APL MEDICAL SYSTEM
K072344 · Bios S.R.L. · GEX · Dec 3, 2007 · General, Plastic Surgery
Device Facts
| Record ID | K072344 |
| Device Name | APL MEDICAL SYSTEM |
| Applicant | Bios S.R.L. |
| Product Code | GEX · General, Plastic Surgery |
| Decision Date | Dec 3, 2007 |
| Decision | SESE |
| Submission Type | Traditional |
| Regulation | 21 CFR 878.4810 |
| Device Class | Class 2 |
| Attributes | Therapeutic |
Intended Use
The APL MEDICAL SYSTEM is a modular multi-laser and multi-use device and are intended for use in aesthetic, cosmetic, and surgical applications. It is used to treat common imperfections and lesions of skin and body , including: - 1. The removal of unwanted hair and to effect stable long-term or permanent hair reduction - 2. Laser skin treatment procedures for the treatment of: - Acne scars . - Wrinkles . - 3. Incision, excision, ablation, vaporization of soft tissue - The non-ablative treatment of facial wrinkles, such as, but not limited to: 4. - Periocular wrinkles . - . Perioral wrinkles - 5. Removal of tattoos and blemishes of the skin, the treatment of various pigmentation lesions - 6. Treatment of capillaries, teleangiectasies and vascular lesions - 7. Treatment of vitiligo and psoriasis The equipment should only be used under medical supervision.
Device Story
APL Medical System is a modular, multi-laser and pulsed light platform for aesthetic and surgical skin treatments. System comprises a base unit containing electronics and a closed-loop water-air cooling circuit, with interchangeable handpieces for specific applications: pulsed light (hair removal, psoriasis, vitiligo, wrinkles), Nd:YAG 1064nm long-pulse (vascular lesions, hair removal), Nd:YAG 1064/532nm Q-switched (pigmented lesions, tattoos), and Erbium 2940nm (skin ablation/exfoliation). Operated by clinicians in medical settings. Clinician selects appropriate handpiece and parameters based on patient skin type and condition. Device output (laser/light energy) targets specific chromophores (melanin, hemoglobin, water) to achieve therapeutic effects like tissue ablation, selective photothermolysis, or vascular coagulation. Microprocessor-controlled system ensures precise delivery of pulse duration and fluence. Benefits include non-invasive or minimally invasive treatment of skin imperfections with minimal damage to surrounding tissues.
Clinical Evidence
No clinical data provided. Bench testing only.
Technological Characteristics
Modular platform with interchangeable handpieces. Energy sources: Flashlamp-pumped Nd:YAG (1064nm, 532nm), Pulsed Light, and Erbium (2940nm). Features internal closed-loop water-air heat exchanger for thermal control. Microprocessor-controlled. Class IIIA aiming beams. Connectivity: Modular handpiece interface. Sterilization: Not specified.
Indications for Use
Indicated for patients requiring aesthetic, cosmetic, or surgical skin procedures, including hair removal, treatment of acne scars, wrinkles (periocular/perioral), pigmentated lesions, tattoos, vascular lesions (capillaries/teleangiectasies), vitiligo, and psoriasis. For use under medical supervision.
Regulatory Classification
Identification
(1) A carbon dioxide laser for use in general surgery and in dermatology is a laser device intended to cut, destroy, or remove tissue by light energy emitted by carbon dioxide.(2) An argon laser for use in dermatology is a laser device intended to destroy or coagulate tissue by light energy emitted by argon.
Predicate Devices
- MSq(M2) LTD - Lovely System (k042000 and k033946)
Related Devices
- K153229 — Superbium · Bios S.R.L. · Apr 12, 2016
- K123777 — MULTILASER SYSTEM · Ultralight Laser Technologies, LLC · Apr 11, 2013
- K113018 — APOLLO V+ MEDICAL PLATFORM, APOLLO IV+ MEDICAL PLATFORM, NICE STATION LIGHT BASED PLATFORM · BEIJING SYNTECH LASER CO., LTD. · Dec 14, 2012
- K221766 — Solon · Lmg Lasers Comercio, Importacao E Exportacao Ltda · Jan 5, 2024
- K212790 — XLase Plus · Biotec Italia, Srl · Aug 15, 2022
Submission Summary (Full Text)
{0}------------------------------------------------
## K 072344
## 510(k) SUMMARY
| Submitter: | Bios Italia s.r.l.<br>Via Guido Rossa, 10/12<br>20090 Vimodrone (MI) - Italy | |
|------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|-------------|
| Contact: | Aldo Casalino | DEC 03 2007 |
| Date Summary Prepared: | June 27, 2007 | |
| Device Trade Name: | APL MEDICAL SYSTEM | |
| Common Name: | Medical Laser System | |
| Classification Name: | Instrument, surgical, powered, laser<br>GEX and FTC<br>21 CFR 878.4810 and 21 CFR 878.4630 | |
| Equivalent Device: | MSq(M2) LTD - Lovely System (k042000 and k033946) | |
| Device Description: | The portable device APL MEDICAL SYSTEM, is a multi-<br>laser and pulsed light system, designated to act on the most<br>diffused imperfections of the skin and body. The<br>equipmentm is conceived as an expandable system for the<br>field of aesthetic medicine, composed of a modular<br>platform, containing the entire electronics of the system and<br>the cooling circuit, and a series of interchangeable hand<br>pieces to be connected on the platform, finalized to treat<br>specific groups of imperfections. A rapid connection<br>enables the quick and easy replacement of the various hand<br>pieces that can be mounted on the modular platform. The<br>following hand pieces are available: | |
| | 1) Hand piece Pulsed Light. With the use of this hand piece<br>the removal of unwanted hair is extremely quick and<br>efficient, as it is possible to treat large areas (up to 9<br>cm2) using a single pulse or "burst" mode This system<br>is effective on varius skin types. On this hand piece it is<br>possible to change easily the guide light, equipped with<br>a specific light filter and to use thus that one most<br>adequate for the individual photo type. Furthermore, it is | |
{1}------------------------------------------------
possible to conduct a combination of methods to attenuate facial skin imperfections related to small perorbital and labial wrinkles. With the use of guide lights within the spectrum 300-380 nm it is possible to treat psoriasis and vitiligo disease with a specific therapy. This technique has the advantage of acting directly on the concerned area while respecting the surrounding tissues.
- 2) Hand piece Nd:Yag Laser at 1064nm, long pulse (ms). With this laser it is possible to treat teleangectasies easyly and quickly with optimal results and without leaving a cicatrix. Furthermore, it is possible to hair removal on small areas with "difficult" skin and hairs.
- 3) Hand piece Nd: Yag Laser 1064 and 532 nm Qs (Qswitched), short pulse (ns). This laser is able to selectively destroy melanosomes through cell lysis. The Qs laser is ideal for the treatment of pigmentated lesions and in general for all superficial pigmentations, since its wavelength is well absorbed by the melanin as well as by the haemoglobin. Furthermore, the pulse width makes this type of laser selective with regards to melanosomes -i.e. benign freckles, ephelis, café-au-lait stains, nervus pilus, seborrheic keratosis and skin chloasm, and tattoos. Fin and superficial vessels are better treated with the Nd: Yag laser 532 nm Qs which causes less pain than other laser types, completing the treatment which may effect the Nd:Yag laser, long pulse.
- 4) Hand piece Erbium Laser at 2.940 nm. Due to the high power and the exclusive guide light, the use of this laser results in the ablation process which is a delicate and uniform exfoliation of the atrophic superficial (5 micron) layer of the skin.
The APL MEDICAL SYSTEM is indicate for: removal of Intended Use: unwanted hair and to effect stable long-term or permanent hair reduction; laser skin procedures for the treatment of acne scars and wrinkles; incision, excision, ablation, vaporization of soft tissue; non-ablative treatment of facial wrinkles, such as, but not limited to periocular wrinkles and perioral wrinkles: removal of tattoos and blemishes of the skin, the treatment of various pigmentation lesions, treatment of capillaries, teleangiectasies and vascular lesions, the treatment of vitiligo and psoriasis.
{2}------------------------------------------------
| Rationale for Substancial<br>Equivalence: | The product specification, functionality, indication for use,<br>and treatment parameters of the APL MEDICAL SYSTEM<br>are the same or very similar to the legally marketed laser<br>MSq(M2) LTD - Lovely System Model: Lovely II<br>Harmony. |
|-------------------------------------------|-----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| | Both systems have the same indication for use. |
| | Both systems comprise a flashlamp pumped laser rod<br>(Nd:YAG) generating light at a wavelength of 1064 nm,<br>which is subsequently delivered to the patient via an optical<br>fiber delivery system, and focusing handpiece. |
| | The APL MEDICAL SYSTEM output characteristics<br>(including pulse duration and fluence) are identical, or very<br>similar, to those of the predicate device. |
| | Both lasers utilize class IIIA aiming beams. |
| | Both lasers are microprocessor controlled devices. |
| | Both systems utilize an internal closed loop water-air heath<br>exchanger circuit for optimal thermal control of laser cavity |
| | The risks and benefits for the APL MEDICAL SYSTEM are<br>comparable to those for the predicate device. Therefore, the<br>introduction of this laser should not raise new questions of<br>Safety and Effectiveness . |
.
Non-Clinical Performance Data: None
Clinical Performance Data: None
:
{3}------------------------------------------------
Image /page/3/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular border with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is an abstract symbol resembling an eagle or bird in flight, composed of three curved lines.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
DEC 0 3 2007
Bios Italia s.r. % Mr. Aldo Casalino General Manager Via Guido Rossa, 10/12 Vinodrone (Milan) MI I-20090 Italy
Re: K072344 Trade/Device Name: APL Medical System Regulation Number: 21 CFR 878.4810 Regulation Name: Laser surgical instrument for use in general and plastic surgery and in dermatology Regulatory Class: II Product Code: GEX, FTC Dated: November 8, 2007 Received: November 27, 2007
Dear Mr. Casalino:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set
{4}------------------------------------------------
Page 2 - Mr. Aldo Casalino
forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus. permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours.
Mark H. Mullman
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATION FOR USE STATEMENT
510(k) Number (if known): K072344
Device Name: APL MEDICAL SYSTEM
Sponsor Name: Bios Italia s.r.l.
Indication for Use:
The APL MEDICAL SYSTEM is a modular multi-laser and multi-use device and are intended for use in aesthetic, cosmetic, and surgical applications.
It is used to treat common imperfections and lesions of skin and body , including:
- 1. The removal of unwanted hair and to effect stable long-term or permanent hair reduction
- 2. Laser skin treatment procedures for the treatment of:
- Acne scars .
- Wrinkles .
- 3. Incision, excision, ablation, vaporization of soft tissue
- The non-ablative treatment of facial wrinkles, such as, but not limited to: 4.
- Periocular wrinkles .
- . Perioral wrinkles
- 5. Removal of tattoos and blemishes of the skin, the treatment of various pigmentation lesions
- 6. Treatment of capillaries, teleangiectasies and vascular lesions
- 7. Treatment of vitiligo and psoriasis
The equipment should only be used under medical supervision.
Do Not Write Below This Line – Continue on Another Page if Needed
Concurrence of CDRH, Office of Device Evaluation (ODE)
| Prescription Use | <span style="font-size: 2em;">☑</span> |
|----------------------|----------------------------------------|
| Over-The-Counter Use | □ |
(Division Sign-Off)
Division of General, Restorative,
and Neurological Devices
510(k) Number: k6723449