← Product Code QSX · K072117

# THROMBIX PATCH THROMBIN HEMOSTASIS PATCH (K072117)

_Vascular Solutions, Inc. · QSX · Oct 5, 2007 · SU · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K072117

## Device Facts

- **Applicant:** Vascular Solutions, Inc.
- **Product Code:** QSX
- **Decision Date:** Oct 5, 2007
- **Decision:** SESE
- **Submission Type:** Traditional
- **Device Class:** Class U
- **Review Panel:** SU
- **Attributes:** Therapeutic

## Intended Use

The Thrombix™ Patch is applied topically and is indicated as an adjunct to manual compression and is indicated for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes.

## Device Story

Thrombix™ Patch is a lyophilized topical hemostat containing bovine-derived thrombin, sodium carboxymethylcellulose (CMC), and calcium chloride. Applied directly over bleeding source at vascular access sites or percutaneous catheter/tube sites; functions as physical barrier combined with adjunctive manual compression. Thrombin facilitates hemostasis by enhancing surface-activated clotting cascade via enzymatic conversion of fibrinogen to fibrin. Used in clinical settings by healthcare providers to manage surface bleeding; benefits patient by accelerating hemostasis at access sites.

## Clinical Evidence

No clinical data; bench testing only. Bench testing included moisture content, thrombin activity, wetting time, and pH. Biocompatibility testing included MEM elution, intracutaneous injection, systemic injection, and Kligman skin sensitization.

## Technological Characteristics

Lyophilized patch containing bovine-derived thrombin, sodium carboxymethylcellulose (CMC), and calcium chloride. Acts as a topical hemostatic barrier. No electronic components or software.

## Regulatory Identification

To temporarily control bleeding and cover external wounds.

## Predicate Devices

- D-Stat® Dry Topical Hemostat ([K030836](/device/K030836.md))
- D-Stat® Dry 3x3 ([K040510](/device/K040510.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

Image /page/0/Picture/0 description: The image contains the logo of the U.S. Food and Drug Administration (FDA). On the left, there is the Department of Health & Human Services logo. To the right of that is the FDA logo, which is a blue square with the letters "FDA" in white. To the right of the blue square is the text "U.S. FOOD & DRUG ADMINISTRATION" in blue.

February 10, 2023

Vascular Solutions, Inc. c/o James Chapman Regulatory Affairs Associate 6464 Sycamore Court Minneapolis, Minnesota 55369

Re: K072117

Trade/Device Name: Thrombix™ Patch, Thrombin Hemostasis Patch Regulatory Class: Unclassified Product Code: QSX

Dear James Chapman:

The Food and Drug Administration (FDA) is sending this letter to notify you of an administrative change related to your previous substantial equivalence (SE) determination letter dated October 5, 2007. Specifically, FDA is updating this SE Letter because FDA has better categorized your device technology under product code QSX.

Please note that the 510(k) submission was not re-reviewed. For questions regarding this letter please contact Julie Morabito, OHT4: Office of Surgical and Infection Control Devices, 240-402-3839, Julie.Morabito@fda.hhs.gov.

> Sincerely, Julie A. Morabito -S Julie Morabito, Ph.D. Assistant Director

DHT4B: Division of Infection Control and Plastic Surgery Devices OHT4: Office of Surgical and Infection Control Devices Office of Product Evaluation and Quality Center for Devices and Radiological Health

{1}------------------------------------------------

Image /page/1/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a stylized eagle with three stripes forming its body and wings. The eagle is facing to the right. The text "DEPARTMENT OF HEALTH & HUMAN SERVICES • USA" is arranged in a circular fashion around the eagle.

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Vascular Solutions, Inc. % Mr. James Chapman Regulatory Affairs Associate 6464 Sycamore Court Minneapolis, Minnesota 55369

OCT 5 - 2007

Re: K072117

Trade/Device Name: Thrombix" Patch, Thrombin Hemostasis Patch Regulatory Class: Unclassified Product Code: FRO Dated: July 30, 2007 Received: August 01, 2007

Dear Mr. Chapman:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally 

{2}------------------------------------------------

Page 2 - Mr. James Chapman

marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115 Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Mark A. Milliken

Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health

Enclosure

{3}------------------------------------------------

# Indications for Use Statement

| 510(k) Number: | K072.117 |
|----------------|----------|
|----------------|----------|

Device Name:

Thrombix™ Patch thrombin hemostasis patch

#### Indications for Use:

The Thrombix™ Patch is applied topically and is indicated as an adjunct to manual compression and is indicated for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

(PLEASE DO NOT WRITE BELOW THIS LINE --CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office off

(Division Sign-Off) Division of General, Restorative, and Neurological Devices

510(k) Number K072117

{4}------------------------------------------------

#### OCT 5 * 2007 510(k) Summary (As required by 21 CFR 807.92(c))

510(k) Number: K072(17 Page 1 of 3

Date Prepared: July 30, 2007

# Submitter Information

| Submitter's Name: | Vascular Solutions, Inc.                     |
|-------------------|----------------------------------------------|
| Address:          | 6464 Sycamore Court<br>Minneapolis, MN 55369 |

Establishment Registration: 2134812

| Contact Person: | James Chapman                |
|-----------------|------------------------------|
|                 | Regulatory Affairs Associate |
|                 | Phone: (763) 656-4380        |
|                 | Fax: (763) 656-4253          |

# Device Information

| Trade Name:          | Thrombix™ Patch, Thrombin Hemostasis Patch |
|----------------------|--------------------------------------------|
| Common Name:         | Topical hemostat                           |
| Classification Name: | Unclassified                               |
| Product Code:        | FRO                                        |
| Regulation:          | Not Applicable                             |

# Predicate Device(s)

The predicate devices are the currently marketed D-Stat® Dry Topical Hemostat (K030836) and D-Stat® Dry 3x3 (K040510).

# Device Description

The Thrombix " Patch thrombin hemostasis patch consists of a lyophilized patch containing bovine derived thrombin as an aid to hemostasis (King Pharmaceuticals license number 0977), sodium carboxymethylcellulose (CMC), and calcium chloride.

The Thrombix" Patch is applied directly over the source of bleeding, creating a physical barrier to blood flow through the application of adjunctive manual compression. The lyophilized components (thrombin, CMC, and calcium chloride) establish an environment in which a natural blood clot can build and form a physical barrier to bleeding.

{5}------------------------------------------------

Image /page/5/Picture/0 description: The image contains handwritten text that appears to be a combination of letters and numbers. The text reads 'K072117'. The characters are written in a casual, slightly slanted style, and the overall impression is that of a handwritten code or identifier.

Vascular Solutions, Inc. Thrombix™ Patch Thrombin Hemostasis Patch 510(k) Summary Page 2 of 3

The thrombin facilitates hemostasis by enhancing the surface-activated clotting cascade through enzymatic cleavage and conversion of fibrinogen to fibrin.

# Intended Use/Indications for Use

The Thrombix" Patch is applied topically and is indicated as an adjunct to manual compression for the control of surface bleeding from vascular access sites and percutaneous catheters or tubes.

# Summary of Non-Clinical Testing

The Non-Clinical Testing included assessment of the physical properties of the Thrombix™ Patch and its ability to achieve its intended use. Bench testing of the physical properties of the Thrombix™ Patch confirmed the suitability of the device for its intended use. The following physical tests were performed:

- Moisture Content Testing ● Thrombin Activity Testing post 4 hours pot-life
- Wetting Time .
- pH .

A biocompatibility assessment of the device was also performed. The purpose of the biocompatibility assessment was to ensure that biocompatibility had been established for the device. The device is biocompatible based on the following testing:

- . MEM Elution
- . Intracutaneous Injection Test
- . Systemic Injection Test
- Kligman Skin Sensitization .

The results of the tests confirmed the suitability of the device for its intended use.

{6}------------------------------------------------

K072117

Vascular Solutions, Inc. Thrombix™ Patch Thrombin Hemostasis Patch 510(k) Summary Page 3 of 3

#### Summary of Clinical Testing

No human clinical testing was required for this device.

#### Statement of Equivalence

The Thrombix" Patch is substantially equivalent to the currently marketed D-Stat® Dry and D-Stat® Dry 3x3 hemostatic bandages based on a comparison of the indications for use and the technological characteristics of the device.

#### Conclusion

The Thrombix" Patch is substantially equivalent to the currently marketed D-Stat® Dry and D-Stat® Dry 3x3 hemostatic bandages based on the technological characteristics and sterilization methods. Bench tests confirmed the suitability of the device for its intended use.

---

**Source:** [https://fda.innolitics.com/device/K072117](https://fda.innolitics.com/device/K072117)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
