The HE4 EIA is an enzyme immunometric assay for the quantitative determination of HE4 in human serum. The assay is to be used as an aid in monitoring recurrence or progressive disease in patients with epithelial ovarian cancer. Serial testing for patient HE4 assay values should be used in conjunction with other clinical methods used for monitoring ovarian cancer.
Device Story
HE4 EIA is a solid-phase, non-competitive, manual enzyme immunoassay (EIA) for quantitative measurement of HE4 antigen in human serum. Input: patient serum sample. Principle: direct sandwich technique using two mouse monoclonal antibodies (2H5 and 3D8) targeting epitopes in the C-WFDC domain of HE4. Process: sample incubated with biotinylated 2H5 MAb in streptavidin-coated microstrips; washed; incubated with HRP-labeled 3D8 MAb; enzyme reaction with HRP substrate/chromogen (hydrogen peroxide and TMB) produces blue color proportional to HE4 concentration. Output: absorbance measured via microplate spectrophotometer at 620 nm; concentration determined via standard calibration curve. Used in clinical/hospital laboratories by technicians. Results assist clinicians in monitoring disease status (progression/recurrence) in epithelial ovarian cancer patients. Serial testing provides longitudinal data to track changes in HE4 levels, aiding clinical decision-making alongside other diagnostic methods.
Clinical Evidence
Retrospective clinical study of 80 women with epithelial ovarian cancer undergoing serial surveillance. Primary endpoint: compare changes in HE4 concentrations with clinical status (progression/non-progression) vs. CA 125. ROC analysis showed AUC 0.725 (HE4) vs 0.709 (CA 125), indicating non-inferiority. Sensitivity/specificity evaluated at various percentage change cut-offs; 25% change cut-off yielded 60.3% sensitivity and 75% specificity for progression. Additional cohort of 1150 specimens used to establish reference ranges and specificity in benign/malignant conditions.
Technological Characteristics
Solid-phase, non-competitive, manual enzyme immunoassay (EIA). Uses two mouse monoclonal antibodies (2H5, 3D8). Microplate spectrophotometer readout (620 nm). Reportable range 15-900 pM. Standard curve-based quantification. No software algorithm; manual data processing.
Indications for Use
Indicated for monitoring recurrence or progressive disease in patients with epithelial ovarian cancer. Used in conjunction with other clinical monitoring methods.
Regulatory Classification
Identification
A tumor-associated antigen immunological test system is a device that consists of reagents used to qualitatively or quantitatively measure, by immunochemical techniques, tumor-associated antigens in serum, plasma, urine, or other body fluids. This device is intended as an aid in monitoring patients for disease progress or response to therapy or for the detection of recurrent or residual disease.
Special Controls
*Classification.* Class II (special controls). Tumor markers must comply with the following special controls: (1) A guidance document entitled “Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA,” and (2) voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards.
K151378 — Lumipulse G HE4 Immunoreaction Cartridges, Lumipulse G HE4 Calibrators · Fujirebio Diagnostics,Inc. · Nov 24, 2015
Submission Summary (Full Text)
{0}------------------------------------------------
Image /page/0/Picture/1 description: The image contains the logo for "FUJIREBIO Diagnostics, Inc". The logo consists of a stylized graphic on the left, followed by the company name in bold, uppercase letters. Below the company name, the words "Diagnostics, Inc" are written in a smaller, non-bold font.
## 510(k) SUMMARY
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The assigned 510(k) number is: K072939
## Submitter Information
JUN - 9 2008
## Address:
Fujirebio Diagnostics, Inc. 201 Great Valley Parkway Malvern, PA 19355
Contact person:
Diana L. Dickson (610) 240-3917 FAX (610) 240-3803 dicksond@fdi.com
Summary preparation date: May 13, 2008
Name of Device
| Trade/Proprietary Name: | HE4 EIA Kit |
|-------------------------|-------------|
|-------------------------|-------------|
Common/Usual Name: · HE4 EIA Test Kit
21 CFR 866.6010 Requiation Number:
Regulatory Class: Class II
To be determined Product Code:
## Predicate Device
ABBOTT ARCHITECT CA 125 II Assay (K042731)
## Device Description
The HE4 EIA is a solid-phase, non-competitive immunoassay based upon the direct sandwich technique based on two mouse monoclonal antibodies, 2H5 and 3D8, directed against two epitopes in the C-WFDC domain of HE4. Calibrators, controls and patient samples are incubated together with biotinylated Anti-HE4 monoclonal antibody (MAb) 2H5 in streptavidin coated microstrips. HE4 present in calibrators or samples is adsorbed to the streptavidin coated microstrips by the biotinylated Anti-HE4 MAb during the incubation. The strips are then washed labeled Anti-HE4 MAb 3D8. After washing, buffered and incubated with HRP Substrate/Chromogen reagent (hydrogen peroxide and 3, 3', 5' tetra-methyl-benzidine) is added to each well and the enzyme reaction is allowed to proceed. During the enzyme reaction a blue color will develop if antigen is present. The intensity of the color is proportionate to the
{1}------------------------------------------------
Image /page/1/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left, followed by the company name in bold, black letters. Below the company name, the words "Diagnostics, Inc" are printed in a smaller, non-bold font.
amount of HE4 present in the samples. The color intensity is determined in a microplate spectrophotometer at 620 nm (or optionally at 405 nm after addition of Stop Solution),
Calibration curves are constructed for each assay by plotting absorbance value versus the concentration for each calibrator. The HE4 concentrations of patient samples are then read from the calibration curve.
## Reportable Range
The reportable range for the HE4 EIA Kit is 15 - 900 pM
## Intended Use
The HE4 ElA is an enzyme immunometric assay for the quantitative determination of HE4 in human serum. The assay is to be used as an aid in monitoring recurrence or progressive disease in patients with epithelial ovarian cancer. Serial testing for patient HE4 assay values should be used in conjunction with other clinical methods used for monitoring ovarian cancer.
## Statement of Substantial Equivalence
The HE4 EIA is an enzyme immunometric assay for the quantitative determination of HE4 in human serum. The assay is to be used as an aid in monitoring recurrence or progressive disease in patients with epithelial ovarian cancer. Serial testing for patient HE4 assay values should be used in conjunction with other clinical methods used for monitoring ovarian cancer.
As there is no FDA cleared or approved device for the HE4 EIA Kit, substantial equivalence for the HE4 EIA test kit was determined by comparing the clinical performance to the ARCHITECT CA 125 II assay as an aid in monitoring patients with epithelial ovarian cancer.
The requlatory submission will be prepared pursuant to Title 21CFR § 866.6010(b) which states Tumor Markers must comply with the following special controls;
- 1. Guidance document entitled "Guidance Document for the Submission of Tumor Associated Antigen Premarket Notifications (510(k)s) to FDA".
- 2. Voluntary assay performance standards issued by the National Committee on Clinical Laboratory Standards (NCCLS).
A comparison of the features of the HE4 EIA Kit and the ARCHITECT CA 125 II assay are as follows:
{2}------------------------------------------------
: 上
Image /page/2/Picture/1 description: The image shows the logo for FUJIREBIO Diagnostics, Inc. The logo consists of a stylized graphic to the left of the company name. The graphic is a combination of black and white shapes with a striped pattern. The text "FUJIREBIO" is in bold, uppercase letters, with "Diagnostics, Inc" in a smaller font size below it.
| | Similarities | |
|-------------------------------|--------------------------------------------------------------|--------------------------------------------------------------|
| | HE4 EIA Kit<br>(Proposed Device) | ARCHITECT CA 125 II<br>(Predicate Device)<br>K042731 |
| Device Type | In vitro diagnostic | In vitro diagnostic |
| Classification | Class II | Class II |
| Regulation Number | 21 CFR 866.6010 | 21 CFR 866.6010 |
| Product Usage | Clinical and Hospital laboratories | Clinical and Hospital laboratories |
| Intended Use | Aid in monitoring patients with<br>epithelial ovarian cancer | Aid in monitoring patients with<br>epithelial ovarian cancer |
| Specimen Collection<br>Method | Routine Phlebotomy Techniques | Routine Phlebotomy Techniques |
| Interpretation of Results | Standard Curve | Standard Curve |
| | Differences | |
|------------------------|-----------------------------------------------|------------------------------------------------------------------|
| | HE4 EIA Kit<br>(Proposed Device) | ARCHITECT CA 125 II<br>(Predicate Device)<br>K042731 |
| Product Code | To be determined | LTK |
| Type of Specimen | Human Serum Only | Human serum or plasma (EDTA,<br>Lithium Heparin, Sodium Heparin) |
| Antigen Detected | HE4 | CA 125 |
| Capture Antibody | 2H5 mouse monoclonal | OC 125 mouse monoclonal |
| Detection Antibody | 3D8 mouse monoclonal | M11 mouse monoclonal |
| Calibrators | 6 Levels (0 - 900 pM)<br>Supplied with Kit | 6 Levels (0 - 1000 U/mL)<br>Supplied as separate Kit |
| Controls | 2 Levels (50 and 400 pM)<br>Supplied with Kit | 3 Levels (40, 300 and 650 U/mL)<br>Supplied as separate Kit |
| Principle of Operation | Manual<br>Enzymatic Immunoassay (EIA) | Automated<br>Chemiluminscent Microparticle<br>Immunoassay (CMIA) |
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image contains the logo for FUJIREBIO Diagnostics, Inc. The logo consists of a stylized graphic on the left, followed by the company name in bold, sans-serif font. The word "FUJIREBIO" is on the top line, and "Diagnostics, Inc" is on the second line in a smaller font size.
## Performance Characteristics
### Precision:
A study was performed with the HE4 EIA Kit modeled after the NCCLS (CLSI) guideline EP5-A2. A panel of four serum samples was assayed, using two lots of reagents, in replicates of two, at two separate times per day for 20 days. The HE4 EIA Kit precision is ≤ 15% total CV.
## Linearity:
A dilution linearity study was conducted for the HE4 EIA Kit modeled affer the NCCLS (CLSI) guideline EP6-A. Serum samples with elevated HE4 values were diluted with HE4 Calibrator A (zero). The HE4 concentration was determined for each dilution and the percent (%) recovery was calculated.
The HE4 EIA Kit was determined to have a mean recovery within 15% of the expected result for diluted samples.
#### Functional Sensitivity:
The functional sensitivity of the HE4 EIA Kit is ≤ 25 pM. The functional sensitivity is expressed as the concentration of an analyte at which the CV is 20%. The NCCLS guideline EP5-A2 was used to design the experiments for determination of functional sensitivity. A study was conducted where a five member sensitivity panel was tested in replicates of 4 in 2 runs on twenty separate days with two lots of reagents. The functional sensitivity determined for the HE4 EIA was found to be < 5 pM.
#### Detection Limit:
The limit of detection of the HE4 EIA Kit is ≤ 15 pM. The limit of detection (LoD) corresponds to the upper limit of the 95% confidence interval and represents the lowest concentration of HE4 antigen that can be distinguished from zero. The NCCLS guideline EP17-A was used to design the LoD experiments. A study was conducted where HE4 Calibrator A (zero) and 4 samples from healthy subjects diluted to 5 pM with Sample Diluent was tested in replicates of 24 per run in 4 runs on two separate days. The LoD was calculated as follows:
LoD (pM) = 5.0 pM x (1.65 x SDg + 1.65 x SDg) / (ODs - ODg)
The Limit of Detection of the HE4 EIA Kit was calculated to be < 2.5 pM.
#### Interference:
The HE4 EIA Kit mean assay specificity is 100 ± 15%. Recovery studies were performed to compare sera containing the following compounds at the indicated concentrations with control sera. The NCCLS guideline EP7-A was used to design the interference experiments. The following substances and concentrations were tested and found not to interfere with the test.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the logo for Fujirebio Diagnostics, Inc. The logo consists of a stylized graphic on the left and the company name on the right. The graphic is a black and white abstract shape. The text "FUJIREBIO" is in bold, black letters, and below it, in a smaller font, is "Diagnostics, Inc."
| Endogenous serum interferences | Test Concentration |
|-------------------------------------|--------------------|
| Triglycerides | 30 mg/mL |
| Billirubin | 0.2 mg/mL |
| Hemoglobin | 10 mg/mL |
| Total Protein | 120 mg/mL |
| Chemotherapeutic drug interferences | Test Concentration |
| Carboplatin | 500 µg/mL |
| Cisplatin | 165 µg/mL |
| Clotrimazole | 0.3 µg/mL |
| Cyclophosphamide | 500 µg/mL |
| Dexamethasone | 10 µg/mL |
| Doxorubicin | 1.16 µg/mL |
| Leucovorin | 2.68 µg/mL |
| Melphalan | 2.8 µg/mL |
| Methotrexate | 45 µg/mL |
| Paclitaxel | 3.5 ng/mL |
## Potentially interfering clinical conditions
The HE4 EIA Kit mean assay specificity is 100 ± 15%. The HE4 EIA Kit was evaluated using specimens with HAMA and Rheumatoid Factor (RF) to further assess the assay specificity. Five specimens positive for HAMA and five specimens positive for RF were evaluated for % recovery with HE4 EIA antigen spiked into each specimen at approximately 50 and 450 pM.
### Monitoring of Disease status in Patients Diagnosed with Ovarian Cancer
The effectiveness of the HE4 EIA as an aid in monitoring of disease status in ovarian cancer patients was determined by assessing changes in HE4 levels in serial serum samples from 80 patients compared to changes in disease status. A study involving a total of 354 pairs of observations was undertaken with an average number of 4.4 observations per patient. A positive change in HE4 was defined as an increase in the value that was at least 25% greater than the previous value of the test. This level of change takes into account the variability of the assay and the biological variability. Sixty percent (60%) or 76/126 of the patient samples with a positive change correlated with the disease progression while seventy-five percent (75%) or 171/228 of the patient serial samples with no significant change in HE4 value correlated with no progression. The total concordance was seventy percent (70% or 247/354). The following table presents the data in a 2 x 2 format.
| Change in Disease State per Sequential Pair | | | |
|---------------------------------------------|-------------|----------------|-------|
| Increase in HE-4<br>concentration | Progression | No Progression | Total |
| >25% | 76 | 57 | 133 |
| <25% | 50 | 171 | 221 |
| Total | 126 | 228 | 354 |
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image contains the logo for Fujirebio Diagnostics, Inc. On the left is an abstract graphic with a white circle in the middle. To the right of the graphic is the company name in bold, black letters. Below the company name is the text "Diagnostics, Inc" in a smaller, non-bold font.
The following table shows the resulting sensitivities and specificities of the HE4 EIA compared to the disease status at various changes in HE4 EIA concentrations. Sensitivity is represented as a concordance of the HE4 EIA to progression of disease. Specificity is represented as a concordance of the HE4 EIA to no progression of disease.
| Percent Change in HE4 Concentration | Sensitivity | Specificity |
|-------------------------------------|-------------|-------------|
| (%) | (%) | (%) |
| 10 | 71 | 62 |
| 25 | 60 | 75 |
| 50 | 43 | 88 |
| 75 | 38 | 92 |
| 100 | 31 | 95 |
{6}------------------------------------------------
Image /page/6/Picture/0 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo features a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the eagle.
# DEPARTMENT OF HEALTH & HUMAN SERVICES
Public Health Service
Food and Drug Administration 2098 Gaither Road Rockville MD 20850
Fujirebio Diagnostics, Inc. c/o Ms. Diana L. Dickson Manager Regulatory Affairs 201 Great Valley Parkway Malvern, PA 19355-1308
JUN - 9 2008
Re: k072939
Trade/Device Name: HE4 EIA Regulation Number: 21 CFR 866.6010 Regulation Name: Tumor-associated antigen immunological test system Regulatory Class: Class II Product Code: OIU Dated: May 29, 2008 Received: May 30, 2008
Dear Ms Dickson:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
{7}------------------------------------------------
Page 2 -
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0450. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21 CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Robert H. Becker
Robert L. Becker, Jr., M.D., Ph.D. Director Division of Immunology and Hematology Devices Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health
Enclosure
{8}------------------------------------------------
# Indications for Use
510(k) Number (if known): K072939
Device Name: HE4 EIA
Indications For Use:
The HE4 EIA is an enzyme immunometric assay for the quantitative determination of HE4 in human serum. The assay is to be used as an aid in monitoring recurrence or progressive disease in patients with epithelial ovarian cancer. Serial testing for patient HE4 assay values should be used in conjunction with other clinical methods used for monitoring ovarian cancer.
Prescription Use V (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of In Vitro Diagnostic Devices (OIVD)
Mona M Chan
Division Sign-Off
Office of in Viro Diagnostic Device Evaiuation and Safety Page 1 of 1
510(k): K072939
Panel 1
/
Sort by
Ready
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.