PORTRAIT PSR

K071786 · Rhytec Incorporated · GEI · Jul 26, 2007 · General, Plastic Surgery

Device Facts

Record IDK071786
Device NamePORTRAIT PSR
ApplicantRhytec Incorporated
Product CodeGEI · General, Plastic Surgery
Decision DateJul 26, 2007
DecisionSESE
Submission TypeSpecial
Regulation21 CFR 878.4400
Device ClassClass 2
AttributesTherapeutic

Indications for Use

The Portrait® PSR³ is indicated for treatment of the following dermatological conditions: Treatment of wrinkles and rhytides . Superficial skin lesions . Actinic Keratosis . Viral Papillomata . Seborrhoeic Keratosis .

Device Story

Portrait® PSR3 is an electrosurgical device for dermatological applications. It uses UHF energy to convert nitrogen gas into plasma within a handpiece. The plasma emerges from the nozzle and is directed onto the skin, causing rapid heating through energy transfer. The device produces individual plasma pulses to achieve predictable tissue effects. It is operated by a clinician. Modifications include adjusting the plasma pulse repetition rate (1 to 2.5 Hz), removing the 12-hour nozzle life limit (now monitored by pulse count), decreasing continuous pulses per foot switch press to 50, and adding a pulse count display. The device is used to treat wrinkles, rhytides, and various skin lesions.

Clinical Evidence

No clinical data provided; the manufacturer states that device modifications did not raise new questions of safety or efficacy and no performance data is required.

Technological Characteristics

Electrosurgical cutting and coagulation device. Uses UHF energy to generate nitrogen plasma. Features include adjustable pulse repetition rate (1-2.5 Hz), pulse counting for nozzle life monitoring, and a display for pulse count per treatment. Operates via foot switch with a limit of 50 continuous pulses.

Indications for Use

Indicated for treatment of dermatological conditions including wrinkles, rhytides, superficial skin lesions, actinic keratosis, viral papillomata, and seborrhoeic keratosis.

Regulatory Classification

Identification

An electrosurgical cutting and coagulation device and accessories is a device intended to remove tissue and control bleeding by use of high-frequency electrical current.

Predicate Devices

Submission Summary (Full Text)

{0}------------------------------------------------ Page 1 of ② ## 510(k) Summary for the Rhytec, Inc. Portrait® PSR3 0 71 786 JUL 26 2007 This 510(k) Summary of Safety and Effectiveness is being submitted in accordance with the requirements of the SMDA 1990 and 21 CFR 807.92. ### 1. General Information | Submitter: | Rhytec, Inc.<br>130 Turner Street<br>Building Two<br>Waltham, MA 02453 | |---------------------------|----------------------------------------------------------------------------------------------------------------------------| | Contact Person: | Robert Zoletti<br>Director, Regulatory Affairs and Quality<br>Rhytec, Inc.<br>Telephone: 781-419-9482<br>Fax: 781-419-9401 | | Summary Preparation Date: | June 22, 2007 | | Names | | | Device Name: | Portrait® PSR3 | | Classification Name: | Electrosurgical cutting and coagulation<br>device | #### 3. Predicate Devices 2. K060948 - Portrait® PSR3 Rhytec, Inc. Portrait® PSR3 #### 4. Device Description The Portrait® PSR³ is an electro-surgical device for use in dermatological applications. UHF energy from the generator converts Nitrogen gas into plasma within the handpiece. The plasma emerges from the nozzle at the distal end of the handpiece and is directed onto the skin to be treated. Rapid heating of the skin occurs as the excited gas gives up energy to the skin. Through the combination within the handpiece of precisely controlled energy and Nittrogen gas, individual plasma pulses are produced that will give predictable tissue effects. Product Code: GEI {1}------------------------------------------------ 0717-86 Page 2 of ② 510(k) Summary for the Rhytec, Inc. Portrait® PSR3 ## 5. Indications for Use The Portrait® PSR³ is indicated for treatment of the following dermatological conditions: - Treatment of wrinkles and rhytides ● - Superficial skin lesions . - . Actinic Keratosis - . Viral Papillomata - Seborrhoeic Keratosis . # 6. Technological Characteristics Modifications include: - Revised the allowable plasma pulse repetition rate from 1 to 4.0 Hz to 1 to . 2.5 Hz as recommended in the treatment parameters in K060948. - Removal of the 12 hour nozzle use life. Nozzle life will continue to be . monitored by the number of pulses used. - Decreased the number of continuous pulses with the foot switch pressed . from 120 to 50. - Added the number of pulses per patient treatment to the display. . ## 7. Clinical and Non-Clinical Data The device modifications did not raise new questions of safety or efficacy, no performance data is required. {2}------------------------------------------------ Image /page/2/Picture/12 description: The image shows the logo for the U.S. Department of Health & Human Services. The logo consists of the department's name in a circular arrangement around a stylized symbol. The symbol is a representation of human figures, with three overlapping profiles. The profiles are connected by flowing lines, creating a sense of unity and movement. Food and Drug Administration 9200 Corporate Boulevard Rockville MD 20850 Rhytec Incorporated % Mr. Robert Zoletti Director, Regulatory Affairs 130 Turner Street, Building Two Waltham, Massachusetts 02453 JUL 26 2007 Re: K071786 Trade/Device Name: Rhytec, Inc. Portrait® Regulation Number: 21 CFR 878.4400 Regulation Name: Electrosurgical cutting and coagulation device and accessories Regulatory Class: II Product Code: GEI Dated: June 27, 2007 Received: July 2, 2007 Dear Mr. Zoletti: We have reviewed your Section 510(k) premarket notification of intent to market the device referenced above and have determined the device is substantially equivalent (for the indications for use stated in the enclosure) to legally marketed predicate devices marketed in interstate commerce prior to May 28, 1976, the enactment date of the Medical Device Amendments, or to devices that have been reclassified in accordance with the provisions of the Federal Food, Drug, and Cosmetic Act (Act) that do not require approval of a premarket approval application (PMA). You may, therefore, market the device, subject to the general controls provisions of the Act. The general controls provisions of the Act include requirements for annual registration, listing of devices, good manufacturing practice, labeling, and prohibitions against misbranding and adulteration. If your device is classified (see above) into either class II (Special Controls) or class III (PMA), it may be subject to such additional controls. Existing major regulations affecting your device can be found in the Code of Federal Regulations, Title 21, Parts 800 to 898. In addition, FDA may publish further announcements concerning your device in the Federal Register. Please be advised that FDA's issuance of a substantial equivalence determination does not mean that FDA has made a determination that your device complies with other requirements of the Act or any Federal statutes and regulations administered by other Federal agencies. You must comply with all the Act's requirements, including, but not limited to: registration and listing (21 CFR Part 807); labeling (21 CFR Part 801); good manufacturing practice requirements as set forth in the quality systems (QS) regulation (21 CFR Part 820); and if applicable, the electronic product radiation control provisions (Sections 531-542 of the Act); 21 CFR 1000-1050. {3}------------------------------------------------ Page 2 -- Mr. Robert Zoletti This letter will allow you to begin marketing your device as described in your Section 510(k) premarket notification. The FDA finding of substantial equivalence of your device to a legally marketed predicate device results in a classification for your device and thus, permitte your device to proceed to the market. lf you desire specific advice for your device on our labeling regulation (21 CFR Part 801), please contact the Office of Compliance at (240) 276-0115. Also, please note the regulation entitled, "Misbranding by reference to premarket notification" (21CFR Part 807.97). You may obtain other general information on your responsibilities under the Act from the Division of Small Manufacturers, International and Consumer Assistance at its toll-free number (800) 633-2011 or (240) 276-3150 or at its Internet address http://www.fda.gov/cdrh/industry/support/index.html. Sincerely yours, Mark A. Milliken Mark N. Melkerson Director Division of General, Restorative and Neurological Devices Office of Device Evaluation Center for Devices and Radiological Health Enclosure {4}------------------------------------------------ #### Indications for Use | 510(k) Number (if known): | K071786 | |---------------------------|------------------------| | Device Name: | Rhytec, Inc. Portrait® | | PSR <sup>3</sup> | | The Portrait® PSR³ is indicated for treatment of the following dermatological conditions: - Treatment of wrinkles and rhytides . - Superficial skin lesions . - Actinic Keratosis ● - Viral Papillomata . - Seborrhoeic Keratosis . Indications for Use are unchanged from K060948 Prescription Use X (Part 21 CFR 801 Subpart D) AND/OR Over-The-Counter Use (21 CFR 801 Subpart C) (PLEASE DO NOT WRITE BELOW THIS LINE-CONTINUE ON ANOTHER PAGE OF NEEDED) Concurrence of CDRH, Office of Device Evaluation (ODE) Mark A. Milken (Division Sign-Off) (Division Sign-On) Restorative, and Neurological Devices 510(k) Number
Innolitics
Guidance Recalls

Panel 1

/
Ready

Predicate graph will load when search results are available.

Embedding visualization will load when search results are available.

PDF viewer will load when search results are available.

Loading panels...

Select an item from Submissions

Click any panel, subpart, regulation, product code, or device to see details here.

Section Matches

Results will appear here.

Product Code Matches

Results will appear here.

Special Control Matches

Results will appear here.

Loading collections...