K072883 · Arbel Medical , Ltd. · GEH · Dec 10, 2007 · General, Plastic Surgery
Device Facts
Record ID
K072883
Device Name
ICE-SENSE
Applicant
Arbel Medical , Ltd.
Product Code
GEH · General, Plastic Surgery
Decision Date
Dec 10, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4350
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The ICE-SENSE ™ device has the following indications for use: Urology - The system may be used to ablate prostatic tissue. - The system may be used for the ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia Oncology - The system may be used for ablation of cancerous or malignant tissue. - The system may be used for ablation of benign tumors. - The system may be used for palliative intervention. Dermatology - The system may be used for the ablation or freezing of skin cancers and other cutaneous disorders. Gynecology - The system may be used for the ablation of malignant neoplasia or benign dysplasia of the female genitalia. General Surgery - The system may be used for the ablation of leukoplakia of mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemangiomas, recurrent cancerous lesions. - The system may be used for the destruction of warts or lesions. - The system may be used for the palliation of tumors of the oral cavity, rectum, and skin. Thoracic Surgery - The system may be used for the ablation of arrhythmic cardiac tissue. - The system may be used for the ablation of cancerous lesions Proctology - The system may be used for the ablation of benign or malignant growths of the anus and rectum - The systems may be used for the ablation of hemorrhoids.
Device Story
ICE-SENSE is a cryosurgical unit for tissue ablation. System utilizes pressurized liquid nitrogen in a closed-loop system to deliver extreme cold to targeted tissue via disposable sharp or blunt probes. Components include main chassis with cooling system, controller with screen/keypad, cryohandle, and foot pedal. Operated by clinicians in various surgical settings. Device destroys unwanted tissue through freezing; passive thawing used. Output is controlled cold temperature at probe tip. Benefits include targeted destruction of malignant, benign, or arrhythmic tissue.
Clinical Evidence
No clinical data provided; bench testing only.
Technological Characteristics
Cryosurgical unit; pressurized liquid nitrogen closed-loop cooling system; disposable sharp/blunt probes; main chassis with controller, screen, keypad, cryohandle, and foot pedal. Operates via extreme cold application; passive thawing; open-loop control.
Indications for Use
Indicated for ablation of prostatic tissue (cancer/BPH), cancerous/malignant/benign tumors, palliative intervention, skin cancers/disorders, female genital neoplasia/dysplasia, various general surgical lesions (warts, hemangiomas, hemorrhoids, etc.), arrhythmic cardiac tissue, and anal/rectal growths. For prescription use.
Regulatory Classification
Identification
(1) Cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories. A cryosurgical unit with a liquid nitrogen cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. (2) Cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories. A cryosurgical unit with a nitrous oxide cooled cryoprobe and accessories is a device intended to destroy tissue during surgical procedures, including urological applications, by applying extreme cold. (3) Cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories. A cryosurgical unit with a carbon dioxide cooled cryoprobe or a carbon dioxide dry ice applicator and accessories is a device intended to destroy tissue during surgical procedures by applying extreme cold. The device is intended to treat disease conditions such as tumors, skin cancers, acne scars, or hemangiomas (benign tumors consisting of newly formed blood vessels) and various benign or malignant gynecological conditions affecting vulvar, vaginal, or cervical tissue. The device is not intended for urological applications.
{1}------------------------------------------------
K072883
:
## 510 (k) SUMMARY
## Arbel Medical Ltd. ICE-SENSE® Device
# Submitter's Name, Address, Telephone Number, Contact Person and Date Prepared
| Manufacturer: | Arbel Medical LTD.<br>New industrial park 7<br>Building 1a<br>P.O.Box 620<br>Yokneam 206923<br>Israel<br>Telephone: +972-4-9090000<br>Facsimile: +972-4-9090001 | | | DEC 1 0 2007 |
|--------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--|--|--------------|
| Contact Person: | Dr. Zvi Ladin PhD.<br>Principal Investigator<br>Boston MedTech Advisors, Inc.<br>990 Washington Street<br>Suite #204<br>Dedham, MA 02026<br>Telephone: (781) 407 0900 x104<br>Facsimile: (781) 407 0901<br>Email: zladin@bmtadvisors.com | | | |
| Date Prepared: | October 4, 2007 | | | |
| Name of Device and Name/Address of Sponsor | | | | |
| Trade/Proprietary Name: | Arbel Medical ICE-SENSE® Device | | | |
| Common Name: | Cryosurgical unit and accessories | | | |
| Classification Name: | Cryosurgical unit and accessories (21 C.F.R. § 878.4810) | | | |
| Manufacturing Facility: | Arbel Medical LTD.<br>New industrial park 7, Building 1a<br>P.O.Box 620<br>Yokneam 206923<br>Radiancy (Israel) Ltd. | | | |
| Establishment<br>Registration Number: | N/A | | | |
| Owner/operator number: | N/A | | | |
{2}------------------------------------------------
## Predicate Devices
Cryomedical Sciences Cryolite (K970995), Cryomedical Sciences Accuprobe 610 (K964366. K973190) and Galil Medical Cryo-Hit (K980913)
### Intended Use / Indications for Use
The ICE-SENSE ™ device has the following indications for use:
#### Urology
- The system may be used to ablate prostatic tissue. .
- . The system may be used for the ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia
Oncology
- The system may be used for ablation of cancerous or malignant tissue. .
- The system may be used for ablation of benign tumors. .
- The system may be used for palliative intervention. .
### Dermatology
- The system may be used for the ablation or freezing of skin cancers and other . cutaneous disorders.
Gynecology
- The system may be used for the ablation of malignant neoplasia or benign . dysplasia of the female genitalia.
General Surgery
- The system may be used for the ablation of leukoplakia of mouth, angiomas, . sebaceous hyperplasia, basal cell tumors of the evelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts, actinic and seborrheic keratoses, cavernous hemangiomas, recurrent cancerous lesions.
- The system may be used for the destruction of warts or lesions. .
- . The system may be used for the palliation of tumors of the oral cavity, rectum, and skin.
Thoracic Surgery
- The system may be used for the ablation of arrhythmic cardiac tissue. .
The system may be used for the ablation of cancerous lesions .
Proctologv
- . The system may be used for the ablation of benign or malignant growths of the anus and rectum
- . The systems may be used for the ablation of hemorrhoids.
### Technological Characteristics
The ICE-SENSE™ Device is used to destroy unwanted tissue by application of extreme cold to the selected site. The device delivers cold temperatures to targeted tissue
{3}------------------------------------------------
by pressurized liquid nitrogen closed system and a disposable sharp or blunt probes. The device consists of a main chassis for the cooling system, a controller with a screen and key pad, a cryohandle that controls and holds the probe and a foot pedal.
#### Substantial Equivalence
The ICE-SENSE™ has the same intended use and indications for use, principles of operation and technological characteristics as its predicate devices. The slight differences between ICE-SENSE™ and its predicate devices do not raise any new issues of safety and effectiveness.
Arbel's ICE-SENSE™ is substantially equivalent to a combination of the above named predicate devices and the minor differences detailed in the discussion do not raise additional concerns of safety or efficacy.
These are cryosurgical devices that use a coolant to provide a very low temperature to the treated tissue by using a designated probe. The main predicate device is the Cryomedical Sciences' Accuprobe 610. The IceSense is equipped with sharp probes for percutaneus procedures as the Cryo-Hit does. It uses passive thawing and is open loop controlled as does the predicate Cryolite device.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. Inside the circle is an abstract symbol resembling a stylized human figure or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
## DEC 1 0 2007
Arbel Medical, Ltd. % Boston Medtech Advisors, Inc. Zvi Ladin, PhD 990 Washington Street, Suite 204 Dedham, Massachusetts 02026
Re: K072883
Trade/Device Name: Arbel ICE-SENSE™ Regulation Number: 21 CFR 878.4350 Regulation Name: Cryosurgical unit and accessories Regulatory Class: II Product Code: GEH Dated: October 4, 2007 Received: October 9, 2007
Dear Dr. Ladin:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug. and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
Page 2 - Zvi Ladin, PhD
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Pretmarket, Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N Mullerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Eyaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known):
K 072883
Arbel ICE-SENSE™
Device Name:
Indications for Use:
The ICE-SENSE ™ device has the following indications for use:
#### Urology
- · The system may be used to ablate prostatic tissue.
- · The system may be used for the ablation of prostate tissue in cases of prostate cancer and benign prostatic hyperplasia
#### Oncology
- · The system may be used for ablation of cancerous or malignant tissue.
- · The system may be used for ablation of benign tumors.
- · The system may be used for palliative intervention. Dermatology
- · The system may be used for the ablation or freezing of skin cancers and other castaneous disorders. Gynecology
- · The system may be used for the ablation of malignant neoplasia or benign dysplasia of the female genitalia.
Prescription Use X (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Melkerson
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
072883 510(k) Number
Page 1 of 2
{7}------------------------------------------------
General Surgery
- · The system may be used for the ablation of leukoplakia of mouth, angiomas, sebaceous hyperplasia, basal cell tumors of the eyelid or canthus area, ulcerated basal cell tumors, dermatofibromas, small hemangiomas, mucocele cysts, multiple warts, plantar warts, hemorrhoids, anal fissures, perianal condylomata, pilonidal cysts actinic and seborrheic keratoses, cavernous hemangiomas, recurrent cancerous lesions.
- · The system may be used for the destruction of warts or lesions.
- · The system may be used for the palliation of tumors of the oral cavity, rectum, and skin.
Thoracic Surgery
- · The system may be used for the ablation of arrhythmic cardiac tissue.
- · The system may be used for the ablation of cancerous lesions
Proctology
- · The system may be used for the ablation of benign or malignant growths of the anus and rectum
- · The systems may be used for the ablation of hemorrhoids
X Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
#### Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Millman
(Division Sign-Off) Division of General, Restorative, and Neurological Devices
Page 2 of 2
**510(k) Number** K072883
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.