SOFTPAP CERVICAL CELL COLLECTOR
Device Facts
| Record ID | K072381 |
|---|---|
| Device Name | SOFTPAP CERVICAL CELL COLLECTOR |
| Applicant | Cytocore, Inc. |
| Product Code | HHT · Obstetrics/Gynecology |
| Decision Date | Jan 30, 2008 |
| Decision | SESE |
| Submission Type | Special |
| Regulation | 21 CFR 884.4530 |
| Device Class | Class 2 |
Indications for Use
For the collection of cervical cytology specimens for Pap test analysis. The SoftPAP™ Cervical Sample Collector should not be used after the first 10 weeks of gestation in pregnant patients.
Device Story
SoftPAP Collector is a multicomponent assembly for collecting exfoliated cervical epithelial cells; consists of disposable balloon assembly and reusable handle. Physician inserts device; balloon inflates against cervix to conformally contact ectocervical, transformation zone, and endocervical surfaces. Handle mechanism expands balloon and controls inflation via air reservoir; button-actuated inflation and release-actuated deflation. Disposable sanitary cover minimizes handle contamination. Device facilitates specimen collection for Pap test analysis; provides conformal surface for cell collection. Used in clinical settings by physicians.
Clinical Evidence
No clinical data; bench testing only.
Technological Characteristics
Multicomponent assembly: reusable handle and disposable balloon assembly. Balloon provides conformal surface for cell collection. Inflation via air reservoir in handle; mechanical deflation mechanism. Includes disposable sanitary cover. Non-electronic, manual instrument.
Indications for Use
Indicated for collection of cervical cytology specimens for Pap test analysis. Contraindicated for use after the first 10 weeks of gestation in pregnant patients.
Regulatory Classification
Identification
An obstetric-gynecologic specialized manual instrument is one of a group of devices used during obstetric-gynecologic procedures to perform manipulative diagnostic and surgical functions (e.g., dilating, grasping, measuring, and scraping), where structural integrity is the chief criterion of device performance. This type of device consists of the following:(1) An amniotome is an instrument used to rupture the fetal membranes. (2) A circumcision clamp is an instrument used to compress the foreskin of the penis during circumcision of a male infant. (3) An umbilical clamp is an instrument used to compress the umbilical cord. (4) A uterine curette is an instrument used to scrape and remove material from the uterus. (5) A fixed-size cervical dilator is any of a series of bougies of various sizes used to dilate the cervical os by stretching the cervix. (6) A uterine elevator is an instrument inserted into the uterus used to lift and manipulate the uterus. (7) A gynecological surgical forceps is an instrument with two blades and handles used to pull, grasp, or compress during gynecological examination. (8) A cervical cone knife is a cutting instrument used to excise and remove tissue from the cervix. (9) A gynecological cerclage needle is a looplike instrument used to suture the cervix. (10) A hook-type contraceptive intrauterine device (IUD) remover is an instrument used to remove an IUD from the uterus. (11) A gynecological fibroid screw is an instrument used to hold onto a fibroid. (12) A uterine sound is an instrument used to determine the depth of the uterus by inserting it into the uterine cavity. (13) A cytological cervical spatula is a blunt instrument used to scrape and remove cytological material from the surface of the cervix or vagina. (14) A gynecological biopsy forceps is an instrument with two blades and handles used for gynecological biopsy procedures. (15) A uterine tenaculum is a hooklike instrument used to seize and hold the cervix or fundus. (16) An internal pelvimeter is an instrument used within the vagina to measure the diameter and capacity of the pelvis. (17) A nonmetal vaginal speculum is a nonmetal instrument used to expose the interior of the vagina. (18) A fiberoptic nonmetal vaginal speculum is a nonmetal instrument, with fiberoptic light, used to expose and illuminate the interior of the vagina.
Special Controls
*Classification.* (1) Class II (special controls). The device, when it is an umbilical clamp with or without a cutter, a uterine tenaculum which is sterile and does not use suction and is intended for single use, a nonmetal vaginal speculum, or a fiberoptic nonmetal vaginal speculum, is exempt from the premarket notification procedures in subpart E of part 807 of this chapter subject to the limitations in § 884.9.(2) Class I for the amniotome, uterine curette, cervical dilator (fixed-size bougies), cerclage needle, IUD remover, uterine sound, and gynecological biopsy forceps. The devices subject to this paragraph (b)(2) are exempt from the premarket notification procedures in subpart E of part 807 of this chapter, subject to the limitations in § 884.9.
Predicate Devices
- CytoCore e2™ Collector (K013285)