offices and in the patients home by prescription or doctors orders.

## Device Story

CADENCE II is a perinatal monitoring system for antepartum fetal assessment. It collects physiological signals including fetal heart rate (FHR1, FHR2), uterine activity (TOCO, IUP), fetal movement (FM, AFM), and direct fetal ECG (DECG). The device processes these inputs to display real-time data, trends, and monitoring parameters on a 5.7" LCD screen. A built-in thermal recorder provides hard-copy documentation. Used in hospitals, clinics, and home settings by trained medical personnel or patients under prescription. Output allows clinicians to evaluate fetal well-being during non-stress tests, facilitating clinical decision-making regarding fetal health.

## Clinical Evidence

No clinical data provided; substantial equivalence is supported by design control activities, safety risk analysis, and verification/validation testing.

## Technological Characteristics

Perinatal monitoring system; 5.7" LCD display; built-in thermal recorder. Inputs: ultrasound (FHR), tocodynamometer (TOCO), remote marker (FM), automatic fetal movement (AFM), DECG, and IUP. Connectivity: standard monitoring interfaces. Sterilization: N/A (non-invasive/external sensors).

## Regulatory Identification

A perinatal monitoring system is a device used to show graphically the relationship between maternal labor and the fetal heart rate by means of combining and coordinating uterine contraction and fetal heart monitors with appropriate displays of the well-being of the fetus during pregnancy, labor, and delivery. This generic type of device may include any of the devices subject to §§ 884.2600, 884.2640, 884.2660, 884.2675, 884.2700, and 884.2720. This generic type of device may include the following accessories: Central monitoring system and remote repeaters, signal analysis and display equipment, patient and equipment supports, and component parts.

## Predicate Devices

- Cadence ([K040903](/device/K040903.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

# Attachment 6

B.

## 510(K) SUMMARY

DEC 2 8 2007

## CADENCE II

#### Submitter's Information A.

| Name:                   | Edan Instruments, Inc             |
|-------------------------|-----------------------------------|
| Address:                | 3/F - B, Nanshan Medical          |
|                         | Equipments Park, Nanhai Rd 1019#, |
|                         | shekou, Nanshan Shenzhen,         |
|                         | 518067 P. R. China                |
| Phone:                  | 86-755-26892220                   |
| Fax:                    | 86-755-26882223                   |
| Contact Person:         | Jiang Yucai                       |
| Official Correspondent: | William Stern                     |
| Date Summary Prepared:  | Oct 17, 2007                      |

Device Information

| Trade/Device Name: | CADENCE II |
|--------------------|------------|
| Regulation Number: | 884.2740   |

Classification Name: System, Monitoring, Perinatal Regulation Class: Class II Product Code: HGM Classification Panel: Obstetrics/Gynecology

### Description of Device

The CADENCE II Fetal Monitor can provide different configurations according to different user requirements, FHR1 (US1), FHR2 (US2), TOCO, FM (remote marker), AFM (automatic fetal movement mark), fetal stimulator (optional), DECG (direct fetal ECG, optional), and IUP

{1}------------------------------------------------

(Intra-uterine Pressure, optional). The user can select the monitors according to requirements.

CADENCE II adopts 5.7" LCD, and the collected data, trends, and monitoring parameters are displayed at the same screen. A built-in thermal recorder is used to record the monitoring information.

#### C. Predicate Device Information

Cadence [K040903] [09/02/2004]

#### D. Indications for Use/Intended Use

The CADENCE II fetal monitor is used to monitor fetal well being during the antepartum period what is commonly called the non stress test. It is to be used by trained medical personnel in hospitals, clinics, physicians offices and in the patients home by prescription or doctors orders.

#### E. Substantial Equivalence

1 ls the product a device? YES-The CADENCE II is a device.

2. Does the new device have the same intended use? YES-The intended use for the CADENCE II is equivalent to that for the Cadence and is as follows:

### Intended Use: Cadence

The Cadence fetal monitor is used to monitor fetal well being during the antepartum period what is commonly called the non stress test. It is to be used by trained medical personnel in hospitals, clinics, physicians offices and in the patients home by prescription or doctors orders.

3. Does the device have technological characteristics the raise new types of safety or effectiveness questions?

NO- The technological characteristics of the CADENCE II raise no new types of safety or effectiveness questions.

4. Does descriptive or performance information demonstrate equivalence?

YES- Edan Instruments, Inc. believes that the information provided

{2}------------------------------------------------

in this submission clearly describes the CADENCE II and demonstrates that it is substantially equivalent to the predicate device Cadence

#### H. Safety Summary

Edan Instruments, Inc. made several modifications to the Cadence cleared under K040903. All design control activities including safety risk analysis and the verification and validation activities conducted as related to the risks proved that the modified CADENCE II is substantially equivalent in intended use, design, principle of operations, performance, and contains the same fundamental scientific technology as the Cadence.

{3}------------------------------------------------

### DEPARTMENT OF HEALTH & HUMAN SERVICES

Image /page/3/Picture/1 description: The image shows the seal of the Department of Health & Human Services (HHS) of the United States. The seal features the department's name encircling a stylized emblem. The emblem consists of three abstract shapes resembling human figures or birds in flight, arranged in a stacked formation.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

DEC 2 8 2007

Edan Instruments, Inc. % Mr. William Stern Multigon Industries, Inc. 1 Odell Plaza YONKERS NY 10701

Re: K073221

Trade/Device Name: CADENCE II Regulation Number: 21 CFR 884.2740 Regulation Name: Perinatal monitoring system and accessories Regulatory Class: II Product Code: HGM Dated: November 7, 2007 Received: December 5, 2007

### Dear Mr. Stern:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{4}------------------------------------------------

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the est on ic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Crant on the of the following numbers, based on the regulation number at the top of this letter.

| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology)          | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology)                      | 240-276-0120 |
| Other           |                                  | 240-276-0100 |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, recuract CDRH s Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 24/0276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.btml

Sincerely yours.

Nancy C. Brogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{5}------------------------------------------------

## Attachment 2

# Indications for Use Statement

510 (k)

Number

(if known)

Kot 32a/

CADENCE II Device Name

The CADENCE II fetal monitor is used to monitor fetal well Indications for Use being during the antepartum period what is commonly called the non stress test. It is to be used by trained medical personnel in hospitals, clinics, physicians offices and in the patients home by prescription or doctors orders.

PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED

Concurrence of CDRH, Office of Device Evaluation (ODE)

Prescription Use YES (Per 21 CFR 801 Subpart D) OR

Over-The-Counter Use _ NO

tony m whay

(Division S Division of Reproductive, Abdominal and Radiological Devices 510(k) Number

---

**Source:** [https://fda.innolitics.com/device/K073221](https://fda.innolitics.com/device/K073221)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
