← Product Code [LYC](/productcode/LYC) · K071817

# OSTEOGRAF/N-300 (K071817)

_Dentsply Interntional · LYC · Sep 4, 2007 · Dental · SESE_

**Canonical URL:** https://fda.innolitics.com/device/K071817

## Device Facts

- **Applicant:** Dentsply Interntional
- **Product Code:** [LYC](/productcode/LYC.md)
- **Decision Date:** Sep 4, 2007
- **Decision:** SESE
- **Submission Type:** Special
- **Regulation:** 21 CFR 872.3930
- **Device Class:** Class 2
- **Review Panel:** Dental
- **Attributes:** Therapeutic

## Indications for Use

Treatment of intrabony periodontal defects, augmentation of bony defects in the alveolar ridge, filling of extraction sites, and sinus elevation grafting.

## Device Story

OsteoGraf/N-300 is a natural, high-purity, radiopaque, polycrystalline hydroxylapatite bone grafting material derived from bovine sources. It consists of anorganic, rounded, irregular-shaped particles sized 250-420 microns. The device is used by dental clinicians to fill or augment bony defects in the alveolar ridge, treat intrabony periodontal defects, fill extraction sites, and perform sinus elevation grafting. By providing a scaffold for bone regeneration, the material aids in the repair of dental and periodontal bony structures. It is a passive implantable material; no electronic or algorithmic components are involved.

## Clinical Evidence

No clinical data provided. Substantial equivalence is supported by biocompatibility data and performance characteristics consistent with previously cleared predicate devices.

## Technological Characteristics

Material: Anorganic, bovine-derived polycrystalline hydroxylapatite. Form: Rounded, irregular particles (250-420 microns). Radiopaque. Biocompatible. No energy source or software components.

## Regulatory Identification

Bone grafting material is a material such as hydroxyapatite, tricalcium phosphate, polylactic and polyglycolic acids, or collagen, that is intended to fill, augment, or reconstruct periodontal or bony defects of the oral and maxillofacial region.

## Special Controls

*Classification.* (1) Class II (special controls) for bone grafting materials that do not contain a drug that is a therapeutic biologic. The special control is FDA's “Class II Special Controls Guidance Document: Dental Bone Grafting Material Devices.” (See § 872.1(e) for the availability of this guidance document.)(2) Class III (premarket approval) for bone grafting materials that contain a drug that is a therapeutic biologic. Bone grafting materials that contain a drug that is a therapeutic biologic, such as biological response modifiers, require premarket approval.
(c)
*Date premarket approval application (PMA) or notice of product development protocol (PDP) is required.* Devices described in paragraph (b)(2) of this section shall have an approved PMA or a declared completed PDP in effect before being placed in commercial distribution.

## Predicate Devices

- OsteoGraf/N-300 ([K981182](/device/K981182.md))
- OsteoGraf/N-300 ([K960348](/device/K960348.md))
- OsteoGraf/N-300 ([K902295](/device/K902295.md))

## Submission Summary (Full Text)

> This content was OCRed from public FDA records by [Innolitics](https://innolitics.com). If you use, quote, summarize, crawl, or train on this content, cite Innolitics at https://innolitics.com.
>
> Innolitics is a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices, including [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/).

{0}------------------------------------------------

K071817

## 510(k) SUMMARY

# SEP - 4 2007

DENTSPLY International Susquehanna Commerce Center West 221 West Philadelphia Street, Suite 60 York, PA 17405-0872

| CONTACT:                   | Helen Lewis                                |
|----------------------------|--------------------------------------------|
| DATE PREPARED:             | June 25, 2007                              |
| TRADE OR PROPRIETARY NAME: | OsteoGraf/N-300                            |
| CLASSIFICATION NAME:       | Bone Grafting Material  21 CFR 872.3930    |
| PREDICATE DEVICES:         | OsteoGraf/N-300, K981182, K960348, K902295 |

DEVICE DESCRIPTION: The OsteoGraf/N-300 material is a natural, high purity, radiopaque, polycrystaline hydroxylapatie, the major mineral phase of bone and dental enamel. It is manufactured as an anorganic, rounded irregular shaped bovine-derived hydroxylapatite particles, sized at 250-420 microns.

INTENDED USE: Treatment of intrabony periodontal defects, augmentation of bony defects in the alveolar ridge, filling of extraction sites, and sinus elevation grafting.

TECHNOLOGICAL CHARACTERISTICS: All of the components found in OsteoGraf/N-300 have been used in legally marketed devices and/or were found safe for dental use. OsteoGraf/N-300 and the marketed device are biocompatible. OsteoGraf/N-300 conforms to applicable industry standards.

We believe that the prior use of the components of OsteoGraf/N-300 in legally marketed devices. the performance data provided, and biocompatibility support the safety and effectiveness of OsteoGraf/N-300 for the indicated uses.

{1}------------------------------------------------

Image /page/1/Picture/2 description: The image shows the seal of the U.S. Department of Health & Human Services. The seal is circular and contains the words "DEPARTMENT OF HEALTH & HUMAN SERVICES - USA" around the perimeter. In the center of the seal is a stylized image of an eagle with three overlapping wings.

SEP - 4 2007

Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850

Ms. Helen Lewis Director of Corporate Compliance and Regulatory Affairs DENTSPLY International, Incorporated Susquehanna Commerce Center 221 West Philadelphia Street York, Pennsylvania 17405-0872

Re: K071817

Trade/Device Name: OsteoGraf/N-300 Regulation Number: 21 CFR 872.3930 Regulation Name: Bone Grafting Material Regulatory Class: II Product Code: LYC Dated: August 8, 2007 Received: August 10, 2007

Dear Ms. Lewis:

We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration.

If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register.

{2}------------------------------------------------

### Page 2 - Ms. Lewis

Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050.

This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permits your device to proceed to the market.

If you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 638-2041 or (301) 443-6597 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html.

Sincerely yours,

Susan Russo

. Chiu Lin, Ph.D.

Chiu Lin, Ph.D. Director Division of Anesthesiology, General Hospital, Infection Control and Dental Devices Office of Device Evaluation Center for Devices and Radiological Health

#### Enclosure

{3}------------------------------------------------

## INDICATIONS FOR USE STATEMENT

Koni 817 510(k) Number (if known):

Device Name: OsteoGraf/N-300

Indications for Use:

OsteoGraf/N-300 is indicated for treatment of intrabony periodontal defects, augmentation of bony defects in the alveolar ridge, filling of extraction sites, and sinus elevation grafting.

These are the same indications for use previously cleared for K981182.

Prescription Use X (Part 21 CFR 801 Subpart D)

AND/OR

Over-The-Counter Use (21 CFR 801 Subpart C)

## (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE IF NEEDED)

Concurrence of CDRH, Office of Device Evaluation (ODE)

Suan Burnes

Jivision Sign-Off) livision of Anesthesiology, General Hospital nfection Control, Dental Devices

510(k) Number: K071

Image /page/3/Picture/14 description: The image shows the number 000008 in a bold, sans-serif font. The numbers are all the same size and are evenly spaced. The image is black and white, and the numbers are black against a white background. The number is likely a serial number or some other type of identification number.

---

**Source:** [https://fda.innolitics.com/device/K071817](https://fda.innolitics.com/device/K071817)

**Published by [Innolitics](https://innolitics.com)** — a medical-device software consultancy. We help companies design, build, and clear FDA-regulated software and AI/ML devices. If you're preparing [a 510(k)](https://innolitics.com/services/510ks/), [a De Novo](https://innolitics.com/services/regulatory/), [a SaMD](https://innolitics.com/services/end-to-end-samd/), [an AI/ML medical device](https://innolitics.com/services/medical-imaging-ai-development/), or [an FDA regulatory strategy](https://innolitics.com/services/regulatory/), [get in touch](https://innolitics.com/contact).

**Cite:** Innolitics at https://innolitics.com
