DISPOSABLE AEM FIXED TIP ELECTRODES AND HANDPIECES
K072789 · Encision, Inc. · GEI · Oct 19, 2007 · General, Plastic Surgery
Device Facts
Record ID
K072789
Device Name
DISPOSABLE AEM FIXED TIP ELECTRODES AND HANDPIECES
Applicant
Encision, Inc.
Product Code
GEI · General, Plastic Surgery
Decision Date
Oct 19, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.4400
Device Class
Class 2
Attributes
Therapeutic, 3rd-Party Reviewed
Indications for Use
AEM Disposable Electrodes and AEM Disposable Handpieces are electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures. The devices are intended for use by qualified medical personnel trained in the use of electrosurgical equipment. AEM instruments incorporate the use of AEM technology and are intended for use with the Encision AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor.
Device Story
AEM Disposable Electrodes and Handpieces are electrosurgical accessories used in open, endoscopic, and laparoscopic surgery. The device connects to an electrosurgical generator via the Encision AEM Monitor. The system utilizes AEM (Active Electrode Monitoring) technology to minimize stray energy injuries; it monitors current returned via a shield circuit near the electrode tip and shuts down the generator if excessive current is detected (insulation failure or capacitive coupling). Electrodes feature an insulated tip, shaft, and shield assembly; handpieces provide cut/coagulation switching (hand or foot-controlled). Surgeons use the device to perform tissue ablation, resection, and coagulation. The system benefits patients by reducing the risk of unintended thermal injury during monopolar electrosurgery.
Clinical Evidence
Bench testing only. Verification included continuity and capacitance measurements, and compatibility testing with a worst-case representative electrosurgical generator and the AEM Monitor.
Technological Characteristics
Monopolar electrosurgical accessories; disposable electrodes and handpieces. Features insulated conductors, shafts, and AEM shield assembly. Connectivity: connects to Encision AEM Monitor and compatible electrosurgical generators. Sterilization: provided sterile, single-use only. Control: hand-switching or foot-switching modes.
Indications for Use
Indicated for ablation, removal, resection, and coagulation of soft tissue requiring hemostasis in open, endoscopic, and laparoscopic surgical procedures. Intended for use by qualified medical personnel trained in electrosurgical equipment.
Regulatory Classification
Identification
An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.
Predicate Devices
Model ES1000 series Monopolar Laparoscopic Electrodes with Electroshield (ENCISION INC.)
Model 30-6000 series Reusable Monopolar Electrodes (Kirwin Surgical Products)
Model 8380 series Unipolar Electrodes (Richard Wolf Medical Instruments Corp.)
Models SW11100, SW11200, SW11202, SW11300 (Reusable) and SW1220, SW12202, SW12300 (Disposable) Electrosurgical Pencils (Shining World Healthcare Co., Ltd.)
Model ES4000 AEM Cord (ENCISION INC.)
Model E2508 Disposable Pencil (Valleylab, Inc.)
Submission Summary (Full Text)
{0}------------------------------------------------
### ENCISIGN
K 072-789
#### 510(k) Summary
DATE: 5/19/07
510(k) Submitter: ENCISION INC. 6797 Winchester Circle Boulder, CO 80301 USA Establishment Registration: 1722040 JU7 1 § 2007
Contact Person.
Judith V. King, VP Regulatory Affairs and Quality Assurance Phone: 303-339-6917 Fax: 303-444-2693 e-mail: jking(a encision.com
Device Name: AEM Disposable Electrodes and AEM Disposable Handpieces
Common name: Device, Electrosurgical, Cutting and Coagulation and Accessories
Classification: CFR Section: 878.4400
Class: II
Product Code: GEI
Predicate Devices:
#### Electrodes
| Trade, Proprietary or Model Name | Manufacturer |
|------------------------------------------------------------------------------------------------------------------------------|-------------------------------------------|
| Model ES1000 series Monopolar Laparoscopic Electrodes with<br>Electroshield | ENCISION INC. |
| Model 30-6000 series Reusable Monopolar Electrodes | Kirwin Surgical Products |
| Model 8380 series Unipolar Electrodes | Richard Wolf Medical<br>Instruments Corp. |
| Models SW11100, SW11200, SW11202, SW11300 (Reusable)<br>and SW1220, SW12202, SW12300 (Disposable) Electrosurgical<br>Pencils | Shining World Healthcare<br>Co., Ltd. |
#### Handpieces
| Trade, Proprietary or Model Name | Manufacturer |
|------------------------------------------------------------------------------------------------------------------------------|---------------------------------------|
| Model ES1000 series Monopolar Laparoscopic Electrodes with<br>Electroshield | ENCISION INC. |
| Model ES4000 AEM Cord (AEM Monitor accessory) | ENCISION INC. |
| Model E2508 Disposable Pencil | Valleylab, Inc. |
| Models SW11100, SW11200, SW11202, SW11300 (Reusable)<br>and SW1220, SW12202, SW12300 (Disposable) Electrosurgical<br>Pencils | Shining World Healthcare<br>Co., Ltd. |
ENCISION 510(k) Notification Summary AEM® Disposable Electrodes and AEM Disposable Handpieces
{1}------------------------------------------------
# ENCISIGN
#### Description of Devices:
The AEM Disposable Electrode and the AEM Disposable Handpiece connect to a compatible electrosurgical generator via an adapter on the ENCISION AEM Monitor. The handpiece has foot switching and hand switching versions. The electrodes and handpieces are single use products, which are provided sterile. They are designed not to be re-sterilized.
The AEM Monitoring System, including the electrodes and handpieces, are designed to minimize the likelihood of stray energy injuries caused by active insulation failure or capacitive coupling. The monitor does this by shutting down the ESU when excessive current is returned via the shield circuit which extends to near the tip of the electrode.
The electrodes, which consist of an insulated tip, shaft with locking knob, and AEM shield assembly, are available in various tip styles. The electrode has an inner insulation between the active conductor and shield tube, as well as a secondary outer insulation on the outside of the clectrode shaft. The electrodes snap into the handpiece.
One version of the handpiece provides switching for the Cut and Coag functions of the ESU. This version has an integral cord. The other version uses foot-switching provided with the electrosurgical generator. The foot-switching handpiece connects to either the existing reusable or disposable AEM Cord from Encision. The electrodes may be removed from the handpiece and replaced with another electrode within the sterile field.
The electrode can rotate freely or be locked in one of multiple orientations relative to the handpiece, as preferred by the surgeon.
#### Intended 【Jse:
AEM Disposable Electrodes and AEM Disposable Handpieces are electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.
The devices are intended for use by qualified medical personnel trained in the use of electrosurgical equipment.
AEM instruments incorporate the use of AEM technology and are intended for use with the Encision AEM Monitoring System and electrosurgical generators having compatibility with the AEM Monitor.
#### Technological Characteristics:
The Encision AEM Disposable Electrodes and AEM Disposable Handpieces incorporate the same technological characteristics as the predicate devices for delivery of the ESU high frequency current, consisting of insulated conductors and shafts with appropriately shaped tips for electrosurgery. The control for ESU Cut and Coag modes on the handswitching handpiece is a switch similar in design and function to the predicate handpiece.
Like the Encision predicate devices, the AFM Disposable Electrodes and AEM Disposable Handpieces include an additional AEM shielding function which diverts stray energy from the shaft of the instrument and is monitored by the Encision AEM Monitor.
#### Non-clinical Performance Testing:
Performance of the devices' AFM technology and delivery of electrosurgical energy has heen verified by bench testing consisting of continuity and capacitance measurements, and
{2}------------------------------------------------
## ENCISIGN
compatibility testing with a worst case representative electrosurgical generator and the AEM Monitor.
Conclusions:
The AEM Disposable Electrodes and AEM Disposable Handpieces are safe and effective and are substantially equivalent to the predicate devices.
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the Department of Health & Human Services - USA. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized image of three overlapping human profiles facing to the right, creating a sense of depth and unity.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
OCT 1 9 2007
Encision, Inc. % Intertek Testing Services Mr. Daniel W. Lehtonen 2307 East Aurora Road Unit B7 Twinsburg, Ohio 44087
Re: K072789
Trade/Device Name: AEM Disposable Electrodes and AEM Disposable Handpieces Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: September 28, 2007 Received: October 1, 2007
Dear Mr. Lehtonen:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. - --------------------------------------------------------------------------------------------------------------------------------------------------------------
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{4}------------------------------------------------
Page 2 - Mr. Daniel W. Lehtonen
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Mark N. Melkerson
Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
Image /page/5/Picture/0 description: The image shows the word "ENCISION" in all caps. The letters are in a bold, sans-serif font and are black. The "O" in "ENCISION" is replaced by a circular logo with a white curved line running through it. There is a registered trademark symbol to the upper right of the last letter N.
### Indications for Use
510(k) Number (if known):
K 072789
Device Name:
AEM Disposable Electrodes and AEM Disposable Handpieces
Indications For Use:
The AEM Disposable Electrodes and AEM Disposable Handpieces are electrosurgical accessories intended, by use of monopolar high-frequency electrical current from compatible electrosurgical generators, for ablation, removal, resection and coagulation of soft tissue where associated hemostasis is required in open, endoscopic and laparoscopic surgical procedures.
The devices are intended for use by qualified medical personnel trained in the use of electrosurgical equipment.
X Prescription Use (Part 21 CFR 801 Subpart D)
OR
Over-The-Counter Use (Part 21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Division Sign Off
Division of General, Restorative, and Neurological Devices
510(k) Number 462289
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only software-only devices, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (software-only devices), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.