APEX 2.0 SOFTWARE FOR QDR X-RAY BONE DENSITOMETERS

K072847 · Hologic, Inc. · KGI · Mar 28, 2008 · Radiology

Device Facts

Record IDK072847
Device NameAPEX 2.0 SOFTWARE FOR QDR X-RAY BONE DENSITOMETERS
ApplicantHologic, Inc.
Product CodeKGI · Radiology
Decision DateMar 28, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 892.1170
Device ClassClass 2

Indications for Use

The APEX for QDR X-Ray Bone Densitometers is indicated for the estimation of bone mineral density (BMD), comparison of measured variables obtained from a given QDR scan to a detabase of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics using the Hologic ODR® X-Ray Bone Densitometers. IVA scans are intended for the visualization or quantitative assessment of vertebral body deformities. IVA also allows the visualization of abdominal aortic calcifications and, if present, clinical correlation may be advised since abdominal aortic calcification may be associated with cardiovascular disease.

Device Story

APEX 2.0 Software functions as an optional software package for Hologic QDR X-Ray Bone Densitometers. It processes X-ray scan data to estimate bone mineral density (BMD), assess fracture risk, perform body composition analysis, and distinguish bone from prosthetic materials. The software enables vertebral deformity assessment (IVA) and visualization of abdominal aortic calcifications. Used in clinical settings by healthcare professionals, the software transforms raw scan inputs into quantitative metrics and visual outputs. These outputs assist clinicians in diagnosing bone health, evaluating cardiovascular risk markers (aortic calcification), and monitoring treatment efficacy. The device provides standardized reference comparisons to aid clinical decision-making.

Clinical Evidence

No clinical data provided; substantial equivalence is based on technological characteristics and intended use compared to the predicate device.

Technological Characteristics

Software-based image processing and analysis tool for QDR X-Ray Bone Densitometers. Operates as an add-on software module. Utilizes established densitometry algorithms for BMD estimation, vertebral deformity assessment, and body composition analysis. Connectivity via integration with Hologic QDR hardware.

Indications for Use

Indicated for patients requiring bone mineral density (BMD) estimation, fracture risk assessment, vertebral deformity assessment, body composition analysis, or discrimination of bone from prosthetics via QDR X-Ray Bone Densitometers. IVA scans indicated for visualization/quantitative assessment of vertebral deformities and abdominal aortic calcifications.

Regulatory Classification

Identification

A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ # G. 510(k) - SUMMARY K072847 MAR 2 8 2008 ### Manufacturing Establishment and Contact Information G.1 #### Manufacturer Name and Address: G.1.1 Hologic, Inc. 35 Crosby Drive Bedford, MA 01730 ### G.1.2 Establishment Registration Number: 1221300 ### G.1.3 Name, Title, and Telephone Number of Contact: Bryan Cowell, RAC Regulatory Affairs Specialist Phone: (781) 999-7300, ex. 7085 Fax : (781) 280-0614 bcowell(@hologic.com #### G.2 Device Identification ### G.2.1 Device Trade Name: APEX 2.0 Software for QDR X-Ray Bone Densitometers ### G.2.2 Common / Usual Name: Software option for Bone Densitometers #### Proposed Intended Use: G.2.3 The APEX for QDR X-Ray Bone Densitometers is indicated for the estimation of bone mineral density (BMD), comparison of measured variables obtained from a given QDR scan to a detabase of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics using the Hologic ODR® X-Ray Bone Densitometers. IVA scans are intended for the visualization or quantitative assessment of vertebral body deformities. IVA also allows the visualization of abdominal aortic calcifications and, if present, clinical correlation may be advised since abdominal aortic calcification may be associated with cardiovascular disease. #### Device Classification G.3 ### G.3.1 Classification: Class II {1}------------------------------------------------ ## G.3.2 Classification Name and Rule Bone Densitometer: 21 CFR 892.1170 - G.3.3 Classification Panel Radiology - G.3.4 Product Code 90 KGI ## G.3.5 Predicate Device(s) - . 510(k) No.: K023398 Trade Name: Discovery Package for Hologic QDR X-Ray Bone Densitometers November 8, 2002 SE Date: Manufacturer: Hologic, Inc. ## G.4 Conclusion Based on the information provided in this 510(k) submission, the Hologic APEX 2.0 Software for QDR X-Ray Bone Densitometers is substantially equivalent to the presently marketed Discovery Package for Hologic QDR X-Ray Bone Densitometers (K023398). No new safety or efficacy questions are raised with the APEX 2.0 Software. {2}------------------------------------------------ Image /page/2/Picture/1 description: The image shows the logo for the Department of Health & Human Services USA. The logo features a stylized eagle with three lines representing its wings. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" is arranged in a circular pattern around the eagle. Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850 Mr. Bryan Cowell Regulatory Affairs Specialist Hologic, Incorporated 35 Crosby Drive BEDFORD MA 01730 MAR 2 8 2008 Re: K072847 Trade/Device Name: APEX 2.0 Software for QDR® X-Ray Bone Densitometers Regulation Number: 21 CFR 892.1170 Regulation Name: Bone Densitometers Regulatory Class: II Product Code: KGI Dated: December 27, 2007 Received: December 28, 2007 Dear Mr. Cowell: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Image /page/2/Picture/10 description: The image is a circular logo with the letters "FDA" in the center. Above the letters, the years "1906-2006" are printed. Below the letters, the word "Centennial" is printed in a stylized font. The logo is surrounded by a dotted border. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Nancy C. Brogdon Nancy C. Brogdon Director, Division of Reproductive. Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## Indication for Use Statement 510(k) Number (if known): K072847 Device Name: APEX 2.0 Software for QDR. X-Ray Bone Densitometers Indication for Use: The APEX for QDR X-Ray Bone Densitometers is indicated for the estimation of bone mineral density (BMD), comparison of measured variables obtained from a given QDR scan to a database of reference values, the estimation of fracture risk, vertebral deformity assessment, body composition analysis, and discrimination of bone from prosthetics using the Hologic QDR® X-Ray Bone Densitometers. IVA scans are intended for the visualization or quantitative assessment of vertebral bone deformities. IVA also allows the visualization of abdominal aortic calcification, and, if present, clinical correlation may be advised since abdominal aortic calcification may be associated with cardiovascular disease. Prescription Use (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) signature Division of Reproductive, Abdominal and Radiological Devices 510(k) Number.
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...