GAME READY CLASSIC SYSTEM, MODEL 550550

K072620 · Coolsystems, Inc. · IRP · Oct 31, 2007 · Physical Medicine

Device Facts

Record IDK072620
Device NameGAME READY CLASSIC SYSTEM, MODEL 550550
ApplicantCoolsystems, Inc.
Product CodeIRP · Physical Medicine
Decision DateOct 31, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 890.5650
Device ClassClass 2
AttributesTherapeutic

Intended Use

The Game Ready Classic System is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain where cold and compression are indicated. It is intended to be used by or on the order of licensed healthcare professionals in hospitals, outpatient clinics, athletic training settings, or home settings.

Device Story

Game Ready Classic System provides intermittent pneumatic compression and controlled cold therapy. System consists of control unit, AC adapter, connector hose, and body-part-specific wraps (sleeve plus heat exchanger). Control unit circulates ice water through wrap for cooling and uses pneumatic pump for intermittent compression. User adjusts compression, cooling, and treatment time. Used in hospitals, clinics, athletic training settings, or home; operated by healthcare professionals or patients. Software monitors fluid flow and pressure, triggering alarms or warnings for errors to ensure user safety. Benefits include pain and edema management for post-surgical or acute injury recovery.

Clinical Evidence

No clinical testing data are submitted, referenced, or relied on to determine substantial equivalence. Bench testing was performed at the component and system level to verify performance against specifications.

Technological Characteristics

DC-powered control unit; pneumatic pump for compression; fluid circulation for cooling. Components include control unit, AC adapter, connector hose, and wraps. Software-based safety monitoring for fluid flow and pressure. No specific material standards or algorithm architecture details provided.

Indications for Use

Indicated for patients with post-surgical or acute injuries requiring reduction of edema, swelling, and pain via cold and compression therapy.

Regulatory Classification

Identification

A powered inflatable tube massager is a powered device intended for medical purposes, such as to relieve minor muscle aches and pains and to increase circulation. It simulates kneading and stroking of tissues with the hands by use of an inflatable pressure cuff.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ K072620 # 510(k) Summary of Safety and Effectiveness # Game Ready Classic System This summary of 510(k) safety and effectiveness information is being submitted in accordance with the requirements of SMDA 1990 and 21 CFR 807.92. The assigned 510(k) number is: OC1 3 1 2007 | A. Submitter: | CoolSystems, Inc.<br>2201 Dwight Way<br>Berkeley, CA 94704 | |----------------|------------------------------------------------------------| | Phone: | 510-868-5378 | | Fax: | 510-984-5330 | | Email: | mbaldwin@gameready.com | | Contact: | Marianne Baldwin, VP, QA/RA/CA | | Date Prepared: | September 14th, 2007 | ### B. Device Names: | Trade Name: | Game Ready Classic System | |----------------------|---------------------------------------------------------------| | Common/usual Name: | Cold/compression therapy system | | Classification Name: | Powered inflatable tube massager /Cold water circulating pack | ### C. Predicate Device: The Game Ready Classic System is substantially equivalent to the Game Ready System, cleared under 510(k) K071050, and currently legally marketed by CoolSystems, Inc. # D. Device Description: The Game Ready Classic System is comprised of the following components: - Control Unit, with AC Adapter (to convert line power to DC input power) . - . Connector Hose (connects Control Unit to Wrap) - Wrap (Heat Exchanger + Sleeve) ● - Optional Carry Bag . - Optional Battery Pack (under development) . The Game Ready Classic System is a DC-powered device that provides intermittent compression and controlled cold therapy by controlling fluid flow and air pressure through a Wrap that is specially designed for a specific body part or location (e.g., shoulder, elbow, knee, ankle, back). A Wrap is comprised of a Sleeve (the outer covering for the Heat Exchanger) and a Heat Exchanger. Cooling is provided by {1}------------------------------------------------ circulating ice water from the Control Unit through the Connector Hose to the Wrap. Compression is provided by an intermittent-cycling pneumatic pump that pumps air from the Control Unit through the Connector Hose to the Wrap. The user can adjust the amount of compression, the amount of cooling, and the treatment time. The Control Unit software is designed to protect both the user and the Control Unit from harm by shutting off the system and sounding an alarm if there is insufficient fluid flow or in case of an over-pressure or under-pressure situation. The software detects other, less-serious error conditions and sounds a warning beep to alert the user, so that the error condition can be corrected. These protections have been carried forward from the original Game Ready to Game Ready Classic. #### E. Intended Use: The Game Ready Classic System is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain where cold and compression are indicated. It is intended to be used by or on the order of licensed healthcare professionals in hospitals, outpatient clinics, athletic training settings, or home settings, ## F. Comparison with the Predicate Device: The Game Ready System Classic System is substantially equivalent to the predicate device in design, function, and indications for use/intended use. #### G. Non-Clinical Testing Bench testing was performed at the component and system level. The testing showed that the Game Ready Classic System met its requirements. Design Verification Testing showed that the Game Ready Classic System performed according to specification. ## H. Clinical Testing No clinical testing data are submitted, referenced, or relied on to determine substantial equivalence. #### ﻨﮯ Conclusions There are no significant differences between the Game Ready Classic System and the predicate device that would adversely affect the use of the proposed device or introduce new potential risks or safety concerns. The Game Ready Classic System is substantially equivalent to the predicate device in design, function, and indications for use/intended use. {2}------------------------------------------------ Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 OCT 3 1 2007 CoolSystems, Inc % Ms. Marianne Baldwin 2201 Dwight Way Berkeley, CA 94704 Re: K072620 > Trade/Device Name: Game Ready Classic System 550550 Regulation Number: 21 CFR 890.5650 Regulation Name: Powered inflatable tube massager Regulatory Class: II Product Code: IRP, ILO Dated: September 14, 2007 Received: September 17, 2007 Dear Ms. Baldwin: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. Image /page/2/Picture/11 description: The image is a black and white logo for the Department of Health & Human Services USA. The logo consists of a stylized eagle with its wings spread, and the words "DEPARTMENT OF HEALTH & HUMAN SERVICES USA" are arranged in a circle around the eagle. The eagle is facing to the right and has a sleek, modern design. {3}------------------------------------------------ Page 2 - Ms. Marianne Baldwin This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely vours. Sincerely yours, for Pete, Mark Melkerson DEP D. RECTOR 10/31/07 Director Division of General Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Indications for Use 510(k) Number (if known): Game Ready™ Classic System Device Name: _ Indications for Use: The Game Ready Classic System is intended to treat post-surgical and acute injuries to reduce edema, swelling, and pain where cold and compression are indicated. It is intended to be used by or on the order of licensed healthcare professionals in hospitals, outpatient clinics, athletic training settings, or home settings. Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) # (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED) | Concurrence of CDRH, Office of Device Evaluation (ODE) | |--------------------------------------------------------| |--------------------------------------------------------| Pelt (Division Sign-Uff) Division of General. Restorative. and Neurological Devic 510/k) Number
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...