ANKYLOS DENTAL IMPLANT SYSTEM/CLAIMS

K073067 · Dentsply International, Inc. · DZE · Feb 15, 2008 · Dental

Device Facts

Record IDK073067
Device NameANKYLOS DENTAL IMPLANT SYSTEM/CLAIMS
ApplicantDentsply International, Inc.
Product CodeDZE · Dental
Decision DateFeb 15, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 872.3640
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The ANKYLOS® Dental Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple unit prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional 2 stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading. Immediate loading is restricted to the anterior mandible, based on 4 interforaminal placed implants, and not indicated for single, unsplinted implants. Patients must be subject for dental treatment with endosseus implant.

Device Story

ANKYLOS® Dental Implant System is an endosseous dental implant used by dental professionals to restore patient chewing function. The system provides a root-form foundation for prosthetic appliances. Implants are surgically placed in the upper or lower jaw arches using either a conventional two-stage process (with optional transmucosal healing) or a single-stage process for immediate loading. Immediate loading is specifically restricted to the anterior mandible, requiring a minimum of four interforaminal implants; it is contraindicated for single, unsplinted implants. The device functions as a mechanical anchor for dental prosthetics. No changes were made to the device design or materials for this submission; the application solely sought clearance for additional marketing claims regarding existing special features.

Clinical Evidence

No clinical data provided. The submission relies on the established safety and effectiveness of the previously cleared ANKYLOS® Dental Implant System (K041509) and its marketing history.

Technological Characteristics

Endosseous dental implant system. Components are identical to those in the legally marketed predicate device. No new materials or technological changes introduced.

Indications for Use

Indicated for patients requiring dental treatment with endosseous implants in the upper or lower jaw arches for single or multiple unit prosthetic attachment to restore chewing function. Immediate loading is restricted to the anterior mandible using 4 interforaminal implants; not indicated for single, unsplinted implants.

Regulatory Classification

Identification

An endosseous dental implant is a prescription device made of a material such as titanium or titanium alloy that is intended to be surgically placed in the bone of the upper or lower jaw arches to provide support for prosthetic devices, such as artificial teeth, in order to restore a patient's chewing function.

Special Controls

*Classification.* (1) Class II (special controls). The device is classified as class II if it is a root-form endosseous dental implant. The root-form endosseous dental implant is characterized by four geometrically distinct types: Basket, screw, solid cylinder, and hollow cylinder. The guidance document entitled “Class II Special Controls Guidance Document: Root-Form Endosseous Dental Implants and Endosseous Dental Implant Abutments” will serve as the special control. (See § 872.1(e) for the availability of this guidance document.)(2) *Classification.* Class II (special controls). The device is classified as class II if it is a blade-form endosseous dental implant. The special controls for this device are:(i) The design characteristics of the device must ensure that the geometry and material composition are consistent with the intended use; (ii) Mechanical performance (fatigue) testing under simulated physiological conditions to demonstrate maximum load (endurance limit) when the device is subjected to compressive and shear loads; (iii) Corrosion testing under simulated physiological conditions to demonstrate corrosion potential of each metal or alloy, couple potential for an assembled dissimilar metal implant system, and corrosion rate for an assembled dissimilar metal implant system; (iv) The device must be demonstrated to be biocompatible; (v) Sterility testing must demonstrate the sterility of the device; (vi) Performance testing to evaluate the compatibility of the device in a magnetic resonance (MR) environment; (vii) Labeling must include a clear description of the technological features, how the device should be used in patients, detailed surgical protocol and restoration procedures, relevant precautions and warnings based on the clinical use of the device, and qualifications and training requirements for device users including technicians and clinicians; (viii) Patient labeling must contain a description of how the device works, how the device is placed, how the patient needs to care for the implant, possible adverse events and how to report any complications; and (ix) Documented clinical experience must demonstrate safe and effective use and capture any adverse events observed during clinical use.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ 510(k) SUMMARY K073067 DENTSPLY International Susquehanna Commerce Center West 221 West Philadelphia Street, Suite 60 York, PA 17405-0872 | CONTACT: | Helen Lewis | |----------------------------|---------------------------------------------| | DATE PREPARED: | October 29, 2007 | | TRADE OR PROPRIETARY NAME: | ANKYLOS® Dental Implant System/Claims | | CLASSIFICATION NAME: | Endosseous dental implant (21 CFR 872.3640) | | PREDICATE DEVICES: | ANKYLOS® Dental Implant System K041509 | FEB 15DEVICE DESCRIPTION: The ANKYLOS® Dental Implant System has been cleared for commercial distribution. The purpose of this application is to obtain clearance for additional marketing claims associated with the special features of the ANKYLOS® Dental Implant System. ## INTENDED USE: The ANKYLOS® Dental Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple unit prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional 2 stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading. Immediate loading is restricted to the anterior mandible, based on 4 interforaminal placed implants, and not indicated for single, unsplinted implants. Patients must be subject for dental treatment with endosseus implant. TECHNOLOGICAL CHARACTERISTICS: All of the components found in the ANKYLOS® Dental Implant System have been used in legally marketed devices and/or were found safe for dental use. No changes have been made to the ANKYLOS® Dental Implant System. Therefore, it was determined that no additional biocompatibility testing was necessary. We believe that the performance data provided and the research and development, and marketing history support the safety and effectiveness of the additional marketing claims for the special features of the unchanged device of the ANKYLOS® Dental Implant System. {1}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is a staff with two snakes coiled around it, representing medicine and healing. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" is arranged in a circular pattern around the caduceus. The logo is black and white. FEB 15 2008 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Helen Lewis Director of Corporate Compliance and Regulatory Affairs DENTSPLY International, Incorporated Susquehanna Commerce Center West 221 West Philadelphia Street, Suite 60 York, Pennsylvania 17405-0872 Re: K073067 Trade/Device Name: ANKYLOS® Dental Implant System/Claims Regulation Number: 21 CFR 872.3640 Regulation Name: Endosseous Dental Implant Regulatory Class: II Product Code: DZE, NHA Dated: February 8, 2008 Received: February 12, 2008 Dear Ms, Lewis: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {2}------------------------------------------------ Page 2 - Ms. Lewis Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Suette Y. Michie MD Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health {3}------------------------------------------------ /cts K073067 ## INDICATIONS FOR USE 510(k) Number (if known): Device Name: ANKYLOS® Dental Implant System/Claims Indications for Use: The ANKYLOS® Dental Implant System is indicated for surgical placement in the upper and lower jaw arches, to provide a root form means for single or multiple unit prosthetic appliance attachment to restore a patient's chewing function. Implants can be placed with a conventional 2 stage surgical process with an option for transmucosal healing or they can be placed in a single stage surgical process for immediate loading. Immediate loading is restricted to the anterior mandible, based on 4 interforminal placed implants, and not indicated for single, unsplinted implants. Patients must be subject for dental treatment with endosseus implant. Prescription Use _ X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Susan Runne (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: KO23067
Innolitics

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...