PIONEER LOWTOP SPINAL ROD SYSTEM

K072187 · Pioneer Surgical Technology · MNI · Oct 12, 2007 · Orthopedic

Device Facts

Record IDK072187
Device NamePIONEER LOWTOP SPINAL ROD SYSTEM
ApplicantPioneer Surgical Technology
Product CodeMNI · Orthopedic
Decision DateOct 12, 2007
DecisionSESE
Submission TypeTraditional
Regulation21 CFR 888.3070
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Pioneer LowTop Spinal Rod System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), spinal tumor, and failed previous fusion (pseudoarthrosis). In addition, this device is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to S1) with removal of the implants after the attainment of a solid fusion.

Device Story

Pioneer LowTop Spinal Rod System is a pedicle screw spinal fixation system; provides immobilization and stabilization of thoracic, lumbar, and sacral spinal segments. Used as an adjunct to fusion in skeletally mature patients. System components include implants and instruments; surgical implantation performed by physicians in clinical/OR settings. Device stabilizes spinal segments to facilitate fusion; implants are removed after solid fusion is achieved for specific severe spondylolisthesis cases. Benefits include structural support for spinal instabilities, deformities, and trauma recovery.

Clinical Evidence

Bench testing only; no clinical data presented. Performance verified through comparison of design, materials, and mechanical testing per recognized ASTM standards against predicate devices.

Technological Characteristics

Pedicle screw spinal fixation system. Materials conform to ASTM standard specifications. System consists of rods, screws, and associated instrumentation for spinal stabilization. Mechanical performance validated via ASTM testing.

Indications for Use

Indicated for skeletally mature patients requiring spinal immobilization/stabilization as an adjunct to fusion for thoracic, lumbar, and sacral spine conditions: degenerative spondylolisthesis with neurological impairment, trauma (fracture/dislocation), deformities (scoliosis, kyphosis, lordosis), spinal tumor, and pseudoarthrosis. Also indicated for severe Grade 3-4 spondylolisthesis at L5-S1 using autogenous bone graft with L3-S1 fixation.

Regulatory Classification

Identification

(1) Rigid pedicle screw systems are comprised of multiple components, made from a variety of materials that allow the surgeon to build an implant system to fit the patient's anatomical and physiological requirements. Such a spinal implant assembly consists of a combination of screws, longitudinal members (e.g., plates, rods including dual diameter rods, plate/rod combinations), transverse or cross connectors, and interconnection mechanisms (e.g., rod-to-rod connectors, offset connectors).(2) Semi-rigid systems are defined as systems that contain one or more of the following features (including but not limited to): Non-uniform longitudinal elements, or features that allow more motion or flexibility compared to rigid systems.

Special Controls

*Classification.* (1) Class II (special controls), when intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute and chronic instabilities or deformities of the thoracic, lumbar, and sacral spine: severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra; degenerative spondylolisthesis with objective evidence of neurologic impairment; fracture; dislocation; scoliosis; kyphosis; spinal tumor; and failed previous fusion (pseudarthrosis). These pedicle screw spinal systems must comply with the following special controls:(i) Compliance with material standards; (ii) Compliance with mechanical testing standards; (iii) Compliance with biocompatibility standards; and (iv) Labeling that contains these two statements in addition to other appropriate labeling information: “Warning: The safety and effectiveness of pedicle screw spinal systems have been established only for spinal conditions with significant mechanical instability or deformity requiring fusion with instrumentation. These conditions are significant mechanical instability or deformity of the thoracic, lumbar, and sacral spine secondary to severe spondylolisthesis (grades 3 and 4) of the L5-S1 vertebra, degenerative spondylolisthesis with objective evidence of neurologic impairment, fracture, dislocation, scoliosis, kyphosis, spinal tumor, and failed previous fusion (pseudarthrosis). The safety and effectiveness of these devices for any other conditions are unknown.” “Precaution: The implantation of pedicle screw spinal systems should be performed only by experienced spinal surgeons with specific training in the use of this pedicle screw spinal system because this is a technically demanding procedure presenting a risk of serious injury to the patient.” (2) Class II (special controls), when a rigid pedicle screw system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion in the treatment of degenerative disc disease and spondylolisthesis other than either severe spondylolisthesis (grades 3 and 4) at L5-S1 or degenerative spondylolisthesis with objective evidence of neurologic impairment. These pedicle screw systems must comply with the following special controls: (i) The design characteristics of the device, including engineering schematics, must ensure that the geometry and material composition are consistent with the intended use. (ii) Non-clinical performance testing must demonstrate the mechanical function and durability of the implant. (iii) Device components must be demonstrated to be biocompatible. (iv) Validation testing must demonstrate the cleanliness and sterility of, or the ability to clean and sterilize, the device components and device-specific instruments. (v) Labeling must include the following: (A) A clear description of the technological features of the device including identification of device materials and the principles of device operation; (B) Intended use and indications for use, including levels of fixation; (C) Identification of magnetic resonance (MR) compatibility status; (D) Cleaning and sterilization instructions for devices and instruments that are provided non-sterile to the end user; and (E) Detailed instructions of each surgical step, including device removal. (3) Class II (special controls), when a semi-rigid system is intended to provide immobilization and stabilization of spinal segments in the thoracic, lumbar, and sacral spine as an adjunct to fusion for any indication. In addition to complying with the special controls in paragraphs (b)(2)(i) through (v) of this section, these pedicle screw systems must comply with the following special controls: (i) Demonstration that clinical performance characteristics of the device support the intended use of the product, including assessment of fusion compared to a clinically acceptable fusion rate. (ii) Semi-rigid systems marketed prior to the effective date of this reclassification must submit an amendment to their previously cleared premarket notification (510(k)) demonstrating compliance with the special controls in paragraphs (b)(2)(i) through (v) and paragraph (b)(3)(i) of this section.

Predicate Devices

Submission Summary (Full Text)

{1}------------------------------------------------ ## K072187 ## 3.0 Summary of Safety and Effectiveness Information | Sponsor: | Pioneer Surgical Technology<br>375 River Park Circle<br>Marquette, MI 49855<br>(906) 226-4812<br>Contact: Jonathan M. Gilbert | |-----------------------------------------|---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------| | | OCT 12 2007 | | Device Name: | Pioneer LowTop Spinal Rod System | | Classification Name: | Spinal Interlaminal Fixation Orthosis, Spinal Intervertebral<br>Body Fixation Orthosis, Spondylolisthesis Spinal Fixation<br>Device System and Pedicle Screw Spinal System, Class II. | | Classification, Name,<br>Number & Code: | Regulation Number: 888.3070, Class II<br>Pedicle Screw System<br>Product Code: MNI and MNH | | Predicate Device: | K070933 – LowTop Pedicle Screw System (SE Date 6/13/07)<br>K070973 – Quantum Spinal Rod System (SE Date 7/3/07) | | Intended Use: | The Pioneer LowTop Spinal Rod System is intended to<br>provide immobilization and stabilization of spinal segments in<br>skeletally mature patients as an adjunct to fusion in the<br>treatment of the following acute instabilities or deformities of<br>the thoracic, lumbar, and sacral spine: degenerative<br>spondylolisthesis with objective evidence of neurological<br>impairment, trauma (i.e., fracture or dislocation), deformities<br>or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), spinal<br>tumor, and failed previous fusion (pseudoarthrosis).<br><br>In addition, this device is a pedicle screw system indicated for<br>the treatment of severe spondylolisthesis (Grade 3 and 4) of<br>the L5-S1 vertebra in skeletally mature patients receiving<br>fusion by autogenous bone graft having implants attached to<br>the lumbar and sacral spine (L3 to S1) with removal of the<br>implants after the attainment of a solid fusion. | | Material: | Materials used to manufacture the implants and instruments of<br>this system are in conformance with ASTM Standard<br>Specifications. | | Performance Data: | Testing per recognized ASTM standards was presented. | | Performance and SE<br>Determination: | Comparisons of device performance data, materials,<br>indications and design/function to predicate devices were<br>provided in making a determination of substantial equivalence. | {2}------------------------------------------------ Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three legs, and the text "DEPARTMENT OF HEALTH & HUMAN SERVICES. USA" arranged in a circle around the eagle. The eagle is black, and the text is also black. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Pioneer Surgical Technology % Mr. Jonathan M. Gilbert VP of Clinical & Regulatory Affairs 375 River Park Circle Marquette, Michigan 49855 OCT 1 2 2007 Re: K072187 Trade/Device Name: Pioneer LowTop Spinal Rod System Regulation Number: 21 CFR 888.3070 Regulation Name: Pedicle screw spinal system Regulatory Class: Class II Product Code: MNH, MNI Dated: October 2, 2007 Received: October 4, 2007 Dear Mr. Gilbert: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ ## Page 2 - Mr. Jonathan M. Gilbert This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market. If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Darbarc Mchup Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ ## 2.0 Indications for Use Statement K072187 510(k) Number (if known): Device Name: Pioneer LowTop Spinal Rod System Indications for Use: The Pioneer LowTop Spinal Rod System is intended to provide immobilization and stabilization of spinal segments in skeletally mature patients as an adjunct to fusion in the treatment of the following acute instabilities or deformities of the thoracic, lumbar, and sacral spine: degenerative spondylolisthesis with objective evidence of neurological impairment, trauma (i.e., fracture or dislocation), deformities or curvatures (i.e., scoliosis, kyphosis, and/or lordosis), spinal tumor, and failed previous fusion (pseudoarthrosis). In addition, this device is a pedicle screw system indicated for the treatment of severe spondylolisthesis (Grade 3 and 4) of the L5-S1 vertebra in skeletally mature patients receiving fusion by autogenous bone graft having implants attached to the lumbar and sacral spine (L3 to S1) with removal of the implants after the attainment of a solid fusion. Prescription Use (Per 21 CFR 801.109) Over-the-Counter Use (PLEASE DO NOT WRITE BELOW THIS LINE - CONTINUE ON ANOTHER PAGE IF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) OR Deubare Bueher fohrm al. Restorative. and Neurologica **510(k) Number** K07287
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...