K073182 · Gyn-Disposables, Inc. · HDC · Nov 28, 2007 · Obstetrics/Gynecology
Device Facts
Record ID
K073182
Device Name
GYN DISPOSABLES TENACULUM 356T
Applicant
Gyn-Disposables, Inc.
Product Code
HDC · Obstetrics/Gynecology
Decision Date
Nov 28, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 884.4530
Device Class
Class 2
Attributes
3rd-Party Reviewed
Indications for Use
Snare, grasp, hold and manipulate cervical and intravaginal tissue.
Device Story
Manual surgical instrument; glass-reinforced polycarbonate construction; hook-like design with two tips, handle with finger holes, and ratcheting lock. Used by physicians in clinical settings to seize and hold cervical or fundal tissue during OB/GYN procedures. Single-use, disposable device; supplied sterile via gamma radiation. Provides mechanical grasping and manipulation of tissue; replaces reusable stainless steel tenacula. Benefits include elimination of cross-contamination risks and reprocessing requirements.
Clinical Evidence
Bench testing only. Performance testing evaluated arm flexibility, tip strength, rack/lock force, and hinge pin strength against acceptance criteria. Biocompatibility testing included seven tests: in vitro cytotoxicity, rabbit pyrogen test, hemolysis, guinea pig dermal sensitization, mouse systemic toxicity, and rabbit intracutaneous/vaginal irritation. All tests passed, meeting FDA G95-1 requirements for Category A (<24 hours) external communicating devices.
Technological Characteristics
Material: Glass-reinforced polycarbonate (RAPRA 43C12/6272). Form factor: Manual instrument with two tips, finger-hole handle, and ratcheting lock. Energy source: None (manual). Sterilization: Gamma radiation per ANSI/AAMI/ISO 1137-2:2006. Single-use disposable.
Indications for Use
Indicated for snaring, grasping, holding, and manipulating cervical and intravaginal tissue during obstetric/gynecologic procedures.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
{1}------------------------------------------------
#### 510(k) Summary
#### Submitted By:
GYN Disposables, Inc. 2671 Appling Road Memphis, TN 38133 Tel: 901.372.6181 Fax: 901.213.0957
NOV 2 8 2007
1
Contact: James K. Patterson, MD, MBA
Device Classification Name: Tenaculum, Uterine
Device Name: GYN Disposables, Inc. Tenaculum 356T
Classification: 884.4530 Obstetric-gynecologic specialized manual instrument ... (15) A Uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus.
Device Class: Class II
Classification Panel: Ob/Gyn Device Panel
Classification Product Code: HDC
Establishment Registration Number: 3006365741
Indication for Use: To snare, grasp, hold and manipulate cervical and intravaginal tissue.
Device Description: GYN Disposables Tenaculum 356T is made of glass reinforced polycarbonate. The GYN Disposable Tenaculum is supplied in a Tyvek pouch and has been sterilized by gamma radiation.
{2}------------------------------------------------
Substantial Equivalence: Atraumatic Tenacula (K930037) and Gynex Iris Hook, Gynex Angle Hook, Gynex Emmett Tenaculum (K980247) are predicate devices.
| | GYN Disposables | Thomasville Medical | Gynex Corp<br>Iris Hook<br>Angle Hook |
|------------------------|-------------------------------------------------------------------------------------------------------|-------------------------------------------------------------------------------------------------------|-----------------------------------------------------------------------------|
| | Tenaculum 356T | Atraumatic Tenaculum | Emmett Tenaculum |
| 510(K) Number | N/A | K930037 | K980247 |
| Product Code | HDC | HDC | HDC |
| Indications for<br>use | Snare, grasp, hold and<br>manipulate<br>cervical and intravaginal<br>tissue | Snare, grasp, hold and<br>manipulate<br>cervical and intravaginal<br>tissue | Snare, grasp, hold and<br>manipulate<br>cervical and intravaginal<br>tissue |
| Design | Two Tips grasp cervical and<br>intravaginal tissue<br>Handle with two finger holes<br>Ratcheting lock | Two Tips grasp cervical and<br>intravaginal tissue<br>Handle with two finger holes<br>Ratcheting lock | Single tip grasp cervical and<br>intravaginal tissue<br>Single Rod |
| Material | Glass Reinforced Polycarbonate | 302 Stainless Steel | 302 Stainless Steel |
| Sterilization | Purchased Sterile (gamma) | Clean only | Clean only |
| Single use | Yes | No | No |
Material Specifications: GYN Disposables Tenaculum 356T is made with glass reinforced polycarbonate. The advanced composite design can easily withstand forces during normal use as evidenced by performance testing.. In these reports non sterile, gamma sterilized non-aged, and gamma sterilized aged product were tested in four ways to determine strength and weakness in their construction. The first report determined the flexibility and the strength of the arms, the second determined the strength of the tips, and the third report determined the strength of the rack and the force needed to lock the instrument during use. The final report determined the strength of the hinge pin. All test reports showed that the instrument met the minimum acceptance criteria. These tests proved that the material does not degrade because of gamma sterilization and extended shelf life. Another test was done to determine the maximum perceived force during normal use by a physician using a stainless steel tenaculum.
Biocompatibility: The material used by GYN Disposbables, Inc. to make the subject device is a glass-reinforced polycarbonate. This material is made into an OB/GYN instrument called a Tenaculum and is gamma sterilized. The material used is classified by the Rubber And Plastic Research Association (RAPRA) as 43C12 for polycarbonate and 6272 as glass-reinforced. Even though this material or its components have been used in other medical applications and these instruments are only used once and only briefly contacts the patient, GYN Disposables has carried out seven biocompatibility tests on the as-molded, as-sterilized finished instrument.
{3}------------------------------------------------
The first test was an in vitro cytotoxicity test using the following method: the test article was put into the extraction media for not less than 24 hours. The media was incubated for 48 hours and examined for biological activity. Relative to positive and negative controls, the test article did not induce cytotoxicity.
The second test was a similar test except that the extraction was intravenously injected into rabbits. The temperatures of the rabbits were measured for up to 3 hours. This test, called a pyrogen test, in essence is a sensitization and irritation indicator for use of the instrument. These tests found that the as-made instrument did not cause any increase in temperature.
The third test was a hemolysis test. Extracts from the instrument were again made and combined with fresh, whole, rabbit blood using a process that takes approximately 2 hours. The average % hemolysis was determined to be 0.0%, meaning that the extract did not cause damage to blood cells.
In addition to these three tests, four in vivo tests of toxicity were also performed-dermal sensitization in a guinea pig, systemic toxicity in a mouse, and two types of irritation testing (one intracutaneous [between the layers of the skin] and one vaginal) both in rabbits. In all four tests the GYN material passed. Duration of contact of tissue for cytotoxicity, and systemic toxicity was 24, 48, and 72 hours. Contact for Intercutaneous irritation test was 72 hours, vaginal irritation 24 hours for five consecutive days, and systemic toxicity for 1, 1.5, 2, 2.5, and 3 hours.
These seven tests meet and exceed the criteria established in Table 1 in the FDA's 510(k) Memorandum #G95 (dated May 1, 1995) as follows. The disposable Tenaculum instrument has minimal contact duration with the patient and therefore meets Category A (<24hours). It also can worse case be considered an "External Communicating Device" since it is used in the cervix and it also worse case can be considered a "Blood Path, Indirect" product since the cervix is often a bloody field during OB/GYN surgery. According to Table 1, these criteria require cytotoxicity, sensitization, system toxicity (acute), and hemocompatibility testing. The testing provided are sufficient to assure the passage of the toxicology testing for this device for the brief and one-time intended use. These tests confirm that the manufacturing and sterilization procedures used by GYN do not affect the underlying biocompatibility of the material used.
Sterilization: Product will be sold sterile using gamma radiation. Sterilization validation was performed by Steris in conjunction with Nelson Laboratories. This report follows the ANSI/AAMI/ISO 1137-2:2006 guidelines to perform product qualification for irradiation sterilization of a health care product. Initial verification dosing, bioburden, bacteriostasis/fungistis, and sterility testing were done and acceptance criteria were met. Accelerated aging validation for one year shelf life has been performed by Ethox using sterilized samples. The product packaging met the requirements of the burst test at the non-aged and 1 yr accelerated aged time point. Product packaging met the requirements of the dye penetration test both non-aged and 1 yr aged.
{4}------------------------------------------------
Packaging: Tyvek
.
and the comments of the comments of the comments of the comments of
and the contraction of the comments of the comments of the comments of
{5}------------------------------------------------
## DEPARTMENT OF HEALTH & HUMAN SERVICES
Image /page/4/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" around the perimeter. Inside the circle is a stylized eagle emblem with three parallel lines forming the body and wings.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
### NOV 2 8 2007
GYN Disposables, Inc. c/o Mr. Robert Mosenkis President CITECH Medical Device Testing and Consulting 5200 Butler Pike PLYMOUTH MEETING PA 19462-1298
Re: K073182
Trade/Device Name: GYN Disposables Tenaculum 356T Regulation Number: 21 CFR §884.4530 Regulation Name: Obstetric-gynecologic specialized manual instrument Regulatory Class: II Product Code: HDC Dated: November 12, 2007 Received: November 13, 2007
Dear Mr. Mosenkis:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{6}------------------------------------------------
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter.
| 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 |
|-----------------|----------------------------------|--------------|
| 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 892.xxxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Nancy C Brogdon
Nancy C. B rogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
# Indications for Use
K013182 510(k) Number (if known):
Device Name: GYN Disposables Tenaculum 356T
Indications For Use: Snare, grasp, hold and manipulate cervical and intravaginal tissue.
Prescription Use (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Hebert Lewis
(Division Sign-Off) Division of Reproductive, Abdominal and Radlological Devices 510(k) Number
Page 1 of 1
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.