K072587 · Cytori Therapeutics, Inc. · MUU · Dec 10, 2007 · General, Plastic Surgery
Device Facts
Record ID
K072587
Device Name
CYTORI AFT SYSTEM
Applicant
Cytori Therapeutics, Inc.
Product Code
MUU · General, Plastic Surgery
Decision Date
Dec 10, 2007
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 878.5040
Device Class
Class 2
Attributes
Therapeutic
Indications for Use
The Cytori AFT System is used in the aspiration, harvesting, filtering and transferring of autologous tissue. The Cytori AFT System is intended for use in the following surgical specialties when the aspiration of soft tissue is desired: · Neurosurgery · Gastrointestinal and Affiliated Organ Surgery · Urological Surgery · Plastic and Reconstructive Surgery · General Surgery · Orthopedic Surgery · Gynecological Surgery · Thoracic Surgery · Laparoscopic Surgery The Cytori AFT System is indicated for use when the fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring is desired.
Device Story
Cytori AFT System is a manual, single-use device for autologous tissue aspiration, harvesting, filtering, and transfer. System components include a tissue collection canister with a stainless steel cannula and a 60cc delivery syringe. The device connects to hospital house vacuum to aspirate subcutaneous fatty tissue through the cannula into the collection canister. The canister contains a 265-micron filter to retain adipose tissue while allowing liquids to pass to a waste trap. Stepped clamps on connection tubing allow manual vacuum control and sealing for safe transport. The collected tissue is transferred via the delivery syringe for re-injection. Used in clinical/surgical settings by physicians. Benefits include efficient tissue harvesting and processing for autologous transfer.
Clinical Evidence
Bench testing only. No clinical data provided.
Technological Characteristics
Manual suction lipoplasty system. Materials: medical grade stainless steel (cannula), polypropylene (handle/syringe barrel), DEHP-free PVC (tubing), polycarbonate (canister), polyisoprene (syringe stopper). Energy source: external house vacuum. Sterilization: ethylene oxide (EtO). Components: 15-36 cm length cannulas, 3.0-4.6 mm diameter, 265-micron filter. Standalone device.
Indications for Use
Indicated for patients undergoing aesthetic body contouring requiring fragmentation, emulsification, and aspiration of subcutaneous fatty tissues. Used across various surgical specialties including plastic, reconstructive, general, orthopedic, gynecological, thoracic, laparoscopic, neurosurgery, gastrointestinal, and urological surgery.
Regulatory Classification
Identification
A suction lipoplasty system is a device intended for aesthetic body contouring. The device consists of a powered suction pump (containing a microbial filter on the exhaust and a microbial in-line filter in the connecting tubing between the collection bottle and the safety trap), collection bottle, cannula, and connecting tube. The microbial filters, tubing, collection bottle, and cannula must be capable of being changed between patients. The powered suction pump has a motor with a minimum of 1/3 horsepower, a variable vacuum range from 0 to 29.9 inches of mercury, vacuum control valves to regulate the vacuum with accompanying vacuum gauges, a single or double rotary vane (with or without oil), a single or double diaphragm, a single or double piston, and a safety trap.
Special Controls
*Classification.* Class II (special controls). Consensus standards and labeling restrictions.
Genesis Biosystems Lipivage (510(k) number unknown)
Submission Summary (Full Text)
{1}------------------------------------------------
#### ADMINISTRATIVE INFORMATION
| Manufacturer Name: | Cytori Therapeutics, Inc.<br>3020 Callan Road<br>San Diego, CA 92121 |
|--------------------|--------------------------------------------------------------------------------------------------------------|
| Official Contact: | Kenneth K. Kleinhenz<br>Sr. Director of Regulatory Affairs<br>Telephone (858) 458-0900<br>Fax (858) 458-0994 |
#### DEVICE NAME
Classification Name: Suction Lipoplasty System Trade/Proprietary Name: ﺑ Cytori AFT System
DEC 1 0 2007
# ESTABLISHMENT REGISTRATION NUMBER 3002642958
## DEVICE CLASSIFICATION AND PRODUCT CODE
As shown in 21 CFR 878.5040 Suction Lipoplasty Systems are defined as devices consisting of collection bottles, cannulas, and connecting tubing for use in aesthetic body contouring procedures. Suction Lipoplasty Systems are classified as Class II. They have been assigned Product Code MUU.
## INTENDED USE
The Cytori AFT System is used in the aspiration, harvesting, filtering and transferring of autologous tissue.
The Cytori AFT System is intended for use in the following surgical specialties when the aspiration of soft tissue is desired:
- · Neurosurgery
- · Gastrointestinal and Affiliated Organ Surgery
- · Urological Surgery
- · Plastic and Reconstructive Surgery
- · General Surgery
- · Orthopedic Surgery
- · Gynecological Surgery
- · Thoracic Surgery
- · Laparoscopic Surgery
The Cytori AFT System is indicated for use when the fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring is desired.
{2}------------------------------------------------
#### DEVICE DESCRIPTION
#### Design Characteristics
The Cytori AFT System is provided in two packaged assemblies: 1). Cytori AFT Tissue Collection Canister and 2). Cytori AFT Delivery Syringe. The Cytori AFT Tissue Collection Canister is a single-use disposable device designed to utilize house vacuum to create suction within the hollow liposuction cannula and remove subcutaneous fatty tissue from the patient and transport the adipose tissue into the collection canister. The Cytori AFT Delivery Syringe is a, sterile, single-use 60cc piston syringe.
#### AFT Tissue Collection Canister
The Cytori AFT Tissue Collection Canister is a sterile, single-use, manual device consisting of a cannula. connection tubing and a tissue collection container that relies on house vacuum for its energy supply. The cannula handle is attached to the tissue collection container via connection tubing. The cannula is a hollow tube with a single opening near the tip to communicate house vacuum to the tissues and subsequently aspirate, harvest and filter subcutaneous fatty tissues from the tissue collection container. The cannula tip is fabricated from medical grade stainless steel and the handle is fabricated from medical grade polypropylene. The connection tubing is fabricated from medical grade polyvinyl chloride (PVC) that is DEHP free. The Collection Canister is fabricated from medical grade polycarbonate to assure robustness in the operating room environment. The tissue collection container contains various capped/sealed luer ports and a 265 micron filter to entrap the tissue while allowing liquid to pass through to a hospital-provided collection trap.
The stainless steel cannula that contacts the patient is provided in various sizes ranging from 15 - 36 cm in length and 3.0 - 4.6 mm in diameter with a single opening near the tip of the cannula. The tip region of the cannula may have a single or multiple openings that range in size from 4 to 12 mm in length distributed uniformly or randomly throughout the end of the cannula. The device is provided with a diameter of 20 mm and may be provided in diameters ranging from 20 to 60 mm. The connecting tubing is provided with an inner diameter of 9.6 mm (3/8"), an outer diameter of 14.3 mm (9/16"), and a wall thickness of 2.2 mm. The tubing that connects the cannula handle to the tissue collection canister is provided in a length of 4 feet and may be provided in lengths ranging from 1 - 8 feet. The bottom of the tissue collection canister is also provided with the same 3/8" inner diameter connection tubing of various lengths. The proximal end of the exiting connection tubing may be provided with a barbed tubing connector to assist in the attachment of like-sized tubing for purposes of connecting the Cytori AFT System to house vacuum and/or assorted waste traps. Connection tubing leading to and from the tissue collection container is provided with a stepped clamp to facilitate manual control of the vacuum supplied to the cannula. The stepped clamps also allow the operator to seal the connection tubing on both sides of the Collection Canister and prevent spillage of the collected fluids/tissues, thereby facilitating safe transport and disposal of the waste materials.
#### AFT Delivery Syringe
The Cytori AFT Delivery Syringe is a polymeric 60cc volume luer-lock style sterile, single-use syringe consisting of a polypropylene barrel with printed graduations, a polypropylene plunger, and a polyisoprene latex free rubber stopper at the end of the plunger.
{3}------------------------------------------------
#### Material Composition
The components of the Cytori AFT System that have direct patient contact are fabricated from surgical stainless steel.
## Sterility
The Cytori AFT System is sterilized by ethylene oxide (EtO) gas.
## In Vitro Testing
Mechanical testing of the Cytori AFT System demonstrates that the device is substantially equivalent to the predicate devices.
## EQUIVALENCE TO MARKETED PRODUCT
The Cytori AFT System shares indications and design principles with the following predicate devices which have been determined by FDA to be substantially equivalent to premarket devices: Shippert Medical Tissu-Trans (K050797), MacroPore Puricel Lipoplasty System (K042261), Exel International Disposable Syringe (K861153) and Genesis Biosystems Lipivage (510(k) number unknown); Class I and II medical devices that were cleared for marketing in the United States under K050797, K042261, and K861153. respectively.
#### Indications for Use
The Cytori AFT System and the predicate devices are substantially equivalent with respect to their indications for use, as they are all indicated for the same surgical procedures of aspirating, harvesting, filtering, and transferring autologous tissues.
## Design and Materials
The design and materials of the Cytori AFT System and the predicate devices [Shippert Medical Tissu-Trans (K050797), MacroPore Puricel Lipoplasty System (K042261), Exel International Disposable Syringe (K861153) and Genesis Biosystems Lipivage (510(k) number unknown)] are substantially equivalent, as they are all single-use, polymer constructed, manually operated systems that utilize manual or external sources of vacuum to withdraw, hold, and/or inject fluids/tissues into the body. The Cytori AFT System is substantially equivalent to the Tissu-Trans (K050797), Puricel Lipoplasty System (K042261), and the Lipivage predicate devices, as they all consist of a polycarbonate tissue collection canister with flexible tubing attached to a stainless steel cannula that contacts the patient. The Cytori AFT System is substantially equivalent to the Tissu-Trans (K050797), Puricel Lipoplasty System (K042261), and the Lipivage predicate devices, as they all share design principles of utilizing house-vacuum to aspirate adipose tissue from the patient and subsequently transport the adipose tissue through a tube into a cylindrical collection chamber that contains a filtering mechanism of pores to allow fluids to pass, but retains the adipose tissue within the chamber. The Cytori AFT System is substantially equivalent to the Tissu-Trans (K050797), Puricel Lipoplasty System (K042261), and the Lipivage predicate devices, as they all consist of a polymeric housing chamber with a filter unit within the chamber. The Cytori AFT System is substantially equivalent to the Tissu-Trans (K050797), Puricel Lipoplasty System (K042261), and the Lipivage predicate devices, as they all have connection ports on the superior ends of the housing chamber for the attachment of suction tubing/stainless steel cannulas that contact the patient and connection ports on the inferior end of the housing chamber to connect to vacuum tubing that draws house vacuum and carries waste to the waste trap. The Cytori AFT System is substantially equivalent to the Tissu-Trans (K050797) and the Lipivage predicate devices, as they are all equipped with a separately packaged plunger device to deliver the tissue back to the same patient.
{4}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo features a stylized caduceus, which is a symbol often associated with healthcare. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the caduceus. The logo is black and white.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 1 0 2007
Cytori Therapeutics, Inc. % Mr. Kenneth K. Kleinhenz Sr. Director of Regulatory Affairs 3020 Callan Road San Diego, California 92121
Re: K072587
Trade/Device Name: Cytori AFT System Regulation Number: 21 CFR 878.5040 Regulation Name: Suction lipoplasty system Regulatory Class: II Product Code: MUU Dated: September 12, 2007 Received: September 14, 2007
Dear Mr. Keinhenz:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{5}------------------------------------------------
## Page 2 - Mr. Kenneth K. Kleinhenz
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark H. Mulkeen
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{6}------------------------------------------------
# Indications for Use
510(k) Number: K072587
Device Name: Cytori AFT System
# Indications for Use:
Cytori AFT System is used in the aspiration, harvesting, filtering and transfering of autologous tissue.
The Cytori AFT System is intended for use in the following surgical specialties when the aspiration of soft tissue is desired:
- · Neurosurgery
- · Gastrointestinal and Affiliated Organ Surgery
- · Urological Surgery
- · Plastic and Reconstructive Surgery
- · General Surgery
- · Orthopedic Surgery
- · Gynecological Surgery
- · Thoracic Surgery
- · Laparoscopic Surgery
The Cytori AFT System is indicated for use when the fragmentation, emulsification, and aspiration of subcutaneous fatty tissues for aesthetic body contouring is desired.
| Prescription Use | X | AND/OR | Over-The-Counter Use |
|-----------------------------|---|--------|------------------------|
| (Part 21 CFR 801 Subpart D) | | | (21 CFR 801 Subpart C) |
(PLEASE DO NOT WRITE BELOW THIS LINE CONTINUE ON ANOTHER PAGE IF NEEDED)
(Division Sign-Off)
Concurrence of CDRH Office of Device Evaluation (ODE), Division of General, Restorative, and Neurological Devices
| 510(k) Number | 1072587 |
|---------------|---------|
|---------------|---------|
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.