← Product Code [LJT](/productcode/LJT) · K072215

# TITANIUM POWERPORT ISP IMPLANTED PORT (K072215)

_C.R. Bard, Inc. · LJT · Nov 1, 2007 · General Hospital · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K072215

## Device Facts

- **Applicant:** C.R. Bard, Inc.
- **Product Code:** [LJT](/productcode/LJT.md)
- **Decision Date:** Nov 1, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 880.5965
- **Device Class:** Class 2
- **Review Panel:** General Hospital
- **Attributes:** Therapeutic

## Indications for Use

The PowerPort™ implanted port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medication, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a PowerLoc™ safety infusion set, the PowerPort™ device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s.

## Device Story

Titanium PowerPort™ isp is an implanted vascular access port with 8 Fr ChronoFlex® polyurethane catheter; designed for long-term, repeated vascular access. Port features triangular body, purple coloring, and three palpation bumps on septum for identification. Accessed via non-coring needle for routine use; requires Bard PowerLoc™ safety infusion set for power injection of contrast media. Device implanted subcutaneously; used by clinicians to deliver medications, fluids, nutrition, or blood products and withdraw blood. Power injection capability allows contrast media delivery at rates up to 5 ml/s, facilitating diagnostic imaging. System benefits include reliable, repeated vascular access while minimizing need for multiple venipunctures.

## Clinical Evidence

No clinical data. Substantial equivalence supported by design verification testing and risk management analysis (FMEA) conducted in accordance with FDA-recognized standards and internal protocols.

## Technological Characteristics

Titanium port body with surface anodization; 8 Fr ChronoFlex® polyurethane catheter. Attached via cathlock compression fitting. Triangular form factor with palpation bumps. Sterilization method not specified. Standalone mechanical device; no software or energy source.

## Regulatory Identification

A subcutaneous, implanted, intravascular infusion port and catheter is a device that consists of a subcutaneous, implanted reservoir that connects to a long-term intravascular catheter. The device allows for repeated access to the vascular system for the infusion of fluids and medications and the sampling of blood. The device consists of a portal body with a resealable septum and outlet made of metal, plastic, or combination of these materials and a long-term intravascular catheter is either preattached to the port or attached to the port at the time of device placement. The device is available in various profiles and sizes and can be of a single or multiple lumen design.

## Special Controls

*Classification.* Class II (special controls) Guidance Document: “Guidance on 510(k) Submissions for Implanted Infusion Ports,” FDA October 1990.

## Predicate Devices

- Titanium PowerPort™ Implanted Port with 8 Fr. ChronoFlex® Polyurethane Catheter ([K060812](/device/K060812.md))
- MRI PowerPort™ Polymeric Implanted Port with 8 Fr. ChronoFlex® Polyurethane Catheter ([K063377](/device/K063377.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

NOV - 1 2007

K072215(A1xA2)

Section 7 - Summary of Safety & Effectiveness

・

Page 24 of 53

## 510(k) Summary

| General<br>Provisions                    | Submitter Name:<br>Address:<br>Telephone Number:<br>Fax Number:<br>Contact Person:<br>Date of Preparation:<br>Registration Number:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                        | Bard Access Systems, Inc. (BAS)<br>[Wholly owned subsidiary of C.R. Bard, Inc.]<br>605 N 5600 W<br>Salt Lake City, UT 84116<br>(801) 595-0700 ext. 5484<br>(801) 595-5425<br>Susan Scott<br>August 8, 2007<br>1720496 BAS<br>2212754 C. R. Bard                                                                                                                                                                                                                                                          |
|------------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|----------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Subject<br>Device                        | Trade Name:<br>Common/Usual Name:<br>Classification Name:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                 | Titanium PowerPort™ isp Implanted Port<br>with 8 Fr. ChronoFlex® Polyurethane Catheter<br>Implanted Infusion Port & Catheter<br>80 LJT - Subcutaneous, Implanted, Intravascular<br>Infusion Port & Catheter                                                                                                                                                                                                                                                                                              |
| Predicate<br>Devices                     | Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Premarket Notification:<br>Trade Name:<br>Common/Usual Name:<br>Classification Name:<br>Premarket Notification:                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                              | Titanium PowerPort™ Implanted Port<br>with 8 Fr. ChronoFlex® Polyurethane Catheter<br>Implanted Infusion Port & Catheter<br>80 LJT - Subcutaneous, Implanted, Intravascular<br>Infusion Port & Catheter<br>K060812, clearance date July 14, 2006<br>MRI PowerPort™ Polymeric Implanted Port<br>with 8 Fr. ChronoFlex® Polyurethane Catheter<br>Implanted Infusion Port & Catheter<br>80 LJT - Subcutaneous, Implanted, Intravascular<br>Infusion Port & Catheter<br>K063377, clearance date Jan 25, 2007 |
| Classification                           | Class II<br>21 CFR §880.5965<br>General Hospital                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Performance<br>Standards                 | Performance standards have not been established by FDA under section 514<br>of the Federal Food, Drug and Cosmetic Act.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                   |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Intended Use                             | PowerPort™ devices are totally implanted vascular access devices designed<br>to provide long-term, repeated access to the vascular system.                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
|                                          |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                           |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Indications<br>for Use                   | The PowerPort™ implanted port is indicated for patient therapies requiring<br>repeated access to the vascular system. The port system can be used for<br>infusion of medication, I.V. fluids, parenteral nutrition solutions, blood<br>products, and for the withdrawal of blood samples.<br>When used with a PowerLoc™ safety infusion set, the PowerPort™ device is<br>indicated for power injection of contrast media. For power injection of<br>contrast media, the maximum recommended infusion rate is 5 ml/s.                                                                                                                                                                                                                                                                                                                                                                                                      |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Device<br>Description                    | The subject Titanium PowerPort™ isp device is a member of the PowerPort™<br>series of power injectable implanted ports. The subject device consists of a<br>titanium port and 8 Fr ChronoFlex® polyurethane catheter that is attached to<br>the port with a cathlock (compression fitting). The subject Titanium<br>PowerPort™ isp device is distinguishable as a member of Bard Access<br>Systems's power injectable port series by the triangular body shape, unique<br>purple coloring, and three palpation bumps on the septum.<br>PowerPort™ implanted ports can be used for routine vascular access using a<br>non-coring access needle. However, for power injection procedures,<br>PowerPort™ ports must be accessed with a Bard PowerLoc™ safety infusion<br>set to create a power injectable system.                                                                                                            |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Technological<br>Characteristics         | Technological characteristics of the subject Titanium PowerPort™ isp device<br>with 8 Fr ChronoFlex® polyurethane catheter are equivalent to those of the<br>Bard Access Systems predicate Titanium PowerPort™ device [K060812].<br>This equivalence extends to basic design, generic materials, construction,<br>and surface anodization treatment. The distinguishing difference exists in the<br>reduced size of the subject port body. The subject port catheter is<br>unchanged from the predicate.                                                                                                                                                                                                                                                                                                                                                                                                                  |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Safety &<br>Performance<br>Tests         | No performance standards have been established under Section 514 of the<br>Food, Drug and Cosmetic Act for this device. However, design verification<br>testing was performed according to protocols based on the<br>recommendations/requirements of applicable FDA guidance and FDA<br>recognized international standards. Verification testing, determined to be<br>applicable to the safety and efficacy of the device, was shown to meet<br>predetermined acceptance criteria listed therein.<br>Risk management, including a failure modes and effects analysis (FMEA), of<br>the subject device was conducted in accordance with an internal protocol<br>based on ISO 14971:2000, Medical Devices - Risk Management for Medical<br>Devices. The analysis did not identify any new types of safety or efficacy<br>questions for the subject Ti PowerPort™ isp device with 8 Fr ChronoFlex®<br>polyurethane catheter. |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |
| Summary of<br>Substantial<br>Equivalence | Based on the indications for use, technological characteristics, and safety and<br>performance testing, the subject Ti PowerPort™ isp device with 8 Fr<br>ChronoFlex® polyurethane catheter meets the minimum requirements that<br>are considered adequate for its intended use and is substantially equivalent in<br>design, materials, sterilization, principles of operation and indications for use<br>to current commercially available power injectable implanted ports cited as<br>predicates.                                                                                                                                                                                                                                                                                                                                                                                                                     |                                                                                                                                                                                                                                                                                                                                                                                                                                                                                                          |

{1}------------------------------------------------

0.2.02 / 72311

Section 7 -- Summary of Safety & Effectiveness

Page 25 of 53

.

{2}------------------------------------------------

Image /page/2/Picture/1 description: The image shows the logo for the U.S. Department of Health and Human Services. The logo consists of a stylized eagle-like symbol with three human profiles incorporated into its design. The text "DEPARTMENT OF HEALTH AND HUMAN SERVICES - USA" is arranged in a circular fashion around the symbol.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

## NOV - 1 2007

Ms. Susan D. Scott Regulatory Affairs Specialist C.R. Bard, Incorporated Bard Access Systems 605 North 5600 West Salt Lake City, Utah 84116

Re: K072215

Trade/Device Name: Titanium PowerPort™ isp Implanted Port with 8 Fr ChronoFlex® Polyurethane Catheter Regulation Number: 880.5965 Regulation Name: Subcutaneous, Implanted, Intravascular Infusion Port and Catheter Regulatory Class: II Product Code: LJT, FPA Dated: October 1, 2007 Received: October 2, 2007

Dear Ms. Scott:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{3}------------------------------------------------

Page 2 – Ms. Scott

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Chiu-Liang, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{4}------------------------------------------------

## Indications for Use Statement

510(k) Number (if known):

Device Name:

Titanium PowerPort™ isp Implanted Port with

8 Fr ChronoFlex® Polyurethane Catheter

Indications for Use:

The PowerPort™ implanted port is indicated for patient therapies requiring repeated access to the vascular system. The port system can be used for infusion of medication, I.V. fluids, parenteral nutrition solutions, blood products, and for the withdrawal of blood samples. When used with a PowerLoc™ safety infusion set, the PowerPort™ device is indicated for power injection of contrast media. For power injection of contrast media, the maximum recommended infusion rate is 5 ml/s.

Prescription Use _ √ (Part 21 CFR §801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR §801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Anthony Ument

. ! Cran-Off) of Anesthesiology, General Hospital, Jann Control, Dental Devices

ber. Kmaals

---

**Source:** [https://fda.innolitics.com/device/K072215](https://fda.innolitics.com/device/K072215)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
