The CARTO® 3 EP Navigation System and accessories are intended for catheterbased atrial and ventricular mapping. The mapping system allows real-time visualization of the catheter as well as display of cardiac maps in a number of different formats. For example, maps may be displayed as cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, and cardiac chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displayed on the System's display screen.
Device Story
CARTO 3 EP Navigation System; catheter-based atrial/ventricular mapping system. Inputs: intracardiac signals from mapping catheters; body surface ECG. Technology: dual-modality location tracking using Sensor Technology (tip) and Active Current Localization (ACL) (shaft electrodes); six external reference patches on patient chest/back. Outputs: real-time 3D electroanatomical maps (activation, propagation, potential, geometry); display of ECG/electrograms. Used in clinical settings by physicians; provides real-time visualization of catheter position and cardiac electrical activity; replaces fluoroscopy-based visualization; facilitates cardiac mapping procedures; improves procedural guidance.
Clinical Evidence
No clinical data. Evidence consists of bench testing, animal studies, and electrical safety testing. System passed all intended criteria.
Technological Characteristics
Catheter-based mapping system; utilizes Sensor Technology and Active Current Localization (ACL). Includes six external reference patches with sensor cables. Hardware/software platform update. Connectivity: system display for real-time mapping and signal visualization. Sterilization: not specified.
Indications for Use
Indicated for patients requiring catheter-based atrial and ventricular mapping for cardiac electrophysiological assessment.
Regulatory Classification
Identification
A programmable diagnostic computer is a device that can be programmed to compute various physiologic or blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices; this device includes any associated commercially supplied programs.
{0}------------------------------------------------
# 5. 510(K) SUMMARY
| Applicant: | Biosense Webster, Inc.<br>3333 Diamond Canyon Rd.<br>Diamond Bar, CA 91765<br>USA<br>Phone: 800-729-9010<br>Fax: 909-839-8804 |
|---------------------------|-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------|
| Date: | November 5, 2007 |
| Contact Person: | Natalie Bennington<br>Manager, Regulatory Affairs |
| Proprietary Device Name: | CARTO® 3 EP Navigation System and Accessories |
| Common Device Name: | Cardiac mapping system and surface reference device |
| Classification Name: | Programmable diagnostic computer<br>(per 21 CFR 870.1245, Product Code DQK) |
| Predicate Device: | CARTO® XP V7 EP Navigation System (K013083)<br>CARTO® RMT V8 EP Navigation System (K060047)<br>RefStar External Reference Patch (K982415, K061468)<br>EnSite System (K060954) |
| Manufacturing Facilities: | System & System Cables<br>Biosense Webster (Israel) Ltd.<br>POB 2009<br>Tirat HaCarmel, 39120<br>Israel<br><br>Accessories (Accessory Cables)<br>Biosense Webster |
.
.
·
{1}------------------------------------------------
## Substantially Equivalent To: 5.1
The Carto® 3 EP Navigation System is substantially equivalent to the CARTO® XP V7 EP Navigation System (K013083, cleared 11/21/01), the CARTO® RMT V8 EP Navigation System (K060047, cleared 6/19/06) and the RefStar External Reference Patch (K982415, cleared 8/10/98 and K061468, cleared 8/14/06), which are all manufactured by Biosense Webster. In addition, the CARTO® 3 System is also substantially equivalent to the EnSite System manufactured by Endocardial Solutions (K060954, cleared 4/21/06).
# Description of the Device Subject to Premarket Notification: 5.2
The CARTO® 3 System is a new platform for the CARTO® Navigation System that incorporates new hardware and software updates.
The CARTO® 3 System is a catheter-based atrial and ventricular mapping system designed to acquire and analyze individual data points, and use this information to display 3D electroanatomical maps of the human heart in real-time. The location information needed to create the cardiac maps and the local electrograms are acquired using a specialized mapping catheter and reference device. The system allows real-time display of cardiac maps based on the received intra cardiac signals from the catheters in a number of different formats. For example, maps may be displayed as cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps and cardiac chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electro grams may also be displayed on the display screen.
Unlike conventional mapping systems that utilize fluoroscopy to visualize the catheter, the CARTO® 3 System uses two distinct types of location technology. In order to locate the tip of the catheter, the System uses "Sensor Technology". In order to locate the electrodes on the catheter shaft, the System uses Active Current Localization (ACL) Technology. Six external reference patches (which are accessories to the system and needed for both tip and electrode localization) are placed externally on the patient's chest and back (three are placed on the chest and three are placed on the patient's back). Sensors at the distal end of each of the six "Patch Sensor Cables" attach to the external reference patches via a stud/snap connector.
When used with the CARTO® 3 System, the intracardiac location of the tip of the navigation catheter (such as the NaviStar Catheter) and the electrodes on the catheter shaft are referenced to the mean calculation obtained from the three sensors located in the Patch Sensor Cables (which are attached to the patient's back) and enable the CARTO® 3 System to construct 3D electrophysiological and electroanatomical maps of the human heart in real-time.
{2}------------------------------------------------
The new features offered with the CARTO® 3 System which differ from the current CARTO® XP V7 EP System include the following:
- · Catheter shaft and electrode visualization.
- · Ability to connect and visualize more than one catheter simultaneously.
- · The CARTO® XP V7 System utilizes an external reference device that incorporates the reference sensor into the adhesive patches. This patch/sensor/cable device is a completely separate device and is sold separately from the CARTO® XP V7 System. The external reference for the CARTO® 3 System, on the other hand, is an integral component of the "Patch Unit" of the system. Only the adhesive patches are sold separately as accessories.
#### 5.3 Indications for Use:
- The CARTO® 3 EP Navigation System and accessories are intended for catheterbased atrial and ventricular mapping. The mapping system allows real-time visualization of the catheter as well as display of cardiac maps in a number of different formats. For example, maps may be displayed as cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps and cardiac chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displayed on the System's display screen.
#### Performance Data: 5.4
The CARTO® 3 System and accessories underwent bench, animal and electrical safety testing. The CARTO® 3 System and accessories passed all intended criteria in accordance with appropriate test criteria and standards.
### 5.5 Overall Performance Conclusions:
Based on the results of the Risk Analysis, the bench, animal and electrical safety studies performed and the comprehensive safety analysis, we concluded that the risk to benefit ratio for the use of the CARTO® 3 System for catheter-based atrial and ventricular mapping has not been negatively affected by the proposed design changes. In summary, the CARTO® 3 System (which incorporates the external reference device within the system) described in this submission is substantially equivalent to Biosense Webster's CARTO® XP V7 EP System, the CARTO® RMT V8 EP System, the RefStar External Reference Patch and to Endocardial Solution's EnSite System and are as safe and effective as the predicate devices for catheter-based atrial and ventricular mapping.
{3}------------------------------------------------
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
NOV 1 9 2007
Biosense Webster, Inc. c/o Ms. Natalie Bennington Manager, Regulatory Affairs 3333 Diamond Canyon Rd. Diamond Bar, CA 91765
Re: K072202
Trade/Device Name: CARTO 3 EP Navigation System (Model FG-5400-00) and CARTO 3 External Reference Patches (Model: D-1283-01) Regulation Number: 21 CFR 870.1425 Regulation Name: Programmable Diagnostic Computer Regulatory Class: Class II (two) Product Code: DQK Dated: November 5, 2007 Received: November 6, 2007
Dear Ms. Bennington:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
{4}------------------------------------------------
# Page 2 - Ms. Natalie Bennington
proceed to the market.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (OS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at 240-276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at 240-276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at 240-276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.
Sincerely yours,
M. Y. Hillebenre
Bram D. Zuckerman, M.D. Director Division of Cardiovascular Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{5}------------------------------------------------
## INDICATIONS FOR USE STATEMENT 4.
510(k) No (if known):
Device Name: CARTO® 3 EP Navigation System and Accessories
# Indications for Use:
The CARTO® 3 EP Navigation System and accessories are intended for catheterbased atrial and ventricular mapping. The mapping system allows real-time visualization of the catheter as well as display of cardiac maps in a number of different formats. For example, maps may be displayed as cardiac electrical activation maps, cardiac electrical propagation maps, cardiac electrical potential maps, and cardiac chamber geometry maps. The acquired patient signals, including body surface ECG and intracardiac electrograms may also be displayed on the System's display screen.
Prescription Use (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
M. A. Willebrenner
Page 1 of 1
Division Sian-C Division of Cardiovascular Devices
510(k) Number
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Guidance alerts
New and changed FDA guidance documents, emailed as a daily digest.
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the ? button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.