ANT ANGIOGRAPHIC SYRINGES, MODEL(S) 100101,102,103,100201,202,203,204,100301,200101,201,202, 203,200301,101,102,202,301
K072696 · Shenzhen Ant Hi-Tech Industrial Co., Ltd. · DXT · May 2, 2008 · Cardiovascular
Device Facts
| Record ID | K072696 |
| Device Name | ANT ANGIOGRAPHIC SYRINGES, MODEL(S) 100101,102,103,100201,202,203,204,100301,200101,201,202, 203,200301,101,102,202,301 |
| Applicant | Shenzhen Ant Hi-Tech Industrial Co., Ltd. |
| Product Code | DXT · Cardiovascular |
| Decision Date | May 2, 2008 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 870.1650 |
| Device Class | Class 2 |
Intended Use
ANT Angiographic Syringes are syringes for the injection of contrast media or saline. This syringe is for single use with US legally marketed angiographic injectors.
Device Story
ANT Angiographic Syringes are single-use, disposable plastic syringes (polypropylene or PET) designed for use with US-marketed angiographic injectors. Available in multiple volumes (10mL to 200mL), these syringes facilitate the delivery of contrast media or saline during angiographic procedures. The device is operated by healthcare professionals in clinical settings. The primary design variation across models involves the nozzle connection to ensure compatibility with specific injector manufacturers (e.g., Medrad). The device functions as a mechanical delivery component; it does not involve electronic processing or software. By providing a sterile, compatible interface for contrast injection, the device enables diagnostic imaging procedures.
Clinical Evidence
Bench testing only. Biocompatibility testing conducted per ISO 10993-1:2003(E) standards, including cytotoxicity, sensitization, irritation/intracutaneous reactivity, systemic toxicity, and haemo-compatibility. All tests met requirements.
Technological Characteristics
Materials: Polypropylene or PET. Dimensions: 10mL to 200mL capacities. Connectivity: Mechanical interface (nozzle) for angiographic injectors. Sterilization: Not specified. Energy source: None (mechanical).
Indications for Use
Indicated for the injection of contrast media or saline using US legally marketed angiographic injectors in patients undergoing angiographic procedures.
Regulatory Classification
Special Controls
*Classification.* Class II (special controls). The device, when it is a non-patient contacting balloon inflation syringe intended only to inflate/deflate balloon catheters and monitor pressure within the balloon, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 870.9.
Predicate Devices
- Disposable CT/MR Syringes for Nemoto Injectors (K051799)
Related Devices
- K082439 — DISPOSABLE ANGIOGRAPHIC SYRINGE, MODEL(S) CT-200-NE, CT-200-MA, CT-200+200-MA, CT-200-LF, CT -100-NE, CT-200NE, CT-50+CT · Shandong Weigao Medical Polymer Co., Ltd. · Mar 23, 2009
- K151960 — Sterile High-pressure Angiographic Syringes for Single-use · Shenzhen Baoan Medical Supplies Co., Ltd. · Jan 7, 2016
- K172694 — Sterile Disposable Syringes · Shenyang Mastech Medical Device Co., Ltd. · Jul 6, 2018
- K162088 — Ecomed Disposable Angiographic Syringes · Ecomed Solutions, LLC · Jan 27, 2017
- K090487 — DISPOSABLE ANGIOGRAPHIC SYRINGES FOR USE WITH NEMOTO ANGIOGRAPHIC POWER, MODEL SY-125, C855-5150 · Coeur, Inc. · Oct 19, 2009
Submission Summary (Full Text)
{0}------------------------------------------------
# 510(k) Summary
This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92.
The Assigned 510(k) Number is:
## 1. Applicant Device Information
Trade/Proprietary Name: ANT Angiographic Syringes Common Name: Angiographic injector and syringe. Classification Name: Injector and syringe, angiographic Device Class: II Product Code: DXT Regulation Number: 870.1650 Intended Use: ANT Angiographic Syringes are syringes for the injection of contrast media or saline. This syringe is for single us with US legally marketed angiographic injectors.
### 2. Submitter Information
### Manufacturer Name:
Shenzhen Ant Hi-Tech Industrial Co., Ltd. Building 11, Lishan Industrial Park, Xinghai Ave, Nanshan District, Shenzhen, China
### Contact Person of the Submission:
Ms. Diana. Hong Mr. Eric. Chen Suite 8D, Zhongxin Zhongshan Mansion, No.19, Lane 999, Zhongshan No.2 Road(S) Shanghai, China 20030 Phone: +86-21-64264467 x 152 Fax: +86-21-64264468 x 809 Email: Diana.hong@mid-link.net Eric.chen@mid-link.net
### 3. Predicate Device
### K051799
{1}------------------------------------------------
Trade/Proprietary Name: Disposable CT/MR Syringes for Nemoto Injectors Common/Usual Name: Syringes for CT and MR Injections Classification Name: Angiographic Injector & Syringe
Submitter Name: Coeur, Inc. Address: 704 Cadet Court Lebanon, TN 37087 Phone: (615) 574-7923 Contact: Debra F. Manning, VP, Q & RA
### 4. Device Description
The applicant device of ANT Angiographic Syringes are plastic, single-use, disposable syringes to be offered in 10 mL, 12 mL, 20mL, 60mL, 60ml, 65ml, 100ml, 115 ml, 125 mL, 130ml, 140ml and 200mL sizes. The syringes will be offered made of polypropylene or PET- both materials, of which, are available in current legally marketed products.
Variants: All variants include 7 series as followings: Series 1: Model 100101, Model 100103, Model 100103, Model 100202, Model 100202, Model 100203, Model 100204, Model 100301, Model 100302, Series 2: Model 200101, Model 200201, Model 200202, Model 200203, Model 200301, Series 3: Model 300101, Model 300102, Model 300202, Model 300301, Model 300302, Series 4: Model 400101, Series 5: Model 500101, Model 500102, Series 6: Model TR0012, Series 7: Model CS0010, Model CS0020,
All 8 series follow same design principle, same material, and same intended use. The only differences are volume and the connection (nozzle) to different angiography injector from different manufacturer. The connection differs to fit with US legally Market Angiographic Injectors from company Medrad. The connector difference is only to fit the injector and does not affect the performance of the syringes.
{2}------------------------------------------------
# 5. Substantially Equivalence Determination
# Comparison Analysis:
The Applicant device has the same classification intended use, sterilizations, performance, biocompatibility, chemical specifications and similar physical and mechanications with the predicate device. The only difference between applicant device and predicate device is some physical specifications variant which is too slight to influence the effectiveness and safety.
# Conclusion:
The applicant device is Substantially Equivalent (SE) to the predicate device in terms of Effectiveness and Safety.
{3}------------------------------------------------
# 6. Effectiveness and Safety Considerations
### Effectiveness:
All the variant models of the applicant device are evaluated regarding the performance.
#### Safety Considerations:
With accordance with the Table 1 Initial Evaluation Tests for Consideration and Table 2 Supplementary Evaluation Tests for Consideration in ISO 10993-1:2003(E), Biological Evaluation of Medical Devices - Part 1: Evaluation and Testing, the necessary tests for Biocompatibility Testing includes: Cytotoxicity, Sensitization, Irritation or Intracutaneous Reactivity, Systemic Toxicity (Acute), Haemo-compatibility.
Per 21 CFR 73.3107, Carbazole violet (Pigment Violet 23) (CAS Reg. No. 6358-30-1, Color Index No. 51319) is exempt from the certification requirement of 721(c) of FD&C Act.
Conclusion: The all conducted Biological Evaluation Tests are in compliance with the standards of ISO 10993, "Biological Evaluation of Medical Devices". The compatibility of all the possible skin-contact component material in the finished product meets the requirement of Biocompatibility
The applicant device is Substantially Equivalent (SE) to the predicate device which is US legally market device. Therefore, the applicant device is determined as safe and effectiveness.
{4}------------------------------------------------
Image /page/4/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is an abstract image of an eagle with three human profiles incorporated into its design. The eagle is facing to the right.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Shenzhen Ant Hi-Tech Industrial Co., LTD c/o Ms. Diana Hong Suite 8D, No. 19, Lane 999 Zhongshan No.2 Road (S) Shanghai, 200030. China
Re: K072696
Ant Angiographic Syringes, Model 100101,102,103,100201 Regulation Number: 21 CFR 870.1650 Regulation Name: Injector and Syringe, Angiographic Regulatory Class: Class II (two) Product Code: DXT Dated: Undated Received: April 28, 2008
Dear Ms. Hong:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA), You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
MAY - 2 2008
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{5}------------------------------------------------
### Page 2 - Ms. Diana Hong
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
Duina R. hichner
Bram D. Zuckerman, M.D. Director
Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number (if known): K072696
Device Name: ANT Angiographic Syringes
Indications for Use:
ANT Angiographic Syringes are syringes for the injection of contrast media or saline. This syringe is for single use with US legally marketed angiographic injectors.
Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Dune D. Achter
(Division Sign-Off) Division or Cardiovascular Devices
Page 1 of ____________________________________________________________________________________________________________________________________________________________________
510(k) Number_K022696
۰۱: