← Product Code [DHA](/productcode/DHA) · K072500

# ONE STEP HCG URINE/SERUM TEST (K072500)

_Guangzhou Wondfo Biotech Co., Ltd. · DHA · Apr 21, 2009 · Clinical Chemistry · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K072500

## Device Facts

- **Applicant:** Guangzhou Wondfo Biotech Co., Ltd.
- **Product Code:** [DHA](/productcode/DHA.md)
- **Decision Date:** Apr 21, 2009
- **Decision:** SESE
- **Submission Type:** Traditional
- **Regulation:** 21 CFR 862.1155
- **Device Class:** Class 2
- **Review Panel:** Clinical Chemistry

## Indications for Use

The Wondfo One Step hCG Urine/Serum in vitro diagnostic test is used for the qualitative determination of human chorionic gonadotropin (hCG) in urine or serum to aid in the early detection of pregnancy. It is intended for professional use only (Clinical Laboratory Use).

## Device Story

The Wondfo One Step hCG Urine/Serum Test is a lateral flow immunochromatographic assay for the qualitative detection of hCG in urine or serum. The device is available as a test strip or cassette. It utilizes mouse monoclonal anti-beta hCG antibodies on a sample pad and mouse monoclonal anti-alpha hCG antibodies on a membrane test line. The test is performed by immersing the device in a sample; hCG, if present at or above 25 mIU/mL, binds to the antibodies, creating a visible red band at the test line. A control line confirms sufficient sample volume. The test is intended for professional use in clinical laboratories to aid in pregnancy detection. Results are read visually by a clinician within 3 to 5 minutes. High concentrations of hCG (>100 IU/mL) may cause a hook effect, potentially leading to false-negative results.

## Clinical Evidence

No clinical data provided; substantial equivalence is based on bench testing of the immunochromatographic assay performance.

## Technological Characteristics

Lateral flow immunochromatographic assay. Components: mouse monoclonal anti-beta hCG antibody (sample pad), mouse monoclonal anti-alpha hCG antibody (test line), goat anti-mouse IgG (control line). Cut-off: 25 mIU/mL. Formats: strip and cassette. Standardized to WHO 3rd International Standard. No electronic components or software.

## Regulatory Identification

A human chorionic gonadotropin (HCG) test system is a device intended for the early detection of pregnancy is intended to measure HCG, a placental hormone, in plasma or urine. A human chorionic goadotropin (HCG) test system is a device intended for any uses other than early detection of pregnancy (such as an aid in the diagnosis, prognosis, and management of treatment of persons with certain tumors or carcinomas) is intended to measure HCG, a placental hormone, in plasma or urine.

## Predicate Devices

- Acon Laboratories, Inc. HCG One Step Urine/Serum Pregnancy Test ([K980736](/device/K980736.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}

1

510(k) SUBSTANTIAL EQUIVALENCE DETERMINATION
DECISION SUMMARY
ASSAY TEMPLATE

A. 510(k) Number:
k072500

B. Purpose for Submission:
New device

C. Measurand:
Human chorionic gonadotropin (hCG)

D. Type of Test:
Qualitative, Immunochromatographic

E. Applicant:
Guangzhou Wondfo Biotech Co., Ltd.

F. Proprietary and Established Names:
One Step HCG Urine/Serum Test

G. Regulatory Information:
1. Regulation section:
21CFR §862.1155-Human Chorionic Gonadotropin (HCG) test system
2. Classification:
Class II
3. Product code:
DHA
4. Panel:
Chemistry (75)

{1}

H. Intended Use:

1. Intended use(s):

See indication for use below.

2. Indication(s) for use:

The Wondfo One Step hCG Urine/Serum *in vitro* diagnostic test is used for the qualitative determination of human chorionic gonadotropin (hCG) in urine or serum to aid in the early detection of pregnancy. It is intended for professional use only (Clinical Laboratory Use).

3. Special conditions for use statement(s):

Professional use only

In the package insert the manufacturer has stated the following limitation:

When hCG is equal to or greater than 100 IU/mL, it may present a negative result because of hook effect.

4. Special instrument requirements:

None

I. Device Description:

The Wondfo One Step hCG Urine/Serum Test will be sold in two formats: test strip and cassette. The test strip kit consists of one test device and a package insert. The cassette kit consists of one test device, a disposable plastic dropper, and a package insert. Each test device contains mouse monoclonal anti-β HCG antibody coated membrane and a pad containing mouse monoclonal anti-α hCG antibody colloidal gold conjugate. The control antibodies are goat anti-mouse IgG.

J. Substantial Equivalence Information:

1. Predicate device name(s):

ACON hCG One Step Urine/Serum Pregnancy Test

2. Predicate 510(k) number(s):

k980736

{2}

3. Comparison with predicate:

|  Similarities and differences  |   |   |
| --- | --- | --- |
|  Item | ACON HCG One Step (Predicate device) k980736 | Wondfo One Step HCG Urine/Serum Test (Candidate device)  |
|  Intended Use | For qualitative detection of HCG in urine or serum to aid in the early detection of pregnancy | Same  |
|  Standardization | WHO 3rdInternational Standard | Same  |
|  Methodology | Lateral flow immunochromatographic assays based on the principle of antigen antibody immunochemistry. | Same  |
|  Formats | Strip | Strip and cassette  |
|  Antibodies | Goat and mouse | Same  |
|  Sample Type | Serum and urine | Same  |
|  Storage temperature | 15-30°C | 4-30°C  |
|  Read time | 3 to 5 minutes | Same. Do not read after 5 minutes  |
|  Cut-off | 25 mIU/mL | Same  |

# K. Standard/Guidance Document Referenced (if applicable):

FDA Guidance Document, Review Criteria for Assessment of Professional Use Human Chorionic Gonadotropin (HCG) In Vitro Diagnostic Devices, November 6, 1996.

# L. Test Principle:

The Wondfo One Step HCG test uses a monoclonal antibody against purified  $\beta$ -hCG. The kit is a rapid qualitative assay based on the principle of double antibodies sandwich immunoassay. When the strip is immersed in a urine or serum sample, the urine/serum migrates upward. The hCG is first captured by monoclonal antibodies to  $\beta$ -hCG in the sample pad and then by monoclonal antibodies to  $\alpha$ -hCG on the T line and anti-mouse IgG on the C line coated in the membrane. If the hCG is at or above  $25\mathrm{mIU/ml}$  red bands appear both in T line and C line which indicates a positive result. If the hCG is below 25 mIU/ml, only the red band in C line should appear. The band at the C line indicates sufficient volume has been added.

# M. Performance Characteristics (if/when applicable):

1. Analytical performance:

a. Precision/Reproducibility:

i.) For Urine Samples:

{3}

Reproducibility of the device was evaluated by testing negative urine samples spiked with hCG (to 500 IU/mL, 200 IU/mL, 100 IU/mL, 5 IU/mL, 1000 mIU/mL, 500 mIU/mL, 100 mIU/mL, 25 mIU/mL, 20 mIU/mL, 15 mIU/mL, and 10 mIU/mL hCG), one positive patient urine sample came from pregnant woman (containing 29 mIU/mL hCG by ELISA), and one negative patient urine sample(from male) 10 times with multiple lots of test strips (total  $\mathrm{N} = 30$ ). The results are summarized below:

|  hCG concentrations tested | N | Positive | Negative  |
| --- | --- | --- | --- |
|  500 IU/mL | 30 | 0 | 30  |
|  200 IU/mL | 30 | 27 | 3  |
|  100 IU/mL | 30 | 30 | 0  |
|  5 IU/mL | 30 | 30 | 0  |
|  1,000 mIU/mL | 30 | 30 | 0  |
|  500 mIU/mL | 30 | 30 | 0  |
|  100 mIU/mL | 30 | 30 | 0  |
|  25 mIU/mL | 30 | 29 | 1  |
|  20 mIU/mL | 30 | 2 | 28  |
|  15 mIU/mL | 30 | 0 | 30  |
|  10 mIU/mL | 30 | 0 | 30  |
|  Positive patient (29 mIU/mL) | 30 | 30 | 0  |
|  Negative patient (0 mIU/mL) | 30 | 0 | 30  |

# ii.) For Serum Samples:

Reproducibility of the device was evaluating by testing negative serum samples spiked with hCG (to 500 IU/ml, 200 IU/ml, 100 IU/ml, 5 IU/ml, 1000 mIU/ml, 500 mIU/ml, 100 mIU/ml, 25 mIU/ml, 20 mIU/ml, 15 mIU/ml, and 10 mIU/ml hCG), one positive serum sample came from pregnant woman (containing 29.4 mIU/mL hCG by ELISA), and one negative serum sample (male serum) 10 times with multiple lots of test strips (total  $\mathrm{N} = 30$ ). The results are summarized below:

|  hCG concentrations tested | N | Positive | Negative  |
| --- | --- | --- | --- |
|  500 IU/mL | 30 | 0 | 30  |
|  200 IU/mL | 30 | 27 | 3  |
|  100 IU/mL | 30 | 30 | 0  |
|  5 IU/mL | 30 | 30 | 0  |
|  1,000 mIU/mL | 30 | 30 | 0  |
|  500 mIU/mL | 30 | 30 | 0  |
|  100 mIU/mL | 30 | 30 | 0  |
|  25 mIU/mL | 30 | 30 | 0  |
|  20 mIU/mL | 30 | 2 | 28  |
|  15 mIU/mL | 30 | 0 | 30  |
|  10 mIU/mL | 30 | 0 | 30  |
|  Positive patient (29.4 mIU/mL) | 30 | 30 | 0  |
|  Negative patient (0 mIU/mL) | 30 | 0 | 30  |

{4}

b. Linearity/assay reportable range:

High dose hook effect was evaluated by spiking high hCG concentrations into negative urine and serum samples and evaluating the test result lines. The concentrations tested ranged from 10 mIU/mL to 500 IU/mL. Each sample was tested in triplicate. Results indicate that for urine/serum samples with HCG &gt; 100 IU/mL (&gt;100,000 mIU/mL) a hook effect may be seen. The sponsor included the following warning in their package insert:

"When hCG is equal to or greater than 100 IU/mL it may present negative result because of hook effect".

c. Traceability, Stability, Expected values (controls, calibrators, or methods):

The Wondfo One Step HCG Urine/Serum Test is traceable to the WHO 3rd International Standard.

d. Detection limit:

The sensitivity of the device was tested by spiking 60 negative urine and 60 negative serum samples (from non-pregnant females or males) with varying concentrations (12.5, 18.75, 25, 50, 100 mIU/mL) of hCG. Results are summarized below (units = mIU/mL).

For Urine Samples:

|  Concentration | Positive | Negative | Total (N)  |
| --- | --- | --- | --- |
|  0 | 0 | 5 | 5  |
|  12.5 | 0 | 5 | 5  |
|  18.75 | 0 | 5 | 5  |
|  25 | 29 | 1 | 30  |
|  50 | 10 | 0 | 10  |
|  100 | 5 | 0 | 5  |
|  Total | 44 | 16 | 60  |

For Serum Samples:

|  Concentration | Positive | Negative | Total (N)  |
| --- | --- | --- | --- |
|  0 | 0 | 5 | 5  |
|  12.5 | 0 | 5 | 5  |
|  18.75 | 0 | 5 | 5  |
|  25 | 29 | 1 | 30  |
|  50 | 10 | 0 | 10  |
|  100 | 5 | 0 | 5  |
|  Total | 44 | 16 | 60  |

{5}

The sponsor claims a cutoff for positive of  $25\mathrm{mIU / mL}$

# e. Analytical specificity:

i.) An interference study was carried out by adding known amounts of potential interfering substances to urine and serum samples that contains 0, 25, and 50 mIU/mL of hCG, and evaluated the test result lines. The results are shown below:

For Urine Samples:

|  Interfering substances | hCG 0 mIU/mL | hCG 25 mIU/mL | hCG 50 mIU/mL  |
| --- | --- | --- | --- |
|  Acetaminophen 20 mg/dL | - | + | +  |
|  Acetylsalicylic acid 20 mg/dL | - | + | +  |
|  Ascorbic acid 20 mg/dL | - | + | +  |
|  Atropine 20 mg/dL | - | + | +  |
|  Caffeine 20 mg/dL | - | + | +  |
|  Centesic acid 20 mg/dL | - | + | +  |
|  Glucose 2 g/dL | - | + | +  |
|  Hemoglobin 20 mg/dL | - | + | +  |
|  Tetracycline 20 mg/dL | - | + | +  |
|  Ampicillin 20 mg/dL | - | + | +  |
|  Albumin 20 mg/dL | - | + | +  |
|  Bilirubin 2 mg/dL | - | + | +  |

For Serum Samples:

|  Interfering substances | hCG 0 mIU/mL | hCG 25 mIU/mL | hCG 50 mIU/mL  |
| --- | --- | --- | --- |
|  Acetaminophen 20 mg/dL | - | + | +  |
|  Acetylsalicylic acid 20 mg/dL | - | + | +  |
|  Ascorbic acid 20 mg/dL | - | + | +  |
|  Atropine 20 mg/dL | - | + | +  |
|  Caffeine 20 mg/dL | - | + | +  |
|  Centesic acid 20 mg/dL | - | + | +  |
|  Glucose 2 g/dL | - | + | +  |
|  Hemoglobin 20 mg/dL | - | + | +  |
|  Tetracycline 20 mg/dL | - | + | +  |
|  Ampicillin 20 mg/dL | - | + | +  |
|  Albumin 20 mg/dL | - | + | +  |
|  Bilirubin 40 mg/dL | - | + | +  |
|  Triglycerides 1200 mg/dL | - | + | +  |

{6}

ii.) A cross-reactivity study was carried out by adding known amounts of potential cross reactants such as LH, FSH, and TSH to a total of 90 negative urine and serum samples and evaluated the test result lines. The cross-reactivity results are shown below:

For Urine Samples:

|  (mIU/mL) | Lot I |   | Lot II |   | Lot III  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  + | - | + | - | + | -  |
|  LH | 100 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 300 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 500 | 1 | 29 | 1 | 29 | 0 | 30  |
|  FSH | 100 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 300 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 500 | 0 | 30 | 1 | 29 | 0 | 30  |
|  TSH | 750 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 1000 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 1250 | 0 | 30 | 0 | 30 | 1 | 29  |

For Serum Samples:

|  (mIU/mL) | Lot I |   | Lot II |   | Lot III  |   |   |
| --- | --- | --- | --- | --- | --- | --- | --- |
|   |   |  + | - | + | - | + | -  |
|  LH | 100 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 300 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 500 | 1 | 29 | 1 | 29 | 1 | 29  |
|  FSH | 100 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 300 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 500 | 1 | 29 | 1 | 29 | 0 | 30  |
|  TSH | 750 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 1000 | 0 | 30 | 0 | 30 | 0 | 30  |
|   | 1250 | 0 | 30 | 1 | 29 | 1 | 29  |

No cross-reactivity was observed for either both urine or serum samples up to the following concentrations:  $\mathrm{LH} = 300\mathrm{mIU / mL}$ ,  $\mathrm{FSH} = 300\mathrm{mIU / mL}$ , and  $\mathrm{TSH} = 1000\mathrm{mIU / mL}$ .

iii.) A pH study was performed to evaluate the device and the sponsor concluded that urine and serum samples with pH 4-9 were not adversely affected and produced the expected results.
iv.) A specific gravity study was performed to evaluate the device and the sponsor concluded that urine samples with densities between 1.000 - 1.040 and serum samples with densities between 1.000 - 1.050 were not adversely affected and produced the expected results.

{7}

f. Assay cut-off:

See 1.d. above

# 2. Comparison studies:

a. Method comparison with predicate device:

A method comparison study was performed comparing the Wondfo One Step hCG Urine/Serum test strips and cassettes formats to the Acon hCG test strips (predicate device). Samples included 282 hCG positive serum/urine samples from pregnant women (aged 20-45), and 300 hCG negative serum/urine samples randomly collected from non-pregnant women and male. Results are summarized below:

Table 1: Comparison between Wondfo test strips and Acon test strips for urine samples

|   | Wondfo hCG test strip  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | - | Total  |
|  Acon HCG test trip(Predicate device) | + | 281 | 1 | 282  |
|   |  - | 2 | 298 | 300  |
|   |  Total | 283 | 299 | 582  |

Table 2: Comparison between Wondfo cassettes and Acon test strips for urine samples

|   | Wondfo hCG cassette  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | - | Total  |
|  Acon HCG test trip (Predicate device) | + | 281 | 1 | 282  |
|   |  - | 2 | 298 | 300  |
|   |  Total | 283 | 299 | 582  |

Table 3: Comparison between Wondfo test strips and Acon test strips for serum samples

|   | Wondfo hCG test strip  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | - | Total  |
|  Acon HCG test trip (Predicate device) | + | 280 | 2 | 282  |
|   |  - | 2 | 298 | 300  |
|   |  Total | 282 | 300 | 582  |

{8}

Table 4: Comparison between Wondfo cassettes and Acon test strips for serum samples

|   | Wondfo hCG cassette  |   |   |   |
| --- | --- | --- | --- | --- |
|   |   |  + | - | Total  |
|  Acon HCG test trip (Predicate device) | + | 280 | 2 | 282  |
|   |  - | 2 | 298 | 300  |
|   |  Total | 282 | 300 | 582  |

Table 5: Agreement rate between Wondfo hCG test results with Acon test strips:

|  Agreement | Wondfo hCG test strip |   | Wondfo hCG cassette  |   |
| --- | --- | --- | --- | --- |
|   | Urine samples | Serum samples | Urine samples | Serum samples  |
|  Positive | 99.6% | 99.3% | 99.6% | 99.3%  |
|  Negative | 99.3% | 99.3% | 99.3% | 99.3%  |
|  Total | 99.5% | 99.3% | 99.5% | 99.3%  |

# b. Matrix comparison:

This test is only applicable to urine or serum samples. Performance with both of these matrices is described in the performance sections above.

# 3. Clinical studies:

a. Clinical Sensitivity:

Not Applicable

b. Clinical specificity:

Not Applicable

c. Other clinical supportive data (when a. and b. are not applicable):

Not Applicable

# 4. Clinical cut-off:

Not applicable

{9}

5. Expected values/Reference range:

As stated in the package insert, “all non-pregnant females should test negative for hCG.”

**N. Proposed Labeling:**

The labeling is sufficient and it satisfies the requirements of 21 CFR Part 809.10.

**O. Conclusion:**

The submitted information in this premarket notification is complete and supports a substantial equivalence decision.

10

---

**Source:** [https://fda.innolitics.com/device/K072500](https://fda.innolitics.com/device/K072500)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
