GYRUS ACMI IPN-2505, INVISIO SMITH PERCUTANEOUS NEPHROSCOPE SYSTEM

K072594 · Gyrus Acmi, Inc. · FGA · Dec 13, 2007 · Gastroenterology, Urology

Device Facts

Record IDK072594
Device NameGYRUS ACMI IPN-2505, INVISIO SMITH PERCUTANEOUS NEPHROSCOPE SYSTEM
ApplicantGyrus Acmi, Inc.
Product CodeFGA · Gastroenterology, Urology
Decision DateDec 13, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 876.1500
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Gyrus ACMI® IPN-2505, Invisio® Smith™ Percutaneous Nephroscope System (which includes the IPN-2505 Invisio® Smith™ Percutaneous Nephroscope and IDC-1500 Invisio® Camera Control Unit) is intended for close visualization of the urinary bladder, renal pelvis and major calyces. The IPN-2505 can be introduced through a percutaneous tract into the kidney and can also be introduced through the urethra to access the bladder. Additional accessories can be used to perform various diagnostic and therapeutic procedures.

Device Story

Rigid endoscope system for urological visualization; utilizes distal-tip CMOS sensor to capture images; transmits signals via internal wiring to IDC-1500 Camera Control Unit (CCU). CCU processes signals using internal PCBs and software to display images on external monitor. Used in clinical settings by physicians for diagnostic and therapeutic procedures in urinary bladder, renal pelvis, and major calyces. Introduced percutaneously into kidney or via urethra into bladder. Provides real-time visualization to assist physician in performing procedures; benefits patient by enabling minimally invasive access to renal and bladder structures.

Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing and comparison of technological characteristics to legally marketed predicate devices.

Technological Characteristics

Rigid endoscope with distal-tip CMOS sensor, LED light source in handle, and electrical connectivity to CCU. Construction materials similar to predicate MRO-20 and MR-6A/MR-6LA. System includes IPN-2505 endoscope and IDC-1500 CCU containing PCBs, power supply, and software for image processing.

Indications for Use

Indicated for visualization of the urinary bladder, renal pelvis, and major calyces in patients requiring diagnostic or therapeutic urological procedures via percutaneous or urethral access.

Regulatory Classification

Identification

An endoscope and accessories is a device used to provide access, illumination, and allow observation or manipulation of body cavities, hollow organs, and canals. The device consists of various rigid or flexible instruments that are inserted into body spaces and may include an optical system for conveying an image to the user's eye and their accessories may assist in gaining access or increase the versatility and augment the capabilities of the devices. Examples of devices that are within this generic type of device include cleaning accessories for endoscopes, photographic accessories for endoscopes, nonpowered anoscopes, binolcular attachments for endoscopes, pocket battery boxes, flexible or rigid choledochoscopes, colonoscopes, diagnostic cystoscopes, cystourethroscopes, enteroscopes, esophagogastroduodenoscopes, rigid esophagoscopes, fiberoptic illuminators for endoscopes, incandescent endoscope lamps, biliary pancreatoscopes, proctoscopes, resectoscopes, nephroscopes, sigmoidoscopes, ureteroscopes, urethroscopes, endomagnetic retrievers, cytology brushes for endoscopes, and lubricating jelly for transurethral surgical instruments. This section does not apply to endoscopes that have specialized uses in other medical specialty areas and that are covered by classification regulations in other parts of the device classification regulations.

Special Controls

*Classification* —(1)*Class II (special controls).* The device, when it is an endoscope disinfectant basin, which consists solely of a container that holds disinfectant and endoscopes and accessories; an endoscopic magnetic retriever intended for single use; sterile scissors for cystoscope intended for single use; a disposable, non-powered endoscopic grasping/cutting instrument intended for single use; a diagnostic incandescent light source; a fiberoptic photographic light source; a routine fiberoptic light source; an endoscopic sponge carrier; a xenon arc endoscope light source; an endoscope transformer; an LED light source; or a gastroenterology-urology endoscopic guidewire, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 876.9.(2) Class I for the photographic accessories for endoscope, miscellaneous bulb adapter for endoscope, binocular attachment for endoscope, eyepiece attachment for prescription lens, teaching attachment, inflation bulb, measuring device for panendoscope, photographic equipment for physiologic function monitor, special lens instrument for endoscope, smoke removal tube, rechargeable battery box, pocket battery box, bite block for endoscope, and cleaning brush for endoscope. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807of this chapter, subject to the limitations in § 876.9.

In combination with the general controls of the FD&C Act, the integrated operating table-electromechanical surgical system is subject to the following special controls: 1. (1) Premarket clinical performance testing, or a combination of premarket clinical performance testing and postmarket surveillance (in accordance with special control (2)), must include the following: 1. (i) Objective performance measures (e.g., rate and number of conversions to other surgical modalities, rate of device related adverse events (including tissue injury, hematoma, and increased blood loss), and their severity, cause, and outcomes) must be reported with relevant descriptive comparator performance measures. 2. (ii) The data must demonstrate the performance of the device for providing accurate and precise control of attached surgical instruments in range of clinical conditions relevant to the device's intended use. 3. (iii) The test dataset must include data collected from a patient population representative of the intended patient population under anticipated conditions of use. 2. (2) Data obtained from postmarket surveillance must demonstrate, in consideration of the premarket data obtained in accordance with special control (1), that the device performs in accordance with special control (1), unless FDA determines, based on the totality of the premarket data, that data from postmarket surveillance is not required to demonstrate that the device performs as intended. Such postmarket surveillance must be conducted per a protocol determined appropriate by FDA to demonstrate that the device performs as intended (in consideration of the premarket data obtained in accordance with special control (1)), and must include initiation, enrollment, and reporting requirements to ensure timely periodic updates to FDA on post-market surveillance progress and outcomes. 3. (3) Animal performance testing must evaluate the extent of port site trauma due to repositioning of table during surgical procedures when utilizing robotic minimally invasive and laparoscopic approaches 4. (4) The device manufacturer must develop, and update as necessary, a device-specific use training program that ensures proper device setup/use/shutdown, accurate control of instruments to perform the intended surgical procedures, troubleshooting and handling during unexpected events or emergencies, and safe practices to mitigate use error. 5. (5) The device manufacturer may only distribute the device to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program. (6) Human factors assessment must demonstrate that the user can correctly use the device system across all intended use environments with the provided instructions and training materials, including patient access during normal operating conditions and emergency situations, and effects arising from the integrated nature of the operating table and robotic surgical arms. (7) Labeling must include: (i) A detailed summary of clinical performance testing conducted with the device, including study population, results, adverse events, and comparisons to any comparator groups identified; (ii) A statement in the labeling that the safety and effectiveness for the representative specific procedures was based on evaluation of the device as a surgical tool and did not include evaluation of outcomes related to the treatment of the patient's underlying disease or condition, unless FDA determines that it can be removed or modified based on clinical performance data submitted to FDA; (iii) Identification of compatible devices; (iv) The list of surgical procedures for which the device has been determined to be safe with clinical justification; (v) Reprocessing instructions for reusable components; (vi) A shelf life for any sterile components; (vii) A description of the device-specific use training program; (viii) A statement that the device is only for distribution to facilities that implement and maintain the device-specific use training program and ensure that users of the device have completed the device-specific use training program; and (ix) A summary of any completed postmarket surveillance data collected as required by special control (2), including updated labeling to accurately reflect outcomes observed in postmarket surveillance. (8) Non-clinical performance testing must demonstrate that the device performs as intended under anticipated conditions of use and must include: (i) Device motion accuracy and repeatability; (ii) System testing; (iii) Instrument reliability; (iv) Crosstalk; (v) Table motion control; (vi) Thermal effects on tissue; (vii) User-device interface performance; (viii) Workspace access testing; and (ix) Performance testing with compatible devices. (9) Software verification, validation, and hazard analysis must be performed. (10) Electromagnetic compatibility and electrical, thermal, and mechanical safety testing must be performed. (11) Performance data must demonstrate the sterility of all patient-contacting device components. (12) Performance data must support the shelf life of the device components provided sterile by demonstrating continued sterility and package integrity over the labeled shelf life. (13) Performance data must validate the reprocessing instructions for the reusable components of the device. (14) Performance data must demonstrate that all patient-contacting components of the device are biocompatible. (15) Performance data must demonstrate that all patient-contacting components of the device are non-pyrogenic. (16) The device manufacturer must submit a report to the FDA annually on the anniversary of initial marketing authorization for the device, until such time as FDA may terminate such reporting, which comprises the following information: (i) Cumulative summary, by year, of complaints and adverse events since date of initial marketing authorization; and (ii) Identification and rationale for changes made to the device, labeling, or device specific use training program, which did not require submission of a premarket notification during the reporting period.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ #### 07259 510(k) Summary of Safety and Effectiveness Gyrus ACMI, Inc. Gyrus ACMI® IPN-2505, Invisio® Smith™ Percutaneous Nephroscope System # General Information DEC 1. 3 2007 | Manufacturer: | Gyrus ACMI, Inc. | |-----------------------------------------------------------|-----------------------------------------| | | 136 Turnpike Rd. | | | Southborough, MA 01772-2104 | | Contact Person: | Graham A.L. Baillie | | | Senior Regulatory Specialist | | | Tel. #: 508-804-2738 | | | Fax #: 508-804-2624 | | Date Prepared: | September 10, 2007 | | Device Description | | | Classification Name: | Endoscope and accessories | | | (21 CFR 876.1500), Class II | | | Surgical camera and accessories | | | (21 CFR 878.4160), Class I | | Trade Name: | Gyrus ACMI® IPN-2505, Invisio® | | | Smith™ Percutaneous Nephroscope | | | System | | Generic/Common Name: | Endoscope, Video Camera and accessories | | Predicate Devices | | | ACMI® DUR-Digital Ureteroscope and Choledochoscope System | K060269 | | ACMI® MROTM -20 Rigid Percutaneous Nephroscopes | K791182 | | ACMI® MR-6A/MR-6LA Autoclavable Ureteroscopes | K052044 | ## Intended Uses The Gyrus ACMI® IPN-2505, Invisio® Smith™ Percutaneous Nephroscope System (which includes the IPN-2505 Invisio® Smith™ Percutaneous Nephroscope and IDC-1500 Invisio® Camera Control Unit) is intended for close visualization of {1}------------------------------------------------ Traditional 510(k) Notification 0(k) Summarv K072594 Pg A22 the urinary bladder, renal pelvis and major calyces. The IPN-2505 can be introduced through a percutaneous tract into the kidney and can also be introduced through the urethra to access the bladder. Additional accessories can be used to perform various diagnostic and therapeutic procedures. ### Product Description The Gyrus ACMI® IPN-2505. Invisio® Smith™ Percutaneous Nephroscope (referred to hereafter as the IPN-2505) is a rigid endoscope that incorporates CMOS (complimentary metal oxide semi-conductor) sensor technology to generate an image. There is a miniature CMOS sensor located in the distal tip, wiring running through the endoscope shaft, a printed circuit board (PCB), a light-emitting diode (LED) light source located in the handle, and an electrical cord that connects the endoscope to the IDC-1500 Camera Control Unit (CCU). The IPN-2505 can be introduced either through the urethra into the bladder or through a percutaneous tract into the abdominal cavity or kidney. The IPN-2505 uses an IDC-1500 CCU that contains printed circuits boards (PCBs), software, power supply and power cables to process and display the images transmitted by the camera. #### Technological Characteristics and Substantial Equivalence The IPN-2505, utilizes features incorporated into the following legally marketed predicate devices: The IPN-2505 incorporates the same basic CMOS video imaging technology located in the endoscope as the predicate ACMI® DUR-Digital Ureteroscope and Choledochoscope System (K060269). The IPN-2505 shaft is dimensionally similar to its predicate ACMI® MRO-20 Rigid Percutaneous Nephroscope (K791182), having the same working channel diameter, similar lengths and diameters. The IPN-2505 utilizes similar materials in its construction as the predicate MRO-20 (K791182) and the predicate MR-6A/MR-6LA Autoclavable Ureteroscope (K052044). Like the predicate MRO-20 the IPN-2505 is indicated to allow close visualization of the renal pelvis and major calyces through a percutaneous tract. In summary, the Gyrus ACMI® IPN-2505, Invisio® Smith™ Percutaneous Nephroscope System is substantially equivalent to the predicate devices and presents no new questions of safety or efficacy. {2}------------------------------------------------ #### DEPARTMENT OF HEALTH & HUMAN SERVICES Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus symbol, which is often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white. #### Public Health Service Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 DEC 1 3 2007 Mr. Graham A.L. Baillie Sr. Regulatory Specialist Gyrus ACMI, Inc. 136 Turnpike Road SOUTHBOROUGH MA 01772-2104 K072594 Re: > Trade/Device Name: Gyrus ACMI® IPN-2505, Invisio® Smith™ Percutaneous Nephroscope System Regulation Number: 21 CFR §876.1500 Regulation Name: Endoscope and accessories Regulatory Class: II Product Codes: FGA, FEC and FED Dated: September 12, 2007 Received: September 14, 2007 Dear Mr. Baille: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA). it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. {3}------------------------------------------------ Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter. | 21 CFR 876.xxxx | (Gastroenterology/Renal/Urology) | 240-276-0115 | |-----------------|----------------------------------|--------------| | 21 CFR 884.xxxx | (Obstetrics/Gynecology) | 240-276-0115 | | 21 CFR 892.xxxx | (Radiology) | 240-276-0120 | | Other | | 240-276-0100 | Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrb/industry/support/index.html. Sincerely vours. Nancy C Brogdon Nancy C. Brogdon Director. Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ Device Name: Gyrus ACMI® IPN-2505, Invisio® Smith™ Percutaneous Nephroscope System 510(k) Number: Ko72594 Intended Use: The Gyrus ACMI® IPN-2505, Invisio® Smith™ Percutaneous Nephroscope System (which includes the IPN-2505 Invisio® Smith™ Percutaneous Nephroscope and IDC-1500 Invisio® Camera Control Unit) is intended for close visualization of the urinary bladder, renal pelvis and major calyces. The IPN-2505 can be introduced through a percutaneous tract into the kidney and can also be introduced through the urethra to access the bladder. Additional accessories can be used to perform various diagnostic and therapeutic procedures. ## (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Prescription Use: ____ X _____________________________________________________________________________________________________________________________________________________ OR Over-the-Counter Use: _ (Per 21 CFR 801.109) Lom Mwhan (Division Sign Off) Division of Reproductive, Abdominal and Radiological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...