The Tactile Guidance System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures. The Tactile Guidance System is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include: Unicondylar knee replacement
Device Story
TGS is a stereotaxic instrument for orthopedic surgery; assists surgeons with presurgical planning and intraoperative navigation. Input: patient CT data. Operation: system uses optical detector and computer to track surgical tools relative to patient anatomy; robotic arm acts as an 'intelligent' tool holder/guide. Software defines spatial boundaries to constrain the robotic arm, ensuring precise preparation of anatomical sites for implants. Used in OR by surgeons; supports standard tools like drills and awls. Benefits: provides haptic/spatial guidance during minimally invasive procedures, improving accuracy of implant placement. Preoperative planning performed on a separate laptop; plans transferred via encrypted USB to TGS.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
System components: optical detector, CPU, Linux-based OS, dual LCD touch screens, isolation transformer, foot control panel, and robotic arm. Connectivity: USB for plan transfer, optical tracking for tool positioning. Software: Voyager application for navigation and spatial boundary definition. Power: UPS. Sterilization: standard surgical tool compatibility.
Indications for Use
Indicated for surgical knee procedures, specifically unicondylar knee replacement, where stereotactic surgery is appropriate and rigid anatomical bony structures can be identified relative to a CT-based model.
Regulatory Classification
Identification
A stereotaxic instrument is a device consisting of a rigid frame with a calibrated guide mechanism for precisely positioning probes or other devices within a patient's brain, spinal cord, or other part of the nervous system.
{0}------------------------------------------------
Ko72806
Image /page/0/Picture/1 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold letters, with the words "SURGICAL CORP." in smaller letters underneath. Above the word "MAKO" is a stylized graphic that resembles a shark fin or a mountain peak. The logo is simple and modern, and the use of bold letters and a strong graphic gives it a sense of power and authority.
Page 1 of 2
JAN 24 2006
lori lauderdale, fronda 33317 ....... 2595 dare road 184 954 327 2044 36 1454 827 0426
## ATTACHMENT 1 510(K) SUMMARY
| Submitter: | |
|----------------------|--|
| Address: | |
| Phone number: | |
| Fax number: | |
| Contact Person: | |
| Date Prepared: | |
| Device Trade Name: | |
| Common Name: | |
| Classification Name: | |
| Classification #: | |
MAKO Surgical Corp. 2555 Davie Road, Fort Lauderdale, FL, 33317 954-927-2044 x. 605 954-927-0446 William F. Tapia September 28 2007 Tactile Guidance System (TGS) Stereotaxic Instrument Class II 21 CFR 882 4560
Substantial Equivalence Claimed To: The TGS is substantially equivalent to the Voyager / Tactile Guidance System - CT (K052851), Uni-Knee Surgetics Navigation System (K062146), Acumen System (K031454, K031337), Vectorvision Uni-Knee (K041899), and Navitrack System-OS Unicondylar Knee Universal (K071714).
Description: The TGS is a stereotaxic instrument that includes an optical detector, computer, dedicated instrumentation, operating software, tools and accessories, and the TGS robotic arm. TGS uses patient CT data to assist the physician with presurgical planning and interpretive navigation. The TGS robotic arm, which is an add-on to the Navigation Module, serves as an "intelligent" tool holder or tool guide used by a surgeon for stereotactic guidance during minimally invasive orthopedic surgical procedures. The TGS robotic arm, an electromechanical arm, is passively constrained by softwaredefined spatial boundaries implemented through the use of the TGS robotic arm and is designed to support a surgeon's preparation of an anatomical site for an orthopedic implant with standard surgical tools such as drills, awls, and 3rd party drill systems.
Summary of Technological Characteristics: The TGS consists of the following basic components:
- High Resolution Color Liquid Crystal Display (LCD) Touch Screen Monitor ﮩ
- Second High Resolution LCD screen that shows an identical image to the primary touch screen.
- Uninterruptible Power Supply (UPS)
- Central Processing Unit (CPU) ব
- 5. Isolation Transformer
- ్. Keyboard and Mouse
- 7. Optical Detector
- Operating Room (OR) Cart 8.
- Foot control panel ഗ്
- 10. Tool and accessories optically tracked surgical tools and accessories
- 11. TGS robotic arm tool platform that is connected to the Voyager Navigation Platform to enable passive stereotactic guidance of any cleared standard surgical tool such as 30 party drill systems (e.g., Anspach's eMax (K011444)) to prepare anatomical sites for placement of orthopedic implants in a minimally invasive manner.
- 12. Voyager software application
- 13. Linux operating system
- 14. Software drivers for video grabber, standard components (keyboard, mouse, monitor, etc.)
- 15. Preoperative planning laptop that runs only the preoperative planning portion of the application. Dicom CDs of CT scans with patient data can be loaded and surgical plans can be created on this laptop and downloaded (in an encrypted format) to a USB memory stick to be transferred to the TGS.
{1}------------------------------------------------
Image /page/1/Picture/0 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold letters, with the words "SURGICAL CORP." in smaller letters underneath. Above the word "MAKO" is a stylized graphic that resembles a mountain range or a wave. The number 6072806 is written in the upper left corner of the image.
Page 2 of 2
2555 davie isad............................................................................................................................................................... ..............................................................................................................................................................................
Intended Use/Indications for Use: The Tactile Guidance System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
The Tactile Guidance System is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include:
- . Unicondylar knee replacement
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES-USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling three human profiles facing right, with flowing lines extending from the bottom of the profiles.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
JAN 2 4 2008
Mako Surgical Corporation % Mr. William F. Tapia Vice President, RA/QA 2555 Davie Road Ft. Lauderdale, Florida 33317
Re: K072806
Trade/Device Name: Tactile Guidance System (TGS) Regulation Number: 21 CFR 882.4560 Regulation Name: Stereotaxic instrument Regulatory Class: II Product Code: HAW Dated: January 16, 2008 Received: January 17, 2008
Dear Mr. Tapia:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{3}------------------------------------------------
Page 2 - Mr. William F. Tapia
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark M Mulhusan
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
Image /page/4/Picture/0 description: The image shows the logo for MAKO Surgical Corp. The logo features the word "MAKO" in large, bold letters, with the words "SURGICAL CORP." in smaller letters underneath. Above the word "MAKO" is a stylized mountain peak. The logo is black and white.
2555 davie road fort louderdale, florida 33317 tel 954 927 2044 fax 954 927 0446
## ATTACHMENT 2
## INDICATIONS FOR USE
510(k) Number (if known): K072806
Device Name: Tactile Guidance System (TGS)
Indications for Use:
The Tactile Guidance System is intended to assist the surgeon in providing software defined spatial boundaries for orientation and reference information to anatomical structures during orthopedic procedures.
The Tactile Guidance System is indicated for use in surgical knee procedures, in which the use of stereotactic surgery may be appropriate, and where reference to rigid anatomical bony structures can be identified relative to a CT based model of the anatomy. These procedures include:
Unicondylar knee replacement .
Prescription Use _ X
OR
Over-the-Counter Use_
(Per 21 CFR 801.109)
(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Minh M. Millkuss
Division of Gen and Neus
K072806
510(k) REVIEW
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.