← Product Code [KDI](/productcode/KDI) · K072232

# GAMBRO POLYFLUX HEMODIALYZER, MODEL: HD-C4 SMALL (K072232)

_Gambro Renal Products, Inc. · KDI · Sep 7, 2007 · Gastroenterology, Urology · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K072232

## Device Facts

- **Applicant:** Gambro Renal Products, Inc.
- **Product Code:** [KDI](/productcode/KDI.md)
- **Decision Date:** Sep 7, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 876.5860
- **Device Class:** Class 2
- **Review Panel:** Gastroenterology, Urology
- **Attributes:** Therapeutic

## Indications for Use

The dialyzer is intended for use in hemodialysis for the treatment of chronic and acute renal failure.

## Device Story

Hemodialyzer for acute/chronic renal failure treatment; utilizes hollow fiber membrane technology. Blood enters inlet port, flows through hollow fibers; hydrostatic/transmembrane pressure drives plasma water and solutes across membrane into dialysate compartment via diffusion and convection. Uremic toxins/waste products removed; dialysate exits via outlet port. Device used in clinical hemodialysis settings; operated by trained healthcare professionals. Output is purified blood returned to patient; clinical decision-making based on patient renal status and dialysis prescription. Benefits include effective waste removal and fluid management for renal failure patients.

## Clinical Evidence

No clinical data. Substantial equivalence supported by in vitro bench testing comparing performance characteristics to the predicate device.

## Technological Characteristics

Hollow fiber hemodialyzer. Membrane material: blend of polyarylethersulfone (PAES) and polyvinylpyrrolidone (PVP). Operation: diffusion and convection across semi-permeable membrane. Single-use. No electronic components or software.

## Regulatory Identification

A high permeability hemodialysis system is a device intended for use as an artificial kidney system for the treatment of patients with renal failure, fluid overload, or toxemic conditions by performing such therapies as hemodialysis, hemofiltration, hemoconcentration, and hemodiafiltration. Using a hemodialyzer with a semipermeable membrane that is more permeable to water than the semipermeable membrane of the conventional hemodialysis system (§ 876.5820), the high permeability hemodialysis system removes toxins or excess fluid from the patient's blood using the principles of convection (via a high ultrafiltration rate) and/or diffusion (via a concentration gradient in dialysate). During treatment, blood is circulated from the patient through the hemodialyzer's blood compartment, while the dialysate solution flows countercurrent through the dialysate compartment. In this process, toxins and/or fluid are transferred across the membrane from the blood to the dialysate compartment. The hemodialysis delivery machine controls and monitors the parameters related to this processing, including the rate at which blood and dialysate are pumped through the system, and the rate at which fluid is removed from the patient. The high permeability hemodialysis system consists of the following devices:(1) The hemodialyzer consists of a semipermeable membrane with an in vitro ultrafiltration coefficient (K uf ) greater than 8 milliliters per hour per conventional millimeter of mercury, as measured with bovine or expired human blood, and is used with either an automated ultrafiltration controller or anther method of ultrafiltration control to prevent fluid imbalance.(2) The hemodialysis delivery machine is similar to the extracorporeal blood system and dialysate delivery system of the hemodialysis system and accessories (§ 876.5820), with the addition of an ultrafiltration controller and mechanisms that monitor and/or control such parameters as fluid balance, dialysate composition, and patient treatment parameters (e.g., blood pressure, hematocrit, urea, etc.). (3) The high permeability hemodialysis system accessories include, but are not limited to, tubing lines and various treatment related monitors (e.g., dialysate pH, blood pressure, hematocrit, and blood recirculation monitors).

## Special Controls

*Classification.* Class II. The special controls for this device are FDA's:(1) “Use of International Standard ISO 10993 ‘Biological Evaluation of Medical Device—Part I: Evaluation and Testing,’ ”
(2) “Guidance for the Content of 510(k)s for Conventional and High Permeability Hemodialyzers,”
(3) “Guidance for Industry and CDRH Reviewers on the Content of Premarket Notifications for Hemodialysis Delivery Systems,”
(4) “Guidance for the Content of Premarket Notifications for Water Purification Components and Systems for Hemodialysis,” and
(5) “Guidance for Hemodialyzer Reuse Labeling.”

## Predicate Devices

- Polyflux HD-C4 (BIG) ([K060195](/device/K060195.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K072232

Image /page/0/Picture/1 description: The image shows the logo for Gambro, a medical technology company. The logo consists of a stylized "G" symbol on the left, followed by the word "GAMBRO" in bold, sans-serif font. A registered trademark symbol appears to the right of the word "GAMBRO".

SEP -7 2007

#### 510(k) Summary

| Submitter:           | Gambro Renal Products, Inc.<br>14143 Denver West Parkway<br>Lakewood, Colorado 80401 |         |
|----------------------|--------------------------------------------------------------------------------------|---------|
| Contact:             | Kae Miller, Manager, Regulatory Affairs<br>Phone: 303-542-5045<br>Fax: 303-876-9264  |         |
| Date prepared:       | August 8, 2007                                                                       |         |
| Device name:         | Gambro Polyflux HD-C4 SMALL Dialyzer for Single Use                                  |         |
| Common name:         | Hemodialyzer                                                                         |         |
| Classification name: | High Permeability Hemodialysis System Accessory (876.5860)                           |         |
| Predicate Devices:   |                                                                                      |         |
| Polyflux HD-C4 (BIG) | Hemodialyzer                                                                         | K060195 |

### Device Description:

This device is intended for use in hemodialysis for the treatment of acute and chronic renal failure.

The intended population of this device is identical to those of the Polyflux HD-C4 (BIG), cleared for marketing in the United States under 510K notification K060195.

The membrane used in this device is a blend of polyarylethersulfone (PAES) and polyviny)pyrrolidone(PVP), which is equivalent to the membrane utilized in the Gambro Polyflux HD-C4 (BIG) single use hemodialyzers cleared for marketing in the United States under 510K Notification (K060195).

Blood enters a blood inlet port where it is distributed to the hollow fibers. The patient's blood traverses the inside of the hollow fibers and exits the device via a blood exit port. By means of a hydrostatic pressure or transmembrane pressure which is created by a combination of positive and negative pressures across the membrane, plasma water along with certain lower and middle molecular weight solutes pass through the membrane and into the dialysate or filtrate compartment of the device. Uremic toxins and waste products are removed from the patient's blood in this device by means of both diffusion and convection through the membrane and into the countercurrent flowing dialysis solution during hemodialysis. The dialysate exits the devices via a dialysate outlet port.

### Indications For Use:

The dialyzer is intended for use in Hemodialysis for the treatments of chronic or acute renal failure.

{1}------------------------------------------------

Image /page/1/Picture/0 description: The image shows the word "GAMBRO" in a bold, sans-serif font. To the left of the word is a circular logo with a stylized "G" shape inside. A small registered trademark symbol is located to the right of the word.

K072232 Page 242

# Technological Characteristics:

The proposed device configurations have the same technological characteristics and are similar in design, function, composition, and operation, to the currently marketed configurations.

### Summary of Non-Clinical Tests:

In vitro testing was conducted to compare the performance of the configurations to the predicate configurations.

# Summary of Clinical Tests:

N/A

# Conclusion:

Testing performed on the Gambro Polyflux Dialyzers indicates that they are safe, effective and perform as well as the predicate devices, when used in accordance with the instructions for use.

{2}------------------------------------------------

Image /page/2/Picture/0 description: The image shows the logo for the Department of Health & Human Services - USA. The logo features a stylized eagle with three lines forming its body and wings. The words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" are arranged in a circular pattern around the eagle.

Food and Drug Administration 9200 Corporate Blvd. Rockville MD 20850

SEP -7 2007

Ms. Kae Miller Manager, Regulatory Affairs Gambro@ Renal Products, Inc. 14143 Denver West Parkway LAKEWOOD CO 80401

Re: K072232

Trade/Device Name: Gambro Polyflux HD-C4 SMALL Dialyzer for Single Use Regulation Number: 21 CFR §876.5860 Regulation Name: High permeability hemodialysis system Regulatory Class: II Product Code: KDI Dated: August 8 2007 Received: August 10, 2007

Dear Ms. Miller:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (Premarket Approval), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21. Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Image /page/2/Picture/10 description: The image is a circular seal with the text "1906-2006 FDA Centennial" arranged around a central FDA logo. The seal is bordered by a ring of stars and text, creating a commemorative design. The text and logo are in a bold, sans-serif font, emphasizing the centennial celebration of the Food and Drug Administration.

Protecting and Promoting Public Health

{3}------------------------------------------------

Page 2 -

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at one of the following numbers, based on the regulation number at the top of this letter:

| 21 CFR 876.xxx | (Gastroenterology/Renal/Urology | 240-276-011. |
|----------------|---------------------------------|--------------|
| 21 CFR 884.xxx | (Obstetrics/Gynecology)         | 240-276-011. |
| 21 CFR 894.xxx | (Radiology)                     | 240-276-0120 |
| Other          |                                 | 240-276-010- |

Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150

or at its Internet address http://www.fda.gov/cdrl/industry/support/index.html.

Sincerely yours,

Nancy C. Hogdon

Nancy C. Brogdon Director, Division of Reproductive, Abdominal, and Radiological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

Image /page/4/Picture/1 description: The image shows the logo for Gambro. The logo consists of a stylized circular graphic to the left of the word "GAMBRO" in all capital letters. A registered trademark symbol appears to the right of the letter "O" in Gambro.

Indications for Use Statement 510(k) number: 072232 (if known) Gambro Polyflux HD-C4 SMALL Dialyzer for Single Use Device Name: The dialyzer is intended for use in hemodialysis for the Indications for Use: treatment of chronic and acute renal failure. Over-The-Counter Use Prescription Use X AND / OR (21 CFR 807 Subpart C) (Per 21 CFR 801 Subpart D)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Annu Sinha

(Division Sign-Off)
Division of Reproductive, Abdominal,
and Radiological Devices
510(k) Number. K072232

---

**Source:** [https://fda.innolitics.com/device/K072232](https://fda.innolitics.com/device/K072232)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
