GBI TSH NEONATAL SCREENING KIT

K072458 · Golden Bridge International, Inc. · JLW · Apr 10, 2008 · Clinical Chemistry

Device Facts

Record IDK072458
Device NameGBI TSH NEONATAL SCREENING KIT
ApplicantGolden Bridge International, Inc.
Product CodeJLW · Clinical Chemistry
Decision DateApr 10, 2008
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 862.1690
Device ClassClass 2
AttributesPediatric

Intended Use

The GBI TSH Neonatal Screening Kit is designed for the quantitative determination of Thyroid Stimulating Hormone (TSH) concentrations in neonatal blood samples that have been collected onto Whatman 903 specimen collection paper. The results are used to screen newborns for congenital hypothyroidism.

Device Story

The GBI TSH Neonatal Screening Kit is an enzyme immunoassay (EIA) for the quantitative measurement of TSH in neonatal dried blood spots. The device uses a 96-well microplate coated with polyclonal goat anti-hTSH antibody. Sample discs punched from Whatman 903 specimen collection paper are incubated in the wells to elute TSH, which is captured by the immobilized antibody. A second, HRP-conjugated monoclonal anti-hTSH antibody is added to form a sandwich complex. After washing, TMB/H2O2 color developer is added, producing a blue color proportional to TSH concentration, which turns yellow upon addition of a stop reagent. The absorbance is measured at 450 nm using a microplate reader. The device is intended for use by clinical laboratory professionals. Results are used to screen newborns for congenital hypothyroidism, enabling early detection and treatment to prevent permanent mental retardation. The assay requires manual processing and interpretation of results against a standard curve.

Clinical Evidence

Bench testing only. Precision evaluated per CLSI EP5-A2 (n=40 per sample, CV 7.9-13.9%). Linearity (r²=0.999) and analytical sensitivity (2.4 µIU/mL) established. Specificity testing showed no cross-reactivity with FSH, LH, or hCG. Interference testing confirmed no adverse effects from lipids, bilirubin, or hemoglobin. Method comparison with predicate (n=995) demonstrated concordance in screening classifications.

Technological Characteristics

Quantitative sandwich ELISA; 96-well microplate format. Reagents: goat polyclonal anti-hTSH capture antibody, HRP-conjugated monoclonal anti-hTSH detection antibody, TMB/H2O2 color reagent. Detection via spectrophotometry at 450 nm. Standardized to WHO 3rd IRP 81/565. Sample type: neonatal dried blood spots on Whatman 903 paper.

Indications for Use

Indicated for the screening of newborns for elevated TSH levels suggestive of congenital hypothyroidism using dried blood spot samples.

Regulatory Classification

Identification

A thyroid stimulating hormone test system is a device intended to measure thyroid stimulating hormone, also known as thyrotrophin and thyrotrophic hormone, in serum and plasma. Measurements of thyroid stimulating hormone produced by the anterior pituitary are used in the diagnosis of thyroid or pituitary disorders.

Predicate Devices

Related Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Golden Bridge International, Inc. Mukilteo, WA 98275 USA APR 1 0 2008 | 1) | Submitter Name: | Golden Bridge International, Inc. | |----|---------------------------------------|----------------------------------------------------------------------------------| | | Address: | 9700 Harbour Place, Suite 129<br>Mukilteo, WA 98725 USA | | | Telephone Number: | 425-493-1801 | | | Contact Person: | Xiaoping Zhang, Vice President | | | Date Prepared: | 03/27/08 | | 2) | Device Trade or<br>Proprietary Name: | GBI TSH Neonatal Screening Kit | | | Device Common or<br>Usual Name: | Thyroid stimulating hormone (TSH) neonatal screening enzyme<br>immunoassay (EIA) | | | Device Classification Name: | Radioimmunoassay, Thyroid Stimulating Hormone | | | Product Code: | JLW | | 3) | Legally Marketed<br>Predicate Device: | Accuwell Thyroid Stimulating Hormone (TSH) ELISA | ### GBI Thyroid Stimulating Hormone (TSH) EIA 510(k) Summary ### 4) Device Description: Summary and Explanation of the Test Congenital hypothyroidism (CH) is one of the most common metabolic disorders resulting in permanent mental retardation if undetected or left untreated soon after birth. Newborns that have been identified and treated for CH within two weeks of birth can be expected to have normal cognitive development. Screening programs for the disorder were developed early in 1974 and have now been established in Western Europe, North America, Japan and parts of Eastern Europe, Asia, South and Central American Academy of Pediatrics first published recommendations for CH in 1993. Advances in screening methods and expanded knowledge of the physiology of thyroid hormones has lead to new recommendations by the Academy in 2006.2 In North America more than five million newborns are screened annually with over 1400 hypothyroid cases detected each year. The incidence of hypothyroidism varies among different populations and ranges from 1 in 3000 to 1 in 4000. The incidence of CH is higher in Hispanics and lower in black individuals. Females have a 2:1 greater incidence of CH than males. A higher frequency of the CH disorder has been observed in individuals with Downs Syndrome. Thyroid stimulating hormone (TSH) is a pituitary-derived glycoprotein with a molecular weight of approximately 28 kDa. TSH is produced by the anterior lobe of the pituitary gland under the control of thyrotropin releasing hormone (TRH) which is secreted by the hypothalamus. TSH acts on the thyroid gland to release the thyroid hormones triiodothyronine (T)) and thyroxine (T). Under normal thyroid function, the level of the thyroid hormones T3 and T4 is inversely related to the concentrations of TSH and TRH in the bloodstream via a negative feedback mechanism. As a result of this negative feedback mechanism between the thyroid gland and the pituitary, TSH is always elevated in cases of primary hypothyroidism and often to very high levels. It is {1}------------------------------------------------ for this reason that the measurement of TSH concentration is a valuable diagnostic tool for the detection of hypothyroidism. Thyroid hormone (TH) levels of the fetus are low during the first half of pregnancy. Maternal status of TH hormones is entirely responsible for the regulation of fetal TH, transmitted through the placenta. The fetal hypothalamic-pituitary-thyroid axis develops during the second half of gestation and is mature at time of birth. A surge of TSH concentration occurs at birth and then declines rapidly within the first few days of life, and continues to decrease more slowly to adult levels. It is recommended to perform screening for CH ideally within the 2nd to 4th days after birth or at time of discharge. Blood samples collected before 24 hours of age, from low birth weight or ill infants can lead to elevated TSH and thus false positive results. Hypothyroid infants may be partially protected by maternal TH levels and so most infants will appear normal at birth. There have also been reported cases of transient hypothyroidism due to maternal hypothyroidism and other maternal abnormalities. Hypothyroidism can develop after birth, and in such cases will yield a normal newborn screening result. Since the incidence of hypothyroidism and the concentrations of TSH have been shown to a variety of factors (e.g., demographics, ethnicity, the sex and age of the infant, low birth weights and premature births) it is important that each newborn screening laboratory examine its test population and determine its own normal range and cut-off values with these factors taken into consideration. Screening algorithms vary among centers from initial TSH with confirmation via T4 initial screen with confirmation via TSH measurements, to both tests (174 and TSH) being performed simultaneously. Regardless of method, repeated positive results on a neonate should be cause for referral for additional confirmation testing. ### Principle of the Assay The GBI TSH Neonatal Screening Kit is an enzyme immunoassay. A highly specific polyclonal goat antihTSH (human) antibody has been immobilized onto each well of the 96-well microplates provided. To begin the assay, sample discs punched from dried whole blood spot standards, controls and neonate specimens are added to the coated wells. An elution buffer is also added. The plate is incubated to elute TSH from the sample disc and to allow capture of the eluted TSH by the antibody immobilized onto the microplate wells. Following incubation the plates are washed to remove the sample discs as well as the eluate. A second antibody, a D-specific anti-hTSH monoclonal that has been conjugated to the enzyme horseradishperoxidase (HRP), is then added to the wells and incubated. The eluted TSH of the sample already captured by the microplate-bound antibody is now also bound by the enzyme-conjugated monoclonal antibody added. An antibody-TSH-antibody bridge, or "sandwich", forms that is bound to the surface of the microplate wells. Any unbound complexes are removed with subsequent plate washings. The final stage of the assay is the detection of the microwell-bound complexes by the addition of a color developing reagent. The enzyme (HRP) portion of the bound "sandwich" reacts with the color developer, 3, 3', 5, 5'-Tetramethylbenzidine (TMB) in the presence of hydrogen peroxide (H2O2). The TMB/ H2O2 liquid is converted from colorless to blue. The degree of color change is directly proportional to the amount of TSH antigen that is bound in the well. The color development is terminated with the addition of a color stopper that converts the blue to yellow. The results are measured with a microplate reader at a wavelength of 450 mm. The absorbance measured is directly proportional to the concentration of TSH in the sample. A standard curve is generated by plotting the light absorbance of each standard versus its known TSH concentrations of TSH in the unknown samples are determined by interpolation from this standard curve. {2}------------------------------------------------ ### Kit Contents ### Materials Supplied | | Quantity per Kit | |----------------------------------------|------------------| | Total Tests: | 576 | | TSH Capture Microplate | 6 X 96 wells | | TSH Elution Buffer | 1 x 70 ml | | Anti-TSH PO Conjugate | 1 x 1.4 ml | | TSH Diluent Buffer | 1 x 70 ml | | PO Wash Buffer (20X) | 1 x 100 ml | | PO Color Reagent* | 1 x 70 ml | | PO Color Stopper | 1 x 70 ml | | TSH Dried Blood Standards and Controls | 1 Set | ### * This reagent is light sensitive. Avoid prolonged exposure to light. NOTE: Do NOT use reagents that are cloudy or discolored as these conditions indicate deterioration. Reagents should be stored in their original containers with the exception of the diluted wash buffer. ### Do NOT exchange reagents from one kit with those of another kit. ### Reagent Descriptions ### TSH Capture Microplate Goat polyclonal anti-alpha hTSH antibody coated 96-well microplates. The microplates are packaged in zipper-lock bags containing a desiccant. Store the unused microplates in zipper-lock bags with desiccant. Storage: Dry at 2-25° C. Expiration: Refer to the expiration date printed on the label ### TSH Elution Buffer A borate buffer containing bovine albumin, surfactant and preservatives. This buffer contains sodium azide as a preservative. Storage: 2-8° C Refer to the expiration date printed on the label. Expiration: ### Anti-TSH PO Conjugate A mouse monoclonal p-specific antibody conjugated to horseradish peroxidase (HRP) in borate buffer containing bovine albumin, surfactant, and preservatives. This buffer contains thimerosal as a preservative; refer to "Warnings and Precautions", below. Store in the original brown container. This conjugate is concentrated and must be diluted before use. Storage: 2-8° C Refer to the expiration date printed on the label. Expiration: ### TSH Diluent Buffer A borate buffer containing bovine albumin, surfactant, and preservatives. This buffer contains thimerosal as a preservative. 2-8° C Storage: Refer to the expiration date printed on the label. Expiration: {3}------------------------------------------------ ### PO Wash Buffer (20X) A concentrated solution of phosphate buffered saline containing a surfactant and preservatives. Wash Buffer Concentrate must be diluted with deionized or distilled water before use. Storage: 2-25° C Expiration: Concentrate: Refer to the expiration date printed on the label. Diluted wash buffer is stable for 1 month when stored at 2-25° C. ### PO Color Reagent* hydrogen coloriess solution of 3,3', 5,5'-Tetramethylbenzidine (TMB) and peroxide. A *This reagent is light sensitive. Store in the original brown container. Do not pour any of the unused Color Developer back into the original container. This reagent should remain colorless. If it has discolored, discard it. 2-8° C Protected from light. Storage: Refer to the expiration date printed on the label. Expiration: ### PQ Color Stopper 、 A solution containing < 1% hydrochloric acid. Storage: 2-8° C. Refer to the expiration date printed on the label. Expiration: ### TSH Dried Blood Standards and Controls Prepared from human whole blood, adjusted to a hematocrit of 55% and spotted onto Whatman 903% specimen collection paper. The six Standards contain concentrations of added TSH at approximately zero, 7.5, 15.0, 30.0, 60.0, and 120.0 ulU/ml serum equivalent. The three Controls contain approximate TSH concentrations of 15, 30 and 80 µIU/m1 serum equivalent. Refer to the labels for the exact concentrations of the Standards and acceptable ranges for the Controls. Store in a zipper-lock bag containing a desiccant. Standards are referenced against WHO 3rd IRP of human TSH 81/565. Storage: Dry at 2-8° C or below. Refer to the expiration date printed on the label. Expiration: ### Materials Required But Not Supplied - 1/8 inch (3 mm) diameter hole punch (sample prep) 1. - Forceps or fine tweezers to pick up the punched sample discs 2. - Plastic, disposable tube and screw-cap; ≥ 15 ml volume (conjugate prep) ి. - Serological pipette to dispense 10 ml volumes (wash buffer prep) 4. - Precision pipettes to accurately dispense 100 ul volumes (reagent adds) 5. - Multi-channel pipettes to dispense 350 ul volumes or an automated plate washer (plate washes) 6. - Microplate reader capable of reading at a wavelength of 450 nm 7. - Plate rotator (sample & test reagent mixing) 8. - Clear, adhesive sealing tape or plate sealers; solid plastic, reusable microplate covers 9. - 10. Graduated cylinders - 11. Deionized or distilled water - 5) Intended Use: The GBI TSH Neonatal Screening Kit is designed for the quantitative determination of Thyroid Stimulating Hormone (TSH) concentrations in neonatal blood samples that have been collected onto Whatman 903 % specimen collection paper. The results are used to screen newborns for congenital hypothyroidism. {4}------------------------------------------------ ### 6) Summary of Technological Characteristics: | Item<br>#: | Characteristic: | Predicate¹²: | New Device: | |------------|-------------------------------------------------------------|------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | 1 | Intended User | Clinical Laboratory Professionals | Clinical Laboratory Professionals | | 2 | Target Population | Neonates only | Neonates only | | 3 | Intended Use | The quantitative determination of Thyroid<br>Stimulating Hormone (TSH) concentrations<br>in neonatal blood samples that have been<br>collected onto Whatman 903® specimen<br>collection paper. | The quantitative determination of Thyroid<br>Stimulating Hormone (TSH) concentrations<br>in neonatal blood samples that have been<br>collected onto Whatman 903® specimen<br>collection paper. | | 4 | Indications for Use | The screening of newborns for elevated TSH<br>suggestive of congenital hypothyroidism. | The screening of newborns for elevated TSH<br>suggestive of congenital hypothyroidism. | | 5 | Assay Type | Enzyme-Linked Immunosorbent assay<br>(ELISA) | Enzyme Immunoassay (EIA) | | 6 | Assay Processing<br>Method | Manual | Manual | | 7 | Chemical Principle | Polyclonal goat anti-hTSH that is bound to<br>the microplate captures eluted specimen<br>TSH; after washing away free TSH,<br>conjugated monoclonal-horseradish<br>peroxidase is added and binds to the captured<br>TSH thus forming an immobilized<br>antibody+TSH+antibody-enzyme<br>"sandwich". After washing away free<br>enzyme, immobilized peroxidase reacts with<br>added color developer containing hydrogen<br>peroxide and 3,3',5,5'-Tetramethylbenzidine<br>(TMB). The TMB is converted from<br>colorless to blue. The color development is<br>terminated by the addition of a stopping<br>reagent.<br>The light absorbance of the reaction endpoint | Polyclonal goat anti-hTSH that is bound to<br>the microplate captures eluted specimen<br>TSH; after washing away free TSH,<br>conjugated monoclonal-horseradish<br>peroxidase is added and binds to the captured<br>TSH thus forming an immobilized<br>antibody+TSH+antibody-enzyme<br>"sandwich". After washing away free<br>enzyme, immobilized peroxidase reacts with<br>added color developer containing hydrogen<br>peroxide and 3,3',5,5'-Tetramethylbenzidine<br>(TMB). The TMB is converted from<br>colorless to blue. The color development is<br>terminated by the addition of a stopping<br>reagent that also converts the blue to yellow.<br>The light absorbance of the reaction endpoint | | 8 | Detection Method | is measured spectrophotometrically at a<br>wavelength of 650nm. The degree of color<br>development is directly proportional to the<br>amount of TSH antigen that has been<br>"sandwiched" onto the microwell. | is measured spectrophotometrically at a<br>wavelength of 450nm. The degree of color<br>development is directly proportional to the<br>amount of TSH antigen that has been<br>"sandwiched" onto the microwell. | | 9 | Special Instrumentation<br>Required | Microplate Spectrophotometer (at 650 nm) | Microplate Spectrophotometer (at 450 nm) | | 10 | Specimen Type | Neonatal whole blood collected onto<br>Whatman 903® filter paper and dried | Neonatal whole blood collected onto<br>Whatman 903 filter paper and dried | | 11 | Specimen Volume<br>Requirement | 1/8 inch punch (from dried blood spot<br>sample) | 1/8 inch punch (from dried blood spot<br>sample) | | 12 | Calibrator (Standard)<br>Configuration | Stripped serum, lysed cells and purified TSH<br>analyte combined, spotted onto Whatman<br>903® filter paper and dried | Stripped serum, lysed cells and purified TSH<br>analyte combined, spotted onto Whatman<br>903 filter paper and dried | | 13 | Calibrator (Standard)<br>Range of Values | 0 - 160 uIU/ml serum equivalents | 0 - 120 ulU/ml serum equivalents | | 14 | Internal Control<br>Configuration | Stripped serum, lysed cells and purified TSH<br>analyte combined, spotted onto Whatman<br>903 filter paper and dried | Stripped serum, lysed cells and purified TSH<br>analyte combined, spotted onto Whatman<br>903® filter paper and dried | | 15 | Internal Control Levels | Low at ≈ 20 uIU/ml<br>Mid at ≈ 40 uIU/ml<br>High at ≈ 80 uIU/ml | Low at ≈ 15 ulU/ml<br>Mid at ≈ 30 uIU/ml<br>High at ≈ 80 uIU/ml | | 16 | Recommended<br>Reference Ranges | From American Academy of Pediatrics:<br>Normal < 20 ulU/ml serum equivalent<br>Borderline 20-40 ulU/ml serum equivalent<br>Hypothyroid > 40 uIU/ml serum equivalent | From American Academy of Pediatrics:<br>Normal < 20 ulU/ml serum equivalent<br>Borderline 20-40 ulU/ml serum equivalent<br>Hypothyroid > 40 uIU/ml serum equivalent | | 17 | Expected Values -<br>Normal Neonate<br>Population | N=1040:<br>Mean = 9.9 ulU/ml<br>S.D. = 5.8 uIU/ml | N=902:<br>Mean = 8.1 uIU/ml<br>S.D. = 4.0 uIU/ml | | 18 | Analytical Sensitivity | 2.9 uIU/ml | 2.4 uIU/ml | | 19 | Precision³ | Within Run Precision:<br>NA<br>NA<br>24.0 uIU/ml cv = 10.0 %<br>51.1 uIU/ml cv = 7.8 %<br>159.2 uIU/ml cv = 4.3 %<br>Between Run Precision:<br>NA<br>NA<br>27.3 uIU/ml cv = 6.1 %<br>40.8 uIU/ml cv = 8.4 %<br>78.5 uIU/ml cv = 6.5 %<br>Over-All (Within-Device) Precision:<br>NA - Not published | Within-Run Standard Deviation (Sr):<br>15 uIU/ml Sr = 0.9 ulU/ml cv= 6.2%<br>30 uIU/ml Sr = 1.3 " cv= 4.6%<br>25+uIU/ml Sr = 2.6 " cv= 8.5%<br>40+ ulU/ml Sr = 4.2 " cv= 9.1%<br>80+ ulU/ml Sr = 6.5 " cv=7.5%<br>Between-Day Standard Deviation (Sdd):<br>15 uIU/ml Sdd = 1.0 ulU/ml cv= 6.8%<br>30 uIU/ml Sdd = 2.3 " cv= 8.1%<br>25+uIU/ml Sdd = 2.6 " cv= 8.5%<br>40+uIU/ml Sdd = 1.2 " cv= 2.6%<br>80+uIU/ml Sdd = 10.3 " cv= 11.9%<br>Within-Device Standard Deviation (ST):<br>15 ulU/ml ST = 1.2 ulU/ml cv= 8.2%<br>30 ulU/ml ST = 2.5 " cv= 8.8%<br>25+ ulU/ml ST = 3.7 " cv= 12.1%<br>40+uIU/ml ST = 4.4 " cv= 9.5%<br>80+uIU/ml ST = 12.2 " cv= 14.1% | | 20 | Interfering Substances | No Interference observed at the highest<br>concentration tested*:<br>Bilirubin - no interference up to 60 mg/dl*<br>Lipids - no interference up to 1000 mg/dl*<br>Hemoglobin - no interference up to 78 g/dl* | No Interference observed at the highest<br>concentration tested*:<br>Bilirubin - no interference up to 20 mg/dl*<br>Lipids - no interference up to 1350 mg/dl*<br>Hemoglobin - no interference up to 80 g/dl* | | 21 | Cross Reacting<br>Substances | No interference found at the highest<br>concentration tested:<br>FSH- no interference up to 500 mIU/ml<br>LH- no interference up to 500 mIU/ml<br>HCG- no interference up to 100,000 mIU/ml | No interference found at the highest<br>concentration tested:<br>FSH- no interference up to 500 mIU/ml<br>LH- no interference up to 500 mIU/ml<br>HCG- no interference up to 100,000 mIU/ml | | 22 | Method Comparison | For N=995:<br>Normal = 974/995 (97.9%)<br>Borderline = 6/995 (0.6%)<br>Positive = 15/995 (1.5%)<br>For (subset) N=832:<br>Mean = 8.1 uIU/ml;<br>Range = 2.9 to 104 uIU/ml. | For N=995:<br>Normal = 969/995 (97.4%)<br>Borderline = 11/995 (1.1%)<br>Positive = 15/995 (1.5%)<br>For (subset) N=832:<br>Mean = 8.9 ulU/ml;<br>Range = 2.4 to 114 uIU/ml<br>y (GBI TSH EIA) = 1.0695x (Predicate) +<br>0.2671. R² = 0.9185 | | | CDC Controls<br>(Enrichment Values in<br>serum equivalents) | N = 294<br>intercept = 0.1505<br>slope = 1.0231 | N = 405<br>intercept = 5.2083<br>slope = 1.0208 | | 23 | CDC Lot 611 –<br>25 uIU/ml | 26.0 uIU/ml serum equivalent | 30.6 uIU/ml serum equivalent | | | CDC Lot 612 –<br>40 uIU/ml | 40.7 uIU/ml serum equivalent | 46.3 uIU/ml serum equivalent | | | CDC Lot 613 –<br>80 uIU/ml | 82.1 uIU/ml serum equivalent | 86.8 uIU/ml serum equivalent | {5}------------------------------------------------ : : . {6}------------------------------------------------ Golden Bridge International, Inc. Mukilteo, WA 98275 USA l Predicate data and descriptive text were obtained from the Accuwell TSH ELISA Directions For Use (DFU) unless otherwise indicated. 2 The predicate DFU declares the use of "Schleicher & Schuell (S&S) 903™" filter paper as the sample substrate. S&S has since been purchased by Whatman and the name of the 903 filter paper has been changed to "Whatman 903". For this reason, table references throughout use the Whatman 903" designation only, even though the predicate DFU actually refers to the paper as S&S 903™ 3 Precision data were not generated in the same way nor reported in precisely the same terms for the predicate vs the new device. The new (GBI TSH) device precision was established and reported according to NCCLS EP5-A2, "Evaluation of Precision Performance of Quantitative Measurement Methods"; Approved Guideline-Second Edition. 4 Predicate values were obtained from the CDC's Newborn Screening Quality Assurance Program (NSQAP) 2006 Annual Summary Report; Volume 24, January 2007. 5 GBI values were obtained from the same data sct that was used to establish assay precision values. {7}------------------------------------------------ ### 7) Clinical and Non-Clinical Data: ### Precision Precision studies were conducted in accordance with NCCLS EP5-A2: Evaluation of Precision Performance of Quantitative Measurement Methods; Approved Guideline-Second Edition. Testing included two sample aliquots per run, one run per day, performed on 20 different days using the GBI published assay procedure. The resulting data were used to estimate repeatability, between-day and withindevice precision as described by the standard. Refer to the table below. Sources of variability included the use of different: operators, days and reagent lot numbers. | Summary Results and Estimates of Precision for GBP TSH EIA Kit | | | | | | | |----------------------------------------------------------------|------------|------|------|------|------|-------| | | Sample ID: | C1 | C2 | C3 | C4 | C5 | | Precision Parameter | Count | 40 | 40 | 40 | 40 | 40 | | | Mean | 14.6 | 28.3 | 30.6 | 46.3 | 86.8 | | | SD | 1.2 | 2.5 | 3.7 | 4.4 | 12.1 | | | %CV | 7.9 | 8.7 | 11.9 | 9.4 | 13.9 | | | Min | 12.5 | 23.5 | 23.3 | 37.5 | 71.4 | | | Max | 18.0 | 34.0 | 40.4 | 59.1 | 122.5 | | Within-Run Standard Deviation: | Sr = | 0.9 | 1.3 | 2.6 | 4.2 | 6.5 | | Between-Day Standard Deviation: | Sdd = | 1.0 | 2.3 | 2.6 | 1.2 | 10.3 | | Within Device Standard Deviation: | ST = | 1.2 | 2.5 | 3.7 | 4.4 | 12.2 | Summary Results and Estimates of Precision for GBI TSH EIA Kit ### Analytical Sensitivity The Analytical Limit at Low Levels (limit of sensitivity) for the GBI TSH EIA Kit was determined by testing the zero standard multiple times (N=20) within a single assay. The resulting data were used to calculate the analytical limit at low levels. Refer to the table below. The analytical sensitivity is defined as the calculated concentration that corresponds to the mean of the absorbance values of the zero standard (N=20) plus two times the standard deviation derived from those same (N=20) absorbance values. These data are provided for example only. Each laboratory should establish appropriate working limits based upon their own patient population and/or data. | Summary Results and Analytical Limits at Low Levels for GBI TSH EL | | |--------------------------------------------------------------------|---------| | | GBI TSH | | Count | 20 | | Mean ABS | 0.0354 | | SD | 0.0063 | | %CV | 18 | | Min | 0.0289 | | Max | 0.0570 | | Mean + (2 *S.D.) ABS: | 0.0480 | | Analytical Limit (uIU/ml) | 2.4 | #### Posults and Analytical Limits of Low Levels for CBI TSH FIA ﮯ {8}------------------------------------------------ ### Linearity, Recovery and Assay Measurable (Reportable) Range The GBI TSH EIA Kit linear range was assessed according to the NCCLS Guideline EP6-A: Evaluation of the Linearity of Quantitative Analytical Methods, 2003. A series of dried whole blood spot samples was prepared to provide the desired range of TSH concentrations for the study (N=19; see table column "Expected Value", below). Two one-eighth inch punches from each concentration level were tested in each of two runs. The mean of the total number of results for a level (n=4) was compared to the expected value. Results are described in the table below. | Linear Regression Analysis | | $y = 0.8997x + 0.6036, R^2 = 0.999$ | | |----------------------------|----------------------------|----------------------------------------|--------------| | Linearity Sample # | Expected Value<br>(uIU/ml) | GBI Assay Result<br>(Mean uIU/ml; N=4) | Recovery (%) | | 1 | 1.5 | 1.6 | 107 | | 2 | 2.7 | 3.1 | 114 | | 3 | 3.9 | 4.1 | 105 | | 4 | 6.2 | 6.0 | 96.8 | | 5 | 8.6 | 8.1 | 93.6 | | 6 | 10.0 | 10.6 | 106 | | 7 | 10.9 | 9.8 | 89.4 | | 8 | 13.3 | 12.4 | 93.0 | | 9 | 15.6 | 14.9 | 95.4 | | 10 | 18.0 | 16.1 | 89.2 | | 11 | 20.3 | 17.9 | 88.1 | | 12 | 22.7 | 20.1 | 88.3 | | 13 | 24.5 | 21.9 | 89.4 | | 14 | 25.0 | 22.8 | 91.1 | | 15 | 39.0 | 37.2 | 95.3 | | 16 | 68.0 | 65.4 | 96.2 | | 17 | 97.0 | 89.5 | 92.3 | | 18 | 126.0 | 113.5 | 90.1 | | 19 | 155.0 | 138.0 | 89.0 | | Average Recovery-Overall % | | | 95.2 | | Minimum % | | | 88.1 | | Maximum % | | | 114 | Results of Assay Linearity and % Recovery Study for GBI TSH ELA Kit The GBI TSH EIA Kit reportable range is approximately 2.4 uIU/ml. The lower limit is the concentration calculated from the analytical sensitivity study (above) and the upper limit is the concentration of the highest standard of the calibration curve. Assay results obtained outside the measuring (reportable) range should be reported as either "less than" or "greater than" the established low or high limit, respectively, as applicable to the individual result. {9}------------------------------------------------ ### Specificity ### Cross Reactivity The following compounds were tested for cross-reactivity. Each compound was added at several concentrations to TSH-free whole blood which had been adjusted to a hematocrit of 55%. The samples were spotted onto Whatman 903 specimen collection paper, air-dried and assayed. Refer to the table below. | Results of a Cross-Reactivity Study of the GBI TSH EIA Screening Assay | | | |------------------------------------------------------------------------|---------------------------------------------------|-------------------------------------------| | Peptide Hormone | Cross-Reactant<br>Concentration Added<br>(µIU/ml) | TSH Concentration<br>Measured<br>(µIU/ml) | | FSH<br>(WHO 2nd IRP HMG) | 125 | None Detected* | | | 250 | None Detected* | | | 500 | None Detected* | | LH<br>(WHO 1st IRP 68/40) | 125 | None Detected* | | | 250 | None Detected* | | | 500 | None Detected* | | HCG<br>(WHO 2nd I.S. 61/6) | 10,000 | None Detected* | | | 50,000 | None Detected* | | | 100,000 | None Detected* | * "None Detected" represents values obtained that were below the limit of detection for the assay (i.e., < 2.4 ulU/ml.) ### Interfering Substances Assay interference due to the presence of hemoglobin, conjugated bilirubin, and lipids was studied using methods described in NCCLS EP7A: Interference Testing in Clinical Chemistry as guidance. Testing was performed to evaluate the effects of interfering substances at the medical decision(s) concentrations using fresh whole blood. The interferents (i.e., hemoglobin, conjugated and unconjugated bilirubin, and lipids) were present in the whole blood specimens at the desired ("worst case") concentrations before specimens were spotted onto filter paper and dried. As such, the sample elution procedure was captured in the subsequent testing process. These prepared dried blood spot specimens were then assayed using the GBI TSH Kit as prescribed herein. (Results are presented below.) An additional study was conducted that challenged the assay at a much higher hemoglobin concentration. During the initial sample elution step of the assay procedure, three different concentrations (physiologic low, mid, high) of hemoglobin were added separately to microplate wells that also separately contained the three different concentrations (physiologic low, mid, high) of a CDC TSH control sample series. The remainder of the GBI TSH assay was performed as prescribed herein. (Results are presented below.) Interference was defined as a test value obtained that was greater than +/- 1 S.D. (within-run) of the matching control value obtained in the same assay; AND, the difference of the test value obtained from the matching control value is clinically significant at that level. No interference with expected values of any clinical significance was observed from any substance tested, at any concentration tested. {10}------------------------------------------------ . | DBS<br>Sample<br>ID# | Interferent | Interferent<br>Concentration | TSH uIU/ml -<br>Expected<br>Value | TSH uIU/ml -<br>Observed<br>(Mean; n=4) | Acceptance<br>Criteria<br>(+/- 1 SD of<br>Control Value) | |----------------------|----------------|------------------------------|-----------------------------------|-----------------------------------------|----------------------------------------------------------| | 25 | Lipid-Control | 0 mg/dL | na | 16.2 | 16.2 +/- 0.9 | | 19 | Lipid | 1350 mg/dL | 16.2 | 18.0 | 15.3 - 17.1 | | 26 | Lipid-Control | 0 mg/dL | na | 46.1 | 46.1 +/- 4.2 | | 20 | Lipid | 1350 mg/dL | 46.1 | 44.2 | 41.9 - 50.3 | | 27 | Lipid-Control | 0 mg/dL | -- | 85.6 | 85.6 +/- 6.5 | | 21 | Lipid | 1350 mg/dL | 85.6 | 83.5 | 79.1 - 92.1 | | 36 | Bili-Control | 0 mg/dL | na | 13.6 | 13.6 +/- 0.9 | | 31 | Conj-Bilirubin | 20 mg/dL | 13.6 | 14.6 | 12.7 - 14.5 | | 28 | Un-Bilirubin | 20 mg/dL | 13.6 | 14.9 | | | 41 | Bili-Control | 0 mg/dL | na | 27.1 | 27.1 +/- 1.3 | | 32 | Conj-Bilirubin | 20 mg/dL | 27.1 | 27.4 | 25.8 - 28.4 | | 29 | Un-Bilirubin | 20 mg/dL | 27.1 | 28.1 | | | 46 | Bili-Control | 0 mg/dL | na | 68.2 | 68.2 +/- 5.4 | | 33 | Conj-Bilirubin | 20 mg/dL | 68.2 | 68.4 | 62.8 - 73.6 | | 30 | Un-Bilirubin | 20 mg/dL | 68.2 | 67.6 | | | 36 | Hb-Control | 17.0 gm/dL | 13.6 | 13.6 | +/- 0.9 | | 37 | Hb | 18.7 gm/dL | 12.1 | 13.1 | 12.7 - 14.5 | | 41 | Hb-Control | 17.0 gm/dL | 27.1 | 27.1 | +/- 1.3 | | 42 | Hb | 18.7 gm/dL | 24.0 | 25.4 | 25.8 - 28.4 | | 46 | Hb-Control | 17.0 gm/dL | 68.2 | 68.2 | +/- 5.4 | | 47 | Hb | 18.7 gm/dL | 60.6 | 63.6 | 62.8 - 73.6 | | Results of Interference Testing of the GBI TSH EIA Screening Assay | | | | | |--------------------------------------------------------------------|--|--|--|--| |--------------------------------------------------------------------|--|--|--|--| ## Results of Additional Hemoglobin Interference Testing of the GB1 TSH EIA Screening Assay | DBS Sample ID# | Hemoglobin<br>Concentration -<br>(gm/dL) | TSH uIU/ml -<br>Expected<br>Value | TSH uIU/ml -<br>Observed<br>(Mean; n=6) | Acceptance<br>Criteria<br>(+/- 1 SD of<br>Control Value) | |------------------|------------------------------------------|-----------------------------------|-----------------------------------------|----------------------------------------------------------| | C1+ Saline Ctrl | 17 | na | 14.8 | 14.8 +/- 0.9 | | C1+ Lysate (Hb) | 40 | 14.8 | 14.3 | (13.9 - 15.7) | | C1+ Lysate (Hb) | 80 | 14.8 | 15.2 | | | C2 + Saline Ctrl | 17 | na | 33.7 | 33.7 +/- 1.3 | | C2 + Lysate (Hb) | 40 | 33.7 | 32.3 | (32.4 - 35.0) | | C2 + Lysate (Hb) | 80 | 33.7 | 31.5 | | | C3 + Saline Ctrl | 17 | na | 53.4 | 53.4 +/- 4.2 | | C3 + Lysate (Hb) | 40 | 53.4 | 50.5 | (49.2 - 57.6) | | C3 + Lysate (Hb) | 80 | 53.4 | 51.9 | | | C4+ Saline Ctrl | 17 | na | 92.1 | 92.1 +/- 6.5 | | C4+ Lysate (Hb) | 40 | 92.1 | 89.9 | (85.6 - 98.6) | | C4+ Lysate (Hb) | 80 | 92.1 | 90.7 | | {11}------------------------------------------------ ### Method Comparison A total of 995 neonatal blood spot specimens, 980 randomly selected from a presumed normal population and 15 from patients confirmed as positive for hypothyroidism, were analyzed in-house with the GBI TSH EIA. Results were compared to those obtained using a commercially available neonatal TSH assay (Predicate). Refer to the Table below. ### Comparison of Result Interpretations Obtained by the GBI TSH EIA versus a Predicate TSH EIA | Population N=995 | | GBI Results | | | | | |--------------------------|------------|-------------|------------|----------|---------------|-----| | Published Cut-off Ranges | | Normal | Borderline | Positive | | | | | | < 20 | 20 - 40 | > 40 | (Row) Totals: | | | Predicate Results | Normal | (<20) N=974 | 967 | 7 | 0 | 974 | | | Borderline | (20-40) N=6 | 2 | 4 | 0 | 6 | | | Positive | (>40) N=15 | 0 | 0 | 15 | 15 | | (Column) Totals: | | | 969 | 11 | 15 | 995 | Of the 995 total blood spot samples tested, 163 results produced non-numeric values (either "less than the limit of sensitivity", n=153; or "greater than the highest standard", n=10) for one or both assays. These samples could not therefore be included in statistical analysis presented below and account for the observed difference in method comparison population "N" (995 - 163 = 832), The method comparison results for the N=832 population were: GBI TSH Mean = 8.9 uIU/ml, with a range of 2.4 to 114 uIU/ml, Predicate TSH Mean = 8.1 uTU/ml, with a range of 2.9 to 104 uIU/ml. The results of linear regression analysis were: y (GBI TSH EIA) = 1.0695x (Predicate) + 0.2671, R2 = 0.9185 {12}------------------------------------------------ Image /page/12/Picture/1 description: The image shows the seal of the Department of Health & Human Services - USA. The seal features a stylized eagle-like symbol with flowing lines, representing the department's mission. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" is arranged in a circular pattern around the symbol. The seal is presented in black and white. Public Health Service Food and Drug Administration 2098 Gaither Road Rockville MD 20850 Golden Bridge International, Inc. c/o Dr. Xiaoping Zhang 9700 Harbour Place, Suite 129 Mukilteo, WA 98275 APR 1 0 2008 Re: k072458 Trade/Device Name: GBI TSH Neonatal Screening Kit Regulation Number: 21 CFR 862.1690 Regulation Name: Thyroid Stimulating Hormone Test System Regulatory Class: II Product Code: JLW Dated: March 10, 2008 Received: March 12, 2008 Dear Dr. Zhang: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in Title 21, Code of Federal Regulations (CFR), Parts 800 to 895. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Parts 801 and 809); and good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820). {13}------------------------------------------------ Page 2 - This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific information about the application of labeling requirements to your device, or questions on the promotion and advertising of your device, please contact the Office of In Vitro Diagnostic Device Evaluation and Safety at (240) 276-0490. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address at http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours. Jean M. Cooper, M.S., D.V.M. Yean M. Cooper, M.S., D.V.M. Director Division of Chemistry and Toxicology Office of In Vitro Diagnostic Device Evaluation and Safety Center for Devices and Radiological Health Enclosure {14}------------------------------------------------ # Indication for Use 510(k) Number (if known): k072458 Device Name: GBI TSH Neonatal Screening Kit Indication For Use: The GBI TSH Neonatal Screening Kit is designed for the quantitativ…
Innolitics

Panel 1

/
Sort by
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...