← Product Code [GEX](/productcode/GEX) · K072424

# COOLTOUCH, MODEL LC215,COOLLIPO (K072424)

_New Star Lasers, Inc. · GEX · Oct 3, 2007 · General, Plastic Surgery · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K072424

## Device Facts

- **Applicant:** New Star Lasers, Inc.
- **Product Code:** [GEX](/productcode/GEX.md)
- **Decision Date:** Oct 3, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 878.4810
- **Device Class:** Class 2
- **Review Panel:** General, Plastic Surgery
- **Attributes:** Therapeutic

## Indications for Use

For use in dermatology for incision, excision, ablation and vaporization with hemostasis of soft tissue. For use in the treatment of fine lines and wrinkles. For treatment of back acne and atrophic acne scars. For treatment of reflux of the great and small saphenous veins associated with varicose veins and varicosities

## Device Story

CoolTouch LC215 is a pulsed Nd:YAG surgical laser system emitting at 1320nm. System comprises a cabinet housing power supply, cooling system, microcontroller, and laser source; connected to fiber optics and a handpiece or JouleTracker. Operated by clinicians in dermatology or surgical settings. Laser energy is delivered to soft tissue to perform incision, excision, ablation, and vaporization with hemostasis. Used for cosmetic dermatological applications (fine lines, wrinkles, acne, scars) and endovenous treatment of saphenous vein reflux. Microcontroller manages laser parameters; clinician controls delivery via handpiece. Output provides therapeutic tissue effects; benefits include precise tissue management and treatment of venous insufficiency.

## Clinical Evidence

No clinical data provided; bench testing only.

## Technological Characteristics

Nd:YAG pulsed surgical laser; 1320nm wavelength. System includes cabinet (power supply, cooling, microcontroller), fiber optics, and handpiece/JouleTracker. Class II device.

## Regulatory Identification

(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.

## Predicate Devices

- CoolTouch CT3S Nd:YAG Laser System

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

KD72424

## OCT 3 * 2007

## Premarket Notification 510(k) Summary CoolTouch LC215 Nd:YAG Laser System

This 510(K) Summary of safety and effectiveness for the CoolTouch LC215 Nd:YAG surgical laser system is submitted in accordance with the requirements of 21CFR 807.92.

| Applicant:                                                    | New Star Lasers, Inc. dba CoolTouch, Inc.                                                                                                                                                                                                                                                                                                       |
|---------------------------------------------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Address:                                                      | 9085 Foothills Boulevard<br>Roseville, CA 95747                                                                                                                                                                                                                                                                                                 |
| Contact Person:                                               | Natalie R. Vollrath                                                                                                                                                                                                                                                                                                                             |
| Telephone:                                                    | (916) 677-1900                                                                                                                                                                                                                                                                                                                                  |
| Fax:                                                          | (916) 677-1901                                                                                                                                                                                                                                                                                                                                  |
| Preparation Date:                                             | August 27, 2007                                                                                                                                                                                                                                                                                                                                 |
| Device Trade Name:                                            | CoolTouch LC215 Nd: YAG Surgical Laser                                                                                                                                                                                                                                                                                                          |
| Common Name:                                                  | Nd: YAG Pulsed Surgical Laser                                                                                                                                                                                                                                                                                                                   |
| Classification Name:                                          | Instrument, Surgical Powered, Laser 79-GEX                                                                                                                                                                                                                                                                                                      |
| Legally Marketed Predicate<br>Device:                         | CoolTouch CT3S<br>Nd:YAG Laser System                                                                                                                                                                                                                                                                                                           |
| Description of the CoolTouch<br>LC215 Nd:YAG Surgical Laser:  | The LC215 Nd: YAG Laser System produces laser emission<br>at 1320nm. The laser consists of three interconnected<br>sections: the cabinet which houses the power supply, the<br>cooling system, the microcontroller and the laser, the fiber<br>optics, and the handpiece or JouleTracker                                                        |
| Intended use of the CoolTouch<br>LC215 Nd:YAG Surgical Laser: | For use in dermatology for incision, excision, ablation and<br>vaporization with hemostasis of soft tissue. For use in the<br>treatment of fine lines and wrinkles. For treatment of back<br>acne and atrophic acne scars. For treatment of reflux of the<br>great and small saphenous veins associated with varicose<br>veins and varicosities |
| Performance Data:                                             | None                                                                                                                                                                                                                                                                                                                                            |
| Conclusion:                                                   | Based on the evaluation of the risks and hazards and<br>including various testing of the modifications, the<br>CoolTouch LC215 Nd:YAG Surgical Laser System is<br>substantially equivalent to the predicate device, the CT3S                                                                                                                    |

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image is a circular seal or logo. The seal contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the circumference of the circle. In the center of the circle is a stylized emblem that resembles three abstract human profiles or faces stacked on top of each other, with flowing lines suggesting movement or connection.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

New Star Lasers, Inc. dba CoolTouch Inc % Ms. Natalie Vollrath Quality Assurance Manager 9085 Foothills Boulevard Roseville, California 95747

OCT 3 * 2007

Re: K072424

Trade/Device Name: CoolTouch LC215 Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX Dated: September 18, 2007 Received: September 19, 2007

Dear Ms. Vollrath:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the

{2}------------------------------------------------

Page 2 - Ms. Vollrath

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled. "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark N. Melkerson
Director
Division of General, Restorative

and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

| 510(k) Number | Pending |
|---------------|---------|
|---------------|---------|

Device Name

## CoolTouch LC215

Indications for Use

The CoolTouch Model LC215 Nd:YAG Surgical Laser is indicated for the following:

> a) for use in dermatology for incision, excision, ablation and vaporization with hemostasis of soft tissue;

b) for use in the treatment of fine lines and wrinkles;

c) for treatment of back acne and atrophic acne scars, and;

d) for treatment of reflux of the great and small saphenous veins associated with varietse veins and varicosities.

> (Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number_

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-the-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE – CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Page 1 of 1

---

**Source:** [https://fda.innolitics.com/device/K072424](https://fda.innolitics.com/device/K072424)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
