← Product Code [EBF](/productcode/EBF) · K072545

# ARTISTE SE FLOWABLE COMPOSITE, MODEL N280 (K072545)

_Pentron Clinical Technologies · EBF · Oct 26, 2007 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K072545

## Device Facts

- **Applicant:** Pentron Clinical Technologies
- **Product Code:** [EBF](/productcode/EBF.md)
- **Decision Date:** Oct 26, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 872.3690
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Artiste SE Flowable Composite product is a self-etch, self-adhesive flowable resin composite indicated for light-cured dental restorations. Applications include: - Self-etch base liner - Restorations of carious lesions - Class 1, III, and V cavity preparations - Pit and fissure sealant.

## Device Story

Artiste SE Flowable Composite is a light-cured, self-etching, self-adhesive dental restorative material. It eliminates the need for separate etchant, primer, or adhesive application steps during cavity preparation. The device is supplied in multi-use syringes or single-dose delivery systems. It is intended for use by dental clinicians in a clinical setting to repair dentition. The composite is composed of approximately 65% barium boro-silicate glass filler by weight, providing strength and radiopacity. The clinician applies the material directly to the prepared site, followed by light-curing to harden the resin. This simplifies the restorative workflow, potentially reducing procedure time and technique sensitivity for the clinician while providing a durable, radiopaque restoration for the patient.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Light-cured, self-etching, self-adhesive flowable resin composite. BisGMA-free formulation. Filler: ~65% barium boro-silicate glass by weight; particle size < 2 microns. Radiopaque. Supplied in syringes or single-dose units.

## Regulatory Identification

Tooth shade resin material is a device composed of materials such as bisphenol-A glycidyl methacrylate (Bis-GMA) intended to restore carious lesions or structural defects in teeth.

## Predicate Devices

- Embrace Wet-Bond Restorative Materials ([K031877](/device/K031877.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K072545

Image /page/0/Picture/1 description: The image shows the logo for Pentron Clinical Technologies, LLC. The logo features the company name in a simple, sans-serif font, with "Pentron" appearing on the top line and "Clinical" on the second line. Below the company name, the text "Technologies, LLC" is displayed in a smaller font size. The tagline "Advancing dentistry one innovation at a time!" is located at the bottom of the image.

OCT 2 6 2007

#### 510(K) SUMMARY 6.0

Pentron Clinical Technologies, LLC. 68 North Plains Industrial Road Wallingford, CT 06492 203-265-7397 Tel: 203-284-4986 Fax: Contact: Greg Moreau ------------------------------------------------------------------------------------------------------------------------------------------------------------------------------

Artiste SE Flowable Composite Trade Name: Bonding Agent Common Name: Classification Name: Material, Tooth Shade, Resin, 21CFR 872.3690, EBF ==============================================================================================================================================================================

Artiste SE Flowable Composite product, as a suitable dental restorative material, performs similar intended functions as its predicate device, Embrace Wet-Bond Restorative Materials (reference K031877). Both devices are intended to provide the dental clinician a safe and effective method to repair dentition using a variety of technique applications. Applications include:

- 트 Self-etch base liner
- 트 Restorations of carious lesions
- Class 1, III, and V cavity preparations 트
- Pit and fissure sealant. I

The subject device is a light-cured self-etching, self-adhesive composite in a BisGMA-free resin formulation. The Artiste SE Flowable Composite requires no application of etchant, primer or adhesive prior to device use once the restorative site has been properly prepared.

Device composition is approximately 65% filler by weight with filler particle size values less than 2 microns. As with other resin products manufactured by Pentron Clinical Technologies, LLC., the predominant filler for Artiste SE Flowable Composite is a barium boro-silicate glass. This filler type provides the subject device with the required strength and radiopacity features.

Product is supplied in various shades either in multi-use syringe or single dose delivery systems. Product packaging includes kit or refill configuration.

A review for safety and effectiveness was performed and found not to have been affected.

510(k) Submission for Artiste SE Flowable Composite

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized caduceus symbol, which features a staff with a snake winding around it, overlaid on three human figures. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES. USA" is arranged in a circular pattern around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

OCT 26 2007

Mr. Greg Moreau Quality Systems, Manager Pentron Clinical Technologies, LLC 68- 70 North Plains Industrial Road Wallingford, Connecticut 06492

Re: K072545

Trade/Device Name: Artiste SE Flowable Composite Regulation Number: 21 CFR 872.3690 Regulation Name: Tooth Shade Resin Material Regulatory Class: II Product Code: EBF Dated: September 6, 2007 Received: September 10, 2007

Dear Mr. Moreau:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

### Page 2- Mr. Moreau

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours.

Neil R. Ogden

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# 5.0 INDICATION FOR USE STATEMENT

# 510(k) NUMBER (IF KNOWN): 《 0 7 2 5 4 5 -

# DEVICE NAME: Artiste SE Flowable Composite

### INDICATION FOR USE:

Artiste SE Flowable Composite product is a self-etch, self-adhesive flowable resin composite indicated for light-cured dental restorations. Applications include:

- 트 Self-etch base liner
- 트 Restorations of carious lesions
- Class 1, III, and V cavity preparations
- I Pit and fissure sealant.

Syper Rives

( Tulsion Sign-Off) Division of Anesthesiology, General Hospital, Infection Control, Dental Devices

510(k) Number: K072545

# (PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED.)

Concurrence of CDRH, Office of Device Evaluation (ODE)

ู้ Prescription Use (Per 21 CFR 801.109)

OR

Over –The-Counter-Use (Optional Format 1-2-96)

510(k) Submission for Artiste SE Flowable Composite

---

**Source:** [https://fda.innolitics.com/device/K072545](https://fda.innolitics.com/device/K072545)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
