K072664 · Ge Medical Systems Ultrasound and Primary Care Dia · KGI · Nov 21, 2008 · Radiology
Device Facts
Record ID
K072664
Device Name
GE LUNAR FEMUR STRENGTH SOFTWARE
Applicant
Ge Medical Systems Ultrasound and Primary Care Dia
Product Code
KGI · Radiology
Decision Date
Nov 21, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 892.1170
Device Class
Class 2
Indications for Use
The GE Lunar Femur Strength analysis software option is an accessory to currently marketed GE Lunar DEXA bone densitometer devices, which are intended to estimate the bone mineral density and body composition (lean and fat tissue mass) of patients when medically indicated by their physicians. The Femur Strength software is intended to measure the hip axis length (HAL) and provides female adult Caucasian and Asian mean reference values of HAL from previously acquired GE Lunar femur scans. The software calculates hip geometry values used to evaluate the structural properties of the hip, such as sub-region lengths, angles, ratios, buckling ratio, section modulus (Z), cortical thickness, cross-sectional area (CSA) and cross-sectional moment of inertia (CSMI). The software calculates a Femur Strength Index (FSI), a strength/stress ratio incorporating geometric properties of the hip such as CSMI, CSA, bone mineral density (BMD), patient age, height and weight. A ScanCheck feature is included to provide online messages to assist the operator or physician in checking that the scan was correctly taken. Color mapping of the variable density of bone images is also provided. The values measured or computed provide additional information about the structure of the hip. They should be used in conjunction with the BMD T-score and other clinical risk factors as an aid in the diagnosis of osteoporosis and medical conditions leading to reduced bone density, and ultimately in the assessment of fracture risk.
Device Story
Software accessory for GE Lunar DEXA bone densitometers; processes previously acquired femur scans. Inputs: femur scan images, patient age, height, weight. Computes hip geometry (HAL, sub-region lengths, angles, buckling ratio, section modulus, CSA, CSMI) and Femur Strength Index (FSI). Includes ScanCheck feature for scan quality verification and color mapping for density visualization. Used in clinical settings by physicians/technicians. Output: quantitative structural metrics and fracture risk indicators. Assists clinicians in fracture risk assessment and osteoporosis diagnosis; does not provide definitive diagnosis or treatment recommendations.
Clinical Evidence
Bench testing only. No clinical tests were required. In vitro precision and accuracy were validated using phantoms, with results conforming to design specifications. Device development conforms to 21 CFR 820 and ISO 13485 quality systems.
Technological Characteristics
Software-based analysis for DEXA bone densitometers. Computes geometric and biomechanical properties (HAL, FSI, CSMI, CSA, Z) from scan data. Features include ScanCheck (automated quality messaging) and color mapping. Operates on GE Lunar DEXA platforms. Conforms to applicable medical device safety standards.
Indications for Use
Indicated for adult female Caucasian and Asian patients undergoing GE Lunar DEXA bone densitometry of the hip to assist clinicians in assessing fracture risk and diagnosing osteoporosis/reduced bone density conditions.
Regulatory Classification
Identification
A bone densitometer is a device intended for medical purposes to measure bone density and mineral content by x-ray or gamma ray transmission measurements through the bone and adjacent tissues. This generic type of device may include signal analysis and display equipment, patient and equipment supports, component parts, and accessories.
Predicate Devices
Advanced Hip Assessment Software Option (K0011917)
Hologic Hip Structural Analysis software (K061561)
Reference Devices
GE Lunar Prodigy Advanced hip assessment Software option
Submission Summary (Full Text)
{0}------------------------------------------------
1072664
NOV 21 2008
510(k) Premarket Notification GE Medical Systems Lunar - GE Lunar Femur Strength Software September 19th, 2007
> Attachment B 510 (k) Summary of Safety and Effectiveness Prepared in accordance with 21 CFR Part 807.92(c).
# GE Healthcare
Image /page/0/Picture/5 description: The image shows the text "Section a):" in a bold, sans-serif font. The text is positioned on the left side of the image. The colon is directly after the letter 'a' and is followed by a parenthesis.
1. Submitter: GE Medical Systems Ultrasound and Primary Care Diagnostics, LLC division of General Electric Company GE Medical Systems Lunar (business name) 726 Heartland Trail Madison, WI 53717
Contact Person: James P. Raskob Safety and Regulatory Engineering Manager Telephone: 608-826-7425; Fax: 608-299-2132
September 19th, 2007 Date Prepared:
- 2. Device Name: GE Lunar Femur Strength Software Option Bone Densitometer, 21 CFR 892.1170, 90-KGI
- 3. Marketed Device: Advanced Hip Assessment Software Option K0011917, Hologic Hip Structural Analysis software K061561 currently in commercial distribution.
- 4. Device Description: The Femur Strength Software Option for GE Lunar DEXA Bone Densitometers analyzes femur BMD scans to measure hip axis length (HAL) and provides a mean reference value of HAL and HAL fracture risk indication. The software calculates hip geometry values used to estimate the structural properties of the hip such as buckling ratio, section modulus (Z), cross-sectional area (CSA) and cross-sectional moment of inertia (CSMI). The software calculates a Femur Strength Index (FSI) based on inputs of CSMI, CSA, bone mineral density (BMD), patient age, height and weight. FSI represents the biomechanical properties of the hip and provides an estimate of the risk of fracture resulting from a fall on the hip.
# 5. Indications for Use:
The GE Lunar Femur Strength Analysis Software option used on GE Lunar DEXA bone densitometer scans of the hip measures the hip axis length (HAL) and provides a female mean reference value of HAL. The software calculates a HAL fracture risk indicator for Caucasian females based on HAL compared to the mean reference value.
The software calculates hip geometry values used to estimate the structural properties of the hip such as sub-region lengths, angles, ratios, buckling ratio, section modulus (Z), cross-sectional area (CSA) and cross-sectional moment of inertia (CSMI). The software calculates a Femur Strength Index (FSI) based on inputs of CSMI, CSA, bone mineral density (BMD), patient age, height and weight. FSI represents the biomechanical properties of the hip and provides an estimate of the risk of fracture resulting from a fall on the hip.
The structural properties of the hip, HAL and FSI can assist the health care professional in assessment of fracture risk.
Overall fracture risk will depend on many additional factors that should be considered before making diagnostic or therapeutic recommendations. The software option does not diagnose
{1}------------------------------------------------
510(k) Premarket Notification
GE Medical Systems Lunar - GE Lunar Femur Strength Software September 19th, 2007
> disease, or recommend treatment regimens, or quantify treatment effectiveness. Only the health care professional can make these judgments.
- 6. Comparison with Predicate Device: The Femur Strength Software Option for GE Lynar DEXA Bone Densitometers is of a comparable type and substantially equivalent to the Hologic Hip Structural Software Option and the GE Lunar Prodigy Advanced hip assessment Software option. It has the same technological characteristics, is comparable in key safety and effectiveness features, it utilizes similar design, construction, and materials, and has similar intended uses as the predicate devices.
# Section b):
- 1. Non-clinical Tests: The device has been evaluated for electrical and mechanical safety, and has been found to conform with applicable medical device safety standards. In vitro precision and accuracy values were computed through a series of tests on phantoms and were within design specifications.
- 2. Clinical Tests: No clinical tests were required to establish safety or effectiveness.
- 3. Conclusion: Intended uses and other key features are consistent with previously cleared bone densitometer Advanced Hip Analysis Software. The design and development process of the manufacturer conforms with 21 CFR 820 and ISO 13485 quality systems. The device conforms to applicable medical device safety standards and compliance was verified through independent evaluation with ongoing factory surveillance. The Femur Strength Software Option for GE Lunar DEXA Bone Densitometers is substantially equivalent to currently marketed devices with respect to safety and effectiveness.
{2}------------------------------------------------
Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized caduceus symbol, which is a staff with two snakes entwined around it, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged in a circular fashion around the symbol. The text is in all caps and appears to be in a sans-serif font. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
Mr. James P. Raskob Safety and Regulatory Engineering Manager GE Medical Systems Lunar 726 Heartland Trail MADISON Wİ 53717
NOV 2 1 2008
Re: K072664
Trade/Device Name: Femur Strength Software Option for GE Lunar DEXA Bone Densitometers
Regulation Number: 21 CFR 892.1170
Regulation Name: Bone densitometer
Regulatory Class: II
Product Code: KGI
Dated: October 21, 2008
Received: October 22, 2008
### Dear Mr. Raskob:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{3}------------------------------------------------
# Page 2
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:
| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-0115 |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology) | 240-276-0115 |
| 21 CFR 894.xxx | (Radiology) | 240-276-0120 |
| Other | | 240-276-0100 |
Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometrics' (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely vours.
Vogue Mr. Zhang
Joyce M. Whang, Ph.D. Acting Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{4}------------------------------------------------
#### 510 (k) Indications for Use
### 510(k) Number (if known): K072664
# Device Name: Femur Strength Software Option for GE Lunar DEXA Bone Densitometers
The GE Lunar Femur Strength analysis software option is an accessory to currently marketed GE Lunar DEXA bone densitometer devices, which are intended to estimate the bone mineral density and body composition (lean and fat tissue mass) of patients when medically indicated by their physicians. The Femur Strength software is intended to measure the hip axis length (HAL) and provides female adult Caucasian and Asian mean reference values of HAL from previously acquired GE Lunar femur scans.
The software calculates hip geometry values used to evaluate the structural properties of the hip, such as sub-region lengths, angles, ratios, buckling ratio, section modulus (Z), cortical thickness, cross-sectional area (CSA) and cross-sectional moment of inertia (CSMI). The software calculates a Femur Strength Index (FSI), a strength/stress ratio incorporating geometric properties of the hip such as CSMI, CSA, bone mineral density (BMD), patient age, height and weight. A ScanCheck feature is included to provide online messages to assist the operator or physician in checking that the scan was correctly taken. Color mapping of the variable density of bone images is also provided.
The values measured or computed provide additional information about the structure of the hip. They should be used in conjunction with the BMD T-score and other clinical risk factors as an aid in the diagnosis of osteoporosis and medical conditions leading to reduced bone density, and ultimately in the assessment of fracture risk.
Prescription Use XXX (Part 21 CFR 801 Subpart D)
AND/OR
Over The Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE)
Angu In Whang
(Division Sign-Off) (Division Sign-Off)
Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.