K071866 · Kcbiomedix, Inc. · HCC · Feb 1, 2008 · Neurology
Device Facts
Record ID
K071866
Device Name
NTRAINER SYSTEM
Applicant
Kcbiomedix, Inc.
Product Code
HCC · Neurology
Decision Date
Feb 1, 2008
Decision
SESE
Submission Type
Traditional
Regulation
21 CFR 882.5050
Device Class
Class 2
Attributes
Therapeutic, Pediatric
Indications for Use
The NTrainer System reinforces non-nutritive suck (NNS) in newborns and infants born prematurely.
Device Story
NTrainer System is a cribside biofeedback device for NICU use to assess and entrain orofacial motor systems in premature infants. Input consists of suck pressure measured via a disposable Soothie pacifier connected to an integrated pressure transducer. In assessment mode (NeoSuck), the system records suck intensity and frequency, displaying waveforms on a screen. In therapy mode (NTrainer), the system uses a low-pressure air pump (approx. 1.5 psi) to provide rhythmic, patterned somatosensory stimulation to the infant's lips, tongue, and jaw via the pacifier nipple. This synthetic input triggers a reflex action, encouraging the infant to adapt their suck pattern, thereby reinforcing NNS development. The device is operated by clinicians to improve oral feeding competency in infants with respiratory disease or brain insult. By entraining the orofacial motor system, the device aims to facilitate the coordinated sucking, breathing, and swallowing required for independent oral feeding.
Clinical Evidence
Clinical study included 30 newborns with moderate-severe Respiratory Distress Syndrome (RDS): 20 in the NTrainer therapy group and 10 in a control group. Therapy consisted of 3-minute sessions per feeding over 7 days. Results showed improvements in NNS burst cycles/minute, mouthing events, NNS bursts/minute, NNS cycles/burst, and a reduction in non-NNS events. No adverse events were reported.
Technological Characteristics
System includes a motor assembly, air cylinder, control electronics, display, and disposable Soothie pacifier. Sensing uses a pressure transducer to detect suck intensity/frequency. Therapy uses low-pressure air (approx. 1.5 psi, <2 cc volume) for rhythmic nipple deflection. Software is a minor level of concern application. Connectivity is standalone.
Indications for Use
Indicated for newborns and infants born prematurely (gestational age >25 weeks) who lack an organized non-nutritive suck (NNS).
Regulatory Classification
Identification
A biofeedback device is an instrument that provides a visual or auditory signal corresponding to the status of one or more of a patient's physiological parameters (e.g., brain alpha wave activity, muscle activity, skin temperature, etc.) so that the patient can control voluntarily these physiological parameters.
Special Controls
*Classification.* Class II (special controls). The device is exempt from the premarket notification procedures in subpart E of part 807 of this chapter when it is a prescription battery powered device that is indicated for relaxation training and muscle reeducation and prescription use, subject to § 882.9.
{0}------------------------------------------------
K071866
Appendix B
# 510(k) Summary
*KC BioMediX, Inc.*
Improving Neonatal Care
510(k) Summary - NTrainer
January 25, 2008 510(k) Owner - KCBioMedix, Inc. Michael Litscher, President & CEO 8960 Commerce Drive Suite 5C Desoto, KS 66018 785.856.2021 Email: mlitscher@kcbiomedix.com 785.832.8234 (fax)
Contact person: Stanley Consulting, LLC Jim Stanley, Principal Consultant 4200 SW Sapelo Drive Lee's Summit. MO 64082 Phone 816.550.4510 Email: stanleyconsulting@comcast.net FEB - 1 2003
- Trade Name NTrainer System .
- . Common Name - There is no common name for this device (or feeding therapy device)
- . Classification Name - Biofeedback device (21 CFR 882.5050, Product Code -HCC)
### Predicate Devices
There are two predicate devices. The Sondrex P.A.L System (K010388) is a pacifier system and method for reinforcing non-nutritive sucking (NNS) of premature infants and the Ultramind Biofeedback System (K980373) is an apparatus and software designed to train the user to control one or more aspects of their psycho-physiological state including the autonomic nervous system.
### Description of the NTrainer Device
A Soothie™ pacifier is inflated by a low pressure air pump that stimulates the baby's lips, tongue and jaw thus allowing the NTrainer System to reinforce non-nutritive suck
{1}------------------------------------------------
(NNS) in newborns and infants. This is accomplished by providing the newborns or infants with a synthetic patterned oral somatosensory input.
There is a central pattern generator in the brain and when normally developed provides the baby with a non-nutritive suck which is an essential building block in an infant's development of the coordinated sucking, breathing and swallowing capability needed for independent oral feed.
The NTrainer represents an integrated cribside application to assess and entrain the orofacial motor systems (lips, tongue and jaw). The baby is provided with a synthetic suck patterned by imposing rhythmic changes in the pacifier nipple which causes a reflex action in baby who adapts the suck pattern.
The NTrainer consists of the motor assembly, air cylinder, control electronics and computer system including a display. The only interface and contact with the baby is the disposable Soothie Pacifier. TM
The application software has two modes of operation, NeoSuck, the assessment mode which defines the baby's state of non-nutritive suck development and, NTrainer, the therapy mode which reinforce the development of non-nutritive suck. NTrainer and NeoSuck application software were originally developed by programmers at the University of Kansas (KU). This software has been reviewed and determined to be software having a "Minor Level of Concern" in accordance with FDA Guidance for the Content of Premarket Submissions for Software Contained in Medical Devices.
Indications for Use: The NTrainer System reinforces non-nutritive suck (NNS) in newborns and infants born prematurely.
### Intended use
Oral feeding competency is one of the most frequent and challenging hurdles facing many premature babies with respiratory disease and / or brain insult. Premature infants often lack a functional non-nutritive suck (NNS) or delayed suck patterning, dyscoordination of suck, swallow, breathe and poor state control. The lengthy intubation / oxygen supplementation procedures cost the baby precious sensory and motor experiences during a critical period of brain development for oromotor pattern generation.
The NTrainer system reinforces non-nutritive suck (NNS) in newborns, from birth (premature) to 1 month of age and in infants greater than 1 month. This is accomplished by providing the premature baby with a synthetic patterned oral somatosensory input. The intended target population is "newborns and infants, born prematurely with a gestation age (GA) greater than 25 weeks and lacking an organized NNS".
{2}------------------------------------------------
One predicate device, the Sondrex P.A.L. System, reinforces non-nutritive sucking of premature infants by playing music while the NTrainer System reinforces non-nutritive suck (NNS) in newborns (premature) and infants by providing the baby with a synthetic patterned oral somatosensory input.
The other predicate device, the Ultramind Biofeedback, is a biofeedback system is used for relaxation training while the NTrainer provides input to reinforce non nutritive suck.
There are no known risks associated with low-level mechanical stimulation, or recording low-level force (pressures) in the orofacial mechanism. Oral stimulation strategies have proven beneficial in developing oral feeding performance in premature infants (Fucile, Gisel, Lau, 2002, 2005; Rocha et al., 2006).
#### Technical Characteristics and Predicate Devices
The Sondrex PAL System predicate is a pacifier system and method of therapeutically treating premature infants to reinforce non-nutritive sucking. A pacifier is connected to a pressure transducer. The pressure transducer serves to detect sucking intensity and the duration of the suck. The pacifier system activates a sound source, such as a music source, upon sucking of a minimum intensity being detected for a minimum period of time. The sound (music) is intended to reinforce the non-nutritive sucking.
The Ultramind Biofeedback System with de-Stress software predicate is an apparatus and program designed to train the user to control one or more aspects of their psychophysiological state including the autonomic nervous system.
The NTrainer System is a pacifier system for assessing and a method of therapeutically stimulating the human orofacial system to reinforce non-nutritive sucking. The system includes a pacifier connected to an air cylinder with an integrated pressure transducer. In the assessment mode the pressure transducer serves to detect sucking intensity and time (frequency) of the suck. In the therapy mode the NTrainer Systems provides patterned, rhythmic stimulation to the infant's orofacial system to stimulate and entrain brain development.
Based on the technical comparison in the Substantial Equivalence Comparison (Section IV), the minor difference between the NTrainer System and the Sondrex PAL System is in the therapy mode. The NTrainer energy source for therapy is very low air pressure that is running about 105 cm H2O or about 1.5 psi with an air volume no greater than 2 cc's, to deflect the nipple cylinder. If the pacifier were to come loose or come off, the local pressure gradient proximal to the aperture on the receiver bulb would drop off steeply within a few millimeters to approach atmospheric pressure.
Technically, in the assessment mode, the NTrainer operates like the PAL. It uses all the same system elements including the measurement of the suction the baby creates while
{3}------------------------------------------------
sucking on the nipple. The difference is the NTrainer converts the signal to a waveform that is shown on a display instead of triggering music to be played.
Based on the technical comparison in the Substantial Equivalence Comparison the minor difference between the NTrainer System and the Ultramind Biofeedback System is in the assessment mode. The Ultramind Biofeedback system measures galvanic skin response in comparison to the NTrainer measuring the suck pressure on the nipple measured through a pressure transducer. The patient contact for the Ultramind Biofeedback System is a finger electrode versus a pacifier nipple for the NTrainer. The pacifier nipple is a commonly used device in the NICU without any known risk
## Non Clinical Performance Data
## Executive Summary -- Bench Tests
There are three key functional tests that were run as part of the assembly and test of each NTrainer: The performance test of the hand piece, the test and calibration of the Pneumatic Generator Assembly (PGA) and the final test of the finished device.
### Handpiece
Using the NeoSuck-RT program the test session was ran to obtain a display showing that the pressure transducer's response was to specifications and it did not leak.
## Pneumatic Generator Assembly
The components of the Pneumatic Generator Assembly (PGA) are used to create the pressure pulse fed to the nipple to entrain the baby. The servo controller and the linear motor were aligned and adjusted to specification. A wave form was captured which confirmed that the system was functioning properly.
### Final System Performance Test
The NTrainer-RT program and NeoSuck-RT program were ran to verify they are operation properly. The displayed wave forms were captured, printed and compared to the requirements. The results met specifications.
In the therapy mode a test sequence was ran to confirm the device is producing the correct frequency and amplitude of the entraining pulses. The expansion of the nipple was sampled and confirmed that the amplitude is correct.
{4}------------------------------------------------
## Clinical Performance Data
The following clinical results are derived from the clinical study that included 30 newborns with moderate-severe Respiratory Distress Syndrome. 20 RDS newborns were provided NTrainer therapy (the RDS Test group) and 10 RDS newborns did not receive therapy as the control group (the RDS Control group).
The RDS Test infants were given a 3 minute therapy session at each feeding. Typically the training was accomplished over a 7 day period of time.
Numerous performance factors were measured to quantify how the improvement was measured. The performance factors included:
- Increase in NNS burst cycles / minute .
- Increase in mouthing event / cycles / minute .
- Increase in NNS bursts / minute ●
- Increase in NNS cycles / burst .
- . Increase in NNS cycles ( % of mouthing)
- t . Reduction in non-NNS events / minute
### Conclusion
In conclusion we have demonstrated that the NTrainer is as safe, as effective and performs better than the Sondrex PAL System in entraining NNS in newborns and infants. Our clinical study showed that entraining is accomplished by providing the baby with a synthetic patterned oral somatosensory input. Additionally there were no adverse events during the study.
The only clinical data found for the Sondrex PAL Systems was in the Journal of Music Therapy: Vol. 42, No. 2, pp. 123 - 139. (Effects of the Pacifier Activated Lullaby on Weight Gain of Premature Infants). During a 2-year time period, 62 infants from a sample of 188 met the criteria for analysis but showed no significance in daily weight gain for the number of PAL trials completed.
{5}------------------------------------------------
Image /page/5/Picture/1 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of a circular seal with the text "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" arranged around the perimeter. Inside the circle is a stylized symbol featuring three human profiles facing right, with flowing lines above them.
Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850
FEB - 1 2008
KCBiomedix, Inc % Mr. James Stanley Stanley Consulting, LLC 4200 Southwest Sapelo Drive Lees Summit, MO 64082
Re: K071866 Trade/Device Name: NTrainer System Regulation Number: 21 CFR 882.5050 Regulation Name: Biofeedback device Regulatory Class: Class II Product Code: HCC Dated: January 25, 2008 Received: January 28, 2008
Dear Mr. Stanley:
We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food. Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.
If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.
Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.
{6}------------------------------------------------
Page 2 - Mr. James Stanlev
This letter will allow you to begin marketing your device as described in your Section 510/k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.
If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Center for Devices and Radiological Health's (CDRH's) Office of Compliance at (240) 276-0120. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). For questions regarding postmarket surveillance, please contact CDRH's Office of Surveillance and Biometric's (OSB's) Division of Postmarket Surveillance at (240) 276-3474. For questions regarding the reporting of device adverse events (Medical Device Reporting (MDR)), please contact the Division of Surveillance Systems at (240) 276-3464. You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at toll-free number (800) 638-2041 or (240) 276-3150 or the Internet address http://www.fda.gov/cdrh/industry/support/index.html
Sincerely yours,
Mark N. Millerson
Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health
Enclosure
{7}------------------------------------------------
## INDICATIONS FOR USE
510(k) Number (if known): K071866
Device Name: NTrainer System
Indications for Use:
The NTrainer System reinforces non-nutritive suck (NNS) in newborns and infants born prematurely.
Prescription Use __X_ (Part 21 CFR 801 Subpart D)
AND/OR
Over-The-Counter Use _ (21 CFR 801 Subpart C)
(PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED)
Concurrence of CDRH, Office of Device Evaluation (ODE)
Mark A. Wilkerson
Sign-Off) Division of General, Restorative, and Neurological Devices
**510(k) Number** K071866
Predicate graph will load when search results are available.
Embedding visualization will load when search results are available.
PDF viewer will load when search results are available.
Loading panels...
Select an item from Submissions
Click any panel, subpart, regulation, product code, or device to see details here.
Section Matches
Results will appear here.
Product Code Matches
Results will appear here.
Special Control Matches
Results will appear here.
Loading collections...
Loading
My Alerts
You will receive email notifications based on the filters and frequency you set for each alert.
Sort by:
Create Alert
Search Filters
Agent Token
Create a read-only bearer token for Claude, ChatGPT, or other agents that can call HTTP APIs.
Copy this now. It will not be shown again.
Connected apps
Apps you authorized through browser sign-in. Disconnecting revokes their access immediately.
Report an issue or suggest a feature
Spot a broken citation, a wrong extraction, a missing document, or something you wish this app did? Tell us — no account needed.
or drag and drop files here
Learn the FDA Browser
Two short videos show you everything — or skip straight to the written tutorial if you'd rather read. You can reopen this any time from the Tutorial button in the top bar.
Part 1 — Search, results, and everyday workflows 16 min
Part 2 — Embeddings: the galaxy map 3 min
1. Search: exact and fuzzy
Type a phrase like "coronary artery calcification" into the search box. You get two kinds of results. Exact results match the literal phrase — prefix searches work ("coronary artery calcificati") but suffix searches do not. Fuzzy results match on the meaning and intent of your phrase rather than the exact words, and are sorted by relevance score. Hover over the Exact or Fuzzy badge on any row to see exactly why it matched.
Use the checkboxes above the results to narrow: SaMD keeps only Software as a Medical Device, AI / ML keeps only devices with AI.
Exact vs. fuzzy search: what's the difference?
Exact matches on the literal phrase (prefix search works, suffix does not). Fuzzy matches on the meaning and intent of the phrase rather than the exact words. Hover over the badge on any row to see why it matched.
You search "coronary artery calcification" and want only software devices with AI. What two filters do you apply?
Narrow by SaMD (Software as a Medical Device), then narrow by AI/ML (devices with AI).
2. The results table
Scroll right in the results table. The intended use is extracted for you — no need to open the PDF. The device story gives a high-level snapshot of what the device does and how it's used. The AI Performance sub-table shows each output name, acceptance criteria, observed values, and development/test dataset descriptions — the same format Innolitics uses for regulatory strategy outputs, and the fastest high-level fingerprint of an AI device. It is AI-generated but has been very reliable in practice.
Where do you find a device's intended use without opening the PDF?
Scroll right in the search results table. The intended use column is extracted for you; no need to dig into the 510(k) summary PDF.
What does the AI Performance sub-table show, and why is it useful?
Output name, acceptance criteria, observed values, development dataset description, and test dataset description. It's the same format we use for regulatory strategy output and Fast 510(k) input, and the fastest high-level fingerprint of an AI device. AI-generated but reliable in practice.
3. Judging fuzzy relevance
Fuzzy results trail off in relevance as you scroll. Use three signals to decide how far down to go: the fuzzy badge explanations, the intended use column, and whether your target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, you're past the relevant zone. A top hit with a low score (~0.4) and a stretched explanation is a hint the closest predicates are far away — the project may be headed for De Novo. Note the fuzzy search is a pattern match: it doesn't handle negation ("not") well, and hardware devices can appear — filter by SaMD/AI ML to cut them.
How do you judge how far down fuzzy search results to go?
Use the relevancy signals: the fuzzy badge explanations, the intended use column, and whether the target output (e.g., Cobb angle) still appears in the AI Performance sub-table. Once it stops appearing, results are trailing off in relevancy.
4. Device detail page: chat and citations
Click a device name to open its detail page: device facts on the left, a chat window on the right. Ask something like "Describe the training data". The answer carries little citation bubbles — click one to jump to the highlighted passage in the source PDF, so you can verify every AI answer against the document. There's also a Download PDF button for sharing.
How do you verify an AI chat answer on the device detail page?
Click the citation bubbles to jump to the relevant highlight in the source document.
Reading rule for every project: how many summaries do you read in full?
At least the three most relevant 510(k) or De Novo summaries, in full. After that, use targeted chat questions to confirm your memory quickly. The tool supports this professional habit — it doesn't replace it.
5. Side-by-side comparison
Select multiple rows in the results table (aim for under ~10), then open the PDF Viewer tab. Ask one question — it goes to all selected devices in parallel, each with citations. This is the fastest way to compare and contrast devices: training data, PCCP scope, how they handled adding new scanners, and so on.
What does the side-by-side PDF viewer mode do?
Select multiple devices, open the PDF viewer tab, and ask one question (e.g., "Describe the training data"). It queries all selected devices simultaneously with citations, so you can compare and contrast quickly.
6. Collections
With rows selected, go to the Collections tab and create a labeled collection (e.g., "Cobb Angle Project"). Reload that selection any time — before a client call, pull up the collection and ask questions across all of its devices at once.
How do you save a set of selected devices for later use?
Select the rows, go to the Collections tab, and create a labeled collection (e.g., "Cobb Angle Project"). You can reload the selection anytime and carry it into the PDF viewer and other tabs that support selections.
7. Product codes and the regulations tree
Click a product code in the results to jump to it in the regulations tree — identification text, sibling product codes, and devices you can open in a PDF viewer on the right. Click a regulation number to see its identification, special controls, and related product codes. You can also search by product code or regulation number at the top of the tree. Always read the special controls if any exist for your device — it broadens your search and sharpens pre-kickoff research.
What can you do from the regulations tree view?
Browse product codes and regulation numbers, read the identification text and special controls, browse sibling product codes, open device PDFs on the right, and search by product code or regulation number at the top of the tree.
8. Chart view
Click Show Chart and segment by regulation number (or product code) to see which regulations dominate your result set. Clicking a regulation takes you into the regulations tree. Great for spotting that most matches are, say, hardware laparoscopic devices — a cue to go back and filter.
How do you see which regulations dominate a search result set?
Click "Show Chart" and segment by Regulation Number. Clicking a regulation takes you to the regulations tree.
9. The predicate graph
Open the Predicates tab for a family-tree view of predicate relationships. Click a node to trace its parents and children; selections from search carry over pre-selected. Commonly predicated devices are worth reading — a lot of people predicated them for a reason. The visual lineage is also handy on client calls, e.g. to show how a predicate family evolved and justify why your predicate still holds.
In the predicate graph, why are commonly predicated devices worth reading?
A lot of people predicated them for a reason. Clicking a node traces parents and children, and selections from search carry over pre-selected.
10. Embeddings: the galaxy map
The Embeddings tab plots every matching document in a 2-D "galaxy map" where semantically similar devices cluster together. Hover or click clusters to explore, and let AI label the clusters for you. Embeddings beat product codes for grouping: two devices can carry different product codes (LLZ vs. QIH) yet do the same thing — the embedding captures the meaning of the intended use and device story. This is also exactly how retrieval-augmented generation (RAG) works under the hood, and it makes a great visual on client calls.
Try it yourself
Head to the search page and work through a few of these AI/ML fuzzy searches to build intuition: perivascular fat on CT · aortic valve calcification opportunistic screening on noncontrast CT · breast cancer prediction on digital pathology slides · autism detection · gestational age prediction · a hearing aid that can also detect a pulse · foundation model based analysis of ECG · large language models · penetration test. Watch how the relevance scores, intended use, and AI Performance tables tell you when results stop being meaningful.