ESOPHAGEAL/RECTAL TEMPERATURE PROBE AND ESOPHAGEAL STETHOSCOPE WITH TEMPERATURE SENSOR

K072621 · Cincinnati Sub-Zero Products, Inc. · BZT · Dec 13, 2007 · Anesthesiology

Device Facts

Record IDK072621
Device NameESOPHAGEAL/RECTAL TEMPERATURE PROBE AND ESOPHAGEAL STETHOSCOPE WITH TEMPERATURE SENSOR
ApplicantCincinnati Sub-Zero Products, Inc.
Product CodeBZT · Anesthesiology
Decision DateDec 13, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 868.1920
Device ClassClass 2

Intended Use

Esophageal/Rectal Temperature Probe, Catalog Numbers 483M-9, 483M-12, and 491B: The CSZ esophageal/rectal temperature probe is intended for use in routine continuous monitoring of the esophageal or rectal temperature as an indicator or core body temperature. The probe is designed for placement in the esophagus or rectum. Esophageal Stethoscope with Temperature Sensor, Catalog Numbers 493M-9, 493M-12, 493M-18, and 493M-24: The CSZ esophageal stethoscope with temperature sensor is intended for use when the esophageal temperature is continuously monitored along with the auscultation of the heart and lung sound as an indicator of core body temperature and cardio-pulmonary performance.

Device Story

Disposable temperature probes utilize thermistors to sense core body temperature. Esophageal/rectal probes provide continuous temperature monitoring; esophageal stethoscopes provide simultaneous temperature monitoring and heart/lung sound auscultation. Signals from the thermistor are transmitted to an external monitoring unit for processing and display. Used in clinical settings by healthcare providers to assess patient core temperature and cardio-pulmonary performance. Output allows clinicians to monitor patient status continuously, facilitating clinical decision-making regarding patient thermal and respiratory stability.

Clinical Evidence

Bench testing only. Compliance with BS EN 12470-4 (Clinical Electronic Thermometers) verified performance requirements.

Technological Characteristics

Disposable probes; thermistor-based temperature sensing; electrical conductors for signal transmission; compatible with external monitoring units. Complies with BS EN 12470-4.

Indications for Use

Indicated for patients requiring continuous core body temperature monitoring via esophageal or rectal placement, or simultaneous core temperature monitoring and heart/lung sound auscultation via esophageal stethoscope.

Regulatory Classification

Identification

An esophageal stethoscope with electrical conductors is a device that is inserted into the esophagus to listen to a patient's heart and breath sounds and to monitor electrophysiological signals. The device may also incorporate a thermistor for temperature measurement.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Image /page/0/Picture/0 description: The image shows the logo for Cincinnati Sub-Zero. The logo is black and white and features the letters "CSZ" in a stylized font. The words "Cincinnati Sub-Zero" are written below the letters in a smaller font. The logo is simple and modern. K072621 Cincinnati Sub-Zero Products, Inc. 12011 Mosteller Road Cincinnati Ohio 45241-1528 Telephone: (513) 772-8810 FAX: (513) 772-9119 WEBSITE: http://www.cszinc.com E-MAIL: cszinc@cszinc.com # 510K Summary | 1. | Owner's Name: | Steve Berke | |----------------------|----------------------|--------------------------------------------------------------------------------------------------| | | Address: | Cincinnati Sub-Zero Products, Inc. (CSZ)<br>12011 Mosteller Rd.<br>Cincinnati, OH 45241 | | | Phone: | 513-772-8810 | | | Fax: | 513-772-9119 | | | Contact: | Fatma Ali, Director of Quality & Regulatory Affairs, CSZ | | Organization Number: | | 84901 | | Establishment: | | New Device | | Registration Number: | | 1516825 | | Operations: | | Manufacturer and Specification Developer | | Date: | | Monday, December 03, 2007 | | 2. | Name of the Devices: | | | | - | Esophageal/Rectal Temperature Probe, and<br>Esophageal Stethoscope With Temperature Sensor Probe | | | | Probes trade/proprietary name: THERMA-TEMP, STERI-PROBE | | | Common name: | Temperature Probe | | | | Classification: Esophageal Rectal and Esophageal Stethoscope, with Electrical Conductors | | | Product Code: | BZT | Image /page/0/Picture/5 description: The image shows a logo with the text "Certified WBENC Women's Business Enterprise". The logo is enclosed in a rounded rectangle. The word "Certified" is at the top, followed by the acronym "WBENC" in a larger font, and then "Women's Business Enterprise" at the bottom. {1}------------------------------------------------ #### Predicate Devices: # Smiths Level 1 Esophageal/Rectal Temperature Probe; and # Esophageal Stethoscope With Temperature Sensor Probe goth devices are originally, listed under Respiratory Support Products, Inc. as referenced in 510K Sumber's: K864043 & K864044 ### Device Description: isposable Temperature Probes using thermistors as temperature sensors. The signal of the sensor s processed and displayed by the monitoring unit. ### Intended Use: ontinuous measurement of core body temperature through the esophagus or the rectum. ## Comparison to the Predicate Device: he CSZ Esophageal/Rectal Probes and Esophageal Stethoscope are substantially equivalent to he Smiths Level 1 Esophageal/Rectal Probes and Esophageal Stethosoped ## Discussion of Non-clinical Tests performed: esting was done in accordance with BS EN 12470-4, Clinical Thermometers. ## Conclusion: The CSZ Esophageal/Rectal Probe and Esophageal Stethoscope have the same intended use and echnological characteristics as the cleared devices (Smiths Level 1 Esophageal/Rectal Probes and Esophageal Stethoscope). The performance testing has shown that the CSZ products included in his pre-market submission meet the requirements of EN 12470-4 for Clinical Electronic hermometers, and therefore maintain the same levels of safety and effectiveness as the predicate evice currently in commercial distribution. certify that, in my capacity as the Director of Quality and Regulatory Affairs of Cincinnati Sub-Zero Products Inc., I believe i of the best of my knowledge, that all risher and regulation Analis of Chication are truthful and accurate nd that no material fact has been omitted. S. Ce. Date: 12 / 3 / 07 or of Quality and Regulatory Affairs 1 Mosteller R. nnati, OH 45 Image /page/1/Picture/18 description: The image is a logo for "WBENC Women's Business Enterprise". The logo is in a black rectangular box with rounded corners. The words "Certified" and "Women's Business Enterprise" are written in a smaller font above and below the large letters "WBENC". {2}------------------------------------------------ Image /page/2/Picture/1 description: The image contains the words "Public Health Service" in a simple, bold, sans-serif font. The text is arranged on a single line, with each word clearly legible. The overall impression is clean and straightforward, suggesting an official or institutional context. Image /page/2/Picture/2 description: The image shows the seal of the Department of Health and Human Services (HHS) of the United States. The seal features a stylized eagle with three stripes forming its wing, symbolizing health, services, and people. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES USA" is arranged around the seal's perimeter. DEC 1 3 2007 Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Ms. Fatma Ali Director of Quality & Regulatory Affairs Cincinnati Sub-Zero Products, Incorporated 12011 Mosteller Road Cincinnati, Ohio 45241 Re: K072621 Trade/Device Name: Esophageal/Rectal Temperature Probe and Esophageal Stethoscope with Temperature Sensor Regulation Number: 21 CFR 868.1920 Regulation Name: Esophageal Stethoscope with Electrical Conductors Regulatory Class: II Product Code: BZT Dated: December 4, 2007 Received: December 6, 2007 Dear Ms. Ali: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Page 2 - Ms. Ali Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801). please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Chiu Liu, Ph.D. Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ # Indications for Use 10(k) Number (if known): Unknown Device Name(s): rsophageal/Rectal Temperature Probe, Catalog Numbers 483M-9, 483M-12, and 491B Essophageal Stethoscope with Temperature Sensor, Catalog Numbers 493M-9, 493M-12, 493M-18, and 493M-24 #### Indications for Use: Esophageal/Rectal Temperature Probe, Catalog Numbers 483M-9, 483M-12, and 491B: The CSZ sesphageal/rectal temperature probe is intended for use in routine continuous monitoring of the esophageal or medal temperature as an indicator or core body temperature. The probe is designed for placement in the esophagus or rectum. Boohageal Stethoscope with Temperature Sensor, Catalog Numbers 493M-9, 493M-12, 493M-18, and 493M-24: The CSZ esophageal stethoscope with temperature sensor is intended for use when the esophageal caperature is continuously monitored along with the auscultation of the heart and lung sound as an indicator of core body temperature and cardio-pulmonary performance. Prescription Use: and/or Over-the-Counter Use: (Part 21CFR801 Subpart D) XXXXX (Part 21CFR801 Subpart C) Arin Bern (Division Sign-Off) Division of Anesthesiology, General Hospital Infection Control, Dental Devices 510(k) Number: __ & 47363 Confidential 9/5/2007
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...