TOPCON 3D OCT-1000 OPTICAL COHERENCE TOMOGRAPHY SYSTEM FOR MEASUREMENT OF RETINAL THICKNESS

K072971 · Topcon Medical Systems, Inc. · OBO · Jan 13, 2009 · Ophthalmic

Device Facts

Record IDK072971
Device NameTOPCON 3D OCT-1000 OPTICAL COHERENCE TOMOGRAPHY SYSTEM FOR MEASUREMENT OF RETINAL THICKNESS
ApplicantTopcon Medical Systems, Inc.
Product CodeOBO · Ophthalmic
Decision DateJan 13, 2009
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 886.1570
Device ClassClass 2

Indications for Use

The Topcon 3D OCT-1000 is a non-contact ophthalmic imaging system for the viewing and axial cross sectional imaging of posterior ocular structures. It is used for in vivo imaging of the retina, retinal nerve fiber laver and optic disc. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases, including but not limited to macular edema and central serous retinopathy. The device is indicated for assessing the area, location, and measurement of retinal thickness, including in patients with retinal pathologies. In addition, the device is indicated to detect separation between identified retinal layers and surfaces, i.e., retinal tissue layers.

Device Story

Non-contact ophthalmic imaging system; uses optical coherence tomography (OCT) via superluminescent diode light interferometry; captures cross-sectional images of posterior ocular structures; software module calculates retinal thickness; used in clinical settings by eye care professionals; output provides visualization and quantitative measurements of retinal layers/thickness; aids diagnosis and management of ocular diseases (e.g., macular edema, central serous retinopathy); benefits patients through objective assessment of retinal pathology and tissue layer separation.

Clinical Evidence

Bench testing only. Accuracy of automated retinal thickness measurements validated by correlating device outputs with manually identified retinal boundaries in populations with healthy and diseased retinas.

Technological Characteristics

Optical coherence tomography (OCT) system; uses superluminescent diode light source; interferometry-based sensing; software module for retinal thickness measurement; non-contact form factor.

Indications for Use

Indicated for in vivo viewing, axial, cross-sectional, and 3D imaging and measurement of posterior ocular structures (retina, retinal nerve fiber layer, macula, optic disc) in patients with ocular diseases affecting the posterior segment, including retinal pathologies, macular edema, and central serous retinopathy.

Regulatory Classification

Identification

An ophthalmoscope is an AC-powered or battery-powered device containing illumination and viewing optics intended to examine the media (cornea, aqueous, lens, and vitreous) and the retina of the eye.

Special Controls

*Classification.* Class II (special controls). The device, when it is an AC-powered opthalmoscope, a battery-powered opthalmoscope, or a hand-held ophthalmoscope replacement battery, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 886.9.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ ## JAN 1 3 2009 #### K072971 510(k) SUMMARY ## Topcon Corporation's 3D OCT-1000 Optical Coherence Tomography System ## Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared Topcon Corporation 75-1 Hasunuma-cho, Itabashi-ku Tokyo, Japan 174 #### Additional Correspondents: Topcon Medical System, Inc. 37 West Century Road Paramus, NJ 07652 Phone: (201) 599-5153 Facsimile: (201) 599-5240 Contact Person: Maureen O'Connell, Regulatory Consultant OR Jonathan S. Kahan Hogan & Hartson, L.L.P. 555 Thirteenth Street, NW Washington, DC 20004 Phone: (202) 637-5794 Facsimile: (202) 637-5910 Date Prepared: October 19, 2007 ## Name of Device and Name/Address of Sponsor Topcon 3D OCT-1000 Optical Coherence Tomography System Topcon Corporation 75-1 Hasunuma-cho, Itabashi-ku Tokyo, Japan 174 18 {1}------------------------------------------------ ### Common or Usual Name AC- Powered Ophthalmoscope #### Classification Name Ophthalmoscope; 21 C.F.R. 886.1570 ## Predicate Devices Topcon Corporation's 3D OCT-1000 Optical Coherence Tomography System Carl Zeiss Meditec, Inc., STRATUSOCT™ with Retinal Nerve Fiber Layer and Macula Normative Database Talia Technology, Ltd., RTA 5 & RTA Model E Retinal Thickness Analyzer Talia Technology, Ltd., RTA Retinal Thickness Analyzer #### Intended Use / Indications for Use The Topcon 3D OCT-1000 is a non-contact ophthalmic imaging system for the viewing and axial cross sectional imaging of posterior ocular structures. It is used for in vivo imaging of the retina, retinal nerve fiber laver and optic disc. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases, including but not limited to macular edema and central serous retinopathy. The device is indicated for assessing the area, location, and measurement of retinal thickness, including in patients with retinal pathologies. In addition, the device is indicated to detect separation between identified retinal layers and surfaces, i.e., retinal tissue layers. ### Technological Characteristics The Topcon 3D OCT-1000 for Measurement of Retinal Thickness uses optical coherence tomography, which relies upon interferometry of superluminescent diode light reflected from the fundus of the eye to obtain cross-sectional images of the retina. The Topcon 3D OCT-1000 for Measurement of Retinal Thickness is identical to the FDA-cleared Topcon 3D OCT-1000 (K063388), with the exception of the addition of a new software module allowing for the measurement of retinal thickness. 19 {2}------------------------------------------------ #### Performance Data Topcon conducted performance testing demonstrating that from an analytical perspective, the device provides accurate measurements. Additional performance testing demonstrated that the accuracy of the measurements reported from the device correlate with measurements calculated using manually identified retinal boundaries in a population with healthy retinas and a population with diseased retinas. The performance data demonstrate that the Topcon 3D OCT-1000 for Measurement of Retinal Thickness is as safe and effective as the predicate devices, and thus, substantially equivalent. #### Substantial Equivalence The Topcon 3D OCT-1000 for Measurement of Retinal Thickness is as safe and effective as the cleared Topcon 3D OCT-1000 Optical Coherence Tomography System, the Carl Zeiss Meditec, Inc., STRATUSOCT™ with Retinal Nerve Fiber Laver and Macula Normative Database, the Talia Technology, Ltd., RTA 5 & RTA Model E Retinal Thickness Analyzer, and the Talia Technology, Ltd., RTA Retinal Thickness Analyzer. The Topcon 3D OCT-1000 for Measurement of Retinal Thickness has the same intended uses and similar indications, technological characteristics, and principles of operation as its predicate devices. The minor differences between the 3D OCT-1000 for Measurement of Retinal Thickness and its predicate devices raise no new issues of safety or effectiveness. Performance data demonstrate that the Topcon 3D OCT-1000 for Measurement of Retinal Thickness is as safe and effective as the predicate devices. Thus, the 3D OCT-1000 for Measurement of Retinal Thickness is substantially equivalent. 20 {3}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/3/Picture/1 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an emblem that resembles an eagle with three stylized lines representing its wings. #### Public Health Service JAN 1 3 2009 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 TopCon Medical Systems, Inc., c/o Jonathan Kahan Partner Hogan & Hartson, LLP Columbia Square 5555 Thirteen St. NW Washington, DC 20004 Re: K072971 Trade Name: TopCon 3D OCT-1000 for Measurement of Retinal Thickness Regulation Number: 21 CFR 886.1570 Regulation Name: Ophthalmoscope Regulatory Class: II Product Code: OBO Dated: December 19, 2008 Received: December 19, 2008 #### Dear Mr. Kahan: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. ﻟﻠﺴﻠﺔ {4}------------------------------------------------ Page 2 – Jonathan Kahan This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Marlina B. Esguerra, M.D. Malvina B. Evdelman. M.D. Director Division of Ophthalmic and Ear, Nose and Throat Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {5}------------------------------------------------ # Indications for Use ## 510(k) Number: K072971 Device Name: TOPCON 3D OCT-1000 for Measurement of Retinal Thickness Indications For Use: The Topcon 3D OCT-1000 is a non-contact, high resolution tomographic and biomicroscopic imaging device. It is indicated for in vivo viewing, axial, cross-sectional and three-dimensional imaging and measurements of posterior ocular structures, including retina, retinal nerve fiber layer, macula, and optic disc. It is intended for use as a diagnostic device to aid in the detection and management of ocular diseases affecting the posterior segment of the eye. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 807 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) A. Osin. (Division Sign-Off) Division of Ophthalmic Ear, Nose and Throat Devices ેને 510(k) Number K072971
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...